Minim Invasive Neurosurg 2003; 46(2): 117-120
DOI: 10.1055/s-2003-39344
Technical Note
© Georg Thieme Verlag Stuttgart · New York

Updating of Neuronavigation Based on Images Intraoperatively Acquired with a Mobile Computerized Tomographic Scanner: Technical Note

N.  Nakao1 , K.  Nakai1 , T.  Itakura1
  • 1Department of Neurological Surgery, Wakayama Medical University, Wakayama, Japan
Further Information

Publication History

Publication Date:
22 May 2003 (online)

Abstract

Image-guided surgery based on preoperatively obtained image data is susceptible to inaccuracy resulting from intraoperative brain shift and distortion. We report a technique of updating neuronavigational data with the aid of a mobile computerized tomographic (CT) scanner. A mobile CT which is readily available in an ordinary operating room was used to acquire intraoperative images. A total of 6 - 7 titanium screws placed on the skull were used as new reference points for updating navigation. Intraoperative CT scanning was performed with a 2 mm slice thickness. After the obtained image data were transferred as Dicom files to the computer workstation of the navigation system through an Ethernet connection, navigational data were updated to registering the new reference points. Under the guidance of the updated navigation, residual lesions were explored, and further resected. Our preliminary experience in 8 patients indicates that interactive image-guidance can stably be updated based on images intraoperatively acquired with a mobile CT scanner. Comparing to intraoperative magnetic resonance imaging, this technique can simply be done in an ordinary operating room without requiring special surgical instruments, thus making it possible to update interactive image guidance on demand during an operation.

References

  • 1 Matula C, Rossler K, Reddy M, Shindler E, Koos W T. Intraoperative computed tomography guided neuronavigation: concepts, efficiency, and work flow.  Computer Aided Surgery. 1998;  3 174-182
  • 2 Wirtz C R, Bonsanto M M, Knauth M, Tronnier V M, Albert F K, Staubert A, Kunze S. Intraoperative magnetic resonance imaging to update interactive navigation in neurosurgery: method and preliminary experience.  Computer Aided Surgery. 1997;  2 172-179
  • 3 Wirtz C R, Knauth M, Staubert A, Bonsanto M M, Sartor K, Kunze S, Tronnier V M. Clinical evaluation and follow-up results for intraoperative magnetic resonance imaging in neurosurgery.  Neurosurgery. 2000;  46 1112-1122
  • 4 Butler W E, Piaggio C M, Constantinou C, Nicklason L, Gonzalez R G, Cosgrove G R, Zervas N T. A mobile computed tomography scanner with intraoperative and intensive care unit applications.  Neurosurgery. 1998;  42 1304-1311
  • 5 Gunnarsson T, Theodorsson A, Karlsson P, Fridricksson S, Bostrom S, Persliden J, Johansson I, Hillman J. Mobile computed tomography scanning in the neurosurgery intensive care unit: increase in patient safety and reduction of staff workload.  J Neurosurg. 2000;  93 432-436
  • 6 Germano I M. The NeuroStation system for image-guided frameless stereotaxy.  Neurosurgery. 1995;  37 348-350
  • 7 Germano I M, Villalobos H, Silvers A, Post K D. Clinical use of the optical digitizer for intracranial neuronavigation.  Neurosurgery. 1999;  45 261-270
  • 8 Smith K R, Frank K J, Bucholz R D. The NeuroStation: A highly accurate, minimally invasive solution to frameless stereotactic neurosurgery.  Comput Med Imaging Graph. 1994;  18 247-256
  • 9 Kelly P J. Volumetric stereotactic surgical resection of intra-axial brain mass lesions.  Mayo Clinic Proc. 1988;  63 1186-1198
  • 10 Berger M S. Ultrasound-guided stereotaxic biopsy using a new apparatus.  J Neurosurg. 1986;  65 550-554
  • 11 Koivukangas J, Louhisalmi Y, Alakuijala J, Oikarinen J. Ultrasound-controlled neuronavigator-guided brain surgery.  J Neurosurg. 1993;  79 36-42
  • 12 Trobaugh J W, Richard W D, Smith K R, Bucholz R D. Frameless stereotactic ultrasonography: Method and applications.  Comput Med Imaging Graph. 1994;  18 235-246
  • 13 Gronningsaeter A, Kleven A, Ommedal A, Aarseth T E, Lie T, Lindsen F, Lamgo T, Unsgard G. SonoWand, an ultrasound-based neuronavigation system.  Neurosurgery. 2000;  47 1371-1380

N. Nakao,M. D. 

Department of Neurological Surgery · Wakayama Medical University

811-1 Kimiidera

Wakayama 641-0012

Japan

Fax: +81-734-47-1771

Email: nnakao@wakayama-med.ac.jp

    >