Minim Invasive Neurosurg 2004; 47(1): 29-31
DOI: 10.1055/s-2003-812535
Original Article
© Georg Thieme Verlag Stuttgart · New York

Fluoroscopic Frameless Computer-Assisted Navigation for Transsphenoidal Surgery: A Clinical Assessment of Accuracy in Spatial Position and Trajectory

G.  K.  Wong1 , W.  S.  Poon1 , S.  C.  Ng1 , A.  Y.  Cheng2
  • 1Division of Neurosurgery, Department of Surgery Prince of Wales Hospital, Chinese University of Hong Kong, Hong Kong
  • 2Faculty of Science and Engineering, City University of Hong Kong, Hong Kong
Further Information

Publication History

Publication Date:
20 April 2004 (online)

Preview

Abstract

Objective: Fluoroscopic navigation technique in transsphenoidal surgery on the one hand provides multi-planar (antero-posterior and lateral views) navigation while on the other hand it does not need any preoperative preparation. To assess the clinical accuracy of the above technique in the transsphenoidal surgery, we performed this study.

Methods: 5 patients undergoing transsphenoidal surgery were assessed. Fluoroscopic images were compared with C-arm X-ray images for difference in spatial position and trajectory.

Results: 26 sets of data were collected for analysis. Mean pointer tip difference was 0.6 mm. The 95 % confidence interval was 1.3 mm. Mean trajectory difference was 1.2 degree. The 95 % confidence interval was 2.7 degree.

Conclusion: With a good clinical accuracy, fluoroscopic navigation offers a distinct advantage to replace the traditional fluoroscopic approach for transsphenoidal surgery for its low X-ray exposure and multi-planar (antero-posterior and lateral views) navigation.

References

W. S. Poon,FRCS 

Division of Neurosurgery · Department of Surgery · Prince of Wales Hospital

Shatin

New Territories

Hong Kong SAR

China

Phone: + 852-2632-2624

Fax: + 852-2637-7974

Email: wpoon @surgery.cuhk.edu.hk