Abstract
The waterjet dissector device has been shown to allow precise tissue dissection under
vessel preservation in liver surgery as well as in first experimental studies in the
brain. In clinical neurosurgery, the instrument seems to reduce the operative blood
loss and the surgical trauma in poorly demarcated or well vascularized tumors. To
evaluate its application characteristics in intrinsic brain tumors, 51 patients harboring
gliomas (WHO grades 1 - 4) were operated on with the waterjet dissector. As this device
is often compared with ultrasonic aspiration, it was in addition directly compared
with the ultrasonic aspirator in twelve cases. Intraoperatively, the waterjet enabled
precise tumor separation from the brain under preservation of the tumor-parenchyma
border. Tumor aspiration was possible in all but one case. Waterjet pressures of 3
to 17 bars were used, and even very small blood vessels were preserved. Complete tumor
mass removal was documented by MRI at three months follow-up in 41 of 43 cases where
total resection was intended. Out of 21 patients with tumors located within or close
to eloquent cortex, only one patient suffered from permanent worsening of his/her
preoperative deficit. In direct comparison with the ultrasonic aspirator, the waterjet
significantly reduced the intraoperative blood loss. Tumor debulking was performed
more efficiently with ultrasonic aspiration. In all, the waterjet device is particularly
suitable for precise glioma-parenchyma dissection and might support complete tumor
resection. For glioma aspiration, the ultrasonic aspirator appears to be more efficient.
However, when only minimal blood loss is intended, the waterjet device might be favorable
even for glioma aspiration. Before the new technique can be considered as an addition
to the neurosurgical armamentarium, further evidence is required that the waterjet
enables less traumatic surgery than conventional instruments.
Key words
Waterjet dissection - glioma - surgical technique - vessel preservation - brain
References
- 1 Summers D A. Waterjetting technology. London: E & FN Spon 1995
- 2
Papachristou D N, Barters R.
Resection of the liver with a water jet.
Br J Surg.
1982;
69
93-94
- 3
Baer H U, Metzger A, Barras J P, Mettler D, Wheatley A M, Czerniak A.
Laparoscopic liver resection in the large white pig - a comparison between waterjet
dissector and ultrasound dissector.
End Surg Allied Technol.
1994;
2
189-193
- 4
Rau H G, Schardey H M, Buttler E, Reuter C, Cohnert T U, Schildberg F W.
A comparison of different techniques for liver resection: blunt dissection, ultrasonic
aspirator and jet-cutter.
Eur J Surg Oncol.
1995;
21
183-187
- 5
Hata Y, Sasaki F, Takahashi H, Ohkawa Y, Taguchi K, Une Y, Uchino J.
Liver resection in children, using a water jet.
J Pediatr Surg.
1994;
29
648-650
- 6
Izumi R, Yabushita K, Shimizu K, Yagi M, Yamaguchi A, Konishi K, Nagakawa T, Miyazaki I.
Hepatic resection using a water jet dissector.
Surg Today.
1993;
23
31-35
- 7
Rau H G, Meyer G, Cohnert T U, Schardey H M, Jauch K, Schildberg F W.
Laparoscopic liver resection with the water-jet dissector.
Surg Endosc.
1995;
9
1009-1012
- 8
Une Y, Uchino J, Shimamura T, Kamiyama T, Saiki I.
Water jet scalpel for liver resection in hepatocellular carcinoma with or without
cirrhosis.
Int Surg.
1996;
81
45-48
- 9
Penchev R D, Losanoff J E, Kjossev K T.
Reconstructive renal surgery using a water jet.
J Urol.
1999;
162
772-774
- 10
Giraud J Y, Villemin S, Darmana R, Cahuzac J P, Autefage A, Morucci J P.
Bone cutting.
Clin Phys Physiol Meas.
1991;
12
1-19
- 11
Kaduk W M, Stengel B, Pohl A, Nizze H, Gundlach K K.
Hydro-jet cutting: a method for selective surgical dissection of nerve tissue.
J Craniomaxillofac Surg.
1999;
27
327-330
- 12
Terzis A J, Nowak G, Rentzsch O, Arnold H, Diebold J, Baretton G.
A new system for cutting brain tissue preserving vessels: water jet cutting.
Br J Neurosurg.
1989;
3
361-366
- 13
Piek J, Wille C, Warzok R, Gaab M R.
Waterjet-dissection of the brain - experimental and first clinical results.
J Neurosurg.
1998;
89
861-864
- 14
Oertel J, Gaab M R, Knapp A, Essig H, Warzok R, Piek J.
Waterjet dissection in neurosurgery: experimental results in the porcine cadaver brain.
Neurosurgery.
2003;
52
153-159
- 15
Piek J, Oertel J, Gaab M R.
Waterjet dissection in neurosurgical procedures - clinical results in 35 patients.
J Neurosurg.
2002;
96
690-696
- 16
Kleihues P, Berger P C, Scheithauer B W.
The new WHO classification of brain tumours.
Brain Pathol.
1993;
3
255-268
- 17
Wagner W, Gaab M R, Schroeder H WS, Tschiltschke W.
Cranial neuronavigation in neurosurgery: assessment of usefulness in relation to type
and site of pathology in 284 patients.
Minim Invas Neurosurg.
2000;
43
124-131
- 18
Oertel J, Kroutilina E, Gaab M R.
Prognosis of glioblastoma in the 70s and today.
Neurosurgery (Abstract).
2001;
49
527
- 19
Jakob S, Kehler U, Retlsche E, Friedrich H J, Arnold H.
Endoscopic use of the water jet dissector in the cerebral ventricle system - an experimental
study.
Zentralbl Neurochir.
2000;
61
14-21
- 20
Mukai H, Yamashita J, Kitamura A, Ito H.
Stereotactic aqua-stream and aspirator in the treatment of intracerebral hematoma:
an experimental study.
Stereotact Funct Neurosurg.
1991;
57
221-227
- 21
Oertel J, Gaab M R, Runge U, Piek J.
Epilepsiechirurgie mit dem Wasserstrahlskalpell.
Z Epileptol (Abstract).
2002;
15
54
Joachim Oertel, M. D.
Department of Neurosurgery · Hannover Nordstadt Hospital
Haltenhoffstrasse 41
30167 Hannover
Germany
Telefon: +49-511-970-1245/1606
eMail: oertelj@freenet.de