Thorac Cardiovasc Surg 1981; 29(2): 108-109
DOI: 10.1055/s-2007-1023454
© Georg Thieme Verlag Stuttgart · New York

Strut Fracture in the New Bjørk Shiley Mitral Valve Prosthesis

O. Brubakk, S. Simonsen, L. Källman, A. Fredriksen
  • Medical Department Sarpsborg Hospital, Sarpsborg, Cardiological Laboratory, Rikshospitalet, Oslo and Central Institute for Industrial Research, Oslo, Norway
Further Information

Publication History

1980

Publication Date:
19 March 2008 (online)

Summary

The case of a patient with the new type Bjørk-Shiley aortic and mitral valve prosthesis is described.

Three months after implant she suffered acute heart failure and died. Post-mortem examination revealed a fractured outlet strut in the mitral valve prosthesis with dislocation of the disc. The fracture was regarded as due to excessive brittleness caused by demonstrated deposition of chromium-tungsten-carbide.

    >