Rofo 2007; 179(9): 896-900
DOI: 10.1055/s-2007-963389
Rapid Communication

© Georg Thieme Verlag KG Stuttgart · New York

Influence of Grid Control and Object Detection on Radiation Exposure and Image Quality using Mobile C-Arms - First Results

Der Einfluss von Gridregelung und Objekterkennung auf Strahlenexposition und Bildqualität bei transportablen C-Bogengeräten - erste ErgebnisseD. Gosch1 , A. Ratzmer2 , P. Berauer2 , T. Kahn1
  • 1Klinik und Poliklinik für Diagnostische und Interventionelle Radiologie, Universitätsklinikum Leipzig AöR
  • 2Ziehm Imaging GmbH, Nürnberg
Further Information

Publication History

received: 10.5.2007

accepted: 1.7.2007

Publication Date:
17 August 2007 (online)

Zusammenfassung

Ziel: Ziel der Studie war, zu untersuchen, inwieweit eine neuartige Dosisleistungsregelung (Gridregelung) die Bildqualität von mobilen C-Bogen verbessern kann. Außerdem sollte die mögliche Dosisreduktion durch automatische Regelung der Pulsfrequenz bei gepulster Durchleuchtung mit 25 Pulsen pro Sekunde durch Objekterkennung ermittelt werden. Material und Methoden: Bei der Gridregelung wird im Gegensatz zu den üblichen Dosisleistungsregelungen, bei denen ein zentrales kreisförmiges Messfeld verwendet wird, ein enges Gitter quadratischer Zellen über das gesamte Röntgenbild gelegt. Das Belichtungssystem nutzt nur die Zellen für die Belichtungssteuerung, die vom abzubildenden Objekt abgedeckt werden. Damit soll unabhängig von Größe, Form und Lage des abzubildenden Objektes eine optimale Belichtung erreicht werden. Das System erkennt auch automatisch Objektbewegungen. Wenn eine Pulsfrequenz von 25 Pulsen pro Sekunde angewählt ist und keine Veränderungen im Bild registriert werden, wird die Pulsfrequenz der gepulsten Durchleuchtung schrittweise reduziert. Damit kann die Strahlenexposition verringert werden. Der Einfluss der Gridregelung auf die Bildqualität wurde mithilfe eines anthropomorphen Phantoms untersucht. Die Dosisreduktion durch Objekterkennung wurde durch Auswerten von Untersuchungsdaten von 146 Patienten aus 5 Ländern ermittelt. Ergebnisse: Die Bilder des statischen Phantoms waren mit Gridregelung unabhängig von der Position des darzustellenden Objektes immer optimal belichtet. Die durchschnittliche Dosisreduktion bei 25 Pulsen pro Sekunde durch Objekterkennung und automatische Pulsregelung betrug 21 %, die maximale Dosiseinsparung lag bei 60 %. Schlussfolgerung: Durch die Gridregelung wird die Bedienung der C-Bogengeräte vereinfacht, da unabhängig von der Objektpositionierung eine optimale Belichtung erreicht wird. Die Objekterkennung kann zu einer Reduktion der Strahlenexposition von Patient und Untersuchungspersonal führen.

Abstract

Purpose: The objective of this study was to examine the extent to which the image quality on mobile C-arms can be improved by an innovative exposure rate control system (grid control). In addition, the possible dose reduction in the pulsed fluoroscopy mode using 25 pulses/sec produced by automatic adjustment of the pulse rate through motion detection was to be determined. Materials and Methods: As opposed to conventional exposure rate control systems, which use a measuring circle in the center of the field of view, grid control is based on a fine mesh of square cells which are overlaid on the entire fluoroscopic image. The system uses only those cells for exposure control that are covered by the object to be visualized. This is intended to ensure optimally exposed images, regardless of the size, shape and position of the object to be visualized. The system also automatically detects any motion of the object. If a pulse rate of 25 pulses/sec is selected and no changes in the image are observed, the pulse rate used for pulsed fluoroscopy is gradually reduced. This may decrease the radiation exposure. The influence of grid control on image quality was examined using an anthropomorphic phantom. The dose reduction achieved with the help of object detection was determined by evaluating the examination data of 146 patients from 5 different countries. Results: The image of the static phantom made with grid control was always optimally exposed, regardless of the position of the object to be visualized. The average dose reduction when using 25 pulses/sec resulting from object detection and automatic down-pulsing was 21 %, and the maximum dose reduction was 60 %. Conclusion: Grid control facilitates C-arm operation, since optimum image exposure can be obtained independently of object positioning. Object detection may lead to a reduction in radiation exposure for the patient and operating staff.

References

  • 1 Wieners G, Pech M, Beck A. et al . Vergleich von Strahlenexposition und Bildqualität eines Siremobil-IsoC3D mit einem 16-Zeilen-Spiral-CT bei Diagnostik und Intervention am humanen Becken.  Fortschr Röntgenstr. 2005;  177 258-264
  • 2 Gebhard F, Kraus M, Schneider E. et al . Strahlendosis im OP - ein Vergleich computerassistierter Verfahren.  Unfallchirurg. 2003;  106 492-497
  • 3 Jones D PG, Robertson P A, Lunt B. et al . Radiation exposure during fluoroscopically assisted pedicle screw insertion in the lumbar spine.  Spine. 2000;  25 1538-1541
  • 4 Rampersaud Y R, Foley K T, Shen A C. et al . Radiation exposure to the spine surgeon during fluoroscopically assisted pedicle screw insertion.  Spine. 2000;  25 2637-2645
  • 5 Theocharopoulos N, Perisinakis K, Damilakis J. et al . Occupational exposure from common fluoroscopic projections used in orthopaedic surgery.  J Bone Joint Surg Am. 2003;  85 1698-1703
  • 6 Lewall D B, Riley P, Hassoon A A. et al . A fluoroscopy credentialling programme for orthopaedic surgeons.  J Bone Joint Surg. 1995;  77 442-444
  • 7 Fuchs M, Modler H, Schmidt A. et al . Messung der intraoperativen Strahlenexposition des Unfallchirurgen.  Unfallchirurg. 1999;  102 371-376
  • 8 Röntgenverordnung. Köln Berlin Bonn München; Carl Heymanns Verlag KG 2002
  • 9 Herrmann K, Helmberger T, Waggershausen T. et al . Erste Erfahrungen mit gepulster Durchleuchtung an einer multifunktionalen Durchleuchtungsanlage.  Fortschr Röntgenstr. 1996;  165 475-479
  • 10 Alonso J A, Shaw D L, Maxwell A. et al . Scattered radiation during fixation of the hip fractures.  J Bone Joint Surg. 2001;  83 815-818
  • 11 Boetticher von H, Lachmund J, Hoffmann W. et al . Optimierung des Strahlenschutzes für das Personal in der Radiologie auf der Grundlage der effektiven Dosis.  Fortschr Röntgenstr. 2006;  178 287-291
  • 12 Eder H, Panzer W, Schöfer H. Ist der Bleigleichwert zur Beurteilung der Schutzwirkung bleifreier Röntgenschutzkleidung geeignet?.  Fortschr Röntgenstr. 2005;  177 399-404
  • 13 Johlin F C, Pelsang R E, Greenleaf M. Phantom study to determine radiation exposure to medical personnel involved in ERCP fluoroscopy and its reduction through equipment and behavior modifications.  AJG. 2002;  97 893-897
  • 14 Pecher G, Koenig H, Pecher S. et al . Reduktion der Strahlenexposition in der interventionellen Angiografie für Patient und Untersucher.  Fortschr Röntgenstr. 1998;  169 505-509

Dr. Dieter Gosch

Klinik und Poliklinik für Diagnostische und Interventionelle Radiologie, Universitätsklinikum Leipzig AöR

Liebigstraße 20

04103 Leipzig

Phone: ++ 49/3 41/9 71 74 16

Fax: ++ 49/3 41/9 71 74 09

Email: gosd@medizin.uni-leipzig.de

    >