Thromb Haemost 2009; 101(03): 563-569
DOI: 10.1160/TH08-09-0601
New Technologies, Diagnostic Tools and Drugs
Schattauer GmbH

International normalised ratio (INR) measured on the CoaguChek S and XS compared with the laboratory for determination of precision and accuracy

Thomas D. Christensen
1   Department of Cardiothoracic and Vascular Surgery & Institute of Clinical Medicine, Aarhus University Hospital, Skejby, Aarhus N, Denmark
,
Torben B. Larsen
2   Department of Biochemistry, Pharmacology and Genetics, Section for Thrombosis and Haemostasis, Odense University Hospital, Sdr. Boulevard 29, Odense, Denmark
,
Claus Jensen
3   UNI•C, The Danish IT Centre for Education and Research, Aarhus N, Denmark
,
Marianne Maegaard
1   Department of Cardiothoracic and Vascular Surgery & Institute of Clinical Medicine, Aarhus University Hospital, Skejby, Aarhus N, Denmark
,
Benny Sørensen
4   Centre for Hemophilia and Thrombosis, Department of Clinical Biochemistry & Institute of Clinical Medicine, Aarhus University Hospital, Skejby, Aarhus N, Denmark
5   Haemostasis Research Unit, Centre for Haemophilia and Thrombosis, Guy’s and St Thomas NHS Trust Foundation & King’s College London School of Medicine, London, United Kingdom
› Author Affiliations
Financial support: Søster og Verner Lipperts Fond, Snedkermester Sophus Jacobsen og hustru Astrid Jacobsens Fond, Jens Anker Andersen Fonden, Helga og Peter Kornings Fond, Arvid Nilssons Fond. The work was independent of the funders.
Further Information

Publication History

Received: 16 September 2008

Accepted after minor revision: 12 February 2008

Publication Date:
24 November 2017 (online)

Summary

Oral anticoagulation therapy is monitored by the use of international normalised ratio (INR). Patients performing self-management estimate INR using a coagulometer, but studies have been partly flawed regarding the estimated precision and accuracy. The objective was to estimate the imprecision and accuracy for two different coagulometers (CoaguChek® S and XS). Twenty-four patients treated with coumarin were prospectively followed for six weeks. INR’s were analyzed weekly in duplicates on both coagulometers, and compared with results from the hospital laboratory. Statistical analysis included Bland-Altman plot, 95% limits of agreement, coefficient of variance (CV), and an analysis of variance using a mixed effect model. Comparing 141 duplicate measurements (a total of 564 measurements) of INR, we found that the CoaguChek S and CoaguChek XS had a precision (CV) of 3.4% and 2.3%, respectively. Regarding analytical accuracy, the INR measurements tended to be lower on the coagulometers, and regarding diagnostic accuracy the CoaguChek S and CoaguChek XS deviated more than 15% from the laboratory measurements in 40% and 43% of the measurements, respectively. In conclusion, the precision of the coagulometers was found to be good, but only the CoaguChek XS had a precision within the predefined limit of 3%. Regarding analytical accuracy, the INR measurements tended to be lower on the coagulometers, compared to the laboratory. A large proportion of measurement of the coagulometers deviated more than 15% from the laboratory measurements. Whether this will have a clinical impact awaits further studies.

 
  • References

  • 1 Hirsh J, Dalen J, Anderson DR. et al. Oral anticoagulants: mechanism of action, clinical effectiveness, and optimal therapeutic range. Chest 2001; 119: 8S-21S.
  • 2 Heneghan C, Alonso-Coello P, Garcia-Alamino JM. et al. Self-monitoring of oral anticoagulation: a systematic review and meta-analysis. Lancet 2006; 367: 404-411.
  • 3 Christensen TD, Johnsen SP, Hjortdal VE. et al. Self-management of oral anticoagulant therapy: a systematic review and meta-analysis. Int J Cardiol 2007; 118: 54-61.
  • 4 Van den Besselaar AM. Accuracy, precision, and quality control for point-of-care testing of oral anticoagulation. J Thromb Thrombolysis 2001; 12: 35-40.
  • 5 Tripodi A, Chantarangkul V, Mannucci P. Near-patient testing devices to monitor oral anticoagulant therapy. Br J Haematol 2001; 113: 847-852.
  • 6 Fairweather RB, Ansell J, Van den Besselaar AM. et al. College of American Pathologists Conference XXXI on laboratory monitoring of anticoagulant therapy: laboratory monitoring of oral anticoagulant therapy. Arch Pathol Lab Med 1998; 122: 768-781.
  • 7 Attermann J, Kynde K, Hasenkam JM. Precision of patients’ measurements of the international normalized ratio (INR) using a patient operated whole blood home coagulometer. Thromb Res 1998; 92: 287-291.
  • 8 Hillarp A, Egberg N, Nordin G. et al. Local INR calibration of the Owren type prothrombin assay greatly improves the intra- and interlaboratory variation. A three-year follow-up from the Swedish national external quality assessment scheme. Thromb Haemost 2004; 91: 300-307.
  • 9 Bruns DE, Huth EJ, Magid E. et al. Toward a checklist for reporting of studies of diagnostic accuracy of medical tests. Clin Chem 2000; 46: 893-895.
  • 10 van den Besselar AM. A comparison of INRs determined with a whole blood prothrombin time device and two international reference preparations for thromboplastin. Thromb Haemost 2000; 84: 410-412.
  • 11 Kjeldsen J, Lassen JF, Petersen PH. et al. Biological variation of International Normalized Ratio for prothrombin times, and consequences in monitoring oral anticoagulant therapy: computer simulation of serial measurements with goal-setting for analytical quality. Clin Chem 1997; 43: 2175-2182.
  • 12 Anderson DR, Harrison L, Hirsh J. Evaluation of a portable prothrombin time monitor for home use by patients who require long-term oral anticoagulant therapy. Arch Intern Med 1993; 153: 1441-1447.
  • 13 Douketis JD, Lane A, Milne J. et al. Accuracy of a portable International Normalization Ratio monitor in outpatients receiving long-term oral anticoagulant therapy: comparison with a laboratory reference standard using clinically relevant criteria for agreement. Thromb Res 1998; 92: 11-17.
  • 14 Poller L, Keown M, Ibrahim SA. et al. Quality assessment of CoaguChek point-of-care prothrombin time monitors: comparison of the European community-approved procedure and conventional external quality assessment. Clin Chem 2006; 52: 1843-1847.
  • 15 Christensen TD. Self-management of oral anticoagulant therapy: a review. J Thromb Thrombolysis 2004; 18: 127-143.
  • 16 Bland JM, Altman DG. Measuring agreement in method comparison studies. Stat Methods Med Res 1999; 8: 135-160.
  • 17 Cromheecke ME, Levi M, Colly LP. et al. Oral anticoagulation self-management and management by a specialist anticoagulation clinic: a randomised crossover comparison. Lancet 2000; 356: 97-102.
  • 18 Hasenkam JM, Kimose HH, Knudsen L. et al. Self management of oral anticoagulant therapy after heart valve replacement. Eur J Cardiothorac Surg 1997; 11: 935-942.
  • 19 Körtke H, Körfer R. International normalized ratio self-management after mechanical heart valve replacement: is an early start advantageous?. Ann Thorac Surg 2001; 72: 44-48.
  • 20 Sawicki PT. A structured teaching and self-management program for patients receiving oral anticoagulation: a randomized controlled trial. Working Group for the Study of Patient Self-Management of Oral Anticoagulation. J Am Med Assoc 1999; 281: 145-150.
  • 21 Watzke HH, Forberg E, Svolba G. et al. A prospective controlled trial comparing weekly self-testing and self-dosing with the standard management of patients on stable oral anticoagulation. Thromb Haemost 2000; 83: 661-665.
  • 22 Braun S, Watzke H, Hasenkam JM. et al. Performance evaluation of the new CoaguChek XS system compared with the established CoaguChek system by patients experienced in INR-self management. Thromb Haemost 2007; 97: 310-314.
  • 23 Guide-lines for near patient testing: haematology.. Near Patient Testing Working Party. General Haematology Task Force of BCSH. Thrombosis and Haemostasis Task Force of BCSH. Clin Lab Haematol 1995; 17: 301-310.
  • 24 Williams VK, Griffiths AB. Acceptability of CoaguChek S and CoaguChek XS generated international normalised ratios against a laboratory standard in a paediatric setting. Pathology 2007; 39: 575-579.
  • 25 Nam MH, Roh KH, Pak HN. et al. Evaluation of the Roche CoaguChek XS handheld coagulation analyzer in a cardiac outpatient clinic. Ann Clin Lab Sci 2008; 38: 37-40.
  • 26 Bauman ME, Black KL, Massicotte MP. et al. Accuracy of the CoaguChek XS for point-of-care international normalized ratio (INR) measurement in children requiring warfarin. Thromb Haemost 2008; 99: 1097-1103.
  • 27 Karon BS, McBane RD, Chaudhry R. et al. Accuracy of capillary whole blood international normalized ratio on the CoaguChek S, CoaguChek XS, and i-STAT 1 point-of-care analyzers. Am J Clin Pathol 2008; 130: 88-92.
  • 28 Bereznicki LR, Jackson SL, Peterson GM. et al. Accuracy and clinical utility of the CoaguChek XS portable international normalised ratio monitor in a pilot study of warfarin home-monitoring. J Clin Pathol 2007; 60: 311-314.
  • 29 Van den Besselaar AM, Barrowcliffe TW, Houbouyan-Reveillard LL. et al. Guidelines on preparation, certification, and use of certified plasmas for ISI calibration and INR determination. J Thromb Haemost 2004; 2: 1946-1953.
  • 30 Cannegieter SC, Rosendaal FR, Wintzen AR. et al. Optimal oral anticoagulant therapy in patients with mechanical heart valves. N Engl J Med 1995; 333: 11-17.