CC BY-NC-ND 4.0 · Asian J Neurosurg 2019; 14(03): 1008-1010
DOI: 10.4103/ajns.AJNS_296_18
Case Report

Precautions for using a filter protection device with a flow reversal method during carotid artery stenting: A case report on in-filter thrombus

Tomotaka Ohshima
Neuroendovascular Therapy Center, Aichi Medical University, Nagakute, Aichi
,
Shigeru Miyachi
Neuroendovascular Therapy Center, Aichi Medical University, Nagakute, Aichi
,
Naoki Matsuo
1   Department of Neurosurgery, Aichi Medical University, Yazakokarimata, Nagakute, Aichi
,
Reo Kawaguchi
1   Department of Neurosurgery, Aichi Medical University, Yazakokarimata, Nagakute, Aichi
,
Masakazu Takayasu
1   Department of Neurosurgery, Aichi Medical University, Yazakokarimata, Nagakute, Aichi
› Author Affiliations

The combination of a flow reversal with two balloons and a filter protection device is one of the safest methods that can be used during carotid artery stenting (CAS). Although major adverse events did not occur under restrict protection, we experienced a case of in-filter thrombus during the procedure. A 68-year-old male presented with temporary right hemiparesis. The radiological examinations revealed 95% stenosis on the origin of the left internal carotid artery (ICA). The patient underwent left CAS of wherein a proximal common carotid artery balloon, an external carotid artery balloon, and a distal ICA filter, with continuous flow reversal to the femoral vein, were used. Although the reversal circulation was established, a massive newly developing thrombus was found in the proximal side of the filter. After an additional systemic heparinization, the thrombus disappeared. The procedure was performed without any complications. Physicians should be aware of the risk of developing intraprocedural thrombosis in a filter protection device. Because the filter protection device is designed for the antegrade flow, it may promote the development of thrombus against the retrograde flow. Thus, the filter protection device should be retrieved first under the flow reversal circumstance to avoid the distal migration of a clot around the filter device.

Financial support and sponsorship

Nil.




Publication History

Article published online:
09 September 2022

© 2019. Asian Congress of Neurological Surgeons. This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes, or adapted, remixed, transformed or built upon. (https://creativecommons.org/licenses/by-nc-nd/4.0/)

Thieme Medical and Scientific Publishers Pvt. Ltd.
A-12, 2nd Floor, Sector 2, Noida-201301 UP, India

 
  • References

  • 1 Macdonald S. New embolic protection devices: A review. J Cardiovasc Surg (Torino) 2011;52:821-7.
  • 2 Harada K, Kakumoto K, Morioka J, Saito T, Fukuyama K. Combination of flow reversal and distal filter for cerebral protection during carotid artery stenting. Ann Vasc Surg 2014;28:651-8.
  • 3 Ohshima T, Bishnori I, Ishikawa K, Goto S, Yamamoto T, Kato Y, et al. The feasibility and safety of separate carotid artery stenting using the restrict protective method for bilateral carotid stenosis. World Neurosurg 2017;102:235-9.
  • 4 Goto S, Ohshima T, Kato K, Izumi T, Wakabayashi T. Efficacy of carotid artery stenting by the universal protection method. J Stroke Cerebrovasc Dis 2018;27:2187-91.
  • 5 Phatouros CC, Higashida RT, Malek AM, Meyers PM, Lempert TE, Dowd CF, et al. Carotid artery stent placement for atherosclerotic disease: Rationale, technique, and current status. Radiology 2000;217:26-41.