Ultraschall Med 2022; 43(02): 204-208
DOI: 10.1055/a-1168-6602
Quality Assurance

Endocavity Ultrasound Transducers: Why High-Level Disinfection Is Necessary

Endokavitäre Ultraschallsonden: Warum ein High-level-Desinfektionsverfahren notwendig ist
Susan Campbell Westerway
1   Ultrasound, Charles Sturt University NSW Australia, NSW, Australia
,
Jocelyne M. Basseal
2   Discipline of Infectious Diseases & Immunology, The University of Sydney Faculty of Medicine and Health, Sydney, Australia
› Author Affiliations

Abstract

As a medical imaging modality, ultrasound is used by a wide cross-section of practitioners including radiologists, obstetricians, gynecologists, gastroenterologists, urologists and cardiologists. The increasing popularity of ultrasound as a diagnostic tool is due not only to the ease of use and portability of systems, but also to the perceived safety aspect of the examination. This latter point needs to be examined. As with any reusable medical device, the ultrasound transducer, also known as a probe, could potentially be a vector for the transmission of pathogenic viruses and fungi between patients if not correctly disinfected after each use. This transmission risk is magnified for an endocavity transducer that has come in contact with the vagina, anal canal or oral cavity, as it could be contaminated with organisms transmitted by blood or mucosal, genital or rectal secretions. Based on the Spaulding system, transducers that come in contact with mucous membranes are classified as semi-critical devices that require high-level disinfection (HLD) after each patient procedure. This HLD process should eliminate all microorganisms except high numbers of bacterial endospores. Only a small number of countries worldwide have implemented transducer reprocessing guidelines that adhere to the Spaulding classification and recommend HLD for endocavity transducers. Overall, there is a lack of conformity among global health agencies regarding the use of HLD for endocavity transducers. This is primarily due to the perception that the infection transmission risk is negligible and that if an endocavity transducer has been covered with a single-use sheath for the procedure, then low-level disinfection provides sufficient protection against pathogen transmission. The objective of this study was to review the published risk of infection transmission from endocavity transducers. By highlighting the outbreaks and case reports that implicate pathogen transmission from transducers, we posit that HLD should be a global standard of practice for the reprocessing of endocavity transducers. It requires substantial time for national health administrations to develop and legislate new recommendations, and for practice changes to be accepted and implemented by healthcare providers. We recommend that Joint Commission International (JCI) and other equivalent organizations enforce the use of HLD of endocavity ultrasound transducers during their accreditation reviews.

Zusammenfassung

Als bildgebendes Verfahren in der Medizin wird Ultraschall von Ärzten verschiedenster Fachrichtungen eingesetzt, darunter Radiologen, Geburtshelfer, Gynäkologen, Gastroenterologen, Urologen und Kardiologen. Die zunehmende Beliebtheit von Ultraschall als diagnostisches Hilfsmittel ist nicht nur auf die einfache Handhabung und Tragbarkeit der Systeme zurückzuführen, sondern auch auf die vermeintlichen Sicherheitsaspekte der Untersuchung. Dieser letztgenannte Punkt muss untersucht werden. Wie bei jedem wiederverwendbaren medizinischen Gerät könnte der Ultraschallkopf, der auch als Sonde bezeichnet wird, potenziell ein Vektor für die Übertragung pathogener Viren und Pilze zwischen den Patienten sein, wenn er nicht nach jedem Gebrauch korrekt desinfiziert wird. Dieses Übertragungsrisiko ist bei einer endokavitären Sonde, die mit der Vagina, dem Analkanal oder der Mundhöhle in Kontakt gekommen ist, erhöht, da diese mit Organismen kontaminiert sein könnte, die durch Blut oder durch Schleimhaut-, Genital- oder Rektalsekrete übertragen werden. Auf der Grundlage des Spaulding-Klassifikationssystems werden Schallköpfe, die mit Schleimhäuten in Kontakt kommen, als semikritische Geräte eingestuft, die nach jedem Patienteneingriff ein High-level-Desinfektionsverfahren (HLD) erfordern. Dieses HLD-Verfahren sollte alle Mikroorganismen mit Ausnahme vieler bakterieller Endosporen eliminieren. Nur wenige Länder weltweit haben Richtlinien für die Aufbereitung von Schallköpfen eingeführt, die sich an die Spaulding-Klassifikation halten und HLD für endokavitäre Sonden empfehlen. Insgesamt besteht unter den globalen Gesundheitsbehörden ein Mangel an Konformität hinsichtlich des Einsatzes der HLD für endokavitäre Sonden. Dies ist in erster Linie auf die Auffassung zurückzuführen, dass das Risiko der Infektionsübertragung vernachlässigbar ist und dass, wenn eine endokavitäre Sonde für den Eingriff mit einer Einweghülle abgedeckt wurde, eine Desinfektion auf niedrigem Niveau einen ausreichenden Schutz gegen die Übertragung von Krankheitserregern bietet. Das Ziel dieser Studie war es, Publikationen zum Risiko der Infektionsübertragung durch endokavitäre Sonden zu überprüfen. Indem wir die Ausbrüche und Fallberichte hervorheben, die eine Erregerübertragung durch Schallköpfe nahelegen, stellen wir fest, dass HLD weltweit Praxisstandard für die Aufbereitung von endokavitären Sonden werden sollte. Die Entwicklung und Verabschiedung neuer Empfehlungen durch die nationalen Gesundheitsbehörden sowie die Akzeptanz und Umsetzung von praktischen Änderungen bei Gesundheitsdienstleistern erfordern einen erheblichen Zeitaufwand. Wir empfehlen, dass die Joint Commission International (JCI) und andere gleichwertige Organisationen während ihrer Akkreditierungsprüfungen den Einsatz der HLD bei endokavitären Ultraschallsonden einfordern.



Publication History

Received: 12 September 2019

Accepted: 14 April 2020

Article published online:
30 July 2020

© 2020. Thieme. All rights reserved.

Georg Thieme Verlag KG
Rüdigerstraße 14, 70469 Stuttgart, Germany

 
  • References

  • 1 Basseal JM, Westerway SC, Juraja M. et al. Guidelines for reprocessing ultrasound transducers. Australas J Ultrasound Med 2017; 20: 30-40
  • 2 AIUM official statements Guidelines for Cleaning and Preparing External- and Internal-Use Ultrasound Probes Between Patients, Safe Handling, and Use of Ultrasound Coupling Gel. American Institute of Ultrasound in Medicine; 2017 accessed 13/12/2017
  • 3 NHS Scotland. Guidance for Decontamination of Semi-Critical Ultrasound Probes; Semi-invasive and Non-invasive Ultrasound Probes. Health Facilities Scotland Decontamination Services. HPS/HFS: Version 1.1. January 2017. Available at (accessed July 2019): https://www.hps.scot.nhs.uk/web-resources-container/guidance-for-decontamination-of-semi-critical-ultrasound-probes/
  • 4 Irish Health Service Executive (HSE) Quality Improvement Division. Decontamination Safety Programme (2017) HSE guidance for decontamination of semi‐critical ultrasound probes; Semi-invasive and Noninvasive Ultrasound Probes QPSD-GL-028-1. https://hse.ie/eng/about/Who/QID/nationalsafetyprogrammes/decontamination/Ultrasound-Probe-Decontamination-Guidance-Feb-17.pdf
  • 5 NHS Wales Shared Services Partnership – Specialist Estates Services. Welsh Health Technical Memorandum 01–06. Decontamination of flexible endoscopes, part C: Operational management 2016. Available at (accessed July 2019): http://www.wales.nhs.uk/sites3/Documents/254/WHTM%2001%2D06%20Part%20C.pdf
  • 6 Public Health Ontario. Best Practices for Cleaning, Disinfection and Sterilization of Medical Equipment/Devices In All Health Care Settings. 3rd edition. May 2013
  • 7 Israel Ministry of Health guidelines. https://www.health.gov.il/hozer/mr03_2017.pdf (accessed July 2019)
  • 8 Spaulding EH. Chemical disinfection and antisepsis in the hospital. J Hosp Res 1957; 9: 5-31
  • 9 Rutala WA, Weber DJ. Healthcare Infection Control Practices Advisory Committee (HICPAC), Centre for Disease Control and Prevention (CDC). Guideline for Disinfection and Sterilization in Healthcare Facilities. USA: Centers for Disease Control; 2008
  • 10 Weber D, Rutala W. Assessing the risk of disease transmission to patients when there is a failure to follow recommended disinfection and sterilization guidelines. American J Infect Control 2013; 41 (Suppl. 05) S67-S71
  • 11 Westerway SC, Basseal JM, Abramowicz J. Medical ultrasound disinfection and hygiene practices: WFUMB global survey results. Ultrasound Med Biol 2019; 45: 344-352 DOI: 10.1016/j.ultrsmedbio.2018.09.019.
  • 12 Health Facilities Scotland. Report on the 2012 National Survey of the Decontamination of Transoesophageal Echocardiography, Transvaginal and Transrectal Ultrasound Probes. 2015
  • 13 Gray R, Williams P, Dubbins PA. et al. De-contamination of transvaginal ultrasound probes: review of national practice and need for national guidelines. Clin Radiol 2012; 67: 1069-1077
  • 14 Nyhsen C, Humphreys H, Nicolau C. et al. Infection prevention and ultrasound probe decontamination practices in Europe: a survey of the European Society of Radiology. Insights Imaging 2016; 7: 841-847
  • 15 Müller T, Martiny H, Merz E. et al. DEGUM Recommendations on Infection Prevention in Ultrasound and Endoscopic Ultrasound. Ultraschall in Med 2018; 39: 284-303
  • 16 Carrico R, Furmanek S, English C. Ultrasound probe use and reprocessing: Results from a national survey among U.S. infection preventionists. Am J Infect Control 2018; 46: 913-920
  • 17 Health Alert Network 2015. CDC/FDA Health Update about the Immediate Need for Healthcare Facilities to Review Procedures for Cleaning, Disinfecting, and Sterilizing Reusable Medical Devices. https://emergency.cdc.gov/han/han00383.asp (accessed May 2019)
  • 18 Improperly sterilized or HLD equipment – a growing problem. The Join Commission. Quick Safety: Issue 33 May 2017. https://www.jointcommission.org/assets/1/23/qs_33a_2017.pdf (accessed May 2019)
  • 19 David CB, Weiner Z, Solt I. The possibility of transmitting infections with vaginal ultrasound probes: why we cannot meet the guidelines. Israel Med A J 2019; 21: 45-49
  • 20 Leroy S. Infectious risk of endovaginal and transrectal ultrasonography: systematic review and meta-analysis. J Hosp Infect 2013; 83: 99-106
  • 21 Leroy S, M'Zali F, Kann M. et al. Impact of vaginal-rectal ultrasound examinations with covered and low-level disinfected transducers on infectious transmissions in France. Infect Control Hosp Epidemiol 2014; 35: 1497-1504
  • 22 NHS Scotland. risk based recommendations for the decontamination of semi-invasive ultrasound probes: Risk of infection following semi-invasive ultrasound procedures in Scotland, 2010 to 2016. Available at (accessed May 2019): http://www.hps.scot.nhs.uk/resourcedocument.aspx?id=6237
  • 23 Kuwata T, Takahashi H, Koibuchi H. et al. Incidence of human papillomavirus contamination of transvaginal probes in Japan and possible contamination prevention strategy. J Med Ultrasonics 2016; 43: 505-508
  • 24 Kac G, Podglajen I, Si-Mohamed A. et al. Evaluation of raviolet C for disinfection of endocavitary ultrasound transducers persistently contaminated despite probe covers. Infect Control Hosp Epidemiol 2010; 31: 165-170
  • 25 Lessa F, Tak S, de Vader R. Risk of infections associated with improperly reprocessed transrectal ultrasound-guided prostate biopsy equipment. Infection Control and Hospital Epidemiology 2008; 29: 289-293
  • 26 Rooks VJ, Yancey MK, Elg SA. et al. Comparison of probe sheaths for endovaginal sonography. Obstet Gynecol 1996; 87: 27-32
  • 27 Storment JM, Monga M, Blanco JD. Ineffectiveness of latex condoms in preventing contamination of the transvaginal ultrasound transducer head. South Med J 1997; 90: 206-208
  • 28 Milki A, Fisch J. Vaginal ultrasound probe cover leakage: implications for patient care. Fertil Steril 1998; 69: 409-411
  • 29 Jimenez R, Duff P. Sheathing of the endovaginal ultrasound probe: is it adequate?. Infect Dis Obstet Gynecol 1993; 1: 37-39
  • 30 Amis S, Ruddy M, Kibbler C. Assessment of condoms as probe covers for transvaginal sonography. J Clin Ultrasound 2000; 28: 295-298
  • 31 Basseal JM, Westerway SC, Hyett J. Analysis of the integrity of ultrasound probe covers used for transvaginal examinations. Infection Disease & Health (in print) https://doi.org10.1016/j.idh.2019.11.003
  • 32 Ferhi K, Roupret M, Mozer P. et al Hepatitis C transmission after prostate biopsy. Urology Case Reports 2013; 797248
  • 33 Casalegno JS, Le Bail CarvalK, Eibach D. et al. High risk HPV contamination of endocavity vaginal ultrasound probes: an underestimated route of nosocomial infection?. PLoS ONE 2012; 7: e48137
  • 34 Ma C, Ting S, Yeung A. et al. Transvaginal ultrasound probe contamination by the human papilloma virus in the emergency department. Emerg Med J 2013; 30: 472-475
  • 35 M’Zali F, Bounizra C, Leroy S. et al. Persistance of microbial contamination on transvaginal ultrasound probes despite low-level disinfection procedure. PLoS One 2014; 9: e93368
  • 36 Lucet J, Heard I, Roueli A. et al. Transvaginal ultrasound probes are HPV-free following low-level disinfection: a cross-sectional multicentre survey of 676 observations. Ultrasound Obstet Gynec 2019; 54: 688-695
  • 37 Benet T, Ritter J, Vanhems P. Risk of human immunodeficiency virus and hepatitis C virus infection related to endocavitary ultrasound probe exposure in France. Infect Control Hosp Epidemiol 2014; 35: 1429-1431
  • 38 Lesourd F, Izopet J, Mervan C. et al. Transmission of hepatitis C virus during ancillary procedures for assisted conception. Human Reprod 2000; 15: 1083-1085
  • 39 Kanagala P, Bradley C, Hoffman P. et al. Guidelines for transoesophageal echocardiographic probe cleaning and disinfection from the British Society of Echocardiography. European Journal of Echocardiography 2011; 12: i17-i23
  • 40 Medicines and Healthcare Products Regulatory Agency UK. Medical Device Alert 28 June 2012. Reusable transoesophageal echocardiography, transvaginal and transrectal probes (transducers) – failure to appropriately decontaminate. Document: MDA/2012/037.2012
  • 41 Noël A, Vastrade C, Dupont S. et al. Nosocomial outbreak of esbl-producing enterobacter cloacae among cardio-thoracic surgical patients: causes and consequences. Journal of Hospital Infection 2019; DOI: 10.1016/j.jhin.2019.01.001.
  • 42 Vesteinsdottir E, Helgason K, Sverrisson K. et al Infections and outcomes after cardiac surgery – the impact of outbreaks traced to transesophageal echocardiography probes. Acta Anaesthesiologica Scandavica 2019; DOI: 10.1111/aas.13360. (accessed May 2019)
  • 43 Levy PY, Teysseire N, Etienne J. et al. A nosocomial outbreak of Legionella pneumophila caused by contaminated transesophageal echocardiography probes. Infection Control and Hospital Epidemiology 2003; 24: 619-622
  • 44 Seki M, Machida H, Yamagishi Y. et al. Nosocomial outbreak of multidrug- resistant Pseudomonas aeruginosa caused by damaged transesophageal echocardiogram probe used in cardiovascular surgical operations. J Infect Chemother 2013; 19: 677-681
  • 45 Bancroft EA, English L, Terashita D. et al. Outbreak of Escherichia coli infections associated with a contaminated transesophageal echocardiography probe. Infection Control and Hospital Epidemiology 2013; 34: 1121-1123