Bacterial infections are a major burden in biliary endoscopy. Besides incomplete manual
cleaning of endoscopes, hard-to-clean components of duodenoscopes are responsible
for a substantial rate of bacterial contamination. A large meta-analysis showed that
15 % of duodenoscopes harbored microorganisms of gastrointestinal or oral origin,
independent from the colony forming unit (CFU) count. This indicates that the current
reprocessing and process control procedures, even with ethylene oxide (ETO), could
be insufficient [1]. Therefore, US Food and Drug Administration (FDA) warnings led companies to redesign
duodenoscopes with the aim of enhancing safety. Disposable caps or caps + elevator
are available options on the market [2]
[3]. Moreover, two disposable duodenoscopes models have recently been approved by the
FDA: the Boston Scientifc EXALT Model D and the Ambu A/S Ambu aScope Duodeno. Studies
on disposable duodenoscope feasibility and possible economic impact on the market
yielded interesting results, although the data are still preliminary [4]
[5]
[6]
[7]
[8].
In this paper, we present an innovative solution of a duodenoscope with a removable
cap, wire, and distal lever mechanism. This elevator system is called the single-use
Albarran module (Karl Storz) ([Fig. 1]). With it, a large, open surface can be created that is easy to brush and clean,
as is the case with any other instrument without a bifurcated channel. Basically,
the module is assembled and removed before and after every procedure, to be reprocessed
separately from the rest of the duodenoscope (even with ETO). Interestingly, the distal
end of the duodenoscope belongs to the removable module ([Fig. 2]). This detail makes the cleaning both the channel and the tip easier. Karl Storz
conducted some tests on duodenoscopes contaminated with organic fluids, such as blood.
After removal of the single-use Albarran module and duodenoscope reprocessing, the
instruments achieved the officially required level of cleaning (i. e. < 6.4 μg/cm2 for protein, < 2.4 μg/cm2 for hemoglobin and < 4 Log10/cm2 for bacteria) [9].
Fig. 1 The image on the left shows the lever mechanism to remove the cap; the image on the
right shows the wire that is inserted in the cap.
Fig. 2 The image shows the distal portion of the endoscope, without wire and cap.
Ten endoscopic retrograde cholangiopancreatographies (ERCPs) were performed in nine
consecutive patients using the single-use Albarran module. Data are summarized in
[Table 1]. The setting was urgent in one case and elective in nine. A native papilla was present
in seven patients (70 %); the bile duct cannulation and successful procedure rates
were both 100 %. Mean procedure time was 30 minutes. Indications for ERCP were biliary
lithiasis in seven cases and cancer palliation in two cases. Sphincterotomy and balloon
dilation were performed in five and four cases, respectively. Mechanical lithotripsy
was performed only once, whereas stone clearing required balloon dilation in six cases.
Brushing for cytology was done in one case and stenting for palliation in two.
Table 1
Elevator performance evaluation[1].
Patient (sex; age)
|
Indication
|
Bile duct cannulation
|
Operative procedure
|
Operative device
|
Stent characteristics
|
Procedure setting/duration (min)
|
Elevator performance
|
M,83
|
Biliary AP
|
Yes
|
Precut, plastic stent placement
|
|
Plastic, 10 Fr, 5 cm
|
Elective/31
|
Good
|
M,76
|
CBD stones
|
Yes (previous sphincterotomy)
|
Stent removal, balloon dilation, mechanic lithotripsy, stent placement
|
CRE Boston Scientific, balloon dilator 12–15 mm, Fogarty catheters
|
NA
|
Elective/34
|
Good
|
F,82
|
Biliary hilar stricture
|
Yes
|
Balloon dilation of main hepatic ducts
|
CONMED balloon dilator 6.00 mm
|
NA
|
Elective/43
|
Good
|
M,53
|
Cancer palliation
|
Yes
|
Stent placement
|
|
Partially covered SEMS 10 mm Ø, 6 cm
|
Elective/19
|
Good
|
M,59
|
CBD stones
|
Yes
|
Balloon dilation
CBD clearance
|
Dormia basket, CRE Boston Scientific, balloon dilator 12–15 mm
|
NA
|
Elective/20
|
Good
|
M,76
|
Biliary AP
|
Yes
|
Stent placement
|
|
Plastic 10 Fr, 5 cm
|
Urgent/43
|
Good
|
M;71
|
CBD stones
|
Yes
|
CBD clearance
|
|
NA
|
Elective/22
|
Good
|
F;42
|
Cancer palliation
|
Yes
|
Balloon dilation
Stent placement
|
CRE Boston Scientific, balloon dilator 10–12 mm
|
Partially covered SEMS
|
Elective/30
|
Good
|
F;82
|
2nd look
|
Yes (previous sphincterotomy)
|
Stent removal
Brushing
Stent placement
|
|
2 plastic stents, 10 Fr, 12 cm
|
Elective/43
|
Good
|
M;41
|
2nd look
|
Yes (previous sphincterotomy)
|
CBD clearance
|
|
NA
|
Elective/15
|
Good
|
NA, not applicable; AP, acute pancreatitis; CBD, common bile duct; PC, pancreatic
cancer; SEMS, self-expandable metallic stent.
1 Special attention was given to ease of accessory insertion, grip, friction, strength,
and width of the movement, rated as good = no problems nor difficulties; fair = any
problem occurred with no interference with the outcome of the procedure; poor = any
problem occurred significant enough to interrupt the procedure or change the elevator.
The aim of this small study was to gain insights into this novel device, with special
attention directed towards identifying any technical problems related to use of the
disposable systems. Notably, we did not find any issues with devices insertion, grip,
friction, strength, or width of the elevator movements, in contrast to the study by
Bang et al. [10] Because of the type of study (case series) and small sample size, it was not possible
to compare the single-use Albarran maneuverability and mechanical features with the
standard module. The only difference from the standard Albarran module was a shorter
width during the elevator stroke, which did not hinder the planned interventions.
In our experience, the disposable device allows the use of all devices commonly employed
in biliary endoscopy. Importantly, no cases of sepsis or infectious events were reported.
Unfortunately, a direct comparison of infectious event rate between the standard and
single-use Albarran module is not possible at this time, due to the small sample size
in the present case series. However, to our knowledge, this is the first description
of single-use Albarran module use during routine endoscopy in real-life settings.
Further studies and randomized controlled trials are needed to address comparisons
with the standard duodenoscopes and to better evaluate its use in clinical practice.