Open Access
CC BY 4.0 · Nuklearmedizin
DOI: 10.1055/a-2784-7468
Original Article

177Lu-based radioligand therapy: A retrospective multicenter analysis to calculate the effective half-life and follow-up dose for the public

177Lu-basierte Radioligandentherapie: Eine retrospektive multizentrische Analyse zur Berechnung der effektiven Halbwertszeit und der Folgedosis für Angehörige und Dritte

Authors

  • Christian Schütze

    1   Clinic for Nuclear Medicine, University Medical Center Göttingen, Göttingen, Germany (Ringgold ID: RIN27177)
  • Markus Borowski

    2   Institute for Radiology and Nuclear Medicine, Braunschweig Municipal Hospital, Braunschweig, Germany
  • Martin Freesmeyer

    3   Clinic for Nuclear Medicine, Jena University Hospital, Jena, Germany (Ringgold ID: RIN39065)
  • Robert Freudenberg

    4   Clinic and Polyclinic for Nuclear Medicine, University Hospital Carl Gustav Carus, Dresden, Germany (Ringgold ID: RIN39063)
  • Heribert Hänscheid

    5   Clinic for Nuclear Medicine, University Hospital Würzburg, Würzburg, Germany
  • Jens Kurth

    6   Clinic for Nuclear Medicine, University Hospital of Rostock, Rostock, Germany (Ringgold ID: RIN39071)
  • Marc Rudi Linde

    7   Institute for Nuclear Medicine, Vivantes Hospital Group, Berlin, Germany (Ringgold ID: RIN9231)
  • Lara-Christiane Militzer

    8   Clinic and Polyclinic for Nuclear Medicine, University Hospital Leipzig, Leipzig, Germany (Ringgold ID: RIN39066)
  • Bernhard Sattler

    8   Clinic and Polyclinic for Nuclear Medicine, University Hospital Leipzig, Leipzig, Germany (Ringgold ID: RIN39066)
  • Tabea Nikola Schmidt

    3   Clinic for Nuclear Medicine, Jena University Hospital, Jena, Germany (Ringgold ID: RIN39065)
  • Paul Soubiran

    7   Institute for Nuclear Medicine, Vivantes Hospital Group, Berlin, Germany (Ringgold ID: RIN9231)
  • Mathias Steffner

    9   Department of Radiation Protection and Medical Physics, Hannover Medical School, Hanover, Germany (Ringgold ID: RIN9177)
  • Lars Stegger

    10   Clinic for Nuclear Medicine, University Hospital Münster, Münster, Germany (Ringgold ID: RIN39069)
  • Carsten Wanke

    9   Department of Radiation Protection and Medical Physics, Hannover Medical School, Hanover, Germany (Ringgold ID: RIN9177)
  • Christian Kühnel

    3   Clinic for Nuclear Medicine, Jena University Hospital, Jena, Germany (Ringgold ID: RIN39065)
 

Abstract

The study investigated routinely used radiopharmaceuticals containing lutetium-177 (177Lu), providing reference values for their effective half-life (T1/2,eff). So far, no guidelines regarding discharging this patients exist, are essential to ensure radiation protection of the public. This study contributes to the development of binding discharge criteria regarding 177Lu therapies.

This retrospective multicenter study comprised eight nuclear medicine departments. Two commercially available products and two in-house preparations were considered. Radiation measurement methods, in terms of measuring device, setup, and time points were reported, as well as dose rate measurements for T1/2,eff calculation.

The study includes 210 dose rate measurement sets with a mean administered activity of (7149 ± 522) MBq. When only measurements up to 48 h p.i. were taken into account the T1/2, eff were shorter (Pluvicto 1.30 d, Lutathera 1.40 d, ihPSMA 1.45 d, ihRRT 1.97 d) compared to 168 h (1.60 d, 1.89 d, 2.10 d, and 2.85 d, respectively). Differences were statistically significant for all compounds except Lutathera. The 90th percentile of the T1/2,eff observed in late measurements (Pluvicto 2.0 d, Lutathera 3.0 d, ihPSMA 3.5 d, ihRRT 4.0 d) can be considered conservative values for discharge calculations.

The T1/2,eff of 177Lu is considerably shorter than its T1/2,phy, owing to the pharmacokinetics of the entire molecule. This highlights the need of considering biological clearance mechanisms into calculations of discharge times for patients treated with 177Lu. If direct measurements are not feasible in clinical routine, reference values support inpatient planning while ensuring radiation protection.


Introduction

Radioligand therapies using lutetium-177 (177Lu)-labelled compounds have become well-established therapies in nuclear medicine. They are particularly important in the management of metastatic castration-resistant prostate cancer using prostate-specific membrane antigen (PSMA) ligands, and in the treatment of neuroendocrine tumors with peptide receptor radionuclide therapy (RRT) with somatostatin analogues [1] [2]. However, despite their widespread clinical use, discharge criteria following administration and the half-lives administered for dose estimation remain inconsistently defined in Germany [3] [4].

According to §122(4) of the German Radiation Protection Ordinance (Strahlenschutzverordnung, StrlSchV), the release of patients from nuclear medicine wards is only permissible if the expected effective dose to members of the public, including relatives and caregivers, does not exceed 1 millisievert (mSv) per calendar year. This limit must be considered cumulatively across all radionuclide therapies administered within a given year.

For thyroid therapies with iodine-131 (131I), which are commonly administered orally using capsules, discharge criteria are outlined in the German guideline “Radiation Protection in Medicine” [5] [6]. These stipulate a minimum inpatient stay of 48 h and give a conservative estimate for the dose rate at patient discharge of 3.5 µSv/h at a distance of 2 m. These criteria are based on conservative assumptions, namely, the continuous presence of another individual at a distance of 2 m until complete physical decay of the administered activity, applying the T1/2, phy for 131I (8.02 d). Importantly, a significant fraction of the administered activity is excreted within the first 48 h after administration [7]. Excretions are collected and safely managed via the hospital’s internal decay system [8]. The majority of thyroid disease patients are not treated more than once per calendar year, thus this criterion does not result in an increased length of stay.

Currently, no fixed discharge criteria are defined for intravenously administered 177Lu-therapies, although these procedures have become standard in recent years [9] [10]. The only available recommendation was published by the German Radiation Protection Commission and relies on the T1/2, phy of 177Lu; however, following this conservative approach could suggest exceedances if multiple administrations are performed within one year. Since only a few studies have addressed this topic to date, robust clinical data are scarce. Therefore, the T1/2, phy remains a provisional parameter for defining discharge conditions [11] [12] [13]. Since 177Lu-therapies are administered in repeated cycles rather than as single administrations, discharge criteria must account for cumulative exposures across treatment courses [14]. Current clinical guidelines recommend up to six cycles of PSMA-targeted therapy or four cycles of PRRT, with intervals of 6–8 weeks (PSMA) or 6–12 weeks (PRRT) between cycles [14] [15]. Both approved radiopharmaceuticals, such as Pluvicto and Lutathera, and in-house preparations manufactured under GMP conditions are established in clinical practice [16] [17]. Consequently, individual discharge decisions must incorporate not only the most recent administration but also the expected cumulative dose from all planned and already completed cycles within the same calendar year, with probably a mixture of different radiopharmaceuticals used.

This regulatory framework underscores the importance of estimating the effective half-life of 177Lu in individual patients to enable reliable dose calculations. The effective half-life, determined by the combined influence of physical decay, tumor burden, and biological clearance, governs the temporal decline of emitted radiation and thereby the potential external exposure to third parties following patient discharge.

In the currently still predominantly palliative setting, in which 177Lu-based therapies are administered, patient acceptance is strongly influenced by the perceived safety, time spent in the hospital and transparency of treatment. Establishing an evidence-based foundation for discharge decision-making is therefore essential. Systematic collection and evaluation of clinical and procedural data could provide the basis for a comprehensive safety concept, which in turn may serve as a formal reference for regulatory approval. Such a framework would not only strengthen the legitimacy of therapeutic practice but also improve transparency and foster trust among patients, caregivers, and stakeholders.

Therefore, the present retrospective multicenter study aims to analyze 177Lu-based therapeutics in clinical routine across several German nuclear medicine departments, with a focus on estimating the effective half-life. Based on this analysis, it is anticipated that recommendations for effective half-life values can be given to support assessments relevant to radiation protection. Specifically, this retrospective study examines the measurement techniques, modelling approaches, and institutional thresholds employed to quantify radiation exposure to third parties and to guide discharge decisions.


Materials and Methods

This retrospective multicenter study was conducted across eight nuclear medicine departments from seven different federal states in Germany, including university hospitals and municipal institutions. All participating centers routinely perform inpatient therapies with 177Lu-labelled radiopharmaceuticals.

Study Design and Ethical Approval

Institutional procedures related to patient discharge following 177Lu therapy were compared across centers using a structured data collection template. The study was approved by the lead Ethics Committee of University Hospital Jena reference number: No. 2024–3421.


Classification by Radiopharmaceutical Type

To account for pharmacokinetic variability, centers differentiated on the type of radiopharmaceutical used:

  • Commercial products:

    1. Pluvicto = [177Lu]Lu-PSMA-617 (Pluvicto, Novartis AG, Basel, Switzerland)

    2. Lutathera= [177Lu]Lu-DOTATATE (Lutathera, Novartis AG, Basel, Switzerland)

  • In-house preparations (using GMP-grade kits and in-house radiolabelling):

    1. ihPSMA = [177Lu]Lu-PSMA-I&T

    2. ihRRT = [177Lu]Lu-DOTATOC

Pharmacokinetic differences, especially in terms of plasma clearance and tumor retention, were not considered when analyzing dose rate decay and estimating T1/2, eff.


Dose Rate Measurement and Kinetic Modeling

Each center reported their local standard operating procedures for calculating the T1/2, eff, with dose rate measurements (DRM) or gamma camera measurements and the subsequent discharge planning, including the following informations:

  • Type of therapy

  • Type and calibration status of measuring device for discharging (handheld or fixed/wall mounted)

  • Geometric setup (distance from patient, anterior/posterior positioning)

  • Further measurements, conducted using a gamma camera

  • Timing and frequency of measurements


Half-life Determination

Measurements of the dose rate were retrospectively evaluated in eight clinics. For this purpose, depending on the data available, different time points were considered. Subgroup separation was achieved through the implementation of a measurement time interval. Some clinics measured the patients until < 48 h after administration due to discharging. In order to apply a uniform adjustment model based on the raw data obtained, mono-exponential functions were fitted to three consecutive measurement points for the calculation of effective half-lives; an early T1/2, eff with the last measurement point after two days (< 48 h) and a late T1/2, eff with the last measurement point between 48 to 168 h (< 168 h). The median and 90th percentile of the effective half-lives for each treatment type were calculated. The 90th percentiles can be taken as practical estimators for the effective half-life when calculating patient discharge times and the resulting effective dose for other persons, see below.

The results were subdivided by treatment type, with the estimates of third-party exposure compared based on physical and T1/2, eff.


Statistical Analysis

The mean value, along with the standard deviation, median, range, and 90th percentile, was calculated for each radiopharmaceutical and time interval. To ascertain whether the effective half-life differed significantly depending on the fitting model (<48 h vs. < 168 h), intraindividual comparisons were performed separately for each radiopharmaceutical and then examined for significant differences. Given the non-normal distribution of the data, the non-parametric Mann-Whitney-U-test was administered. The level of significance was defined as α = 0.05 . All analyses were conducted using the R programming language (Version 4.4.1, R Core Team, Vienna, Austria) and Integrated Development Environment RStudio (Version 2024.12.0, Posit-Software, PBC).



Results

All participating centers contributed a total of n=210 sets of dose rate measurements, divided into n=25 ihRRT, n=116 ihPSMA, n=14 Lutathera, and n=55 Pluvicto datasets following the administration of 177Lu-labelled radiopharmaceuticals including the dose rate measurements ([Table 1]). The mean activity of 177Lu regardless of therapy type was (7149 ± 522) MBq (median: 7399 MBq, range: 5077–7853 MBq). Two centers used handheld devices, three employed fixed mounted and another three ceiling-mounted detectors. Measurement distances ranged from 1 to 2 m, with most centers opting for a 2 m anterior setup. Three centers performed both anterior and posterior measurements for calculating a mean of both or used additional information from whole body scans using gamma cameras. The number of measurements per patient ranged from 3 to 7, timing varied considerably: while some centers employed fixed daily schedules (e.g., 9:00, 13:00 and 20:00), others defined time points relative to the actual administration (e.g., 12 h, 36 h, 48 h p.i.). Three centers routinely conducted measurements later than 48 h after administration, reflecting logistic limitations related to the discharge workflow. A small number of outliers were identified, and following individual case review, it was determined that these were attributable to tumor burden.

Table 1 Clinical site comparison for intra-therapeutic dose rate measurements.

Site Nr.

Type

DRM type

DRM status

Distance patient to DRM

Method of measurement

Time-points p.i.

1 = ihPSMA

2 = Pluvicto

3 = ihRRT

4 = Lutathera

oc = officially calibrated certified by authorities

ivc = inhouse verication and calibration

a = anterior

p = posterior

s = standing

l = lying

0–6 h

12–20 h

24–30 h

36–42 h

44–48 h

72–96 h

144–168 h

* The actual distance between the ceiling-mounted measuring device and the patient is approximately 2.6 m. During commissioning, calibration and correction was performed for distances up to 2.0 m. An annual calibration control was performed.

+ Only DRM measurements performed for discharge, separate time-point measurements for calculating the effective half-life with gamma camera measurements.

1

1, 3

mounted

Berthold LB 123

oc

2 m

a + p/s

x

x

x

x

x

x

x

2

1, 3

mounted

Berthold LB 123

ivc

2 m

a + p/s

x

x

x

3

1

handheld

Thermo Scientific FH-40+

ivc

2 m

a/s

x

x

x

x

4

1, 2, 3, 4

wall mounted

Step Sensor DLMon

ivc

2 m*

a/l

x

x

x

x

x

5

1, 2, 3

wall mounted

Step Sensor DLMon

ivc

2 m*

a + p/l

x

x

x

x

x

6

1

wall mounted

Automess 6150AD 18+

ivc

2 m*

a/l

x

x

x

x

x

7

2, 3, 4

Mounted

Berthold UMO-LB123+

oc

2 m

a/s

x

x

x

8

2, 3, 4

handheld

Thermo Scientific FH-40

oc

2 m

a + p/s

x

x

x

For time windows < 48 h ([Fig. 1]), the effective half-life showed the following distributions: Pluvicto exhibited the shortest mean effective half-life of (1.30 ± 0.91) d (median: 1.00 d; range: 0.47 – 6.26 d; 90th percentile: 2.35 d), followed by Lutathera with (1.40 ± 0.75) d (median: 1.11 d; range: 0.58 – 3.06 d; 90th percentile: 2.35 d), ihPSMA with (1.45 ± 0.76) d (median: 1.20 d; range: 0.35 – 4.84 d; 90th percentile: 2.58 d), and ihRRT with (1.97 ± 1.23) d (median: 1.67 d; range: 0.40 – 6.19 d; 90th percentile: 2.44 d).

Zoom
Fig. 1 Effective half-lives in days of the four radiopharmaceuticals divided in measurements < 48 h and < 168 h.

In contrast, for time windows < 168 h ([Fig. 1]), higher mean effective half-life were observed across all groups: Pluvicto reached (1.60 ± 0.63) d (median: 1.41 d; range: 0.75 – 3.14 d; 90th percentile: 2.57 d), Lutathera showed (1.89 ± 1.14) d (median: 1.86 d; range: 0.66 – 3.66 d; 90th percentile: 2.59 d), ihPSMA reached (2.10 ± 0.98) d (median: 2.04 d; range: 0.35 – 6.65 d; 90th percentile: 3.08 d), and ihRRT demonstrated the longest values with (2.85 ± 1.12) d (median: 2.72d; range: 0.90 – 5.82 d; 90th percentile: 3.88 d).

Significant differences in T1/2, eff were observed between several therapy types and measurement time intervals ([Table 2]). Notably, comparisons involving Pluvicto < 48 h and ihRRT < 168 h with other subgroups yielded the lowest p-values, indicating pronounced variability as a function of both therapy and timing.

Table 2 P-values from pairwise comparisons between therapy and measurement time intervals (< 48 h vs.  < 168 h) based on effective half-life.

ihRRT < 48 h

ihRRT < 168 h

ihPSMA < 48 h

ihPSMA < 168 h

Lutathera < 48 h

Lutathera 168 h

Pluvicto < 48 h

Pluvicto < 168 h

*Statistically significant differences (p < 0.05)

ihRRT < 48 h

0.002*

0.018*

0.187

0.854

0.728

0.001*

0.294

ihRRT < 168h

< 0.001*

0.007*

< 0.001*

0.010*

< 0.001*

< 0.001*

ihPSMA < 48 h

< 0.001*

0.549

0.038*

0.035*

0.072

ihPSMA < 168h

0.003*

0.569

< 0.001*

0.003*

Lutathera < 48 h

0.124

0.418

0.190

Lutathera < 168h

0.019*

0.263

Pluvicto < 48 h

0.003*

Pluvicto < 168h

[Fig. 2] shows that the T1/2, eff, which had been chosen with regard to the lowest number of datasets for one of the radiopharmaceuticals investigated in this work to provide a statistically sound basis for generic estimators, of all compounds was substantially lower than the T1/2, phy of 177Lu. The median deviation ranged from –54 % to –74 %, with the smallest median deviation observed for ihRRT (–54 %), followed by ihPSMA (–67 %), Lutathera (–72 %), and Pluvicto (–74 %).

Zoom
Fig. 2 Deviation of the measured effective half-lives from measurements < 168 h compared to the physical half-lives (Teff).

Based on the 90th percentiles of the measured T1/2, eff, the following values are recommended as conservative estimates for radiation protection and dose calculation for third parties as also shown in [Table 3]: 4.0 d for ihRRT, 3.5 d for ihPSMA, 3.0 d for Lutathera, and 2.0 d for Pluvicto. These values reflect upper-range kinetics in 90 % of treated patients and can be used to estimate expected dose exposures in non-monitored environments.

Table 3 Estimated cumulative exposure for different numbers of therapy cycles per year based on dose rate and time of discharge.

Type

Recommended as conservative estimates for Teff [d]

Follow-up dose for third parties per application [mSv]

Max. number of therapies per year for exposure < 1 mSv for third parties

* mixture of different Types

ihRRT

4.0

0.31

3

ihPSMA

3.5

0.17

5

Lutathera

3.0

0.10

10

Pluvicto

2.0

0.07

13

all*

3.0

0.12

8


Discussion

One of the main findings of this study is the substantial heterogeneity in dose rate measurement protocols among the participating centers. Differences were observed in terms of measurement devices, source-detector distances, anatomical orientation (anterior vs. posterior), frequency, and timing of measurements. While some of this variability reflects local infrastructure and logistical constraints, it also introduces inconsistencies in the estimation of patient-specific T1/2, eff and, consequently, in the extrapolated exposure to relatives and caregivers. However, no statistically significant differences were observed between the facilities in terms of the half-life determined. This can be attributed to the fact that, even in cases where a system is not calibrated, the temporal component, i.e. the T1/2, eff, can be accurately determined provided the same setup is utilized for all patient-specific measurements.

From a methodological point of view, it is important to differentiate between absolute measurement accuracy and relative measurement consistency. While the former is essential when comparing dose rates across institutions or against regulatory thresholds, the latter is sufficient for modeling the time-dependent decline in dose rate. This, for instance, applies for estimating the T1/2, eff provided that all measurements in a given patient are performed under identical geometric and instrumental conditions. In such cases, the lack of calibration only affects the absolute values, not the temporal decay pattern, allowing for reliable curve fitting and relative kinetic interpretation.

Bi-exponential modeling of dose rate decay is a well-established approach for determining T1/2, eff, especially when both fast and slow clearance components are relevant. Particularly in cases of high tumor retention with a long half-life, pronounced bi-exponential kinetics may occur, which cannot be adequately represented by a simple exponential function and should be fit accordingly. However, the accuracy and stability of the fitted parameters are strongly dependent on the availability of late-phase data points. This underlines the importance of integrating at least one data point in the terminal elimination phase to avoid over- or underestimation of residual activity and associated dose rates. Given the frequent discharge of patients after 48 h in Germany, this poses a methodological challenge and supports the development of standardized fitting protocols, potentially supplemented by population-based decay constants derived from larger datasets.

Although this variability did not universally reach statistical significance, it is clinically relevant in the context of radiation protection, as it affects the predictability of external dose contributions and thus the precision of discharge planning.

A key finding of this multicenter study is the consistent observation that the T1/2, eff derived from patient-specific dose rate measurements are, on average, substantially shorter than the T1/2, phy of 6.7 d. This effect was observed across all therapeutic administrations and reflects the influence of their pharmacokinetic properties.

To quantify the effect of observation duration on calculated half-life, intraindividual comparisons were performed between measurements taken within the site performing measurements during the first 48 h and those extending measurement points beyond 48 h after administration. Mann-Whitney-U-tests revealed statistically significant differences in T1/2, eff between these time intervals for ihRRT (p = 0.002), ihPSMA (p < 0.001) and Pluvicto (p = 0.003). Only for Lutathera the difference did not reach statistical significance (p = 0.124). These results support the notion that shorter acquisition windows (< 48 h) systematically underestimate the T1/2, eff, particularly for radiopharmaceuticals with slower tissue clearance.

Given this, the use of early time-point measurements alone may be insufficient for accurate dosimetric estimation or radiation protection assessment. Nonetheless, the consistent deviation pattern across compounds may permit conservative extrapolation using percentile-based half-life estimates, as proposed in this study.

This observation is physiologically plausible, as the T1/2, eff incorporates both physical decay and biological clearance mechanisms – primarily renal excretion and, to a lesser extent, hepatobiliary elimination. Consequently, assuming a purely physical decay model would systematically overestimate the exposure of third parties and lead to overly conservative discharge planning. The recommended half-lives show that, except for ihRRT and very close to the ihPSMA, the other preparations allow the maximum number of therapy cycles according to the guidelines 1 mSv per year for the public depending on the release time. This would not be the case when only using basic T1/2, phy. Considering the variability in delivery conditions, necessary activity adjustments, and patient-specific circumstances, it may in practice become necessary to combine different radiopharmaceuticals in one period of cycles (e.g., ihPSMA with Pluvicto, or ihRRT with Lutathera). This clinical reality further underscores the need for defining a combined recommended half-life value, which would provide a more robust and universally applicable basis for both clinical decision-making and regulatory frameworks. Due to the multicenter retrospective data situation, data from patients with higher tumor volumes could not be explicitly evaluated and analyzed. This was not explicitly taken into account during data collection. One limitation at that point was that the maximum cycles per year were conservatively estimated based on constant uptake and T1/2, eff in all cycles, which is unlikely to be the case under real-world conditions. The finding supports the clinical relevance of individualized or protocol-based T1/2, eff estimation to improve the precision of dose calculations, without compromising safety. In routine settings, using institution-specific or population-based T1/2, eff values derived from validated measurements could provide a pragmatic and evidence-based alternative to worst-case assumptions based solely on physical decay.

The proposed extrapolated T1/2, eff are not intended to replace individualized intratherapeutic measurements, which remain essential for accurate patient-specific dosimetry and radiation protection. To better assess the cumulative exposure potential for third, extended dose rate measurements beyond the current standard – ideally up to 14 d post- administration – may offer valuable insights into the long-term effective clearance behavior of 177Lu labelled radiopharmaceuticals. First approaches already exist for PRRT [18]. Such data would allow for a more accurate risk assessment of residual external and internal radiation burdens within the home environment, where patient movement, proximity, and shielding conditions vary greatly. When combined with the results of the present study, these extended measurements could serve as the foundation for a differentiated risk-based approach to discharge planning.

In addition to the patient’s external radiation exposure, the excretion of the administered activity must also be taken into account as part of the hospitalization decision. High renal elimination is characteristic of almost all low-molecular compounds and peptide therapeutics, which leads to a correspondingly high concentration of activity in the urine. The activity that is not bound in organs and tumours leaves the body as a fast clearance component with a half-life of a few hours, meaning that 70% or more is often excreted after one day [12] [19]. However, depending on the percentage of bound activity and the half-life of the slow clearance component, more than half of the activity may still be excreted after 48 hours in individual cases in patients with high tumour uptake. Especially in patient populations with metastatic castration-resistant prostate cancer (mCRPC), contamination events through urine have been documented in the clinical setting, as documented by recent conference reports, among others [13] [19]. Such contamination poses the risk of significant local skin doses, which can occur in particular in direct contact with contaminated surfaces or materials and are described in the literature with individual doses in the range of several 10 mSv [19] [20]. In addition to the limitation of the effective dose of 1 mSv for third parties required in § 80 StrlSchG, compliance with the limit value of 50 mSv for the skin dose is therefore also coming into focus. Renal excretion is less relevant for antibody-based radiopharmaceuticals, so that this risk is comparatively reduced. Nevertheless, this aspect underlines the need to consider not only the external radiation exposure, but also the contamination and skin dose risk in a differentiated manner when assessing the discharge criteria.

Therefore, rather than aiming for rigid national thresholds or singular discharge values, a more productive approach may lie in harmonizing the methodology:

  • Defining minimum measurement requirements (e.g., ≥ 3 time points, ideally with one > 48 h post-administration)

  • Standardizing geometric conditions

  • Providing recommended half-life for different therapies, stratified by radiopharmaceutical

Such an approach would preserve flexibility for clinical workflow, while ensuring regulatory compliance and enhancing radiation protection by basing assessments on real patient kinetics rather than theoretical decay models.


Conclusion

This multicenter study demonstrates that the T1/2, eff of 177Lu-based therapies is significantly shorter than the T1/2, phy of 177Lu, with median deviations ranging from –54 % to –74 % depending on the radiopharmaceutical. An important implication of our findings is that post 48 h measurements, which showed comparatively higher values of half-life, should be integrated into future dose estimations to improve accuracy and clinical relevance. In the future, patient-specific measurement of half-life remains “gold standard”; alternatively, the use of population-based half-life can be an alternative for discharge planning with lesser effort but in many cases longer assumed half-life compared to measured half-life. However, in clinical scenarios where direct measurement is not feasible, reference values based on the 90th percentile of the T1/2, eff distribution – stratified by therapeutic indication – may provide a conservative yet pragmatic framework to ensure compliance with dose limits, such as the 1 mSv/year threshold for third-party exposure. Nevertheless, it must be emphasized that these extrapolated values cannot substitute for individualized intratherapeutic assessment. Further studies for long-term kinetic measurements beyond 1–2 weeks, including different patient subgroup parameter (e.g. the glomerular filtration rate, tumor load) are warranted.


Abbreviations

a: anterior
DRM: Dose Rate Meter
GMP: Good Manufacturing Practice
ihPSMA: In-house Lu-177-PSMA-I&T
ihRRT: In-house Lu-177-DOTATOC
ivc: inhouse verication and calibration
l: lying
Lu-177: Lutetium-177
Lutathera: Lu-177-DOTATATE (Fa. Novartis)
mSv: Millisievert
n: sample size
oc: officially calibrated certified by authorities
p: posterior
p.i.: post injectionem
Pluvicto: Lu-177-PSMA-617 (Fa. Novartis)
PSMA: Prostate-Specific Membrane Antigen Therapy
RRT: Receptor Radionuclide Therapy
s: standing
StrlSchV: Strahlenschutzverordnung (German Radiation Protection Ordinance)
µSv/h: Mikrosievert per hour



Contributorsʼ Statement

Christian Schütze: Conceptualization, Data curation, Formal analysis, Investigation, Methodology, Project administration, Resources, Supervision, Validation, Visualization, Writing - original draft, Writing - review & editing. Markus Borowski: Conceptualization, Data curation, Investigation, Methodology, Resources, Supervision, Writing - review & editing. Martin Freesmeyer: Resources, Supervision, Writing - review & editing. Robert Freudenberg: Conceptualization, Formal analysis, Methodology, Supervision, Writing - review & editing. Heribert Hänscheid: Data curation, Formal analysis, Investigation, Resources, Supervision, Writing - review & editing. Jens Kurth: Conceptualization, Data curation, Formal analysis, Investigation, Methodology, Resources, Supervision, Writing - review & editing. Marc Rudi Linde: Investigation, Resources, Writing - review & editing. Lara-Christiane Militzer: Investigation, Resources, Writing - review & editing. Bernhard Sattler: Conceptualization, Data curation, Investigation, Methodology, Resources, Supervision, Writing - review & editing. Tabea Nikola Schmidt: Data curation, Formal analysis, Validation, Writing - review & editing. Paul Soubiran: Investigation, Resources, Writing - review & editing. Mathias Steffner: Data curation, Methodology, Writing - review & editing. Lars Stegger: Data curation, Investigation, Resources, Writing - review & editing. Carsten Wanke: Conceptualization, Data curation, Methodology, Resources, Writing - review & editing. Christian Kühnel: Conceptualization, Data curation, Formal analysis, Investigation, Methodology, Project administration, Resources, Supervision, Validation, Visualization, Writing - original draft, Writing - review & editing.

Conflict of Interest

Jens Kurth: Member Advisory Board Novartis, Eli Lilly


Correspondence

Dr. Christian Schütze
Clinic for Nuclear Medicine, University Medical Center Göttingen
Göttingen
Germany   

Publication History

Received: 26 September 2025

Accepted after revision: 17 December 2025

Article published online:
29 January 2026

© 2026. The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution License, permitting unrestricted use, distribution, and reproduction so long as the original work is properly cited. (https://creativecommons.org/licenses/by/4.0/).

Georg Thieme Verlag KG
Oswald-Hesse-Straße 50, 70469 Stuttgart, Germany


Zoom
Fig. 1 Effective half-lives in days of the four radiopharmaceuticals divided in measurements < 48 h and < 168 h.
Zoom
Fig. 2 Deviation of the measured effective half-lives from measurements < 168 h compared to the physical half-lives (Teff).