Abstract
Purpose: To evaluate the feasibility of transperineal MR-guided prostate biopsy using a stereotactical
targeting system originally designed for anorectal usage.
Materials and Methods: A commercially available DynaTRIM MR targeting system (Invivo corp., Gainesville
FL, USA) originally designed only for anorectal application was used on a 70 cm wide-bore,
whole-body 3 Tesla MR-system (Ingenia, Philips Healthcare, Best, NL). Transperineal
biopsy was performed following mulitparametric MR imaging for targeting of the lesion.
Results: The anorectal device allowed for correct localization and successful MR-guided transperineal
biopsy of the targeted lesion.
Conclusion: MR-guided transperineal biopsy is feasible using a commercially available anorectal
stereotactic biopsy device. This may lead to a broader acceptance of this approach
for targeted prostate biopsies.
Zusammenfassung
Ziel: Machbarkeit der transperinealen Prostata-Biopsie mit einer MR-gestützten Stereotaxie-Einheit,
welche primär für die transrektale Anwendung konzipiert wurde.
Material und Methoden: Eine kommerziell verfügbare DynaTRIM-MR-Stereotaxie-Einheit (Invivo corp., Gainesville
FL, USA), initial konzipiert für die transrektale Probeentnahme, wurde an einem 70 cm
wide-bore 3 Tesla Ganzkörper-MRT (Ingenia, Philips Healthcare, Best, NL) eingesetzt.
Nach Akquisition einer multiparametrischen MRT zur Tumorlokalisation und Interventionsplanung
erfolgte eine transperineale Stanzbiopsie.
Ergebnisse: Die transrektale Stereotaxie-Einheit ermöglichte die korrekte Lokalisation und eine
technisch erfolgreiche transperineale Probeentnahme mittels MR-gesteuerter Navigation.
Schlussfolgerung: Eine transperineale Stanzbiopsie ist mit einer kommerziell verfügbaren transrektalen
Stereotaxie-Einheit unter MR-Navigation durchführbar. Dies kann zu einer besseren
Verbreitung der transperinealen Biopsie-Methode führen.
Key words
prostate - interventional MR - technology assessment