Key words
radiation effects - radiation safety - CT fluoroscopy - biopsy - interventional procedures
Introduction
Nowadays, minimally invasive CT-guided interventions play an important role at many
institutions and are routinely performed. CT fluoroscopy was first introduced in the
clinical routine by Kato and Katada in 1993 [1]. Images can be acquired with real-time control, high geometric precision, and without
significant artifacts, resulting in increased accuracy, reduced intervention times,
and improved biopsy results [2]
[3]
[4]
[5]
[6]. However, interventions performed under CT fluoroscopy guidance entail relevant
additional radiation exposure for the personnel involved in the intervention. In particular,
manipulations performed in the direct beam path of the CT unit can result in significant
skin doses for the examiner [4]
[7]
[8]. CT fluoroscopy-guided interventions are currently a fixture in the clinical routine
due to the wide range of possible applications. In the last decades, the interventional
possibilities and indications of CT have continued to expand, e. g. to the areas of
interstitial brachytherapy and radiofrequency ablation [9]
[10]
[11]
[12]
[13]
[14].
Objective
Due to the increase in CT fluoroscopy-guided interventions and their complexity, it
is necessary to conduct a detailed analysis of the radiologist’s hand dose exposure.
The hand dose range and the risks for the operator must be determined. This raises
the following questions: Is the annual hand dose limit of 500 mSv for persons occupationally
exposed to radiation reached? Which influencing factors result in a higher hand dose?
Materials and Methods
Patient Data
In the prospective study for examining the radiologist’s hand dose, 138 CT-guided
interventions involving 131 patients were able to be completed between November 2009
and December 2010. The following interventions were included: CT-guided core biopsy,
CT-guided drainage, and CT-guided pain therapy (periradicular therapy and neurolysis/block
of the celiac plexus). The study was approved by the Ethics Committee of the Medical
Faculty of the Otto-von-Guericke University of Magdeburg.
Inclusion and exclusion criteria for the CT-guided interventions
Inclusion criteria
-
All interventions had to be performed on a 16-slice multidetector CT unit (Aquilion
16, Toshiba, Japan) using CT fluoroscopy.
-
The preliminary diagnostic CT examination had to show a clearly defined target lesion
or target region (in the case of periradicular therapy or celiac plexus neurolysis).
-
Re-interventions were possible.
-
Interventionalists were required at least 6 months of experience with CT-guided interventions.
Exclusion criteria
-
Complex interventions (e. g. CT-guided brachytherapy or radiofrequency ablation),
-
puncture of multiple lesions during the same procedure,
-
active participation of multiple interventionalists in the puncture procedure,
-
misplacement of the EDD-30 sensor or periinterventional shifting of the sensor,
-
lack of documentation regarding the needle tip position/drainage system or lack of
documentation regarding the contrast agent distribution during interventional pain
therapies,
-
change of the default slice thickness or tube voltage.
Measurement methods
The hand dose was measured with an immediately readable electronic dosimeter, the
EDD-30 (Unfors, Sweden). The detection threshold of the sensor is a dose rate of 10
nGy/s (range: 10 nGy/s – 0.6 mGy/s) [15]. The EDD-30 sensor was attached to the back of the interventionalist’s dominant
hand. It was positioned between the head of the 2nd and 3rd metacarpal bone ([Fig. 1]).
Fig. 1 EDD-30 sensor affixed to the back of the right hand combined with bilaterally affixed
TLD plates.
Validation of the dose measurement via the thermoluminescence detector (TLD) measurement
method
In an initial validation study of the hand dosimeter, a simultaneous measurement with
TLD plates (Hp 0.07) involving 31 patients was performed. A plate was adhered directly
to the right and to the left of the EDD-30 sensor ([Fig. 1]).
Technical parameters
All interventions were performed on a 16-slice CT unit (Aquilion 16, Toshiba, Japan).
The default technical parameters for CT fluoroscopy were 120 kV, 90 mA, a 4 mm slice
thickness, and a tube rotation time of 0.5 s. In the case of six obese patients (BMI
> 30 kg/m²), the tube current intensity had to be increased (to 200 mA for five patients
and 300 mA for one patient) due to a decrease in image quality. A combination of discontinuous/intermittent
and continuous fluoroscopy was used. Radiation protection gloves were not worn. A
standard needle holder (14 cm) was used for 13 retroperitoneal punctures. The following
radiation protection measures for the interventionalists were taken: Radiation protection
apron with thyroid protection (0.35 mm Pb equivalent), radiation protection glasses,
and a lead blanket for the patient.
Recorded parameters
The following variables were documented: Tumor size (anterior-posterior size, width
and craniocaudal diameter), location, name of the interventionalist, level of experience,
degree of difficulty (1 – 3), body-mass index of the patient, and scan parameters,
such as tube current time product (mAs), dose-length product (DLP), and fluoroscopy
time.
Moreover, the distance between the skin surface and the needle tip (penetration depth)
was measured. In the case of samples being taken multiple times, the distance was
multiplied by the number of samples taken (or measured again) and then evaluated as
a “summed puncture depth”. This was not taken into consideration for punctures using
a coaxial needle since the distance to the area to be punctured would not have to
be fully covered again.
Classification of the different interventionalists
The eleven interventionalists were classified according to two groups (two different
levels of experience):
-
low level of experience (< 5 years of experience with CT fluoroscopy interventions),
-
high level of experience (> 5 years of experience with CT fluoroscopy interventions).
Classification according to degree of difficulty
The interventions were categorized according to three degrees of difficulty with the
lesion size and the position in relation to other risk structures being taken into
consideration. Two radiologists retrospectively and independently categorized the
interventions (based on the acquired CT images) according to three degrees of difficulty.
In the case of differences in the degree of difficulty (due to varying reader evaluation),
the higher degree of difficulty was used.
The classification according to degree of difficulty on a scale of 1 – 3 was based
on the following criteria:
-
lesion > 2 cm and risk structures at a distance of > 3 cm in relation to the lesion
and the path of the puncture channel,
-
lesion < 2 cm and adjacent risk structures > 2 cm away, angled path of puncture; lung
puncture with a distance of > 2 cm from the pleura,
-
lesion < 1 cm in a space orientation and risk structures directly adjacent to or along
the puncture channel.
Despite the size of the target area: 5 × 5 × 5 mm³, periradicular therapy was classified
as easy (degree of difficulty 1).
Data Analysis and Statistics
A descriptive data analysis was first performed. Since many of the analyzed variables
have an asymmetrical distribution, the median and the extreme values were always provided
in addition to the mean and standard deviation. Group differences and correlations
between the variables were also examined with nonparametric tests. The following statistical
methods were used for this purpose: The group differences (different procedures and
locations) of the measured hand dose values were analyzed with a Kruskal-Wallis test.
To evaluate the influencing factors with respect to the measured hand dose values,
a multivariate analysis and a Spearman-Rho rank correlation coefficient test were
performed. Univariate variance analyses were used for the simultaneous analysis of
degree of difficulty and level of experience with respect to hand dose and fluoroscopy
time. The classification of the degree of difficulty by the two readers was checked
with a Cohen’s Kappa analysis. A significance level of α = 0.05 (two-sided) was used
for all tests.
Results
The distribution of the 138 CT-guided interventions completed between November 2009
and December 2010 is shown in [Fig. 2]. The median body-mass index of the included patients was 24.7 kg/m² (standard deviation:
7.5 kg/m²).
Fig. 2 Distribution of the interventions.
Comparison of the reference TLD measurements and the EDD-30 measurements
The mean deviation was –7.33 % (95 % tolerance interval for the individual measured
value pairs: 0.678 – 1.175). The hand dose measured with the EDD-30 was slightly lower
than the doses measured with the TLD.
Dose measurements
The measured hand dose values ranged from 0.001 – 3.02 mSv and the median was 0.22 mSv
(mean: 0.315 mSv, standard deviation: 0.391 mSv).
The hand dose median for lung interventions (0.32 mSv; n = 41) was slightly higher
than that of interventions in the region of the liver (0.248 mSv; n = 25) and in the
retroperitoneum (0.17 mSv; n = 53) and the spine (0.014 mSv; n = 19); (p = 0.01).
The maximum single hand dose value (3.02 mSv) was reached during a retroperitoneal
puncture of the pancreas. Without taking the anatomic location into account, the median
hand dose value was 0.23 mSv (0.005 – 3.04 mSv) in the biopsy group (n = 99), 0.26 mSv
(0.027 – 0.979 mSv) in the drainage group (n = 23), and 0.01 mSv (0.001 – 0.077 mSv)
in the pain therapy group (n = 16); (p = 0.011) ([Fig. 3]).
Fig. 3 Different procedures and their hand doses (illustrated median values).
The median fluoroscopy time was 24.2 s (mean: 30.7s; 4.6 – 140.2s; standard deviation:
23.4 s).
The longest median fluoroscopy time of 32.5 s was reached during liver puncture followed
by interventions involving the lungs (25.8 s), retroperitoneum (23.5 s), and spine
(17.9 s). The median scan time was 25.5 s (5.6 – 140.2 s) in the biopsy group, 20.2 s
(4.6 – 66.4 s) in the drainage group, and 17.7 s (8.8 – 48.3 s) in the pain therapy
group; (p = 0.015) ([Fig. 4]).
Fig. 4 Different procedures and their fluoroscopy times (illustrated median values).
The mean dose-length product for fluoroscopy was 123.19 mGy×cm (standard deviation:
104.73 mGy×cm).
Classification according to level of experience and degree of difficulty
The experienced interventionalists (level of experience 2) conducted a total of 97
interventions (64 core biopsies, 18 drainages, 11 periradicular therapies, and 4 celiac
plexus blocks). In contrast, the inexperienced interventionalists (level of experience
1) only performed 42 interventions (36 core biopsies, 5 drainages, and 1 celiac plexus
block). [Fig. 5] shows the classification of the interventions according to the three degrees of
difficulty and the level of experience. The interrater reliability for the classification
according to degree of difficulty (measure of the agreement between the two radiologists)
or the Cohen’s kappa value was 0.739 and is therefore to be evaluated as “considerable
agreement” [16].
Fig. 5 Distribution of the interventions depending on degree of difficulty and level of
experience.
Statistical evaluation of the influencing factors
We used the Spearman-Rho rank order correlation test for the three target variables,
hand dose, fluoroscopy time, and dose-length product (DLP), to analyze the influencing
factors. The following table shows the correlations of important variables and the
associated significance ([Table 1]). The physical relationships of fluoroscopy time, dose-length product, and the mAs
product were confirmed with correspondingly strong correlations (r ≥ 0.9). However,
the fluoroscopy time and the tube current time product only showed a weak linear relationship
to the interventionalist’s hand dose (r = 0.411 and r = 0.418; in each case p = 0.01).
A high degree of difficulty resulted in a longer fluoroscopy time, a greater dose-length
product, and a higher hand dose for the interventionalist (r ≤ 0.344; in each case
p = 0.001). The more often a lesion was punctured, the greater the effect on the hand
dose and the fluoroscopy time (p = 0.023 and p = 0.029). The needle diameter (needle
thickness) showed a significant negative relationship in regard to the hand dose (r = –
0.25; p = 0.003). This can be explained by the fact that thinner biopsy needles were
primarily used for lung interventions which resulted in higher hand doses. The number
of puncture attempts and the summed puncture depth showed significant but only weak
correlations to the fluoroscopy time (p = 0.029; p = 0.001) and hand dose (p = 0.023;
p = 0.004) ([Table 1]). In the Spearman-Rho analysis no significant correlation between the level of experience
and the hand dose could be shown. A significant correlation also couldn’t be shown
via two-factor univariate variance analysis of the level of experience and the degree
of difficulty (p = 0.946). In the case of an average degree of difficulty (2), the
inexperienced interventionalists had a lower average hand dose value (0.313 mSv) compared
to the experienced interventionalists (0.345 mSv); p = 0.45, while the opposite was
true in the case of easy interventions (degree of difficulty 1) (0.213 vs. 0.165 mSv;
p = 0.82). However, the level of experience did have a significant influence with
respect to the fluoroscopy time (p = 0.03). The body-mass index resulted in a higher
tube current setting in six cases but did not have a significant effect on the hand
dose in the analysis (p = 0.308). Multivariate analysis was only able to confirm the
tube current time product (p = 0.001), the degree of difficulty (p = 0.004), and the
summed puncture depth (p = 0.045) as significant influencing factors on the hand dose.
In the biopsy group, the craniocaudal lesion size in particular showed the greatest
correlation with the hand dose (r = – 0.352; p = 0.001) and the fluoroscopy time (r = –
0.258; p = 0.01). The negative correlations indicate that smaller tumor diameters
resulted in a longer fluoroscopy time and a higher hand dose.
Table 1
Spearman-Rho-Analysis regarding influence factors.
|
variable
|
|
DLP
|
fluoroscopy time
|
hand dose
|
|
fluoroscopy time
|
correlation coefficient
|
0.923[1]
|
1.000
|
0.4111
|
|
significance (2-sided)
|
0.001
|
0.001
|
0.001
|
|
mAs
|
correlation coefficient
|
0.9051
|
0.9431
|
0.4181
|
|
significance (2-sided)
|
0.001
|
0.001
|
0.001
|
|
degree of difficulty
|
correlation coefficient
|
0.3291
|
0.2851
|
0.3441
|
|
significance (2-sided)
|
0.001
|
0.001
|
0.001
|
|
level of experience
|
correlation coefficient
|
– 0.110
|
– 0.148
|
0.132
|
|
significance (2-sided)
|
0.201
|
0.084
|
0.123
|
|
needle diameter
|
correlation coefficient
|
– 0.075
|
– 0.014
|
– 0.2501
|
|
significance (2-sided)
|
0.379
|
0.873
|
0.003
|
|
number of punctures
|
correlation coefficient
|
0.155
|
0.186[2]
|
0.1942
|
|
significance (2-sided)
|
0.069
|
0.029
|
0.023
|
|
summed puncture depth
|
correlation coefficient
|
0.3791
|
0.3421
|
0.2471
|
|
significance (2-sided)
|
0.001
|
0.001
|
0.004
|
Significant results are in bold.
1 The correlation is significant (two-sided) at α = 0.01.
2 The correlation is significant (two-sided) at α = 0.05.
Evaluation of the hand dose with respect to the use of a needle holder
The Mann-Whitney test was used to evaluate the reduction in radiation exposure due
to the use of a needle holder with a coaxial needle in retroperitoneum interventions.
This showed a lower median hand dose value of 0.08 mSv (n = 13) compared to a value
of 0.23 mSv (n = 40) without the use of a needle holder. The significance level was
almost met with p = 0.052.
Discussion
Hand dose measurements of the TLD versus the EDD-30
A maximum dose rate of 0.6 mGy/s can be recorded with the immediately readable EDD-30
dosimeter, which does not seem particularly high compared to the dose values in the
direct beam path (up to 10.4 mGy/s) [15]
[17]. For this reason, the 31 simultaneous comparison measurements using TLD plates were
performed. However, identical measured values could not be expected due to the test
setup since the plates were not in the exact same position as the EDD-30 sensor ([Fig. 1]). The slightly lower measured values can be explained by a directional dependence
of the EDD-30 sensor [15]. Other electronic dosimeter models had satisfactory measurement accuracy in clinical
use compared to the TLD but the applied dose can be underestimated in the direct beam
path [18].
Hand dose values
The hand dose median was 0.22 mSv regardless of the procedure and location. This means
that it would take over 2000 CT interventions to reach the annual limit value. In
the study by Buls et al. (2003), the median hand dose for the right hand was 0.76 mSv
which is significantly higher than our measured values [19]. However, our values were significantly higher than those of Stoeckelhuber et al.
(2003) (0.0295 mSv per case) [20]. An explanation for these lower hand dose values is the use of radiation protection
gloves and long needle holders. The biopsy needles of our interventionalists were
controlled by grasping a side wing of the biopsy system and an angled puncture technique
with a “floating CT table” was often used, thus preventing manipulations in the direct
beam path. In contrast to these results, the comprehensive dose measurements involving
interventionalists’ fingers by Pereira et al. (2011) yielded very different results.
Despite the additional use of radiation protection gloves, extremely high dose values
(up to 36.29 mSv) were measured. It must be mentioned that the TLD measurements were
taken at eleven measurement points on the hand (proximal and distal on every finger
and on the wrist). Therefore, the proximity to the direct beam path differed greatly
from our measurement method [21]. The hand dose values measured at the wrist were lower (0.07 – 0.57 mSv) and comparable
to our measurements. Moreover, the back of the hand is usually turned away from the
gantry in our measurement, which also explains the lower hand dose values. The substantial
variation of the measured hand dose values is consequently dependent on manipulations
in the immediate vicinity of the direct beam path and the use of continuous fluoroscopy.
In the direct beam path a dose rate of up to 10.4 mGy/s (120 kV, 90 mA) can be expected.
This value drops to 27 µGy/s (120 kV, 50 mA) at a distance of 10 cm [17]. Moreover, there is no clearly defined fixed point for hand dose measurements so
that there can be significant deviations in the measurements of other authors. More
complex interventions, such as CT-guided brachytherapy or radiofrequency ablation,
were not included in this study. Higher hand dose values would be expected in these
cases.
Influencing factors
In addition to the physical influences to be expected (fluoroscopy time, tube current
time product, needle size, lesion size, and summed puncture depth), the degree of
difficulty of the intervention could also have a major effect on the hand dose. For
the procedures with an easy or average degree of difficulty (biopsy, drainage, and
pain therapy), there was surprisingly no significant difference in hand dose between
the inexperienced and experience interventionalists. A possible explanation for this
could be the greater respect of the inexperienced interventionalists for the direct
beam path or their avoidance of continuous fluoroscopy. Furthermore, level of experience
2 as defined by us (> 5 years of experience) could be too high. However, the inexperienced
interventionalists needed on average approximately 40 % more scan time even though
they only performed one difficult intervention (degree of difficulty 3) ([Fig. 5]).
Biological effect of radiation exposure on the hand
Careless use of X-rays already had significant consequences at the beginning of the
X-ray era. Accordingly, radiodermatitis of hands exposed to radiation (so-called “Roentgen
hand”) was documented for the first time in 1896 [22]
[23].
Deterministic radiation effects are not expected in the case of the measured hand
doses. Since there is no limit value for the stochastic radiation effect, every X-ray
quantum is in principle carcinogenic so that the risk estimate of a chronic suberythemal
hand dose is difficult to determine [24]
[25]. From the Life Span Study of atomic bomb survivors, an increase in the incidence
of basal cell carcinoma due to the additional radiation exposure was able to be confirmed
[26]
[27]. The hand dose values measured in the present study yielded the following risk estimates
(two alternative calculations) on the basis of the ICRP 103 publication [28] and according to Preston et al. [27]: Calculation of the cumulative total hand dose (30 years professional experience):
0.22 mSv (median hand dose per intervention) × 3 (interventions/day) × 250 (work days)
× 30 (years) = 4.95 Sv. The nominal risk coefficient for the skin is 670 cases per
10 000 persons per Sv. This results in 4.95 Sv × 0.067 persons/Sv. Consequently, the
risk of incidence of “non-melanocytic” skin cancer according to these calculations
is 33.1 % (with a very low mortality rate of 0.07 % and a case fatality rate of 0.002
[28]). An alternative calculation according to the relative risk model on the basis of
the Life Span Study analysis (based on Table 27 from Preston et al. 2007 [27]) shows the following: The excess relative risk is 0.17 per Sv resulting in 4.95
Sv × 0.17/Sv = 0.84 as an excess relative risk. This value would have to be multiplied
by the lifetime risk of developing basal cell carcinoma (incidence approximately 100
per 100,000 people × 80 years × 0,84 = 7 % [29]). This would therefore result in a lower additional risk compared to the first calculation
(7 %; mortality rate: 0.014 %, case fatality rate: 0.004 [27]
[28]
[29]
[30]). Despite these calculations of the risk of incidence, the annual hand dose limit
of 500 mSv for persons occupationally exposed to radiation was not reduced by the
German Commission on Radiological Protection or in the amendment of the X-Ray Ordinance
2011. In the statement of the German Commission on Radiological Protection regarding
the topic of “dose limits for occupational skin exposure to ionizing radiation”, the
above calculations are dismissed and surface personal dose measurement is greatly
overestimated [31]
[32].
With the following additional radiation protection tools/measures, the hand dose can
be greatly minimized: Use of a needle holder, fluoroscopy with low tube current intensity
(possibly with incremental adjustment, start thorax: 10 mA; abdomen: 30 mA) or lower
tube voltage (≤ 120 kV), low collimation (≤ 4 mm to lower the scattered radiation),
restrictive use of continuous fluoroscopy, “angular beam modulation”, and, if possible,
use of radiation protection gloves [17]
[20]
[33]
[35]. However, radiation protection gloves can give a false sense of security thus resulting
in higher hand dose values [35].
Conclusion
An interventionalist would have to perform over 2000 CT interventions (at a hand dose
median of 0.22 mSv) to reach the annual hand dose limit value of 500 mSv. In addition
to the physical parameters, the summed puncture depth and the degree of difficulty
have a significant effect on the hand dose. The estimates regarding the additional
radiation risk posed by CT fluoroscopy interventions on the radiologist's hand dose
only allow ambigious conclusions. Additional radiation protection measures or tools
should therefore be used.