Keywords
lupus anticoagulants - guidelines - dilute Russell viper venom time - activated partial
thromboplastin time
Lupus anticoagulants (LA) are somewhat akin to subatomic particles and extrasolar
planets, in that they are “detected” by inference, after exclusion of other possible
causes of test findings. In the diagnostic setting, this makes LA detection both a
challenge and a frustration, and guidelines from various expert groups have been published
and updated over recent decades to outline current opinion on best practice. Problems
such as antibody heterogeneity, reagent and analyzer variability, differing interpretation
strategies, and absence of gold standards continue to conspire against generation
of unequivocally ideal diagnostic strategies. Consequently, it is perhaps unsurprising
that incomplete agreement exists between contemporaneous groups and their guidelines,
even with some degree of author crossover. Indeed, it has been suggested that consensus
guidelines on antiphospholipid antibody testing are particularly prone to biases inherent
the process, mainly because they rely heavily on expert opinion.[1]
The International Society on Haemostasis and Thrombosis (ISTH) has updated their guidelines
in late 2009[2] (ISTH 2009), and the British Committee for Standards in Haematology (BCSH) has published
an update of their previous guidelines in early 2012[3] (BCSH 2012). The Clinical and Laboratory Standards Institute (CLSI) will publish
its first LA guideline in 2014[4] (CLSI 2014). In this article, blood samples will be followed on their journey through
the convoluted maze of preanalytical variables, testing with different reagents and
the myriad issues impacting on interpretation, with reference to main commonalities
and contrasts from the above guidelines, which are summarized in [Table 1].
Table 1
Summary of commonalities and contrasts between recent ISTH, BCSH, and CLSI guidelines
for lupus anticoagulant detection
Area of recommendation
|
ISTH 2009
|
BCSH 2012
|
CLSI 2014
|
Sample preparation
|
Double centrifugation
|
Double centrifugation
|
Double centrifugation
|
Assays to use
|
dRVVT and aPTT
|
dRVVT plus aPTT or others
|
dRVVT and aPTT and/or others
|
Testing order
|
Screen-mix-confirm
|
Screen-mix-confirm
|
Screen-confirm-mix
|
Ratio derivation
|
NPP denominator
|
NPP denominator
|
RI mean denominator
|
RI/cutoffs
|
99th percentile
|
97.5th percentile (if Gaussian)
|
97.5th percentile (if Gaussian)
|
Calculations for phospholipid-dependence
|
% correction of screen by confirm, or
LA ratio (screen/confirm)
|
% correction of screen by confirm, or
LA ratio (screen/confirm)
|
% correction of screen by confirm, or
LA ratio (screen/confirm)
|
Mixing test
|
Perform on 1:1 mixture with NPP;
Interpret with ICA or mixing test-specific cutoff
|
Perform on 1:1 mixture with NPP
|
Perform on 1:1 mixture with NPP;
Interpret with ICA or mixing test-specific cutoff
|
Testing patients on VKAs
|
Undiluted plasma if INR < 1.5;
Mix with NPP if INR > 1.5 < 3.0
|
Screen and confirm on 1:1 mixture with NPP;
TSVT + ET or PNP
|
Screen and confirm on 1:1 mixture with NPP;
TSVT + ET or PNP
|
Testing patients on UFH
|
Interpret with caution
|
Not recommended
|
Can detect LA in some cases where heparin neutralizer is effective
|
Interpretive reporting
|
Recommended
|
Recommended
|
Recommended
|
Abbreviations: aPTT, activated partial thromboplastin time; BCSH, British Committee
for Standards in Haematology; CLSI, Clinical and Laboratory Standards Institute; dRVVT,
dilute Russell viper venom time; ET, Ecarin time; ICA, index of circulating anticoagulant;
INR, international normalized ratio; ISTH, International Society on Thrombosis and
Haemostasis; LA, lupus anticoagulant; NPP, normal pooled plasma; PNP, platelet neutralization
procedure; RI, reference interval; TSVT, Taipan snake venom time; UFH, unfractionated
heparin; VKA, vitamin K antagonist.
Preanalytical Variables
Residual platelet material in test plasma can shorten clotting times and mask the
presence of LAs, particularly if plasma is frozen/thawed before analysis.[5] There is full consensus that blood for LA testing, collected into 0.109 mol/L trisodium
citrate, must be rendered platelet poor by a process of double centrifugation, to
achieve a final platelet count of < 10 × 109/L. CLSI 2014 indicates that this may potentially be achieved after the first centrifugation
step; however, this approach should be validated by the laboratory, and the second
centrifugation step is necessary if the requisite platelet reduction is not achieved
by a single centrifugation. In practice, this will depend on many variables, including
centrifugation type and speed, as well as operator technique. Thus, a single centrifugation
step may be inefficient for achieving the target of < 10 × 109/L in all samples.[6] The guidelines also differ on the recommended centrifugation technique, with ISTH
2009 recommending differential speeds on first and second spins. BCSH 2012 and CLSI
2014 discourage ultracentrifugation (> 5,000g) in the second step because it may generate microparticles.[5]
[7] The previously fashionable filtration through 0.22 μm cellulose acetate filters
is no longer recommended because it introduces variables such as loss of von Willebrand
factor and some other coagulation factors.[8]
Preliminary routine coagulation screening, including prothrombin time, activated partial
thromboplastin time (aPTT), and thrombin time, are crucial before LA testing to exclude
undiagnosed coagulopathies or undisclosed anticoagulant therapy. However, local regulations
may prohibit some laboratories from undertaking testing on patient samples additional
to that specifically requested by the clinician. This creates diagnostic challenges
and clinicians should be encouraged to order these tests simultaneously to requesting
LA testing. CLSI 2014 further recommends employing a LA-unresponsive aPTT reagent
in routine hemostasis practice to reduce serendipitous findings of LA in asymptomatic
patients, and permit interpretation of more sensitive assays unencumbered by the possibility
of alternative abnormalities if coagulation screening is normal. Thus, LA assay data
can be interpreted at face value without the need to consider the presence of abnormalities
that may mask, mimic, or coexist with LAs.
BCSH 2012, which incorporates clinical management guidelines, recommends that primary
thromboprophylaxis is not used for patients with incidentally detected antibodies
because thrombosis rates are low, even with persistent antibodies.[9]
[10]
[11] ISTH 2009 assigns moderate appropriateness to further testing for LA when an elevated
routine aPTT is “accidentally” encountered in asymptomatic patients. Interestingly,
a recent study revealed that asymptomatic patients positive for all three of LA, anticardiolipin
antibodies, and anti-β2 glycoprotein I antibodies are at high risk of incurring a thromboembolic event,[12] and it may be that this debate already requires revisiting.
Generating Reference Intervals/Cutoff Levels
Generating Reference Intervals/Cutoff Levels
Reagent variability necessitates use of locally derived reference intervals (RIs)
specific to the reagent-analyzer pairings in use.[13]
[14]
[15] In common with many other analytes, RIs for LA assays have historically been derived
from the RI mean ± 2 standard deviations (SD). This is not inappropriate since data
from normal donor populations for clotting tests are commonly Gaussian or near Gaussian,
or can be made so by data transformation,[14]
[16] although this requires confirmation with each assessment to enable data to be suitable
for parametric statistics.[17] The upper limit operates as the cutoff (97.5th percentile) for determining positivity
and initiating confirmatory tests, while the RI mean clotting time can be employed
to generate normalized ratios.
ISTH 2009 recommends application of the 99th percentile to determine the cutoff, which
equates to the RI mean + 2.3 SD for normally distributed data. This introduces potential
to reduce the frequency of false-positive results and thereby increase specificity,
yet it is a statistical inevitability that this will reduce sensitivity. The recommendation
has proven controversial, partly because the guideline states that a 99th percentile
value can be derived from a minimum of 40 donors, yet this is not the case and a minimum
of 120 has been previously recommended.[18] An even greater number of normal individuals is required to generate a 99th percentile
value from nonnormally distributed data. BCSH 2012 discusses inaccuracy of RI generation
with small sizes, which impinges whether you attempt to calculate 97.5th or 99th percentile.
More practical advice is given in the form of validating previously established cutoffs,
such as from the reagent manufacturer or from a different analyzer, using just 20
to 60 normal donors. Although accepting that no statistical tool is perfect, CLSI
2014 maintains that generating a RI from its mean ± 2 SD remains a valid and achievable
proposition for diagnostic laboratories. Reference is made to a separate CLSI publication
on clinical laboratory RIs,[18] echoing the BCSH 2012 advice of validating established cutoffs with reduced donor
numbers. An important point is made that an elevated LA screening test in isolation
is not a false-positive result but merely a prolonged clotting time requiring confirmatory
tests to reveal whether or not the initial abnormality is due to a LA. Statistical
outliers or other causes of prolonged clotting times will usually generate an equally
elevated confirmatory test result for that assay, and crucially, not generate a false-positive
interpretation. A false-positive screening test obtained from the use of the 97.5th
percentile will not generate false-positive composite interpretation, yet a false-negative
resulting from application of the 99th percentile could lead to failure to secure
appropriate diagnosis and treatment.
Which Screening Tests to Use and How Many to Use?
Which Screening Tests to Use and How Many to Use?
It has long been recognized that no single assay system will detect all LAs and therefore
multiple assays, using differing analytical principles, are required to achieve acceptable
detection rates.[19]
[20] However, ISTH 2009 indicates that the risk of false-positive results is increased
to unacceptable levels if more than two screening tests are performed, and restricts
assay choice to only dilute Russell viper venom time (dRVVT) for its specificity and
aPTT with low phospholipid concentration because of its sensitivity. ISTH 2009 further
supports this view according to potential inconsistency between techniques used for
additional test methods. Although contentious, these viewpoints are not without logic
and warrant further consideration. Aside from the logistics and costs of performing
numerous tests on every patient, it is likely that some patients will generate an
elevated screening test with at least one test/reagent type, with the chance on this
increasing the more tests performed. Such a result could be due to a genuine LA unreactive
in other reagents, a “weak” LA, a discrete analytical error or merely because the
patient is a natural statistical outlier for that reagent/analyzer pairing as for
conventional laboratory testing. Here, it is important to reflect that to identify
a cutoff, be it 97.5th or 99th percentile, some normal individuals (2.5 or 1% for
these examples) will be outliers. As different normal individuals may be outliers
in different tests, there is an additive effect for “false-positive” identification.
As a statistical example, performance of 10 screening tests, with only one needing
to be positive for potentially defining positivity for the condition, could theoretically
identify 12.5% of tested (normal) individuals as being positive for the condition
using a 97.5th percentile cutoff. In addition, ethnic differences have also been described
for dRVVT ratios that can lead to apparent false-positive screening tests.[21] These non-LA examples, however, would not normally lead to false-positive composite
interpretation, for the reasons described above, and the issue becomes more that of
performing unnecessary confirmatory tests.
The ISTH 2009 recommendation has its basis in the considerable body of evidence indicating
that pairing dRVVT and aPTT achieves good detection rates, which additionally serves
to nurture common diagnostic practices. CLSI 2014 recommends dRVVT and LA-responsive
aPTT (LAR-aPTT) as the first-line screening assays but does not exclude employing
additional tests of different principles during initial analysis or as second-line
assays. BCSH 2012 indicates that dRVVT should be one of two tests used in LA screening,
the other being an aPTT with proven LA sensitivity, modified aPTT or a dilute prothrombin
time (dPT). Parity of recommendation for dRVVT is not surprising because this test
procedure has repeatedly been shown to be sensitive to β2 glycoprotein I-dependent antibodies and highly correlate with thrombosis.[22] Familiarity, ease of performance, and cost are likely to ensure aPTT remains the
most common partner test for dRVVT. Nevertheless, it is important to recognize that
considerable variability in diagnostic performance exists for reagents from different
manufacturers for both tests.[2]
[4]
[5]
[13]
[14]
[17]
[20] In addition, although most commercial dRVVT reagents behave similarly, potentially
suggesting common manufacturing practices, occasional deviations in behavior occur.[23]
The different wording to describe “suitable” aPTT reagents is interesting. ISTH 2009
is the only document to specify dilute phospholipid, whereas phrasing in the other
two guidelines can encompass routine aPTT reagents without dilute phospholipid but
recognized to be (relatively) LAR. Both are potentially fallible. Diluting a phospholipid
preparation of LA-unresponsive composition may not perform as well as an undiluted
LAR preparation,[19]
[24] yet the latter risks missing weaker antibodies if the phospholipid is too concentrated.[22]
[25] CLSI 2014 deliberately avoids the use of “LA-sensitive/LA-insensitive” terminology,
based on the concept that “LA sensitivity” is difficult to appropriately define due
to the lack of any “gold standard” LA detection method. Finally, although the dRVVT/aPTT
pairing cannot guarantee detection of all antibodies, external quality assurance scheme
reports continue to validate the pair as an optimal approach sufficient for LA detection/exclusion
in most samples assessed.[26]
[27]
The Case for Other Tests
That two[3]
[4] of three expert guideline panels promote tests other than dRVVT and aPTT warrants
further coverage. This stems from not inconsiderable experience of their application
in the diagnostic setting and recognition that these assays can detect clinically
significant antibodies undetectable in dRVVT and aPTT as well as many that are.[28]
The kaolin clotting time (KCT) is often quoted as being one of the most sensitive
assays for LA because of its extremely low phospholipid concentration.[5] Technical difficulties such as meticulous sample handling, long clotting times,
and variation in residual lipid mean that reproducibility is poor relative to other
tests and these issues provide the reasons that ISTH 2009 does not recommend this
assay. Being unrestricted in length, the CLSI 2014 guideline considers each available
assay in more detail, and also covers best practice and limitations. In experienced
hands, the KCT is deemed a sensitive LA assay, providing that results are interpreted
in light of assay design, which of course holds true for all assays. Bearing in mind
that all recent guidelines emphasize the necessity of confirmatory assays, an important
limitation for the KCT is that it lacks a widely available confirmatory test, thereby
compromising specificity by not fulfilling an indispensable diagnostic criterion.[28]
The silica clotting time (SCT) was first described as a similar test to KCT, with
the advantage of being compatible with automated analyzers employing photo-optical
clot detection. ISTH 2009 does not recommend kaolin as activator in aPTT-based tests
because its particulate nature compromises this detection principle and requires constant
mixing to prevent rapid sedimentation. Low-opacity/slow-settling KCT reagents are
in fact commercially available, although sample-related technical difficulties and
the lack of confirm test issue persists. SCT was subsequently revised as a paired
system with dilute phospholipid in the screen and concentrated phospholipid in the
confirm,[29] ostensibly converting it into an automate-friendly dilute aPTT. This paired system
has demonstrated good diagnostic performance when used in conjunction with a dRVVT
system.[29]
[30] ISTH 2009 also discourages the use of aPTT reagents employing ellagic acid as the
contact activator, citing insensitivity to LA. However, inconsistent sensitivity of
aPTT reagents to LA is largely due to variation in phospholipid concentration or composition
and ellagic acid-containing reagents can be suitable for LA detection.[17]
[25]
[31] Consequently, neither BCSH 2012 nor CLSI 2014 restricts aPTT reagents based on activator
alone. The BCSH 2012 recommendation to employ a LAR-aPTT or modified aPTT can reasonably
be taken to include use of KCT, SCT, or dilute aPTT.
dPT is no longer endorsed by ISTH because of the variability in thromboplastin reagents.
That premise is particularly true for tissue-derived reagents although LA sensitivity
and reagent variability are improved when recombinant reagents are employed.[5]
[22]
[28] Several studies have shown that dPT detects clinically significant antibodies, some
unreactive with dRVVT and aPTT.[28]
[32]
[33]
[34]
[35] A persistent problem has been that different laboratories adopt a variety of modifications
with their locally employed thromboplastin, such as the dilutions employed for screen
and confirm tests, resulting in significant variability in diagnostic performance.[36] The relatively recent availability of a standardized commercially available reagent
kit has led to the suggestion that LA detection is improved when dRVVT and aPTT are
accompanied by dPT.[34]
[37]
Assays based on snake venoms such as Textarin and Ecarin are not recommended by ISTH
2009 as standardized commercial assays are said to be unavailable. Although this is
currently true for Textarin, preparations of Taipan and Ecarin venom for use in LA
assays have been available for some years from at least one manufacturer.[38]
The variation in epitope specificities, even for antibodies to domain I of β2 glycoprotein I,[39] contributes to the fact that the dRVVT and aPTT combination alone cannot deliver
diagnostic certainty. New or alternative assays are commonly evaluated against dRVVT
and/or aPTT, and are thus disadvantaged because they may be sensitive to different
antibody subpopulations. Comparison against the dRVVT/aPTT combination ostensibly
creates a selection bias. The less commonly employed assays can detect LA that will
not manifest in dRVVT and aPTT, and at least some of these antibodies are clinically
significant.[28]
[32]
[33]
[34]
[35]
[40]
[41]
[42]
[43] It is impractical and probably unnecessary for all laboratories to adopt extensive
LA-assay repertoires, but there is a case for the use of additional assays in select
patients or circumstances.[3]
[4]
[28]
Mixing Tests
There is a clear continuum of views across the three guidelines concerning mixing
tests. ISTH 2009 advocates performing the mixing test immediately after finding an
elevated screening test. It is implicit that the confirmatory test is only undertaken
if the mixing test reveals inhibition. BCSH 2012 indicates that mixing tests improve
specificity, but introduce a dilution factor that may make weak antibodies appear
negative. By way of clarification, it states that in the absence of other causes of
prolonged clotting times, samples with negative mixing tests but clear positive screen
and confirm tests on undiluted plasma can be considered LA positive.[44]
[45]
[46] It is therefore implicit that all three of the screen, mixing test, and confirmatory
test medley are undertaken.
CLSI 2014 goes further and reprioritizes the order of testing to screen, confirm,
and mix, the latter only being undertaken if initial testing is not clear cut because
its limitations can lead to false-negative reporting.[5]
[20]
[22]
[28]
[38]
[44]
[45]
[46]
[47]
[48]
[49]
[50] Omitting the mixing test is recommended only where there is no evidence of other
causes of elevated clotting times, which can be obtained in many cases from the CLSI
2014 recommendation of performing a routine coagulation screen that includes a LA-unresponsive
aPTT. In addition, the confirmatory test (on undiluted plasma) must not only mathematically
correct an elevated screening test result but should also return into the RI.[51] Otherwise, the elevated results could derive from an abnormality undetected in coagulation
screen, or be a LA sufficiently potent to achieve some degree of resistance to the
swamping effect of confirm reagent. In such circumstances, the mixing test limitation
can be advantageous if the dilution reduces the antibody's effect and facilitates
overwhelming it with the confirm reagent. Where a coexisting abnormality is suspected,
performing screen and confirm tests on 1:1 mixtures with normal pooled plasma (NPP)
to correct it can reveal the presence of a LA. All guidelines agree that mixing tests
should be performed on a 1:1 mixture of test plasma and NPP.
No guideline before ISTH 2009 gave definitive advice for interpreting mixing tests,
so it was especially welcome that this publication fully addresses the issue. Two
options are given. One entails using a mixing test-specific RI, which will be narrower
than that for undiluted plasma, and thus have a lower cutoff due to clotting times
of normal samples at the extremes being compensated upon mixing with NPP.[22]
[46] The lower cutoff increases sensitivity in comparison to assessing against the cutoff
for undiluted plasma. The alternative is application of the index of circulating anticoagulant
(ICA)[52] as shown below:
The ICA is advantageous because it assesses test plasma clotting times relative to
the NPP in which it is diluted. A mixing test-specific cutoff can do so if applied
to ratios and not to clotting times. CLSI 2014 concurs with these recommendations,
whereas BCSH 2012 does not cover the issue. However, no mathematical method for assessing
mixing test data can perfectly discriminate between factor deficiencies and inhibitors,
and cutoffs vary between test types and reagents.[53]
[54]
Ratios
It is common practice to convert clotting times for screen and confirm tests to normalized
ratios via NPP values. This improves inter- and intra-assay variation by mitigating
differences in operator and/or analyzer performance, reagent quality and stability
issues, and NPP clotting time variation with different reagents.[13]
[14]
[55] Ratios also compensate for the fact that NPP clotting times with paired screen and
confirm reagents can differ.[20] BCSH has recommended this practice since 1991[3]
[20]
[56] and it is adopted by ISTH 2009.
The CLSI 2014 approach differs, in that the recommendation is to normalize against
the RI mean clotting time and not against the NPP value. The rationale here is that
not all NPPs, or batches of the same NPP, generate the same clotting times with different
reagents for the same test type,[13]
[14]
[55]
[57] so that false-positive and false-negative results can ensue if an NPP value is toward
or beyond an extreme of a RI.[55] Establishing the RI and its mean by analyzing normal donors over several days accounts
for innate technique variability,[18] and running an NPP as a normal control will identify sudden analytical difficulties.
The use of a mean RI value also simplifies the process of automated calculations,
for example, in laboratory information systems.
Confirmatory Tests
All three guidelines agree that confirmatory tests used to demonstrate phospholipid
dependence must be based on the screening test(s) that was abnormal. It is not explicitly
stated in ISTH 2009, yet this is an update of ISTH 1995[19] where it is a clear recommendation and thus remains current. This is a crucial point,
as some laboratories apply only the mixing test in response to an elevated aPTT and
are at risk of both loss of specificity and increased false-negative interpretations
unless dRVVT testing is also positive.[44]
[45]
[46]
[47]
[48]
[49]
[57]
[58]
ISTH 2009 indicates that confirmatory tests must be performed by increasing the phospholipid
concentration, with bilayer or hexagonal (II) phase phospholipid. BCSH 2012 does not
specify the type, and suggests employing high phospholipid concentration, platelet
neutralization procedure (PNP), or LA-insensitive reagent. ISTH 2009 advises against
freeze/thawed platelets because of reagent batch inconsistencies, which perhaps should
have specified that this relates particularly to locally prepared reagents.[59] All confirmatory tests are versions of their parent screening test rendered LA unresponsive
and the specific reference to such reagents in BCSH 2012 alludes to techniques pairing
LAR and LA-unresponsive aPTT reagents. CLSI 2014 gives details of recommended available
confirmatory tests for each screening test, including considerations of limitations,
and ostensibly concurs.
Another welcome first for ISTH is a recommended calculation to demonstrate or exclude
phospholipid dependence, the percentage correction of screen by confirm, as shown
below for ratios:
BCSH has championed this calculation since 1991, with a correction of 10% or more
being indicative of the presence of a LA, whereas ISTH 2009 indicates that the cutoff
should be locally derived. CLSI 2014 offers alternatives specific to particular assays.
The hexagonal phase phospholipid neutralization test and PNP confirmatory test commonly
used for aPTT are assessed using deltas,[60]
[61] while the normalized screen to confirmatory ratio, as below, is recommended for
dRVVT, SCT, and dPT when used with paired screen and confirm reagents. BCSH 2012 suggests
that this as an alternative to percentage correction of ratio, and CLSI 2014 suggests
the latter as another option for dRVVT.
Integrated Testing
Integrated testing involves performing screen and confirm assays (i.e., low and high
phospholipid concentration) on every patient, and then mathematically assessing for
correction irrespective of whether the screen is elevated beyond a predetermined cutoff.[29]
[50]
[51]
[62]
[63] Detection of weaker LAs is a particular advantage of this approach, where the prolongation
of a patient's basal clotting time is insufficient to exceed the upper reference limit
of a screening test, yet the screen and confirm discrepancy reveals the antibody.[49]
[64]
[65] Some integrated tests are performed on a 1:1 mixture with NPP to simultaneously
correct factor deficiencies and demonstrate inhibition. However, they inevitably risk
diluting some antibodies such that even screen and confirm discordance does not manifest
because the dilution compromises screening test elevation such that a significant
difference is not apparent.[49]
[65] CLSI 2014 reserves its definition of integrated tests for those that incorporate
dilution in NPP, while ISTH 2009 includes those performed on undiluted test plasma
and a separate mixing test. The specifics are not covered in BCSH 2012.
The Integrated Test and Mixing Study Debate
A statement in ISTH 2009 that performing screen and confirm tests in parallel on undiluted
plasma does not, in principle, require subsequent performance of a mixing test and
has prompted useful and revealing debate.[48]
[49]
[51]
[62]
[63]
[66] The principle arises from the fact that the traditional testing order of screen-mix-confirm
is circumvented and the presence of a LA can be immediately apparent without performing
the mixing test. Some have interpreted that statement to mean that mixing tests are
no longer mandated by ISTH if integrated testing is undertaken and expressed concerns,[48]
[63] resulting in clarification that this was not the intended message.[66]
LAs with a degree of resistance against the “overwhelming” effect of confirmatory
reagents can lead to a confirm result above the confirmatory test RI. Many of these
will nonetheless have generated a screen result sufficiently higher than the confirm
to achieve a positive interpretation.[48]
[51] However, the possibility of a coexisting or alternative abnormality must be considered,
and performance of both screen and confirm on 1:1 mixtures with NPP can aid discriminating
between potent LA, LA + coexisting abnormality, or non-LA abnormality.[20]
[38]
[51] There are occasions when screen and confirm for a genuine LA are elevated to a similar
extent such that integrated testing on undiluted plasma generates a false-negative
interpretation, implying a nonphospholipid-dependent abnormality. Again, screen and
confirm mixing tests can be invaluable in this scenario by reducing the antibody's
effects and revealing a screen and confirm discrepancy.[48]
[51]
[62]
Although considered rare, the LA cofactor effect cannot, by definition, be demonstrated
without mixing tests, and integrated testing alone will not detect this phenomenon.[62]
[66] Personal experience suggests that these LAs are usually evident in undiluted plasma,
and if not clear cut, will receive mixing tests anyway. Such samples can be problematic
for those laboratories that do not perform mixing studies.[67]
It has also been demonstrated that test and reagent type can influence the diagnostic
efficacy of mixing tests.[63]
[68] Assays with higher specificity are less likely to require mixing tests to clearly
demonstrate the presence of a LA.
The CLSI 2014 recommendations for when mixing tests can be omitted or not, crystallize
this debate, and they affirm that the decision can only be made on a case-by-case
basis. It is not a choice of performing mixing tests on everyone or no one, it is
a case of scrutinizing raw data to see if mixing tests will tell you anything you
do not already know.
Testing Anticoagulated Patients
Testing Anticoagulated Patients
All three guidelines concur that testing for LAs in patients receiving vitamin K antagonists
(VKA) and therapeutic doses of unfractionated heparin (UFH) should be postponed until
anticoagulation has been discontinued for a suitable period of time. BCSH 2012 indicates
that brief discontinuation of VKA therapy is not a high-risk strategy in most cases,
where decisions on advisability of long-term anticoagulation are required. Bridging
with low-molecular-weight heparin (LMWH) can be employed. However, the reality for
diagnostic laboratories is that they are often asked to test for LAs in anticoagulated
patients, often without knowing a patient is receiving them until laboratory results
suggest so, and guidance is required.
Vitamin K Antagonists
ISTH 2009 indicates that dRVVT and aPTT can be performed on undiluted plasma if the
international normalized ratio (INR) is less than 1.5. If the INR is between 1.5 and
3.0, undertaking the tests on 1:1 mixtures with NPP can be considered, with recognition
of the dilution effect. Integrated tests and procedures employing “VKA-insensitive”
snake venoms are not recommended as it is suggested that they require further critical
evaluation.
BCSH 2012 and CLSI 2014 advocate performing screen and confirm tests on 1:1 mixtures
with NPP but do not restrict this practice to plasmas with INRs ≤ 3.0. Careful data
interpretation of mixing tests has been shown to reveal LAs in some patients with
INRs > 3.0.[38] BCSH 2012 states that an elevated screen on the mixture can be taken as grounds
to suspect an inhibitor and the confirm will demonstrate phospholipid dependence.
Both documents make an important point that negative testing in mixing studies on
these patients does not exclude the presence of a LA because of the dilution effect.
CLSI 2014 gives worked examples of a probable negative and definite positive set of
results. Testing undiluted plasma can lead to false-positive or false-negative results,
and is thus not recommended,[13]
[17]
[20]
[69] although some workers contend that it can be reliable.[70]
[71] Both documents suggest that Taipan snake venom time (TSVT) can be a useful secondary
screening test for patients on VKAs, accompanied by Ecarin time (ET) or a PNP as confirmatory
test.[20]
[28]
[33]
[38]
[41]
[72]
[73]
Heparins
Both ISTH 2009 and CLSI 2014 urge caution when interpreting results on patients receiving
UFH, while BCSH 2012 specifies that LA testing should not be undertaken on patients
receiving therapeutic doses as it may cause erroneous results. CLSI 2014 discusses
reagent-integral heparin neutralizers (as found in most commercial dRVVT reagents)
in more detail, indicating that they quench heparin only up to a specified level,
and it must be demonstrated that this has occurred before result interpretation is
undertaken. It also discusses heparin-neutralizing compounds that are added to plasma
before analysis and gives worked examples of unquenched UFH and a genuine LA after
full UFH neutralization. BCSH 2012 and CLSI 2014 indicate the possibility of false-positive
interpretations when a PNP is used as a confirmatory test. LMWH has less effect on
LA assays, particularly dRVVT reagents, as the heparin neutralizers are usually capable
of covering prophylactic doses.
New Anticoagulants
Only CLSI 2014 gives evidenced detail of the potential interferences by the newer
anticoagulants, information on which is still emerging. Direct thrombin inhibitors
(DTI), such as dabigatran and argatroban, interfere with all LA assays and introduce
a significant risk of false-positive results.[17]
[74] Although the multiple acquired factor deficiency of VKA anticoagulation can often
be corrected in a mixing test and reveal a LA, there is no such correction with DTIs
or factor Xa (FXa) inhibitors as there is usually no underlying factor deficiency.[17] The FXa inhibitor rivaroxaban interferes with dRVVT analysis for LA more so than
aPTT and has a variable, reagent-dependent effect on prothrombin time-based assays.[17]
[75]
[76]
[77] TSVT coupled with ET has been shown to be a sensitive test for detecting LAs in
patients receiving rivaroxaban because both venom fractions are prothrombin activators
and thus unaffected by FXa inhibition.[76]
Reporting
All three guidelines state that the numerical results should be accompanied by an
interpretation indicating whether the findings are consistent with the presence or
absence of an LA (or more correctly the inability to detect an LA). ISTH 2009 discourages
use of comments such as “borderline” while CLSI 2014 suggests that the use of “indeterminate”
can be useful where a panel of assay results does not permit clear distinction between
LA detection or lack thereof. Each document supports the recommendation to retest
in no less than 12 weeks when a LA is initially detected.
Conclusion
Commonalities between the three guidelines show that we have come a long way toward
wider understanding, consensus, and common practices since guidelines were first published
over two decades ago. Few would argue against inclusion of dRVVT in any diagnostic
repertoire and we have in depth understanding of performance and limitations of all
tests available to us. The ISTH 2009 recommendation to employ only dRVVT and aPTT
is not surprising since there is a wealth of evidence, unlikely to be surpassed, that
this pairing constitutes a sensitive and specific diagnostic strategy. BCSH 2012 and
CLSI 2014 consider that there is sufficient evidence for other tests that they should
not be discounted in their entirety. Proponents of these assays support them on the
basis of published evidence and direct clinical experience.[1] There is more debate to be had about the place of mixing tests in the LA analytical
hierarchy, although it does seem that they can be omitted in certain circumstances.
The immediate future probably holds little more than refinements of what we already
have until such time as the pathological antibodies causing the in vitro LA phenomenon
have been fully characterized. If and when those assays are available to diagnostic
laboratories, the days of challenging, fascinating, and frustrating searches for ghosts
may be done.