Assessment of the Hepatic Veins in Poor Contrast Conditions using Dual Energy CT: Evaluation of a Novel Monoenergetic Extrapolation Software Algorithm
Beurteilung von Lebervenen bei schwachen Kontrastbedingung in der Dual-Energy-CT: Evaluation eines neuartigen monoenergetischen ExtrapolationsalgorithmusCorrespondence
Publication History
20 December 2013
21 March 2014
Publication Date:
22 April 2014 (online)
Abstract
Purpose: To evaluate a novel monoenergetic post-processing algorithm (MEI+) in patients with poor intrahepatic contrast enhancement.
Materials and Methods: 25 patients were retrospectively included in this study. Late-phase imaging of the upper abdomen, which was acquired in dual-energy mode (100/140 kV), was used as a model for poor intrahepatic contrast enhancement. Traditional monoenergetic images (MEI), linearly weighted mixed images with different mixing ratios (MI), sole 100 and 140 kV and MEI+ images were calculated. MEI+ is a novel technique which applies frequency-based mixing of the low keV images and an image of optimal keV from a noise perspective to combine the benefits of both image stacks. The signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) of the intrahepatic vasculature (IHV) and liver parenchyma (LP) were objectively measured and depiction of IHV was subjectively rated and correlated with portal venous imaging by two readers in consensus.
Results: MEI+ was able to increase the SNR of the IHV (5.7 ± 0.4 at 40keV) and LP (4.9 ± 1.0 at 90keV) and CNR (2.1 ± 0.6 at 40keV) greatly compared to MEI (5.1 ± 1.1 at 80keV, 4.7 ± 1.0 at 80keV, 1.0 ± 0.4 at 70keV), MI (5.2 ± 1.1 M5:5, 4.8 ± 1.0 M5:5, 1.0 ± 3.5 M9:1), sole 100 kV images (4.4 ± 1.0, 3.7 ± 0.8, 1.0 ± 0.3) and 140 kV images (2.8 ± 0.5, 3.1 ± 0.6, 0.1 ± 0.2). Subjective assessment rated MEI+ of virtual 40 keV superior to all other images.
Conclusion: MEI+ is a very promising algorithm for monoenergetic extrapolation which is able to overcome noise limitations associated with traditional monoenergetic techniques at low virtual keV levels and consequently does not suffer from a decline of SNR and CNR at low keV values. This algorithm allows an improvement of IHV depiction in the presence of poor contrast.
Key points:
• The evaluated new image-based algorithm for virtual monoenergetic imaging allows calculating low virtual keV images from dual energy datasets with significantly improved contrast-to-noise ratios.
• The image based novel monoenergetic extrapolation algorithm applies frequency-based mixing of the low keV images and an image of optimal keV from a noise perspective to combine the benefits of both image stacks.
• When compared to traditional monoenergetic images, the novel monoenergetic algorithm has improved contrast-to-noise ratios for both low and high virtual keV images.
• Contrast-enhanced dual energy images with poor contrast conditions can be significantly improved, e.g. late phase imaging of the liver.
Citation Format:
• Schabel C, Bongers M, Sedlmair M et al. Assessment of the Hepatic Veins in Poor Contrast Conditions using Dual Energy CT: Evaluation of a Novel Monoenergetic Extrapolation Software Algorithm. Fortschr Röntgenstr 2014; 186: 591 – 597
#
Zusammenfassung
Zielsetzung: Die Evaluierung eines neuartigen monoenergetischen Nachverarbeitungsalgorithmus (MEI+) bei Patienten mit eingeschränkter intrahepatischer Kontrastierung.
Material und Methoden: 25 Patienten wurden retrospektiv in der Studie eingeschlossen. Computertomografien des oberen Abdomen in später Kontrastmittelphase, welche im Dual-Energy-Modus (100/140 kV) akquiriert wurden, wurden als Modell für eine schwache intrahepatische Kontrastierung verwendet. Es wurden linear gewichtete Mischbilder (MI), traditionelle monoenergetische (MEI), reine 100 und 140 kV sowie MEI+ Bilder berechnet. MEI+ ist eine neuartige Technik, welche frequenzbasiert rauschbehaftete virtuelle niedrig-keV Bilder mit Bildern optimaler keV (aus Perspektive des Bildrauschens) kombiniert und so die Vorteile beider Bilddatensätze vereint. Signal-zu-Rauschen (SNR) und Kontrast-zu-Rauschen (CNR) intrahepatischer Gefäße (IHV) und des Leberparenchyms (LP) wurden objektiv gemessen. Subjektiv wurde die Abgrenzbarkeit der IHV von zwei Untersuchern im Konsensus beurteilt und mit portal-venösen Bildern verglichen.
Ergebnisse: MEI+ verbesserte SNR von IHV (5,7 ± 0,4 bei 40keV) und LP (4,9 ± 1,0 bei 90keV) sowie CNR (2,1 ± 0,6 bei 40keV) erheblich, verglichen mit MEI (5,1 ± 1,1 bei 80keV; 4,7 ± 1,0 bei 80keV; 1,0 ± 0,4 bei 70keV), MI (5,2 ± 1,1 M5:5; 4,8 ± 1,0 M5:5; 1,0 ± 3,5 M9:1), reinen 100 kV Bildern (4,4 ± 1,0; 3,7 ± 0,8; 1,0 ± 0,3) und 140 kV Bildern (2,8 ± 0,5; 3,1 ± 0,6; 0,1 ± 0,2). Die subjektive Beurteilung schätzte MEI+ bei 40 keV gegenüber allen anderen Bildern als überlegen ein.
Schlussfolgerung: MEI+ ist ein vielversprechender Algorithmus für die monoenergetische Extrapolation, welcher die Limitation eines hohen Bildrauschens bei niedriger virtueller keV überwindet und keine Abnahme von SNR und CNR bei niedrigen keV verursacht. MEI+ ermöglicht eine signifikante Verbesserung der IHV-Darstellung bei eingeschränkter Kontrastbedingung.
#
Introduction
The reliable visualization of the hepatic vasculature is essential for liver imaging and pre-interventional planning but can be challenging with CT due to a variety of reasons. For example, in patients with acute or chronic liver disease and severely altered hemodynamics of the portal venous system, the timing of optimal contrast enhancement can be difficult [1] [2]. Furthermore, renal dysfunction accompanied by the need for saving contrast media can result in poor portal venous contrast enhancement [3] [4], thus demanding optimization of examination protocols or improved post-processing capabilities.
Dual-energy CT (DECT) has recently been revived with the introduction of dual-source CT. The ability to simultaneously acquire two CT scans with different photon energies allows algorithm-based monoenergetic extrapolation [5], leading to virtual monoenergetic images (MEI) of selectable photon energies. Commercially available software allows the creation of MEI with virtual photon energies in a range of 40 to 190 keV. Due to a reduction of beam hardening artifacts, MEI with high virtual photon energies can be used to reduce metal artifacts [6] [7] [8] [9] [10] [11]. In MEI with low virtual photon energies, the Hounsfield numbers of materials with a high atomic number like iodine and calcium are increased, leading to an improved iodine contrast [12] [13] [14]. However, since the image noise increases even more than the iodine contrast, the contrast-to-noise ratio (CNR) of traditional MEI with very low virtual photon energies is lower than in the source CT images [15].
To obtain better results and to overcome noise limitations associated with traditional monoenergetic techniques [15], we investigated a novel technique (MEI+, Monoenergetic+, Siemens Healthcare, Forchheim, Germany) which applies a frequency-based mixing of the low keV images (which contain the high contrast signal) and an image of optimal keV from a noise perspective (typ. around 70keV) to combine the benefits of both images stacks – the low noise and the improved contrast.
The aim of this study was to evaluate the potential of MEI and MEI+ to increase iodine contrast for the visualization of liver vasculature in comparison to regular CT images. Since patients with severely altered hemodynamics of the portal venous system are infrequently seen in clinics, clinically indicated additional late-phase scans of the upper abdomen with poor depiction of intrahepatic vessels were chosen as a model for patients with intrinsically substandard hepatic image quality and more generally for situations requiring boosting of iodine-dependent contrast in order to achieve diagnostic image quality.
#
Materials and Methods
This retrospective study was approved by the local ethics committee. In our institution, abdominal CT scans were frequently acquired in dual-energy mode since virtual non-contrast images can be calculated from these. Patients were scanned using a 2nd generation dual-source CT system (SOMATOM Definition Flash with a Stellar detector, Siemens Healthcare, Forchheim, Germany). The scanning parameters for abdominal DE scans at our institution were: collimation 2x0.6x64 with pitch of 0.6, tube voltages 100/140 kV with tin-filter for 140 kV, reference effective tube currents 196/151 mAs. Depending on the patient’s weight and kidney function, 90 to 120 ml of contrast medium (400 mg Iomeprol/ml, Imeron 400, Bracco, Konstanz, Germany) and 40 ml of saline chaser were injected through an antecubital vein catheter at a flow rate of about 2.2 ml/s using a dual-syringe injector (CT Stellant, Medrad, Indianola, Pennsylvania, USA).
The most recent examinations with additional late-phase scans of the upper abdomen (typically indicated for the characterization of liver or kidney lesions) were reviewed. The latest 25 examinations with poor visualization of the intrahepatic vasculature (portal venous system and hepatic veins) were selected and included in this study. Using a software prototype (syngo.IPIPE 1.10.0.0, Siemens Healthcare), MEI and MEI+ were created at virtual photon energies of 40, 50, 60, 70, 80, 90, 100, 110, 120, 130, 150, 170 and 190 keV. Furthermore, polychromatic mixed images (MI) composed of a linear combination of the scans of low (100 kV) and high (140 kV) energies with ratios of 1:9 (M0.1), 3:7 (M0.3), 5:5 (M0.5), 7:3 (M0.7) and 9:1 (M0.9) were created.
Objective image evaluation
All image series were imported into a custom MATLAB program, which enabled the propagation of identical region of interest (ROI) measurements to multiple series. Multiple ROIs were defined within hepatic veins, portal veins and liver parenchyma using MEI+ of 40 keV. The mean CT numbers and image noise (pixel standard deviation) were recorded. Averaged signal-to-noise ratios (SNR) for intrahepatic vascular structures (IHV) and liver parenchyma (LP) as well as contrast-to-noise ratios (CNR) of the IHV and LP were calculated:
SNR = CT numbers/noise
CNR = (CT numbers IHV – CT numbers LP)/mean noise[1]
#
Subjective image evaluation
According to the results of the objective CNR analysis, exemplary series of each patient (MEI/MEI+ 40/70/190 keV, M0.3, M0.5, M0.7, M0.9 and sole 100 kV images) were reviewed by two readers in consensus. A subjective 5-point Likert scale was used to assess the depiction of IHV and the accordance with true portal-venous imaging (1: no visibility ([Fig. 1E, F]), 2: poor depiction ([Fig. 1A]), 3: moderate depiction ([Fig. 2A, B]), 4: good depiction ([Fig. 1C, D]; [Fig. 2C, D]), 5: very good depiction ([Fig. 1B])).
#
Statistical evaluation
Statistical analysis was performed with Prism 5 (GraphPad, La Jolla, USA). Measurements were treated as paired samples and statistically compared using Wilcoxon matched pair tests for ordinal and paired t-tests for continuous data. P-values < 0.05 were regarded as significant.
#
#
Results
25 patients (14 females, 11 males, age 62.9 ± 16.2 years) were included in this study. The reviewed late-phase scans were started 193.6 ± 44.8 s after beginning the application of contrast media (93.6 ± 11.1 ml). The mean CTDI recorded was 10.47 ± 3.00 mGy.
Objective image analysis revealed that MEI+ showed the best SNR values for both vascular structures (5.7 ± 1.3 at 40 keV) and liver parenchyma (4.9 ± 0.7 at 80 keV) compared to MEI (IHV 5.1 ± 1.1 at 70 keV and LP 4.7 ± 1.0 at 80 keV), MI (IHV 5.2 ± 1.1 and LP 4.8 ± 1.0 both for M0.5) and sole 100 / 140 kV scans (IHV 4.4 ± 1.0 and LP 3.7 ± 0.8 both at 100 kV). MEI datasets showed optimal SNR values at 80 keV, while the SNR values decreased for low and high virtual keV values due to a strong increase in image noise. In contrast, enhanced attenuation of iodine outweighed the increase in image noise in MEI+ resulting in increased SNR values even in poorly contrasted liver veins ([Fig. 3]). Nevertheless, the best objective measure for the depiction of vascular structures within liver parenchyma was CNR that was found to be superior in MEI+ at 40keV (2.1 ± 0.6) by a more than 2-fold increase compared to MI (CNR 1.0 ± 0.3 M0.9), MEI (1.0 ± 0.4 at 70 keV) and sole 100 kV images (1.0 ± 0.3) ([Fig. 4]). Both MEI and MEI+ showed an inverse contrast for virtual keV settings above 110 keV, while MEI+ images showed a significantly higher absolute CNR (MEI+ CNR -0.6 ± 0.3 vs. MEI -0.2 ± 0.1). Compared to polychromatic images of linear-weighted mixing of dual energy scans, MEI and MEI+ yielded few benefits at keV settings close to routine tube potential. However, the more the keV differed from the routine tube potential (120 kV), the better MEI+ performed (MEI+ at 40 keV). MEI could not show an objective benefit compared to MI, since the best CNR was measured at 70 keV and decreased significantly at low and high keV compared to MEI+, MI0.9 and sole 100 kV scans.
In the absence of iodine contrast media, exemplarily assessed within the gallbladder, MEI+ showed similar SNR values (1.1 ± 0.5 at 90 keV) like MI (1.0 ± 0.4 M0.5) and MEI (0.8 ± 0.4 at 80 keV) but again less of an increase in image noise than MEI for low and high virtual keV values ([Table 1]).
SNR: Signal-to-noise ratio; CNR: Contrast-to-noise ratio; LP: Liver parenchyma; IHV: intrahepatic vasculature, GB: Gallbladder; MEI: Monoenergetic extrapolated images; MEI+: Novel MEI; MI: Linear-weighted mixed images (fraction 100kV: 140 kV projections).
SNR: Signal-zu-Rausch-Verhältnis; CNR: Contrast-zu-Rausch-Verhältnis; LP: Leberparenchym; IHV: Inrahepatisches Gefäßsystem, GB: Gallenblase; MEI: Monoenergetisch extrapolierte Bilder; MEI+: neuartiges MEI; MI: Linear gewichtete Mischbilder (Anteil 100 kV : 140 kV Projektionen).
The subjective image assessment rated MEI+ images at 40 keV best compared to all other evaluated series (MEI 40/70/190 keV, MEI+ 70/190 keV, M0.3, M0.5, M0.7, M0.9 and sole 100 kV images) (p < 0.0001). MEI images were rated highest at 70 keV and did not show significant differences to MEI+ datasets (p = 0.15), which was consistent with our objective image analysis. Furthermore, MEI did not show significant improvement of the depiction of intrahepatic vessels (p = 0.07) compared to images only reconstructed from 100 kV projections. MEI at 40 keV, MEI/MEI+ at 190 keV, MI (M0.3) and sole 140 kV scan were rated not diagnostic in terms of intrahepatic vessel depiction ([Fig. 5])
#
Discussion
To the best of our knowledge, this is the first study evaluating the novel MEI+ algorithm, which facilitates frequency-based mixing of low keV images (which contain the high contrast signal) and an image of optimal keV from a noise perspective (typ. round 70keV) to combine the benefits of both images stacks – the low noise and the improved contrast.
25 patients who had undergone a clinically indicated dual energy late-phase scan of the upper abdomen were retrospectively included in this study. MEI, MEI+, linearly weighted mixed images (MI) and sole 100/140 kV images were calculated in order to assess the depiction of intrahepatic vessels objectively and subjectively. Objective analysis of MEI showed the best SNR values in the intrahepatic vasculature (IHV) and liver parenchyma (LP) at virtual 80 keV images, which was consistent with previous studies [5] [15]. Nevertheless these images suffered from a predominant increase in image noise at low virtual keV values overcoming iodine’s increased attenuation and consequently causing a decline of the SNR that was also consistent with previous findings [5] [15]. In contrast, the novel MEI+ only suffered a mild increase of image noise at < 70 keV and thus a significantly increased SNR compared to MEI, MI and sole 100 and 140 kV scans. The CNR of the IHV and LP as an objective measure for the depiction of vessels within the liver showed a more than 2-fold increase of CNR values at virtual 40keV. Subjectively, MEI+ at 40 keV was rated as the best way to depict intrahepatic vessels while having just a mild increase of noise levels and showed good accordance with portal venous phase imaging ([Fig. 6]).
Previous studies showed that liver imaging was able to benefit by monochromatic extrapolation of CT images [16] [17] in terms of increased sensitivity for the detection of lesions and greater diagnostic confidence [17]. Nevertheless, 70 keV was recommended due to increased noise levels to achieve the best overall image quality [15] [17]. MEI+ was able to overcome this limitation by still providing only mild image noise at 40 keV. Therefore, MEI+ images at 40 keV could be regarded as ideal for the depiction of intrahepatic vessels [17]. Furthermore, this algorithm-based post-processing is a promising candidate for calculated saving of contrast media, from which patients with reduced kidney function might benefit [3] [4] [18] [19]. These improved post-processing capabilities at low virtual keV values might also increase the sensitivity for the detection of soft-tissue lesions, e. g. in melanoma. Reduced contrast media CT angiographies might also be feasible, which should be evaluated in further studies.
One limitation of this study was its retrospective design. Late-phase imaging of the upper abdomen was used as a model for the low contrast situation or missed portal venous phase. Nevertheless, patients had received the appropriate weight-adjusted amount of contrast media, which redistributed to tissues and started to be renally eliminated at the time of the late-phase examination. Furthermore, our patients did not have severe impairment of portosystemic blood flow and intrahepatic lesions were not assessed. The assumption that the sensitivity for the detection of intrahepatic lesions would improve was based on the hypothesis that a further increase of iodine-based contrast enhancement with just a mild increase in image noise would improve sensitivity, which was described as beneficial before [17]. A true limitation of any monoenergetic extrapolation is a possible amplification of beam hardening artifacts which was not evaluated in our study, but might cause problems with metal implants in patients, e. g. following back surgery.
In conclusion, MEI+ is a very promising algorithm for monoenergetic extrapolation, which is able to overcome noise limitations associated with traditional monoenergetic techniques at low virtual keV levels [5] [13] [15]. This algorithm allows an improvement of IHV depiction in the presence of poor contrast.
#
#
Acknowledgement
The authors dedicate this article to Prof. Dr. Claus D. Claussen in occasion of his retirement.
1 mean noise = (noise IHV + noise LP)/2
-
References
- 1 Vignaux O, Gouya H, Augui J et al. Hepatofugal portal flow in advanced liver cirrhosis with spontaneous portosystemic shunts: effects on parenchymal hepatic enhancement at dual-phase helical CT. Abdominal imaging 2002; 27: 536-540
- 2 Vignaux O, Legmann P, Coste J et al. Cirrhotic liver enhancement on dual-phase helical CT: comparison with noncirrhotic livers in 146 patients. Am J Roentgenol American journal of roentgenology 1999; 173: 1193-1197
- 3 Garcia P, Genin G, Bret PM et al. Hepatic CT enhancement: effect of the rate and volume of contrast medium injection in an animal model. Abdominal imaging 1999; 24: 597-603
- 4 Rengo M, Bellini D, De Cecco CN et al. The optimal contrast media policy in CT of the liver. Part I: Technical notes. Acta radiologica 2011; 52: 467-472
- 5 Yu L, Leng S, McCollough CH. Dual-energy CT-based monochromatic imaging. Am J Roentgenol American journal of roentgenology 2012; 199: S9-S15
- 6 Lewis M, Reid K, Toms AP. Reducing the effects of metal artefact using high keV monoenergetic reconstruction of dual energy CT (DECT) in hip replacements. Skeletal radiology 2013; 42: 275-282
- 7 Tanaka R, Hayashi T, Ike M et al. Reduction of dark-band-like metal artifacts caused by dental implant bodies using hypothetical monoenergetic imaging after dual-energy computed tomography. Oral surgery, oral medicine, oral pathology and oral radiology 2013; 115: 833-838
- 8 Guggenberger R, Winklhofer S, Osterhoff G et al. Metallic artefact reduction with monoenergetic dual-energy CT: systematic ex vivo evaluation of posterior spinal fusion implants from various vendors and different spine levels. European radiology 2012; 22: 2357-2364
- 9 Lee YH, Park KK, Song HT et al. Metal artefact reduction in gemstone spectral imaging dual-energy CT with and without metal artefact reduction software. European radiology 2012; 22: 1331-1340
- 10 Bamberg F, Dierks A, Nikolaou K et al. Metal artifact reduction by dual energy computed tomography using monoenergetic extrapolation. European radiology 2011; 21: 1424-1429
- 11 Zhou C, Zhao YE, Luo S et al. Monoenergetic imaging of dual-energy CT reduces artifacts from implanted metal orthopedic devices in patients with factures. Academic radiology 2011; 18: 1252-1257
- 12 Venema HW. Virtual monochromatic spectral imaging with fast kilovoltage switching should not be used as standard CT imaging modality. Radiology 2011; 260: 916-917 author reply 917
- 13 Yu L, Christner JA, Leng S et al. Virtual monochromatic imaging in dual-source dual-energy CT: radiation dose and image quality. Medical physics 2011; 38: 6371-6379
- 14 Yuan R, Shuman WP, Earls JP et al. Reduced iodine load at CT pulmonary angiography with dual-energy monochromatic imaging: comparison with standard CT pulmonary angiography--a prospective randomized trial. Radiology 2012; 262: 290-297
- 15 Apfaltrer P et al. Value of monoenergetic low-kV dual energy CT datasets for improved image quality of CT pulmonary angiography. European journal of radiology (2013), European journal of radiology 2014; 83: 322-330
- 16 Cui Y, Gao SY, Wang ZL et al. Which should be the routine cross-sectional reconstruction mode in spectral CT imaging: monochromatic or polychromatic?. The British journal of radiology 2012; 85: e887-e890
- 17 Lv P, Lin XZ, Chen K et al. Spectral CT in patients with small HCC: investigation of image quality and diagnostic accuracy. European radiology 2012; 22: 2117-2124
- 18 Moos SI, van Vemde DN, Stoker J et al. Contrast induced nephropathy in patients undergoing intravenous (IV) contrast enhanced computed tomography (CECT) and the relationship with risk factors: a meta-analysis. European journal of radiology 2013; 82: e387-e399
- 19 Davenport MS, Khalatbari S, Cohan RH et al. Contrast material-induced nephrotoxicity and intravenous low-osmolality iodinated contrast material: risk stratification by using estimated glomerular filtration rate. Radiology 2013; 268: 719-728
Correspondence
-
References
- 1 Vignaux O, Gouya H, Augui J et al. Hepatofugal portal flow in advanced liver cirrhosis with spontaneous portosystemic shunts: effects on parenchymal hepatic enhancement at dual-phase helical CT. Abdominal imaging 2002; 27: 536-540
- 2 Vignaux O, Legmann P, Coste J et al. Cirrhotic liver enhancement on dual-phase helical CT: comparison with noncirrhotic livers in 146 patients. Am J Roentgenol American journal of roentgenology 1999; 173: 1193-1197
- 3 Garcia P, Genin G, Bret PM et al. Hepatic CT enhancement: effect of the rate and volume of contrast medium injection in an animal model. Abdominal imaging 1999; 24: 597-603
- 4 Rengo M, Bellini D, De Cecco CN et al. The optimal contrast media policy in CT of the liver. Part I: Technical notes. Acta radiologica 2011; 52: 467-472
- 5 Yu L, Leng S, McCollough CH. Dual-energy CT-based monochromatic imaging. Am J Roentgenol American journal of roentgenology 2012; 199: S9-S15
- 6 Lewis M, Reid K, Toms AP. Reducing the effects of metal artefact using high keV monoenergetic reconstruction of dual energy CT (DECT) in hip replacements. Skeletal radiology 2013; 42: 275-282
- 7 Tanaka R, Hayashi T, Ike M et al. Reduction of dark-band-like metal artifacts caused by dental implant bodies using hypothetical monoenergetic imaging after dual-energy computed tomography. Oral surgery, oral medicine, oral pathology and oral radiology 2013; 115: 833-838
- 8 Guggenberger R, Winklhofer S, Osterhoff G et al. Metallic artefact reduction with monoenergetic dual-energy CT: systematic ex vivo evaluation of posterior spinal fusion implants from various vendors and different spine levels. European radiology 2012; 22: 2357-2364
- 9 Lee YH, Park KK, Song HT et al. Metal artefact reduction in gemstone spectral imaging dual-energy CT with and without metal artefact reduction software. European radiology 2012; 22: 1331-1340
- 10 Bamberg F, Dierks A, Nikolaou K et al. Metal artifact reduction by dual energy computed tomography using monoenergetic extrapolation. European radiology 2011; 21: 1424-1429
- 11 Zhou C, Zhao YE, Luo S et al. Monoenergetic imaging of dual-energy CT reduces artifacts from implanted metal orthopedic devices in patients with factures. Academic radiology 2011; 18: 1252-1257
- 12 Venema HW. Virtual monochromatic spectral imaging with fast kilovoltage switching should not be used as standard CT imaging modality. Radiology 2011; 260: 916-917 author reply 917
- 13 Yu L, Christner JA, Leng S et al. Virtual monochromatic imaging in dual-source dual-energy CT: radiation dose and image quality. Medical physics 2011; 38: 6371-6379
- 14 Yuan R, Shuman WP, Earls JP et al. Reduced iodine load at CT pulmonary angiography with dual-energy monochromatic imaging: comparison with standard CT pulmonary angiography--a prospective randomized trial. Radiology 2012; 262: 290-297
- 15 Apfaltrer P et al. Value of monoenergetic low-kV dual energy CT datasets for improved image quality of CT pulmonary angiography. European journal of radiology (2013), European journal of radiology 2014; 83: 322-330
- 16 Cui Y, Gao SY, Wang ZL et al. Which should be the routine cross-sectional reconstruction mode in spectral CT imaging: monochromatic or polychromatic?. The British journal of radiology 2012; 85: e887-e890
- 17 Lv P, Lin XZ, Chen K et al. Spectral CT in patients with small HCC: investigation of image quality and diagnostic accuracy. European radiology 2012; 22: 2117-2124
- 18 Moos SI, van Vemde DN, Stoker J et al. Contrast induced nephropathy in patients undergoing intravenous (IV) contrast enhanced computed tomography (CECT) and the relationship with risk factors: a meta-analysis. European journal of radiology 2013; 82: e387-e399
- 19 Davenport MS, Khalatbari S, Cohan RH et al. Contrast material-induced nephrotoxicity and intravenous low-osmolality iodinated contrast material: risk stratification by using estimated glomerular filtration rate. Radiology 2013; 268: 719-728