Thromb Haemost 2002; 87(02): 277-281
DOI: 10.1055/s-0037-1612985
Letters to the Editor
Schattauer GmbH

Field Study of Lyophilized Calibrant Plasmas for Fresh Plasma INR Determination

A. M. H. P. van den Besselaar
1   Haemostasis and Thrombosis Research Center, Department of Haematology, Leiden University Medical Center, Leiden; The Netherlands
,
L. L. Houbouyan-Réveillard
2   Hôpital Ambroise Paré, Service Central d’immuno-hématologie, Boulogne-Billancourt, France
› Author Affiliations
Further Information

Publication History

Received 13 December 2000

Accepted after resubmission 09 November 2001

Publication Date:
13 December 2017 (online)

Summary

An alternative approach to INR estimation is for laboratories to calibrate their own local system using calibrant plasmas supplied by manufacturers or reference laboratories. The purpose of the present study was to investigate the within-laboratory variability of a calibrant plasma procedure using various sets of lyophilized plasmas. INR had been assigned to 13 calibrant plasmas in a previous multi-center study. Each of 10 other (“field”) laboratories measured PTs in the 13 calibrant plasmas and in 15 local fresh coumarin plasmas, using three different thromboplastin reagents. Each fresh coumarin PT was converted to INR using a calibration procedure with a set of 4 calibrants (1 normal + 3 abnormals). The abnormals of each set were either coumarin or artificial and were used with different assigned INR. When the INR had been assigned with a thromboplastin brand identical to the thromboplastin in the field laboratory, the procedure was named “reagent-specific” calibration. Otherwise the procedure was named “dissimilar” calibration. Using “reagent-specific” calibration procedures, relatively homogeneous INR were obtained for the fresh coumarin plasmas, whatever type of calibrant was used. In contrast, discrepant INR were obtained when “dissimilar” cross-species calibration procedures were used. The study was limited to 9 laboratories using the same type of coagulometer and one using a different type.

 
  • References

  • 1 Kitchen S, Jennings I, Woods TAL, Preston FE. Local calibration of international normalised ratio improves between laboratory agreement: results from the UK national external quality assessment scheme. Thromb Haemost 1999; 81: 60-5.
  • 2 Poller L, Triplett DA, Hirsh J, Carroll J, Clarke K. The value of plasma calibrants in correcting coagulometer effects on international normalized ratios. Am J Clin Pathol 1995; 103: 358-65.
  • 3 Hubbard AR, Margetts SML, Barrowcliffe TW. International normalized ratio determination using calibrated reference plasmas. Br J Hemat 1997; 98: 74-8.
  • 4 Houbouyan LL, Goguel AF. Long-term French experience in INR standardization by a procedure using plasma calibrants. Am J Clin Pathol 1997; 108: 83-9.
  • 5 Stevenson KJ, Craig S, Dufty JMK, Taberner DA. System ISI calibration: a universally applicable scheme is possible only when coumarin plasma calibrants are used. Br J Haemat 1997; 96: 435-41.
  • 6 Poller L, van den Besselaar AMHP, Jespersen J, Tripodi A, Houghton D. A comparison of artificially-depleted, lyophilized coumarin and fresh coumarin plasmas in thromboplastin calibration. Br J Haemat 1998; 101: 462-7.
  • 7 Poller L, Barrowcliffe TW, van den Besselaar AMHP, Jespersen J, Tripodi A, Houghton D. Minimum lyophilized plasma requirement for ISI calibration. Am J Clin Pathol 1998; 109: 196-204.
  • 8 Poller L, van den Besselaar AMHP, Jespersen J, Tripodi A, Houghton D. The European Concerted Action on Anticoagulation (ECAA): field studies of coagulometer effects on the ISI of ECAA thromboplastins. Thromb Haemost 1998; 80: 615-23.
  • 9 Chantarangkul V, Tripodi A, Cesana B, Mannucci PM. Calibration of local systems with lyophilized calibrant plasmas improves the interlaboratory variability of the INR in the Italian external quality assessment scheme. Thromb Haemost 1999; 82: 1621-6.
  • 10 Brien WF, Baskerville JC, Taberner DA, Crawford L. Calculation vs. calibration curve for INR determination. Am J Clin Pathol 1999; 111: 193-201.
  • 11 Van den Besselaar AMHP, Houbouyan-Reveillard LL, Aillaud MF, Denson KWE, Droulle C, Johnston M, Kitchen S, Lindahl TL, Marren M, Martinuzzo ME, Tripodi A, Vergnes C. Multicenter evaluation of lyophilized and deep-frozen plasmas for assignment of the international normalized ratio. Thromb Haemost 1999; 82: 1451-5.
  • 12 WHO Expert Committee on Biological Standardization. Guidelines for thromboplastins and plasma used to control oral anticoagulant therapy. WHO Technical Report Series No; 889 Geneva: 1999
  • 13 Poller L, van den Besselaar AMHP, Jespersen J, Tripodi A, Houghton D. The importance of “like to like” ISI calibrations with freeze dried plasmas. J Clin Pathol 1998; 51: 275-9.
  • 14 Hubbard AR, Margetts SML, Weller LJ, Macnab J, Barrowcliffe TW. An international collaborative study on the INR calibration of freeze-dried reference plasmas. Br J Haematol 1999; 104: 455-60.
  • 15 Kitchen S, Roberts E, Watson C, Makris M. Local calibration of INR systems: results with different calibrants. Br J Haemat 2001; 113 (Suppl. 01) 15 (abstract 57).
  • 16 Johnston M, Adcock D. The use of frozen plasma calibrants to standardize the international normalized ratio. Thromb Haemost 1999; (abstract): 544.