Keywords
Apical transportation - centering ability - rotary systems - reciprocating motion
Introduction
The success rate of endodontic treatment has increased significantly over the last
few years, mainly due to the development of new instruments and techniques that have
led to safer and more efficient clinical performance.[1] However, failures may occur at any stage of treatment, thereby compromising the
desired result.[1]
[2] Endodontic retreatment continues to be the primary alternative for resolution of
these cases, in which the filling material occupying the root canal must be removed
to enable new instrumentation and filling to be performed.[3]
[4]
[5]
Systems that use a reduced number of instruments for mechanical preparation of root
canals are a contemporary trend, making the treatment faster and consequently diminishing
both operator and patient fatigue.[6]
[7] According to this concept, it would be expected that these systems would also be
used for filling material removal, thereby obtaining success as relevant as that achieved
when this procedure is performed by traditional methods.[8]
[9]
[10]
ProTaper Next system (Dentsply Maillefer, Ballaigues, Switzerland) is composed of
five instruments—X1 (17.04), X2 (25.06), X3 (30.075), X4 (40.065), and X5 (50.06)—and
it can perform mechanical preparation in a short period of time, in comparison with
other rotary systems, due to the reduced number os instruments.[5] Manufactured with M-Wire technology, the instruments have an innovative design,
due to their rectangular cross-sectional shape that allows only two cutting edges
to touch the root canal walls simultaneously, while the other two rotate freely, resulting
in a rotary movement outside of their center of mass.[1] This kinematics of movement generates less stress on the instruments, providing
a corridor through which residues resulting from instrumentation may circulate.[11]
[12]
Due to these characteristics, studies have reported that ProTaper Next may be effective
in performing endodontic retreatment.[13]
[14] However, undesirable changes in the original trajectory of the root canal submitted
to retreatment with this system still require more in-depth evaluations.[15]
[16]
Therefore, the purpose of this study was to evaluate the shaping ability (apical transportation
and centering ability) of the ProTaper Next system after root canal retreatment in
comparison with the ProTaper Universal (Dentsply Maillefer) and Reciproc (VDW, Munich,
Germany) systems. The null hypothesis tested was that there would be no difference
in the shaping ability of the systems.
Materials and Methods
Sample Collection
In this study, 40 recently extracted human mandibular molars were used after obtaining
approval from the Research Ethics Committee (Protocol CAAE No. 52219315.1.0000.5020).
Sample size calculation was performed with the aid of the Sealed Envelope software
(Sealed Envelope Ltd., https://sealedenvelope.com/), thereby obtaining a minimum number of 10 randomly selected root canals per experimental
group for a power of 80% and type I error (level of significance) of 5%. The teeth
selected had completely formed roots and two independent mesial root canals. In addition,
the selected teeth should have a mesial root with an angle of curvature of 30 degrees
and radius of curvature of 10 mm, which were calculated following the methods of Schneider[17] and Pruett et al,[18] respectively. Radiographic and tomographic examinations confirmed the anatomic findings
necessary for inclusion or exclusion of the teeth from the final sample. Then, the
teeth were disinfected in a 0.5% chloramine-T solution and stored in a receptacle
containing distilled water, in which they were kept at 5°C until they were used.
Initially, the distal roots were sectioned with a diamond disc (KG Sorensen, Cotia,
São Paulo, Brazil), and the mesial roots length standardized at 16 mm. Coronal access
was achieved with a spherical diamond burr No. 1015 (KG- Sorensen, São Paulo, SP,
Brazil) mounted on a high-speed hand piece (Extra Torque 605C; Kavo, Joinville, SC,
Brazil) under copious water cooling. A glide plane was created with an Endo Z burr
(Dentsply Maillefer; Ballaigues, Switzerland) and a high-speed hand piece under copious
water cooling.
Biomechanical Preparation and Root Canal Filling
With the purpose of standardizing the position of the roots during instrumentation,
and later, when the tomographic examination was performed, the teeth were embedded
in colorless self-polymerizing acrylic resin (Jet Classic, São Paulo, SP, Brazil),
forming blocks that were afterward fitted onto a plastic base.
The mesial root canals were initially negotiated with sizes 10 and 15 K-type files
(Dentsply Maillefer), establishing the working length at 15 mm. Mechanical preparation
was performed with the ProTaper Universal (Dentsply Maillefer) system, in accordance
with the sequence of instruments recommended by the manufacturer, which are as follows:
SX (19.04), S1 (18.02), S2 (20.04), and F1 (20.07). The instruments were coupled to
a 6:1 contra-angle reducer (ENDO 6:1; Sirona Dental Systems GmbH, Bensheim, Germany),
driven by an electric motor (X-SMART Plus, Dentsply-Maillefer) in rotary movement,
at a speed of 300 rpm and torque of 2 Ncm, in accordance with the manufacturer’s instructions.
The root canals were irrigated with 1 mL of 2.5% sodium hypochlorite solution (Rio
Química; São José do Rio Preto, SP, Brazil) with the help of a disposable plastic
syringe (Ultradent Products Inc., South Jordan, UT, United States) and yellow NaviTip
needle (Ultradent Products Inc.), 3 mm short of the working length, after each gradual
insertion of the instruments. At the end of the instrumentation, the excess of solution
was aspirated with a Capillary Tip (Utradent Products Inc.), and the root canals were
again irrigated with 2 mL of 17% EDTA (Biodinâmica, Ibiporã, PR, Brazil) for 1 minute,
followed by 2 mL of distilled water. Thereafter, the root canals were dried with sterile
absorbent paper points (Dentsply-Maillefer) and filled with gutta-percha ProTaper
F1 cones (Dentsply Maillefer) and AH Plus sealer (Dentsply, Petrópolis, RJ, Brazil)
by means of a thermoplastification technique. Subsequently, the samples were stored
in an oven (ECB1; Odontobrás, Ribeirão Preto, SP, Brazil) at 37°C, and 100% relative
humidity, for 15 days. At the end of this period, the samples were randomly distributed
into four groups containing 10 mesial roots each, totaling 20 root canals (n = 20).
Filling Material Removal
Each experimental group was submitted to a different filling material removal protocol,
as follows: PTU Group–instrument F2 (25.08) of the ProTaper Universal (Dentsply Maillefer)
system in rotary movement at 300 rpm and torque of 2 Ncm; R25 Group–instrument R25
(25.08) of the Reciproc (VDW) system in reciprocating movement; X2 Group–instrument
X2 (25.06) of the ProTaper Next (Dentsply Maillefer) system in rotary movement at
300 rpm and 2 Ncm torque; and X3 Group–instrument X2 (25.06) of the ProTaper Next
(Dentsply-Maillefer) system in the cervical and middle thirds, followed by the instrument
X3 (30.07) for apical finishing, both in rotary movement at a speed of 300 rpm and
2 Ncm torque.
All the instruments were used in accordance with the recommendations of their respective
manufacturers, in a 6:1 contra-angle reducer (ENDO 6:1, Sirona Dental Systems GmbH),
driven by an electric motor (X-SMART Plus, Dentsply Maillefer). The instruments were
gradually inserted into the root canals, with an in-and-out motion, light apical pressure,
and amplitude of movement not exceeding 3 mm. At each insertion and removal of the
instruments from the root canal, they were cleaned with sterilized gauze. Each instrument
was used for reinstrumentation of a maximum of two teeth, with exception of R25, which
was used in only one tooth. The same irrigation protocol used during biomechanical
preparation of the root canals was repeated during the reinstrumentation procedure.
All the above-described procedures (initial preparation and retreatment) were performed
by only one operator, specialist in endodontics.
Apical Transportation
Before biomechanical preparation, all teeth were submitted to the Cone Beam Computed
Tomography (CBCT) assessment (CS 8100 3D, Carestream Health Inc., Rochester, NY, USA)
for acquisition of apical third images, and analysis of the original configuration
of the root canal trajectories. After initial preparation and filling, all samples
went through the same process, and again after filling material removal, resulting
in taking a total of three datasets of tomographic images. For this purpose, each
set resin block/base set was adapted to a wax platform (that had low density, thus
not interfering in performing tomography), with the mesial root parallel to the horizontal
plane. Then, the samples were placed on the CBCT scanner, according to the following
specifications: source of X-rays with value tension of 60 to 90 kVp, valve current
2 to 15 mA, and focal point of 0.7 mm.
For apical transportation analysis, images at 1 mm, 3 mm, 5 mm, and 7 mm of the apical
third were selected, generating a total of eight axial images of each 1 mm for each
canal. The transportation values were calculated by using the OsiriX Imaging software
(OsiriX Imaging Software, http://dwww.osirix-viewer.com), in which the postreinstrumentation and prereinstrumentation diameters of the mesial
root canal wall values were calculated by using the following equation: AT = (X1–X2)–(Y1–Y2).
The value of X1 corresponded to the mesial dentin wall thickness postinstrumentation;
X2 was the mesial dentin wall thickness postreinstrumentation; Y1 was the distal dentin
wall thickness postinstrumentation; and Y2 was the distal dentin wall thickness postreinstrumentation.[19] When the value of AT was equal to zero, there was no transportation; when it presented
a negative value, transportation occurred in the distal direction; and when the value
was positive, transportation in the mesial direction was considered.
Centering Ability
The centering ability of the instruments was also calculated for the 1 mm, 3 mm, 5
mm, and 7 mm of the apical third, compared with the values obtained during the transportation
analysis, using the equation described by Gambill et al[19]: CA = X1–X2/Y1–Y2 or CA = Y1–Y2/X1–X2. When the values were close to 1 (one), they
indicated optimum centering ability; and when close to 0 (zero), they indicated lower
ability of the instrument to maintain centralization in the root canal.
The images analysis and calculations for determining apical transportation and centering
ability were performed in a blind manner by a single and previously calibrated examiner.
Statistical Analysis
The values obtained (apical transportation and centering ability) were analyzed as
independent factors, considering both mesial canals and apical distance. The Kolmogorov–Smirnov
test was initially applied to test the normality of the sample. Then, the data were
submitted to the nonparametric Kruskal-Wallis and Dunn multiple comparison tests (p < 0.05). Statistical analysis was performed with the aid of GraphPad InStat software
for Windows (GraphPad Software; La Jolla, CA, United States).
Results
Apical Transportation
The graphical representation (mm) for apical transportation and its direction may
be visualized in [Fig. 1A]
[1B].
Fig. 1 Mean values of apical transportation and its direction, and centering ability. (A) Graphic representation (mm) of mean apical transportation values. (B) Graphic representation of apical transportation direction. (C) Graphic representation of mean centering ability values. There was no statistically
significant difference among groups (Kruskal–Wallis, the Dunn multiple comparisons
test, p < 0.05). n = 20. M–mesial; D–distal; MB–mesiobuccal; ML–mesiolingual.
There was no statistically significant difference among groups (p > 0.05), considering the different factors evaluated (apical distance and mesiobuccal
or mesiolingual root canals). Furthermore, a total of 13 root canals, distributed
throughout the four experimental groups (PTU = 6; R25 = 3; X3 = 2; X2 = 2), had no
apical transportation. Considering the direction of apical transportation, the highest
rates occurred in the distal/inner direction (57.8%) when compared with the outer
mesial/outer direction (42.2%).
Centering Ability
None of the instruments tested had complete centering ability (= 1.0). The group that
had the centering ability closest to the ideal value was the group in which filling
material removal was performed with instrument X2, followed by PTU, R25, and X3 groups.
However, there was no statistically significant difference among groups (p > 0.05) ([Fig. 1C]).
Discussion
The present study aimed to evaluate the shaping ability of the ProTaper Next system
after filling material removal in comparison with the ProTaper Universal and Reciproc
systems. Based on the results obtained, the null hypothesis was accepted, as the different
systems presented similar performance with regard to apical transportation and centering
ability after root canal retreatment.
Systems that use a reduced number of instruments for mechanical preparation have been
constantly used for root canal retreatment.[5] Therefore, changes in the original trajectory after root canal filling removal,
using these systems, must also be evaluated. To the best of our knowledge, there is
little information related to the apical transportation caused by these instruments
and their centering ability when they are used for filling material removal.[5] Just as the action of the instruments might promote undesirable changes in the original
trajectory of the root canal during its preparation,[2]
[6] the same might occur during its retreatment.[5]
The mesial canals of mandibular molars represent a challenge to clinicians, because
they are generally atretic, and their roots invariably present an accentuated degree
of curvature.[20] For this reason, mandibular molars with two independent mesial canals, and a 30-degree
angle of curvature—considered severe—were used in the present study. From a clinical
point of view, it is relevant to simulate these conditions because these are the conditions
that clinicians are frequently faced within their daily clinical practice.[21]
In this study, there was no significant difference among groups regarding transportation,
irrespective of the apical portion or the mesial canal (buccal or lingual) assessed.
These findings are in agreement with several other studies; [22]
[23]
[24] however, the authors would like to point out that these studies were not performed
in root canals submitted to retreatment, as it was performed in our study.
Fan et al[25] have reported that apical transportation greater than 0.300 mm is clinically unacceptable,
as it hinders the following stages of endodontic treatment, mainly root canal filling.
Although apical transportation occurred in almost of all samples in the present study,
the values ranged from 0.150 mm (mesial) to −0.225 µm (distal), ensuring favorable
conditions to root canal obturation.
In a recent study, Nevares et al[5] demonstrated that ProTaper Next and Reciproc systems had similar capacity for filling
material removal during retreatment. These systems also presented equivalent levels
of apical transportation, with clinically acceptable values.[25] These results corroborated the findings of our study. However, in the study conducted
by Nevares et al,[5] reinstrumentation of root canals in the ProTaper Next group was initially performed
with the instrument X3 (30.075) in the cervical and middle thirds, followed by the
instrument X2 (25.06) for apical finishing. In the present study, in the X2 Group,
there was no preflaring of the root canal (cervical and middle thirds enlargement)
with the instrument X3. Only the instrument X2 was used until working length was reached.
Whereas, in X3 Group, the instrument X2 was used in the cervical and middle thirds,
followed by the instrument X3 for apical finishing. Therefore, it is valid to state
that no enlargement of the cervical and middle thirds is needed when using the instrument
X2 for root canal reinstrumentation.
Although apical preparation was performed with an instrument with larger diameter
and taper (X3–30.075), the apical transportation was similar to the other groups.
The same might be said in relation to the centering ability of the instruments. Larger
diameter instruments tended to present greater difficulty in maintaining their long
axis centralized with the root canal long axis.[6] However, our findings demonstrated that the performance among the experimental groups
was similar, even in the group where a larger instrument was used for apical preparation.
Moreover, it is important to point out that none of the instruments tested demonstrated
complete centering ability.
In addition, a discrete tendency toward apical transportation in the distal/inner
direction was observed. Of the 320 points assessed in the root canals, 174 presented
transportation in this direction. Differently from that observed in the present study,
other research work reported a tendency toward higher transportation values in the
mesial direction, as the distal wall of the root canal acts in the antifurca direction.[21] This force presses the instrument against the mesial wall of the root canal, which
is situated contrary to the curvature.[23] However, we were able to note that the direction of transportation ranged according
to the apical portion evaluated. In the first and third millimeters of the root canal,
the trend toward transportation generally occurred in the mesial direction, whereas
transportation occurred in the distal direction in the fifth and seventh millimeters.
You et al[2] demonstrated that this situation—variation in the direction of transportation within
one and the same canal—may frequently occur, depending on the angle and radius of
curvature of the root. In addition, several studies have reported that root canal
preparation depends more on the root canal anatomy than the instruments used.[2]
[6]
[24]
Solvents such as eucalyptol and chloroform are frequently used in endodontic retreatment,
because they promote gutta-percha dissolution, facilitating the action of instruments
during root canal reinstrumentation.[8]
[26] On the other hand, these substances are considered toxic, in addition to leading
to formation of a softened gutta-percha-based residue that is difficult to remove,
because it strongly adheres to the root canal walls, particularly in the apical portion.[8] For this reason, no type of solvent was used before the action of the instruments
in this study. The presence of residual gutta-percha in the apical portion may present
difficulty in apical transportation analysis, thereby, compromising the accuracy of
the results.[5] Therefore, filling material removal in this study was performed according to the
protocol of other recent studies that did not use solvents.[4]
[14]
[27] Furthermore, it is worth emphasizing that this protocol was shown to be safe, since
no instrument was fractured during reinstrumentation, even without presoftening of
the gutta-percha.
Conclusions
Despite the limitations of this in vitro study, the authors were able to affirm that
all the systems tested, including ProTaper Next, presented clinically acceptable apical
transportation values after filling material removal. With regard to their centering
ability, none of the instruments were capable of maintaining their long axis centralized
in the root canal, therefore, presenting similar performance.