Keywords aerosol - ATP luminescent measurements - cross-transmission - infection disease transmission
- oral health - SARS-CoV-2
Introduction
The cross-contamination and dissemination of pathogens in a dental operatory environment
is a concern during dental procedures. The recent pandemic caused by coronavirus (SARS-CoV-2)
that on February 11, 2020, World Health Organization named as “coronavirus disease”[1 ] has increased this concern. Dental professionals must be aware that droplet and
splatter have proven to be the main spread routes of oral pathogens and, among them,
SARS-CoV-2.[1 ]
Oral droplet and splatter are biological substances (made of a combination of particles,
gases, vapors, biological fragments, or microorganisms that are or become airborne)
with a diameter between 0.5 and 10 µm that have the potential to penetrate and lodge
in the smaller passages of the lungs.[2 ]
[3 ] These particles are disseminated during standard dental treatments (e.g., ultrasonic
scaling, professional mechanical tooth cleaning, etc.) and are potential sources of
cross-transmission of microorganisms.[4 ] In this regard, a recent study has shown that instruments, such as rotating tools,
ultrasonic scalers, and piezo tools, produce a greater amount of aerosols/spray during
dental procedures that can widely spread the saliva of a potentially infected patient,
thereby exponentially increasing the risk of contamination/contagion.[5 ]
These considerations have gained relevance since SARS-CoV-2 represents a high risk
for cross-transmission of microorganisms for oral health professionals (OHP), which
can become potential carriers of the disease due to the unique nature of oral health
procedures, that involve aerosol generation and the close proximity of the professionals
to the highly contaminated oropharyngeal region of symptomatic and asymptomatic patients.[6 ] Besides, as reported by Zemouri et al,[7 ] patients could also generate droplets and splatter by coughing, sneezing, and talking;
therefore, the transmission probability is driven by indoor air quality, patients
infectiousness, and the masks used by OHPs. Actually, Gou et al concluded that SARS-CoV-2
has been widely distributed in the air (as an aerosol) and on object surfaces with
transmission distances close to 4 m in hospital facilities such as intensive care
units. This poses a high infection risk for medical staff and other close persons.
Most particularly, such risk is the highest in small indoor rooms.[8 ]
Measuring the concentration of adenosine triphosphate (ATP) by bioluminescence assay
is a technique routinely used in hospitals and the food industry[9 ]
[10 ]
[11 ]
[12 ] for fast and accurate quantification of biological contamination on surfaces. ATP
systems are able to detect dried organic materials and dead microorganisms[9 ] and to determine the efficacy of environmental cleanliness procedures even with
low microbial counts,[13 ]
[14 ] but do not detect viruses. This measurement technique has been successfully used
in a recent dentistry study which showed increased biocontamination levels on operator’s
goggles, masks, arms, and patient’s goggles after standard dental procedures, demonstrating
the ability of generated droplet and splatter to spread infection among operators
and patients.[4 ]
Cross-transmission of microorganisms can be prevented by preventing the contamination.[15 ] This can be achieved either at a macro level increasing ventilation or at the source
of contamination by applying an aerosol suction device, like the OBF, which was initially
developed for astronauts in space. In the absence of gravity, particles of saliva
or blood containing pathogens are released into the environment[16 ]
[17 ] as also reported in patients in isolated missions[17 ] both probably due to reduced ventilation, as suggested by Zemouri et al,[7 ] which studied the analogous situation in dental clinics due to the indoor air quality.
The perioral suction device named Oral BioFilter (OBF) consists in an ergonomic lip
retractor that supports an extraoral suction device which connects to the dental chair’s
high-volume evacuator (HVE) suction system. This device acts as a filter that prevents
airborne microparticles from expanding beyond the work area, thus harming the OHP
as well as the patients. As a consequence, OBF improves everyone’s safety by keeping
the dental workplace hygiene and cleanliness.[16 ]
[18 ]
The aim of the present study is to assess the effectiveness of the OBF device to reduce
cross-transmission of microorganisms by analyzing the contamination patterns produced
by droplet and splatter during professional dental hygiene with ultrasonic scaling,
followed by professional mechanical tooth cleaning (PMTC) with high-speed contra-angle
handpiece.
Materials and Methods
Study Design
The study was performed during the routine clinical practice in patients undergoing
a programmed professional dental hygiene with ultrasonic scaler followed by a PMTC.
All participants gave written informed consent to the procedure, which was approved
by the University of Barcelona Bioethics Committee, as a part of the first author
PhD Thesis project, and was conducted in accordance with the Helsinki Declaration
of 1975, as revised in 2013.
The standard professional dental hygiene procedure was performed by a single right-handed
hygienist at a height of 1.55 m with no assistant, using an ultrasonic scaler (Satelec
Acteon group, Merignac, France) during 10 minutes, followed by polishing with polishing
paste and NSK contra-angle (M 25 S-MAX, NSK, Nakanishi, Tochigi-ken, Japan). As a
pretreatment standard procedure, patients’ oral cavity was rinsed with 0.2% povidone
iodine for 1 minute to reduce pathogen load stemming from droplet and splatter forming
procedures in dental settings.[19 ]
[20 ] The instrumental table was located 1 m from the patient’s mouth, on the right of
the patient and in front of the hygienist. Patients were lying down in the dentist’
chair (FEDESA JS 500, FEDESA, Madrid, Spain), with a distance between the operator’s
face shield and the patient’s mouth of 30 cm ([Fig. 1 ]). The procedure was carried out with a dental chair’s HVE suction system, which
provided at the HVE hose while saliva ejector was also active a relative negative
pressure of 143 mbar measured by a vacuum meter (Thyracont VD81, Thyracont, Passau,
Germany) and an air flow of 240 liters per minute (DURR ROTA 84.4000 SW N4 10× flowmeter,
Dürr systems, Bietigheim-Bissingen, Germany). The sole difference between the control
and the OBF group was the static device connecting the HVE system with the patient’s
mouth, during a dental hygiene two hands procedure, with the OHP keeping the intraoral
mirror in the left hand while scaling with the right hand.
Fig. 1 (A ) Image of Oral BioFilter device. (B ) Representative image of professional hygienic procedure.
Patients were randomized 1:1 to use or not the OBF device. The study primary outcome
was the total number of RLU from every location of dental environment: operator’s
face shield, operator’s back of the surgical glove, patient’s safety goggles, and
instrumental table. Secondary outcomes were percentage contamination differences at
the different locations between control and OBF groups.
Sample Collection and Illuminometer Assay
The analyzed surfaces were the instrumental table, the operator’s right hand protection
(long cuff sterile latex surgical gloves), the patient’s safety goggles, and the operator’s
face shield. They were cleaned with 70% alcohol disposable gauze just before the dental
procedure and after taking the samples. Operator’s gloves were changed and discarded
between patients.
Samples were obtained immediately after dental procedure finalization from the aforementioned
surfaces from the patient as well as the operator using a chemically impregnated reagent
cotton swab (ATP Test Swab UXL100, 3M, Minnesota, United States) for ATP assay. Areas
measured consisted in 100 cm2 on the surface of the face shield, patient dental goggles, OHP back of right-hand
glove, and instrumental table.
All samples were obtained by a second operator, also provided with the adequate protective
equipment. The ATP bioluminescence assessment was performed as described by Sanna
et al.,[11 ] by using a Lumitester (3M Clean-Trace LM1, 3M, Minnesota, United States) according
to the manufacturer’s instructions.
Biocontamination levels, expressed in relative light units (RLUs) which in turn are
a function of ATP concentration, were recorded. An RLU level <150 units was considered
as clean; RLU levels ranging from 151 to 300 were considered as acceptable, and an
RLU level >300 was considered as a failure of the cleaning protocol, a threshold based
on the report by Ho et al[21 ] as well as the manufacturer’s recommendations.
Statistical Analysis
Variables were described by using summary statistics as median, minimum, and maximum
values and splatted by group (OBF and control). The Shapiro–Wilk test was used to
check normality of transformed data.
The potentially contaminated areas (100 cm2 ) were measured on the surface of the operator’s face shield and back of the right-hand
glove, the patient’s goggles, and the instrumental table. After the natural logarithm
transformation, a centered distribution was reached (Shapiro–Wilk p = 0.5447 for OBF and p = 0.8553 for control groups).
Primary outcome, total RLU, was analyzed by comparing the means of logged data, using
a two-sided two-sample t -test. Conclusions were drawn in terms of percentage of reduction as ratio of (GM
− 1) × 100, where GM was the geometric mean and their confidence intervals.
Power calculation was performed taking into account a previous pilot study (data not
shown), in which a coefficient of variation on the original scale of 12 was obtained,
then a two-sided two-sample t -test with 20 subjects per group and significance level of 0.05 achieved 94% power
to detect a ratio of geometric means of 0.08 or reduction of 92% when using OBF.
Secondary outcomes as RLU of logged data for the different locations were analyzed
in the same way. Proportion of patients from whom different locations reported events
(clean, acceptable, and failure) were analyzed by using Fisher’s exact test. A natural
logarithm transformation was applied for total RLUs and for every location due to
the extremely asymmetrical distribution of data.
Randomization list and power calculation were performed with PASS 2020 (PASS 2020
Power Analysis and Sample Size Software (2020). NCSS, LLC. Kaysville, Utah, USA css.com/software/pass)
and data management and statistical analysis with STATA 16 (StataCorp. 2019. Stata
Statistical Software: Release 16. College Station, Texas, United States: StataCorp
LLC).
Results
The study was carried out on May and June 2020. In total, 40 patients were included
16 males and 24 females, with a mean age (standard deviation) of 45.6 (16.4) years
and 40.6 (13.7) years, respectively.
The mean RLU values and the difference between groups (total and by each location)
are shown in [Table 1 ], indicating the percentage of reduction using OBF with respect to control. For the
whole dental environment, the percentage of reduction in RLUs achieved with OBF was
98.4%; this reduction ranged from 97.4 to 99% in the 95% of the cases ([Table 1 ]). Regarding the analysis by locations, the use of OBF showed different percentages
of reduction in RLUs ranging from the maximum, a 99.6%, in the OHP face shield, to
a minimum (83%) in the instrumental table ([Table 1 ]).
Table 1
Comparison between groups for total relative light unit and by locations for logged
data
Location
Mean
Difference
Reduction (%)a
95% CI
p -Value
All
Abbreviations: CI, confidence interval; GM, geometric means; OBF, Oral BioFilter.
a Reduction: (ratio of GM − 1) × 100.
OBF
6.62
−4.15
98.4
97.4–99.0
<0.001
Control
10.77
Face shield
OBF
4.33
−5.46
99.6
99.2–99.8
<0.001
Control
9.79
Surgical glove
OBF
5.31
−3.69
97.5
94.7–98.8
<0.001
Control
9.00
Patient safety goggles
OBF
5.63
−2.76
93.7
86.3–97.1
<0.001
Control
8.39
Instrumental table
OBF
4.42
−1.77
83.0
66.1–91.6
<0.001
Control
6.19
Regarding the three cleaning levels according to the RLUs values, high differences
were also found favoring OBF group in all measured locations ([Table 2 ]). The control group reported a very high percentage of failures, being the 100%
in the surfaces nearer to the patient’s mouth and decreasing to 70% in the instrumental
table. In contrast, the higher percentage of failures in the OBF group was found in
the patient’s goggles (40%), while the OHP face shield showed a complete absence of
particles as revealed by RLU counts.
Table 2
Counts of occasions with clean, acceptable, and failure events by locations
Location
Clean
(RLU < 150)
Acceptable
(RLU = 150–300)
Failure
(RLU > 300)
p -Value
Face shield
Abbreviations: OBF, Oral BioFilter; RLU, relative light unit.
Note: Data are expressed as counts (%).
OBF
15 (75)
5 (25)
0
<0.001
Control
0
0
20 (100)
Surgical glove
OBF
8 (40)
5 (25)
7 (35)
<0.001
Control
0
0
20 (100)
Patient safety goggles
OBF
6 (30)
6 (30)
8 (40)
<0.001
Control
1 (5)
0
19 (95)
Instrumental table
OBF
15 (75)
4 (20)
1 (5)
<0.001
Control
4 (20)
2 (10)
14 (70)
The results of the present study are consistent in proving that the use OBF device
reduces the bioluminescent droplet and splatter particles and splatter produced by
dental procedures. The grouped results showed a 98.4% of whole dental environment
contaminant reduction.
Discussion
In general dental practice, the use of face seal masks was recommended to reduce the
exposure to aerosolized microorganisms such as Mycobacterium tuberculosis .[22 ] Influenza or SARS-CoV virus may be also shed into the environment and environmental
surfaces and transferred to hands of patients and health care providers[23 ] as well as form airborne particles.[3 ]
[7 ] The pandemic has made more evident the need for masks since SARS-CoV-2 has been
consistently detected in the saliva of COVID-19 patients.[4 ]
[24 ]
[25 ] In two recent reports, Zemouri et al[7 ]
[12 ] have reported that the highest transmission possibilities estimated in a dental
clinic were for measles virus (100%), followed by coronaviruses (99.4%), influenza
virus (89.4%), and M. tuberculosis (84%).
The detection of biological contamination in dental clinics provides useful information
to promote directed hygienic measures and avoid biohazard transmission. The ATP levels
(ATP bioluminescence assay) detected on surfaces and equipment showed a positive correlation
with bacterial contamination in dental practice.[12 ] Bioluminescence has not proven to date a direct correlation between viable microbial
counts and detected ATP levels.[9 ]
[13 ] However, ATP-bioluminescence is a valuable tool for determining the efficiency of
cleanliness procedures as well as to distinguish between live and dead cells,[13 ] but it is not efficient in detection environmental, not cell-bound ATP and viruses.
Using ATP bioluminescence, Watanabe et al reported significant higher values of contamination
and splatters in surfaces produced by droplet and splatter in operator masks, goggles,
chest, and gowned right arm after dental procedures.[4 ] Zemouri et al[7 ]
[12 ] estimated that, even if the contamination due to droplet and splatter may be low
in a dental clinic, the highest level of contaminated particles is to be found within
80 cm around the head of the patient because the droplet and splatter produced by
oral health treatments may contain a similar number of pathogens as compared with
coughing and sneezing. Although the exact numbers of viral pathogens in droplet and
splatter are not known to date, Zemouri et al[7 ] as well as Prather et al[26 ] suggest that SARS-CoV-2 is spreading in droplet and splatter exhaled by highly contagious,
nonsymptomatic individuals.
It is important to remark that although our study has focused in assessing the novel
OBF device efficacy for cross-contamination reduction; personal protective equipment
such as face shield, mask, goggles, and surgical gloves; and protective measures are
still fundamental elements to mitigate infectious risks. Several studies have shown
how such protective gear coupled with a reasonable dental equipment use are important
to prevent infections.[27 ]
[28 ]
In our study, the impact of using a OBF device was assessed by using the same biological
detection assay used by Watanabe et al,[4 ] in patients after a standard dental hygiene procedure. Since patients both groups
of patients were submitted to the same standard procedure that includes a 0.2% povidone
iodine mouthwash for 1 minute, the ATP bioluminescence measurement was aimed to detect
and localize the biological cross-contamination, be that constituted by living or
dead cells and particles. We observed a significant decrease, around 98.4%, in whole
dental environment contamination with the use of OBF. Our results corroborate the
risk of biological cross-contamination by droplet and splatter described by Watanabe
et al[4 ] and Zemouri et al[7 ]
[12 ] in operator’s and patient’s protective gear, thus demonstrating the utility of bioluminescence
to assess the reduction of droplet and splatter spreading by using OBF. The limitation
of our study is that bioluminescence, despite being able to identify living microorganisms,
such as bacteria, dinoflagellates, and fungi, does not detect the presence of viruses.[13 ]
Despite reports of measurable droplet and splatter reduction have encouraged the HVE
use in routine dental practice while working with four hands,[29 ] other authors did not find neither significant droplet and splatter counts decrease
nor environmental contamination reductions unless the HVE aspirator was subject to
modifications.[30 ] In our study, the OBF device as provided by the manufacturer was attached to the
standard HVE hose while permitting the OHP to operate with two hands without assistant.
A much improved reduction of droplet and splatter dissemination was evidenced, when
compared with the control group, that do not used the HVE device because is incompatible
with the use of an intraoral mirror in two hands work. Moreover, we wanted to corroborate
the dispersion distance of particles Watanabe et al found in their study[4 ] and to analyze the effectiveness of the OBF device against the total spread of droplet
and splatter particles. For this reason, further analysis of the droplet and splatter
reduction in four hands work, where the operator’s dental assistant will handle the
HVE and the saliva ejector, while the operator will continue to use ultrasound and
the intraoral mirror are planned.
Watanabe et al reported that the surface with the highest contamination levels was
in the patient’s safety goggles,[4 ] while Zemouri et al found the highest bacterial contamination on the patient’s chest
area.[12 ] Previous studies highlighted that eye infections are common among dentists,[31 ] which risk being also signaled for SARS-CoV-2 both for general population[32 ] and for dentists. Moreover, blood or saliva contained in the splatters could produce
infection and corneal epithelial exfoliation, conjunctivitis, and keratitis.[33 ] With the use of OBF, the droplet and splatter reduction achieved 97.3% in patient’s
goggles and 99.6% in OHP’s face shield, thus providing an extra protective measure
against ocular cross-contamination.
The OHP’s hand is the surface closest to patient’s mouth; therefore, a contamination
of this surface is common in clinical practice. The use of OBF promotes a reduction
of contamination on the surgical gloves of 97.5% versus control group (without OBF).
This result demonstrates that OBF could also enhance the prevention of biological
cross-contamination in high-risk airway management procedures[34 ] as well as in orofacial aerosol-generating procedures also usual in head and neck
surgery,[35 ] otolaryngology,[36 ] and anesthesia.[37 ] Although this noticeable decrease represents an improvement of the cross-contamination
control, for safety purposes, adopting updated guidelines on infection prevention[7 ]
[12 ]
[38 ]
[39 ] is still mandatory.
It is interesting to note that the lowest percentage of the contamination reduction
achieved with the OBF device was found in the instrumental table (83% reduction of
biological contamination) located at a distance of 1 m from the patient which agrees
with the 80 cm distance reported by Zemouri et al.[7 ] At first glance, despite being statistically significant, such reduction highlights
the difference between 70% unacceptably contaminated surface cases (control group)
down and just 1% (OBF group; p < 0.001). On the other hand, particles sized between 1 and 4 microns expelled by
patients just by breathing have the potential of lingering in the air[7 ]
[26 ] and may be displaced by airflows and deposited on equipment.[40 ] This result is relevant as the material placed on the instrument table should be
preserved in the cleanest possible condition after being unpacked, so further measures
have to be taken to avoid cross contamination on this area.
The significant difference among RLU measures between both groups (OBF vs. control;
6.62 vs. 10.77 [difference = −4.15]; p < 0.001) confirms that the major effect of OBF is for the OHPs. When proper infection
control measures are used, the chance for the patient to become contaminated by microorganisms
from the previous patient is substantially limited. Therefore, only in the event that
these standard measures failed due to human error, we can consider that the device
adds extra protection to the next patient. The device does not remove the infection
risk but reduces the cross-contamination risk by limiting the droplet and splatters.
Taken together, these first results show a significant reduction of cross-contamination
on oral health operatory surfaces by using the OBF device as well as relevant information
about which specific surfaces require additional hygienic measures to ensure patient’s
and OHPs protection.
In conclusion, to our knowledge, this is the first study that uses a specific device
designed to reduce the biological cross-contamination in routine oral health practice.
It is important to remark that OBF device provides a much improved protection, but
it is no substitute to other protective measures such as face shield and mask, goggles,
and surgical gloves which must still be considered mandatory.