Vet Comp Orthop Traumatol 2021; 34(04): 294-302
DOI: 10.1055/s-0041-1725015
Clinical Communication

Zurich Cementless Dual Mobility Cup for Canine Total Hip Prosthesis: Implant Characteristics and Surgical Outcome in 105 Cases

Authors

  • Otto I. Lanz

    1   Department of Small Animal Clinical Science, VA-MD Regional College of Veterinary Medicine, Blacksburg, Virginia, United States
  • Ida Forzisi

    2   Clinica Veterinaria Vezzoni, Cremona, CR, Italy
  • Aldo Vezzoni

    2   Clinica Veterinaria Vezzoni, Cremona, CR, Italy
 

Abstract

Dislocation after total hip arthroplasty (THA) remains a troublesome complication, and a source of frustration for the owner and the surgeon. The dislocation rate of hip prostheses in dogs is reported to range from 4 to 15%, representing the most common short-term complication. This is especially true in large and giant breed dogs, usually requiring revision surgery. With the increase in the number of THA being performed in veterinary surgery, reducing or preventing complications such as postoperative THA dislocation will be of paramount importance. The Zurich cementless dual mobility (DM) system allows impingement-free range of angulation of 80 to 132 degrees between the ceramic head and the polyether ether ketone (PEEK) cup when combined with the range of the PEEK cup in the outer metal cup. In this article, we review the use of the DM cup in THA in large and giant breed dogs, in terms of its history, biomechanics, outcomes and complications based on 105 cases.


Introduction

The concept of dual mobility (DM) was originally developed in the 1970s by Dr G. Bousquet for human total hip arthroplasty (THA) procedures, with the purpose of reducing the risk of dislocation of the hip prosthesis in the postoperative period.[1] The original acetabular cup consisted of two bearings: a smaller ball inserted inside a larger polyethylene hemisphere, which articulated with a smooth metal surface. Primary movement occurs at the inner bearing with the DM cup. The secondary articulation, between the liner and the acetabular shell, engages during activities that exceed the normal range of motion (ROM) of the hip, when the femoral component impinges on the rim of the liner.[2]

Dislocation following a THA continues to be the major complication associated with the THA procedure, with rates of luxation ranging from 1.9 to 5.8% in humans.[3] [4] In 2010, the Australian Joint Registry reported that dislocations accounted for 14.4% of THA revision procedures.[5] The dislocation rate of hip prostheses in dogs is reported to range from 4 to 15%, representing the most common short-term complication.[6] [7] [8] [9] [10] [11] [12] [13] [14] [15] This dislocation rate complication is especially true in large and giant breed dogs, and necessitates revision surgery, leading to further costs to the owner, prolonged recovery times, or, in the worst-case scenario, explantation.[13] [15] The rate of dislocation when using a DM cup prosthesis in the initial THA procedure in people is reported to be between 0 and 1.1%.[16] [17] [18] In the veterinary field, the only documented use of a DM cup is reported by Guillaumot and colleagues.[19] [20] In his second publication, a cemented DM cup prosthesis was implanted in dogs, and no dislocation was experienced in 50 consecutive cases.[20] The overall rate of dislocation following a THA depends on several aspects, such as appropriate positioning of the implants, orientation and offset of the acetabular and femoral implants and the preservation of the local soft tissues.[3] [21] [22] It has been shown that a large diameter head can decrease the rate of luxation.[23] [24] A larger diameter head improves stability by not only affecting the jump distance, which is the degree of lateral translation of the centre of the femoral head required before dislocation occurs, but also reducing impingement by altering the head-to-neck ratio, thus increasing the ROM before impingement and dislocation.[25] There may be decreased wear when using a DM cup, as there are two interfaces, and thus less friction and sliding at each interface. This has been proven in the laboratory, but not yet clinically.[26] [27] Long-term in vivo wear studies are needed to determine wear in comparison with standard acetabular cups. In humans, DM cups are used in complex primary or revision THA.[28] [29] [30] THA for displaced fractures of the femoral neck has a higher rate of dislocation than THA performed for osteoarthritis.[30] [31] In dogs, the DM cup is potentially useful in cases with higher risk of dislocation, as is currently seen in cases with chronic hip luxation, nonreparable femoral head fractures, unsuccessful femoral head and neck excisional arthroplasty, large and giant breed dogs and young, giant breed dogs with already severe hip dysplasia with hip laxity (luxoid hips). The cementless hip prosthesis is now widely used in both the human and veterinary fields, especially in young patients, where a biological fixation by osseointegration guarantees a longer duration of the implants.[6] [14] [15] [32] [33] [34] Therefore, the development of a cementless DM cup was motivated by the need to have a DM cup for the Zurich cementless prosthetic system to be used in dogs with a known increased risk of dislocation following a THA procedure. The adoption of carbon fibre ring (CFR)/polyether ether ketone (PEEK, Evonik, Germany) cup insert instead of highly crosslinked polyethylene since January 2016 in the Zurich system allowed the outer bearing of the DM cup to be made of PEEK, which is more resistant to wear when compared with polyethylene.[35]

The purpose of this study is to review the use of the Zurich cementless DM system in THA in terms of its history, biomechanics, outcomes and complications based on 105 cases.


Materials and Methods

Implant Characteristics

The Zurich cementless DM system has two main components, the cup and the head ([Fig. 1A]). The cup is composed of an outer and an inner shell, both made from a titanium alloy. The outer shell is perforated and coated by porous titanium and hydroxyapatite for rapid bone ingrowth and increased compliance. The inner shell's articulation surface is coated by titanium nitrate. The head of the DM articulation has two components. The outer head component is made of PEEK with a carbon fibre reinforced PEEK ring (PEEK/CFR-PEEK) on the inside. The outer surface of the PEEK head component articulates with the titanium nitrate surface of the cup and the carbon fibre reinforced PEEK ring inside it articulates with the second component of the head, which is made of ceramic. The ceramic (VERILOX, CeramTec, Germany) head is preassembled with the PEEK head by snap-fitting that allows for free rotation between the two but also provides strong retention in the axial direction. The outer head component has a nominal diameter of 25 mm and the ceramic head has a 19 mm diameter.

Zoom
Fig. 1 Technical drawings (complete description in the text). (A,1–4) The four components of the Zurich dual mobility cup system. (A, 5) Ceramic (Ceravet) head of the articulation. (A, 6) Neck of the prosthesis. (A, 7) Spike guided into its receiving hole in the outer shell for concentric assembly of the shells, by either the manufacturer or in surgery for the revision-style cup; G1, small gap at the outer surface of the polyether ether ketone (PEEK) head component; G2, small gap at the primary articulation surface between ceramic head and carbon fibre ring-PEEK ring. (B) Eighty degrees of impingement-free range of angulation between ceramic head and PEEK head. (C) Impingement-free range increased to 132 degrees when combined with range of PEEK head in the outer metal cup; (C, 8) The jumping distance (yellow arrow) is the degree of lateral translation of the femoral head centre required for the hip to dislocate.

The articulation ([Fig. 1A]) is attached to the neck of the prosthesis via a standard conical coupling. The spike is guided into its receiving hole in the outer shell facilitating concentric assembly of the shells, by either the manufacturer or in surgery for the revision-style cup. The inner titanium nitrate coated surface of the inner shell is spherical in shape. The outer surface of the PEEK component is aspherical in shape leaving a small gap, G1, at the pole resulting in an annular contact at the articulation.

This articulation geometry is shown to reduce the wear rate by approximately a factor of three, when compared with sphere-in-sphere articulation with a polar contact.[36] At the primary articulation surface, the spherical ceramic head makes an annular contact with the CFR-PEEK ring leaving a small gap, G2, at the pole. The ceramic head is snap-fitted into the PEEK head by the manufacturer. The centre of rotation C1, between the ceramic head and the PEEK head, is positioned deeper into the cup than the centre of rotation C2, between the PEEK head and the cup. This theoretically provides a slight bias in forces tending to align the PEEK cup over the average joint force. The friction moment between the ceramic head and the PEEK head is smaller than that between the PEEK head and the cup, so most of the motion is expected to happen at the ceramic head-CRF-PEEK ring interface, which results in the lowest wear of known articulations, except for perhaps ceramic–ceramic pairing.[34] The impingement-free range of angulation between the ceramic head and the PEEK head of 80 degrees ([Fig. 1B]) is increased to 132 degrees ([Fig. 1C]) when combined with the range of the PEEK head in the metal cup. The ceramic head diameter is 19 mm; the nominal diameter of the PEEK head is 25 mm; the nominal outer cup diameter is 29.5 mm and 32.5 mm; and the nominal neck diameter is 5.5 mm. During extreme flexion or extension of the hip, the femoral neck moves the PEEK head inside the acetabular component, which has a highly polished articulating surface of titanium nitrate ([Fig. 2]).

Zoom
Fig. 2 (A) Dual mobility cup and dual mobility head. (B) The neck inserted into head of the dual mobility head. The neck is first inserted into the ceramic head followed by insertion of the peg of the femoral prosthesis into the neck.

Surgical Technique

The technique of performing a primary THA using the DM cup is similar to that of a standard Zurich THA. Full exposure of the acetabulum needs to be obtained prior to reaming the acetabulum, with the femur retracted in a caudoventral direction following excision of the head and neck. The outer shell of the cup is implanted between 40 and 45 degrees of lateral opening, and 15 and 20 degrees of retroversion. Lateral opening is evaluated visually with the impaction bar aligned to 45 degrees, provided that the pelvis is in true lateral position, and checked with an image intensifier. Due to the variability in pelvis conformation in dogs of different breeds, the cup retroversion is aimed to follow the anatomy of the acetabulum, centring the cup between the cranial and caudal pillars of the acetabulum. To impact the DM cup, the impactor for the 26.5 revision cup is used. The femur is prepared in the standard fashion, paying particular attention to the preoperative planning measurements and preservation of the surrounding soft tissues, ensuring that appropriate offset is obtained. Fixation of the cup when performing a revision THA may require fixation with screws. In such an instance, a disassembled DM cup is used to fix the outer shell with 2.0 or 2.4 mm cortical screws ([Fig. 3A] and [B]). The inner shell is impacted next inside the outer shell. The inner shell is aligned in a perfectly coaxial direction by the spike in the pole of the inner shell, which in turn aligns with a receptacle in the centre of the outer shell. Before implanting the femoral stem, the clearance between the cup and the femoral neck osteotomy is checked. This is performed by inserting a 23 mm trial cup into the DM cup (outer shell) while keeping the positional bar vertical and contacting the femoral osteotomy. The femur is simultaneously retracted with a Hohmann retractor or with pointed reduction forceps on the trochanter with distal distraction. A clearance of ∼2 cm should be obtained to allow reduction of the prosthesis with a medium neck, which is indicated in large dogs where a 29.5 mm or a 32.5 mm cup is required. When the clearance is less than 2 cm, the femoral osteotomy is modified in a distal direction, and the clearance is re-evaluated. Reduction of the DM head requires a different technique compared with the reduction of a standard cup and 16 mm or 19 mm heads. Because of the overall larger diameter of 25 mm (ceramic head plus CFR-PEEK head), the jumping distance necessary for reduction is greater when compared with the standard method, making reduction more difficult ([Fig. 1C]). To simplify the reduction, it is recommended that the preassembled head and CFR-PEEK head be impacted on the correct neck first. The neck, head and CFR-PEEK head are inserted into the cup. The femur is then levered, so that the peg of the femoral prosthesis is reduced into the neck. To summarize, the correct order for ease of reduction is to first impact the preassembled DM head (ceramic head plus CFR-PEEK head) on the appropriate neck, and then insert both inside the cup. The femur is retracted laterally and distally with a pointed bone holding forceps fixed at the greater trochanter, and the peg of the femoral stem is guided until it enters the recess of the neck. To complete the insertion of the peg of the femoral stem inside the recess of the neck, a narrow Hohmann retractor, inserted caudally to the acetabulum, is used to lever the femur laterally and distally, allowing the peg of the femoral prosthesis to enter the recess of the neck. With this method, accidental entrapment of soft tissues inside the DM cup during reduction is avoided. The usual tests for stability of the reduction are performed, followed by the closure of the anatomical layers.

Zoom
Fig. 3 Dual Mobility Cup. (A) The inner shell (gold in colour) is impacted inside the outer shell (white in colour). (B) The inner shell is aligned in the coaxial direction by a spike in the pole of the inner shell, aligning the centre of the outer shell.

Standard radiographic views are performed to evaluate the implant orientation and position: ventrodorsal view with the hip flexed to evaluate the deepness of insertion of the cup, lateral view of the pelvis with superimposed ilial bodies and tuber ischii to measure the angle of lateral opening and to evaluate retroversion of the cup, mediolateral view of the femur with the stifle positioned at 90 degrees to evaluate the stem anteversion, and an oblique view of the pelvis with the acetabular cup viewed fully laterally (radiographic beam tangent to the equator of the cup) with the femur perpendicular to the pelvis and slightly externally rotated, which allows the X-ray beam to be perpendicular to the screws of the stem. The latter most radiographic projection is obtained with fluoroscopic guidance for accurate positioning.


Indications for Dual Mobility Cup

There are no extensive published data on the use of DM cups in dogs, except for the aforementioned publication by Guillaumot and colleagues.[20] However, it is the authors' experience that large and giant breed dogs have an increased risk of dislocation following a THA procedure compared with smaller dogs using standard cups and heads. The DM cup is currently available in two dimensions, 29.5 and 32.5 mm, because of the space required for the ceramic head plus the CFR-PEEK inner liner. Due to these limited sizes, its use is indicated only in large and giant breed dogs, for whom those sizes of cups are appropriate. In those dogs, especially when affected by preoperative joint instability, the risk of luxation should be greatly reduced when compared with the published rate of luxation following a THA procedure without using the DM protheses.



Results

A total of 105 THA were performed using the Zurich cementless total hip system with the DM cup between April 2018 and March 2020 at the Clinica Veterinaria Vezzoni, Cremona, Italy, by the same surgeon (AV). There were 53 males and 34 females for a total of 87 dogs, with 18 dogs having bilateral THA performed.

Median body weight was 43 kg (range: 30–64 kg), and median age was 37 months (range: 6–116 months). Breeds included are as follows: Bernese Mountain dog (n = 14), German Shepherd dog (n = 14), Rottweiler (n = 14), Cane Corso (n = 8), Labrador retriever (n = 7), Mixed breed (n = 6), Newfoundland (n = 5), Leonberger (n = 3), Maremma Shepherd (n = 3), Central Asia Shepherd dog (n = 2), Dogue de Bordeaux (n = 2) and 1 each for the following 9 different breeds – Akita Inu, Alaskan Malamute, Broholmer, Bull Mastiff, Golden retriever, Bobtail, Rhodesian Ridgeback, Russian terrier and Swiss Shepherd dog. The 29.5 cup was used in 97 cases (92%) and the 32.5 in 8 cases (8%). A short neck was used in 13 cases (12%), a medium neck in 86 cases (82%), a long neck in 4 cases (4%) and an x-long neck in 2 cases (2%). The radiographic and clinical follow-up examinations had a median follow-up time of 5 months with a range of 2–24 months and were available in 85 of 87 (98%) dogs ([Figs. 4] and [5]). Two cases in this study using the DM cup were performed as revisions of previously performed Zurich cementless THA surgeries that later dislocated. The first case involved a 15-month-old male intact Dogue de Bordeaux that dislocated 2 months following surgery (26.5 mm acetabular cup with a 19 mm head). A 29.5 mm DM cup was used to revise this case and at the 6-month follow-up evaluation, no postoperative complications were noted and the implants radiographically appeared stable. The second case involved a 3-year-old female intact Newfoundland that had a previously performed Helica THA that failed 4 months following surgery due to failure of the Helica stem. Initially the Helica stem was solely replaced with Kyon stem and Helica 18 mm neck/head unit suitable for Kyon stem peg, but the prosthesis dislocated 15 days following surgery. A second revision was performed with a 29.5 mm DM cup. Radiographs performed 4 months following the second revision revealed no abnormalities.

Zoom
Fig. 4 Pelvic radiographs of a 29-month-old male Bernese Mountain dog, 39.6 kg, with bilateral total hip replacement using dual mobility cups. Left hip was done at 11 months of age and the right hip at 14 months of age.
Zoom
Fig. 5 Same dog as in [Fig. 4]. Follow-up with oblique pelvic views tangent to the implants: left after 18 months and right after 15 months reveal normal bilateral total hip replacements.

One case involving a Cane Corso had a Zurich cementless THA procedure (26.5 mm cup and 19 mm head) performed in the left hip, which dislocated in the postoperative period and was revised by replacing the medium neck with a long neck. A Zurich cementless THA using a 29.5 mm DM cup was later performed in the right hip with no postoperative complications encountered.

Four cases involved dogs with traumatic hip luxations. This group consisted of a: 3-year-old male mongrel dog (38 kg) after failed attempts of using an Ehmer sling, 15-month-old female intact Bernese Mountain dog (35 kg) 1 month after a failed Slocum sling, 5-year-old female Bernese Mountain dog (40 kg) with a chronic traumatic hip luxation and a 9-year-old female Golden retriever (46 kg) with a chronic traumatic hip luxation.

The final group of dogs was categorized with conditions known to have a high rate of postoperative dislocation following THA surgery. This consisted of 7 cases with capital physeal fractures, 1 case with an unsatisfactory femoral head and neck excisional arthroplasty, and 22 cases with luxoid hips in large breed dogs.

There was one dislocation that occurred 27 months following surgery in a 1-year-old male intact Bernese Mountain Dog that had a left THA using a 29.5 mm DM cup, medium neck, and a large stem. There was evidence of caudal ventral subluxation observed on pelvic frog view radiograph (normal in the extended radiograph) 9 months following surgery but the dog was asymptomatic. The dog became acutely lame at 27 months following surgery and radiographs at this time revealed dorsal dislocation of the implant. This case was successfully revised inserting a new 25 mm DM head, and a long neck.

One dislocation of the prosthesis was encountered (1 of 105 cases) in the luxoid group of dogs. No other complications were noted in the short-term follow-up period in the remaining dogs. There were no notable changes in any of the remaining dogs (104 cases) on the follow-up radiographs with regard to wear or observed changes in the dimensions of the components of the DM cup or head in any of the cases. As well, there was no evidence of femoral loosening or screw loosening in any of the cases (105 cases).


Discussion

While THA is an extremely successful operation in dogs as in people, dislocation remains one of the most frustrating complications. DM articulation lowers the risk of dislocation by creating a favourable femoral head-to-neck ratio, which decreases impingement.[2] [20] [30] [37]

There are several mechanisms responsible for the increased stability afforded by the DM components. Several authors have noted that stability is improved with an increased femoral head-to-neck ratio, a concept known as the McKee-Farrar principle, which increases the jump distance required for the femoral head to separate from the acetabular component.[2] [24] [33] [38] [39] The design of the DM cup provides a favourable head-to-neck ratio by allowing movement between the metal component and the outer polyethylene or PEEK bearing, the size of which determines the effective diameter of the femoral head. The increased jump distance decreases the risk of luxation when the femoral head or neck impinges on the acetabular component. The incidence of impingement is also decreased by increasing the ROM of the dual-bearing head within the acetabular component. With the Zurich cementless DM cup, the ROM is 145 degrees compared with 135 degrees with the standard cup and 19 mm head. The smaller, inner head provides most of the effective movement.

The most severe complication described in humans is intraprosthetic dislocation, which differs from the more common extra-articular dislocation. Intraprosthetic dislocation is also referred to as retentive failure, which is unique to DM cups and originally described in humans.[30] [39] These types of dislocation are extremely difficult to reduce using closed techniques, and require an open reduction. The newer models of DM cups have vastly decreased the rates of dislocation in comparison with the original design. Most reports of the newer designs describe a rate of luxation of 0%.[16] [29] [40] Intraprosthetic dislocation was not observed in any of our cases, since the ceramic head and CFR-PEEK head are preassembled by the manufacturer. One case of wear and breakage of the polyethylene inner liner (15 mm internally and 20 mm externally) of a cemented DM cup was described by Vedrine and colleagues in a young hunting English Setter 4 months after surgery.[41] According to the authors, the use of a long neck with flanges led to the failure of the ultrahigh molecular weight polyethylene, promoting abnormal wear of the component during extreme ROM. This may explain the unusual complication encountered. However, as the CFR-PEEK head in Zurich DM cup degrades due to internal wear, its ability to constrain the ceramic head may decrease, which increases the risk of intraprosthetic dislocation. To reduce this risk, the carbon fibre reinforced ring is manufactured inside the PEEK inner liner, where the load of the ceramic head is mainly concentrated[35] ([Fig. 6]). Also, the abnormal wear between the PEEK head and the metal acetabular component could lead to wear of the PEEK head, but long-term follow-up of these cases will be required to evaluate this possibility.

Zoom
Fig. 6 (A) The inner components of the dual mobility system consist of the ceramic head (VERILOX) and carbon fibre ring-polyether ether ketone outer head. (B) The head components are preassembled by the manufacturer and are inserted after the outer titanium shell is impacted in the acetabulum.

Abnormal wear causes the production of wear particles, which may lead to failure in bone integration to the prosthesis. Aseptic loosening may be associated with particulate debris caused by degradation of bearing surfaces, but an association between aseptic loosening and modern DM components with highly crosslinked polyethylene is not described in humans.[26] [30] [42] [43] Another theoretical concern is that PE wear may be proportional to the surface area of the articulation, which is higher in DM cups compared with single-bearing design cups.[26] [42]

A DM cup by nature allows wear to occur on both surfaces of the head (inside and outside), and on the articulating head, which potentially could increase the wear of polyethylene. However, in vitro studies demonstrate that PEEK is more resistant to wear compared with polyethylene.[35] The polyethylene wear in dogs leads to reactive bone resorption around the implants, and can ultimately be seen as early as 3 to 4 years after a THA procedure.[15] [44] [45] This is termed ‘wear disease’. PEEK could also lead to aseptic loosening and ‘wear disease’; however, this complication has not been observed in our cases with PEEK liners, even after 4 years following THA procedures using DM cups (personal communication).

Is it possible that a 25 mm single-bearing primary THA head provides the same stability as a DM THA? A correlation between the size of the femoral head and stability is reported in the human orthopaedic literature.[24] [30] In one retrospective study comparing DM cups and single-bearing 36 mm heads in primary THA for patients at high risk for luxation, the rates of luxation and revision in the DM cup patients were much lower when compared with patients receiving a single-bearing 36 mm head.[25]

A potential disadvantage for some, but not all, DM cups used in humans is that they lack screw holes.[28] [30] [31] [32] As with all monolithic cups, the inability to visualize the floor of the acetabulum during the impaction process and not being able to anchor the outer shell with screws may compromise overall fixation of the DM cup. This is less likely to be a cause of failure in primary THA procedures than with revision procedures, where there may be significant bone loss associated with the acetabulum. The Zurich cementless DM cup is available also as a revision cup, with the two components of the shell to be assembled in surgery after the outer shell is secured to the acetabulum using 2.0 or 2.4 mm cortical screws. In our cases, one Zurich cementless DM revision cup (29.5 mm) was used in a Dogue de Bordeaux, which luxated after having a Zurich cementless 26.5 mm cup and 19 mm head. The revision DM cup was stable at the 6-month follow-up evaluation.

Large and giant breed dogs, especially young dogs (<1 year of age) with very unstable hips (luxoid hips), were shown in a previous study evaluating the standard Zurich cementless hip system to have an incidence of dislocation of 8.5% at 2 months (personal communication). In our preliminary experience with the Zurich cementless DM cup, we observed one dislocation of the prosthesis in a dog with a known increased risk of dislocation, but no other dogs in this group had a dislocation. Guillaumot and colleagues reported zero cases of dislocation with the cemented DM cup in the postoperative period.[20] In our cases, dislocation was observed 27 months following surgery in one dog with luxoid hips. This case dislocated due to repeated episodes of subluxation between the PEEK/CFR head and the cup of the DM system. This repeated subluxation leads to deformity of the PEEK/CFR liner causing the inner ceramic head to be locked inside the PEEK/CFR liner, leading to the loss of mobility within the DM cup. The loss of full mobility of the DM cup leads to impingement and dislocation of the prosthesis. Soft tissue entrapment can also lead to loss of mobility of DM cups leading to dislocation. Careful attention not to entrap soft tissues in the DM cup during reduction of the prosthesis is imperative to prevention of dislocation. The authors recommend preassembling the neck to the head, already inserted in the PEEK/CFR liner, prior to inserting the head and PEEK/CFR liner in the outer shell. Reduction is then performed by toggling the femoral peg into the neck, completing the reduction.

In the available follow-up of our cases (103 THA procedures in 85 dogs), no other complication was reported. Nevertheless, longer-term follow-ups are required to establish the longevity of this new cementless DM cup, which represents a limitation of this clinical study.


Conclusion

Dual mobility cups may be an alternative to standard single-bearing cups in dogs reported to be at high risk for luxation in both revision and primary arthroplasty. The complication rate, including dislocation, is relatively low in the cases reported in this study (1/103, 1%).



Conflict of Interest

None declared.

Acknowledgments

The authors would like to thank Slobodan Tepic and Stephen Bresina who developed this Dual Mobility System, Reto Hitz who did the drawings of [Fig. 1] and Kyon (CH), producer of the Zurich Cementless THR, who provided the photos of the implants.

Authors' Contributions

O.L. and A.V. conceptualized and designed the study. I.F. and A.V. acquired, analyzed and interpreted the data. All authors drafted, revised and approved the submitted manuscript and are publically accountable for relevant content.



Address for correspondence

Aldo Vezzoni
SCMPA, Dipl. ECVS, Clinica Veterinaria Vezzoni
Via delle Vigne 190, Cremona, CR 26100
Italy   

Publication History

Received: 23 June 2020

Accepted: 15 January 2021

Article published online:
31 March 2021

© 2021. Thieme. All rights reserved.

Georg Thieme Verlag KG
Rüdigerstraße 14, 70469 Stuttgart, Germany


Zoom
Fig. 1 Technical drawings (complete description in the text). (A,1–4) The four components of the Zurich dual mobility cup system. (A, 5) Ceramic (Ceravet) head of the articulation. (A, 6) Neck of the prosthesis. (A, 7) Spike guided into its receiving hole in the outer shell for concentric assembly of the shells, by either the manufacturer or in surgery for the revision-style cup; G1, small gap at the outer surface of the polyether ether ketone (PEEK) head component; G2, small gap at the primary articulation surface between ceramic head and carbon fibre ring-PEEK ring. (B) Eighty degrees of impingement-free range of angulation between ceramic head and PEEK head. (C) Impingement-free range increased to 132 degrees when combined with range of PEEK head in the outer metal cup; (C, 8) The jumping distance (yellow arrow) is the degree of lateral translation of the femoral head centre required for the hip to dislocate.
Zoom
Fig. 2 (A) Dual mobility cup and dual mobility head. (B) The neck inserted into head of the dual mobility head. The neck is first inserted into the ceramic head followed by insertion of the peg of the femoral prosthesis into the neck.
Zoom
Fig. 3 Dual Mobility Cup. (A) The inner shell (gold in colour) is impacted inside the outer shell (white in colour). (B) The inner shell is aligned in the coaxial direction by a spike in the pole of the inner shell, aligning the centre of the outer shell.
Zoom
Fig. 4 Pelvic radiographs of a 29-month-old male Bernese Mountain dog, 39.6 kg, with bilateral total hip replacement using dual mobility cups. Left hip was done at 11 months of age and the right hip at 14 months of age.
Zoom
Fig. 5 Same dog as in [Fig. 4]. Follow-up with oblique pelvic views tangent to the implants: left after 18 months and right after 15 months reveal normal bilateral total hip replacements.
Zoom
Fig. 6 (A) The inner components of the dual mobility system consist of the ceramic head (VERILOX) and carbon fibre ring-polyether ether ketone outer head. (B) The head components are preassembled by the manufacturer and are inserted after the outer titanium shell is impacted in the acetabulum.