J Knee Surg 2022; 35(14): 1563-1570
DOI: 10.1055/s-0041-1728785
Original Article

Is Valgus Cut Angle Based on Radiographic Measurements in Total Knee Arthroplasty Really Inaccurate? A Comparison of Two- and Three-Dimensional Measurements

LiMing Liu*
1   Center for Joint Surgery, Southwest Hospital, Third Military Medical University, Chongqing, China
,
Kai Lei*
1   Center for Joint Surgery, Southwest Hospital, Third Military Medical University, Chongqing, China
,
Xin Chen
1   Center for Joint Surgery, Southwest Hospital, Third Military Medical University, Chongqing, China
,
HuaQuan Fan
1   Center for Joint Surgery, Southwest Hospital, Third Military Medical University, Chongqing, China
,
Liu Yang
1   Center for Joint Surgery, Southwest Hospital, Third Military Medical University, Chongqing, China
,
1   Center for Joint Surgery, Southwest Hospital, Third Military Medical University, Chongqing, China
› Author Affiliations

Funding This study was funded by National Key R&D Program of China (no.: 2016YFB1101404) and Guangdong Provincial Key R&D Program (no.: 2018B090906003).
Preview

Abstract

Radiographs are widely used to measure distal femoral valgus cut angle (VCA) in total knee arthroplasty (TKA), but its accuracy is controversial. This study used three-dimensional (3D) reconstruction models to verify the accuracy of VCA measurements on radiographs, and explore the correlation of VCA with hip–knee–ankle (HKA) angle and lateral femoral bowing angle (FBA). A total of 444 osteoarthritis knees of 444 patients from August 2016 to June 2018 was included retrospectively. On radiographs, two-dimensional VCA (VCA-2D) was measured between the femoral mechanical axis and the distal femoral anatomical axis, and HKA was measured between the femoral mechanical axis and the tibial mechanical axis. On the coronal projection of computed tomography 3D models, the anatomical landmarks used for VCA-3D measurements were the same as those on the radiographs, FBA was measured between the proximal and distal femoral anatomical axis. The distributions of VCA-2D and VCA-3D were evaluated by means and variances. The correlation between HKA and VCA and between FBA and VCA was explored. There was a statistical difference between VCA-2D and VCA-3D (p < 0.001), but the deviation was very small (0.15 ± 0.69 degrees), 83.3% of the deviations were less than 1 degree. VCA would increase both in 2D and 3D with increasing of FBA and HKA varus. There was no statistically significant difference between VCA-2D and VCA-3D in patients with moderate varus knees (0–8 degrees of varus) and mild bowing femurs (FBA <5 degrees). Overall, the deviation caused by using radiography to measure VCA was negligible. VCA measurements using radiographs were accurate in patients with moderate varus knees and mildly bowed femurs. This study reflects level of evidence III.

Ethical Approval

Ethical approval was obtained from the local ethical committee. All procedures performed were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Declaration of Helsinki and its later amendments or comparable ethical standards.


* These authors contributed to the work equally and should be regarded as co–first authors.




Publication History

Received: 16 July 2020

Accepted: 12 March 2021

Article published online:
29 April 2021

© 2021. Thieme. All rights reserved.

Thieme Medical Publishers, Inc.
333 Seventh Avenue, 18th Floor, New York, NY 10001, USA