Automatic-Scoring Actigraph Compares Favourably to a Manually-Scored Actigraph for Sleep Measurement in Healthy Adults

1   Te Huataki Waiora, School of Health, University of Waikato, Hamilton, Waikato, New Zealand
2   Joint Support Group, Human Performance Cell, New Zealand Army, Upper Hutt, Wellington, New Zealand
,
1   Te Huataki Waiora, School of Health, University of Waikato, Hamilton, Waikato, New Zealand
,
Nicholas D. Gill
1   Te Huataki Waiora, School of Health, University of Waikato, Hamilton, Waikato, New Zealand
,
Jennifer L. Zaslona
3   Sleep/Wake Research Centure, Massey University, Wellington, Wellington, New Zealand
,
4   Sport, Performance, and Nutrition Research Group, School of Allied Health, Human Services, and Sport, La Trobe University, Melbourne, Australia
› Author Affiliations
 

Abstract

Introduction Actigraphy has been used widely in sleep research due to its non-invasive, cost-effective ability to monitor sleep. Traditionally, manually-scored actigraphy has been deemed the most appropriate in the research setting; however, technological advances have seen the emergence of automatic-scoring wearable devices and software.

Methods A total of 60-nights of sleep data from 20-healthy adult participants (10 male, 10 female, age: 26 ± 10 years) were collected while wearing two devices concomitantly. The objective was to compare an automatic-scoring device (Fatigue Science Readiband™ [AUTO]) and a manually-scored device (Micro Motionlogger® [MAN]) based on the Cole-Kripke method. Manual-scoring involved trained technicians scoring all 60-nights of sleep data. Sleep indices including total sleep time (TST), total time in bed (TIB), sleep onset latency (SOL), sleep efficiency (SE), wake after sleep onset (WASO), wake episodes per night (WE), sleep onset time (SOT) and wake time (WT) were assessed between the two devices using mean differences, 95% levels of agreement, Pearson-correlation coefficients (r), and typical error of measurement (TEM) analysis.

Results There were no significant differences between devices for any of the measured sleep variables (p ≥0.05). All sleep indices resulted in very-strong correlations (all r ≥0.84) between devices. A mean difference between devices of <1 minutes for TST was associated with a TEM of 15.5 minute (95% CI =12.3 to 17.7 minutes).

Conclusion Given there were no significant differences between devices in the current study, automatic-scoring actigraphy devices may provide a more practical and cost-effective alternative to manually-scored actigraphy in healthy populations.


#

Introduction

The measurement and quantification of sleep in population research and clinical settings is of increasing importance due to its integral role in physical and mental health.[1] Diverse methods of monitoring and researching sleep have been extensively investigated and validated in the literature.[2] Although polysomnography (PSG) is regarded as the ‘gold standard’ of sleep measurement, it is a somewhat intrusive and expensive form of assessment.[3] [4] Additionally, PSG typically requires an individual to sleep in an unfamiliar laboratory-based clinic while sleep is being assessed via the use of multiple electrodes to monitor neurophysiological and cardiorespiratory variables, which may be difficult and invasive for many individuals and may compromise the ecological validity of the data attained outside of a strictly pathological sleep assessment.[2] [5] Over the last decade, many emerging sleep-monitoring devices, such as commercially-available wearables, have demonstrated promising capability for tracking sleep and wake episodes.[1] A popular method of minimally-invasive sleep monitoring is via wrist actigraphy, where wearable devices allow for continuous monitoring of sleep movement during sleep with either automatic or manual-scoring options available.[3]

Although various products are now available on the market, actigraphy generally involves a device being housed in a wristwatch, bracelet, or ring that contains an accelerometer capable of sensing movement along each of three axes.[1] [6] The tri-axial accelerometer samples multiple times per second and with each limb movement, the accelerometer estimates metrics of sleep and wake including total sleep time (TST), total time in bed (TIB), sleep efficiency (SE%), wake after sleep onset (WASO), and sleep onset latency (SOL), as well as sleep and wake times. Data are then stored in the device memory to be downloaded and either automatically or manually scored.[1]

The advantage of actigraphy over traditional PSG is that it can record continuously for 24 hours a day for days, weeks, or even longer,[7] and can easily be utilized to monitor sleep-wake patterns in home-based or ecologically valid settings. As a result, it has also been proposed that actigraphy could be adapted for use in primary care settings to improve sleep health in the community.[8] To date, actigraphy has been used widely in sleep research to provide continuous monitoring of rest/wake activity rhythms in varying environments; including residential care patients,[9] elite athletes,[10] shift workers,[11] [12] and in operational settings such as firefighting[13] and the military.[14]

Manually scored actigraphy has historically been used in sleep research settings.[15] [16] Despite this, it can be difficult to make conclusions on the overall reliability of manually-scored actigraphy data given variations in methods of scoring, different brands of hardware, varying software, and inter/intra-scorer reliability.[15] Studies have shown that high inter-rater agreement for manually scored data (e.g., α = 0.975 for rest onset, and α = 0.998 for rest offset) can be achieved with clearly defined scoring criteria by trained researchers.[17] However, a limitation of manually scored actigraphy is the possibility of human error and the time requirement of analyzing large groups of participant data.[5] [7] Recent advances in technology have seen the emergence of automatically-scored, commercially-available actigraph devices[5] [14] and the accuracy and relability of these devices has improved considerably.[1] [18] These developments include devices specifically tailored to detect periodic limb movements and the introduction of new algorithms.[19] Most sleep-wake scoring algorithms are based on a combination of linear compilations of activity levels (in predefined windows around the scored minute) and smoothing or other logical decisions.

Many of the commercial wearable devices on the market, such as the Readiband™ (RB), contain tri-axial accelerometers that record the frequency and intensity of limb movement that can be converted to sleep-wake periods using a built-in automated scoring algorithm. In a study by Chinoy et al.[1] evaluating seven consumer sleep-tracking devices, the RB performed comparably to other devices and displayed high intraclass correlations (>0.93) for overall epoch-by-epoch sensitivity.[1] In one previous study of 50 adults who wore an automatically scored RB device and a manually scored device (ActiGraph GT3X + ) for 7-nights, sleep onset, sleep duration and wake time were compared.[3] The RB performed similarly to the manually scored device when measuring these sleep metrics, during an unfamiliar laboratory night stay and when worn at home in a familiar environment. It was concluded that the RB could be used in the same capacity as the ActiGraph for the collection of sleep metrics.[3]

With the emergence of new sleep-monitoring technologies, it is important to understand what differences exist between manually and automatically-scored devices to enable decisions regarding whether the data obtained is comparable. Therefore, the aim of the current study was to investigate the differences between sleep metrics from a device using manual scoring of sleep metrics to a commercially available actigraphy device that uses automatic scoring, by evaluating 60-nights of sleep data from 20-healthy adult participants wearing the two devices concurrently.


#

Methods

Participants

A total of 20 healthy adults (10 male, 10 female, age: 26 ± 10 years [mean ± SD]), participated in the current study. Participation in the study was voluntary and all participants provided written consent before taking part, with inclusion dependent on being free from any diagnosed sleep disorders. Ethical approval for the study was obtained from the University's Human Research Ethics Committee (HREC) (Health) #2018–0. Sample size was calculated for the current study using an a priori analysis based on an expected r-value of 0.8, a precision value of ± 0.2 and 95% confidence levels using a web-based calculator.[20] This calculation resulted in an estimated n = 18 participants.


#

Experimental Design

Participants were required to wear two different wrist actigraphs, the automatically scored Readiband™ (AUTO) and the manually scored Micro MotionLogger® (MAN), and have sleep recorded for a 3-day/night period concurrently, similar to the procedures of Dennis et al.[18] In the current study, both the MAN and AUTO devices were tightly secured together with the MAN on top of the AUTO, using electrical tape so that the devices could not move independently of each other. Devices were initialised before being worn on whichever wrist felt comfortable,[21] and data commenced recording in 1-minute epochs.[18] Total sleep time (TST), sleep efficiency (SE), time in bed (TIB), sleep onset latency (SOL), wake after sleep onset (WASO), wake episodes (WE), sleep onset time (SOT), and wake time (WT) were all assessed ([Table 1]). The devices were removed for any water submersion activities and placed back on the wrist immediately post activity. Participants were instructed to maintain their usual sleep habits and general daily activity patterns during the 3-day monitoring period, before actigraphs were removed and data downloaded.

Table 1

Definitions of each sleep variable measures to be compared and validated between the Fatigue Science Readiband™ (AUTO) and Micro Motionlogger® (MAN) actigraphy devices.

Sleep indices

Units

Description

Total Sleep Time (TST)

Minutes

Total time spent asleep

Total Time in Bed (TIB)

Minutes

Total time spent in bed

Sleep Efficiency (SE)

%

Total sleep time divided by total time in bed

Sleep Onset Latency (SOL)

Minutes

Time taken for sleep onset

Wake Episodes per Night (WE)

Number count

Total number of awakenings per night

Wake After Sleep Onset (WASO)

Minutes

Time spent awake after sleep onset per night

Sleep Onset Time (SOT)

Time of day (p.m.)

Time fell asleep at night

Wake Time (WT)

Time of day (a.m.)

Time woken in morning


#

Automatic Scoring Actigraphy

The AUTO actigraph (Readiband™ version-5, Fatigue Science, Honolulu, USA), has been previously used in sleep research,[8] [10] [22] and records data at a sample rate of 16 Hz. The AUTO uses a patented algorithm to automatically score sleep data derived from raw acceleration signals via specialized Readiband Sync™ software.[14] [23] The AUTO device has shown accuracy in distinguishing sleep from wakefulness ∼82% of the time when epoch scoring against PSG.[24] The RB has also been approved by the US Federal Drug Administration for measurement of sleep.[23]


#

Manually Scored Actigraphy

The MAN actigraphy (Micro MotionLogger®, Ardsley, New York, USA) uses a tri-axial accelerometer which has also been compared with PSG, and distinguished sleep from wakefulness accuracy ∼80% of the time.[24] Data were collected using the device's zero-crossing mode ad recorded in 1-minute epochs.[25] Using the manufacturer's software (Action-W version 2.7.3045, Ambulatory Monitoring Inc., Ardsley, New York, USA), sleep and wakefulness were estimated based on activity count using the Cole-Kripke algorithm.[26] Manual scoring of the sleep data involved one technician scoring all 60 night's sleep files individually for ‘start time’ and ‘end time’ of the rest interval, and for any wake periods throughout the rest interval for each participant.[7] Points were placed on the computer file to mark the intervals the participants were in bed and the times the device was removed. To then assess the reliability of manual selection of rest intervals, a randomly selected 33% (20 sleep files) were double scored by a second independent trained researcher. Any discrepancies of more than 15 minutes for either ‘start time’ or ‘end time’ of the rest interval were flagged and re-analyzed by both technicians. If agreement could not be reached on any files, a third independent researcher would have been used for scoring; however, this did not occur. A total of four files were re-analyzed by both researchers with a final accuracy rate of 87.9% achieved between the two researchers, this threshold has previously been described as acceptable.[27]


#

Statistical Analysis

Simple group descriptive statistics are shown as means ± standard deviations unless stated otherwise. A paired t-test was used to compare AUTO and MAN metrics using a Statistical Package for Social Science (V. 22.0, SPSS Inc., Chicago, IL), with statistical significance set at p < 0.05. Inter-device agreements for AUTO and MAN were examined using Pearson correlation coefficients (r) with 95% confidence intervals (95% CI) and interpreted using thresholds of <0.30: poor, >0.30: fair, >0.50: moderately strong, >0.80: very strong.[28] The mean differences/bias and upper and lower limits of agreement (1.96 standard deviations or 95% of a normally distributed population) between devices were determined in absolute values for TST, SE, TIB, SOL, WE, and WASO. Between-device typical error of measurement (TEM) was determined using a customized excel spreadsheet.[29] Consistent with previous research, we defined an a priori difference between the two devices of <30 minute for TST, and <5% for SE as satisfactory.[30]


#
#

Results

There were no significant differences between devices (AUTO and MAN) for any of the measured sleep variables (p > 0.05, [Table 2]). There was a mean difference between devices of less than 1-minute over the 60 nights of data for TST and 1.1% for SE ([Table 2]), with very strong correlations between devices for both these measures ([Table 3]).

Table 2

Mean ± SD values for both the automatically scored Readiband™ (AUTO) and the manually scored Micro Motionlogger® (MAN) actigraphy devices, for all measured sleep variables and p-values for each comparison.

Sleep Indices

AUTO

MAN

P-Value

Total Sleep Time (min)

438.6 ± 87.5

439.1 ± 90.6

0.974

Sleep Efficiency (%)

91.1 ± 5.1

92.2 ± 5.2

0.240

Total Time in Bed (min)

459.1 ± 96

465.3 ± 92.6

0.717

Sleep Onset Latency (min)

19.4 ± 14.5

16.7 ± 11.9

0.145

Wake Episodes per Night (Number)

8.0 ± 4.9

9.7 ± 5.0

0.061

Wake After Sleep Onset (min)

30.0 ± 23.3

32.6 ± 20.7

0.528

Sleep Onset Time (Time of day)

21:49 ± 0:48

21:59 ± 0:42

0.231

Wake Time (Time of day)

5:49 ± 1:01

5:55 ± 1.02

0.634

Table 3

Typical error of measurement (TEM), mean difference, range of difference and Pearson correlations for each sleep metric between automatically scored Readiband™ (AUTO) and manually scored Micro Motionlogger® (MAN) actigraphy devices.

TEM (95% CI)

Mean difference ( ± SD)

Range of mean difference (1.96xSD)

Pearson correlation coefficient (95% CI)

Total Sleep Time (min)

15.5 (12.3 to 17.7)

0.53 ± 20.6

-39.7 to 40.8

0.97 (0.96 to 0.98)

Sleep Efficiency (%)

2.7 (2.3 to 3.4)

1.1 ± 2.9

-4.6 to 6.8

0.84 (0.74 to 0.90)

Total Time in Bed (min)

29.6 (25.0 to 36.2)

6.2 ± 29.4

-51.4 to 63.9

0.95 (0.92 to 0.97)

Sleep Onset Latency (min)

6.4 (5.4 to 7.8)

-2.6 ± 6.5

-15.4 to 10.1

0.90 (0.83 to 0.94)

Wake Episodes per Night (No)

1.4 (1.2 to 1.7)

1.6 ± 1.4

-1.2 to 4.5

0.96 (0.93 to 0.97)

Wake After Sleep Onset (min)

12.6 (10.7 to 15.4)

2.5 ± 12.6

-22.2 to 27.2

0.84 (0.75 to 0.90)

The variables of SOL, TIB, and WE resulted in very strong correlations between devices and a mean difference of <6.2 minute ([Table 3]). Comparison between devices for these variables also resulted in TEM's of <29.6 minute ([Table 3]). The remaining variables: SE and WASO, also resulted in very strong correlations between devices, with TEM values of 2.7%and 12.6 minute, respectively ([Table 3]). Level of agreement (Bland-Altman) plots showing ± 95% limits of agreement between AUTO and MAN for key sleep variables of TST, SOL, and SE are displayed in [Fig. 1].

Zoom Image
Fig. 1 Level of agreement (Bland-Altman) plots showing ± 95% limits of agreement between automatically scored Readiband™ (AUTO) and manually-scored Micro Motionlogger® (MAN) for a) total sleep time (TST); b) sleep latency (SL); c) sleep efficiency (SE).

#

Discussion

This study examined the differences between a commercially available, automatic-scoring actigraph when compared with a manually scored actigraph in healthy adult participants while wearing both devices concurrently. The aim of the study was to simply compare the metrics coming from the two devices and not to evaluate the overall validity or accuracy of actigraphy as a method of monitoring sleep. The correlation between these manually and automatically scored devices was very strong for all sleep variables with no significant differences in any of the measured sleep variables between devices. The automatically scored device performed comparably to the manually scored device in the current study, suggesting a practical alternative to achieve similar levels of accuracy without the time demand or expertise of a trained technician required to score the actigraphy trace.

Werner et al.[29] stated that a difference between two devices of < 30 minute for TST and a difference < 5% for SE can be considered satisfactory. Indeed, results from the current study were under < 30 minute for TST and < 5% for SE, with mean differences between devices of less than 1-minute over the 60 nights for TST and 1.1% for SE. Accordingly, based on the suggestions of Werner et al. (2008), the differences between devices for these identified key sleep metrics in the current study can be deemed acceptable, and the AUTO can be considered an appropriate alternative for use in both practical and research settings.

In the current study, TST, TIB, SOL, WASO, WE and SE for AUTO and MAN indicated no significant difference and all Pearson correlation coefficients were very strong. Dunican et al.[3] compared the automatically-scored RB to a different manually-scored actigraph (ActiGraph GT3X + ) during a laboratory observation night, and when worn at home for 7-nights in a healthy adult population. Dunican et al.[3] reported that TST showed no difference between the devices in the laboratory condition (p = 0.58), but a longer sleep duration (38 ± 61-minute, p < 0.001) and differences for time at lights-out for RB in the at-home condition. For SE, the RB estimated 5–12% less (p <0.001), with longer SOL (22–36 minute, p < 0.05) than the ActiGraph GT3X+ device in both conditions.[3] It was concluded that the differences between the at-home and laboratory condition between devices were due to inaccuracies of the RB reporting time at lights out compared with ActiGraph's requirement to self-report (e.g., using a marker button on the watch), and highlighted the challenge of accurately defining these metrics due to different assessment methods.[3] When compared with the current study, our results indicated no significant difference between devices, suggesting that there were similarities between the proprietary AUTO algorithms and MAN scoring. Thus, the results align with the conclusion of Dunican et al.,[3] who stated that the RB automated algorithm may be used in the same capacity as a manually scored actigraph for the collection of key sleep measures.

Previous research has identified the inter-device reliability of the AUTO device used in the current study.[5] In the study by Driller et al.,[5] participants wore two RB devices concomitantly for 77 nights of sleep where sleep data was assessed. The Driller et al. study[5] found no significant differences between devices for any of the measured sleep variables (p < 0.05). Mean differences of 2.1 and 0.2 minute for TST and SL were associated with a low TEM between devices (9.5 and 3.8 minute, respectively). Interestingly, the non-significant differences between devices observed in the Driller et al.[5] study for all sleep metrics, are remarkably similar to those observed in the current study. Driller et al.[5] also reported very high intraclass correlations between devices indicating the RB to have acceptable inter-device reliability. In comparison, while the current study had slightly higher TEMs for some of the measures, this is somewhat expected, as the current study included two different brands of device, with different algorithms and methods of scoring (manual versus automatic). It is therefore promising that the differences were similar when comparing the inter-device reliability (RB versus RB) and the comparison of two different devices (MAN versus AUTO). RB has been shown to be reliable (inter-device), and RB compares favorably to MAN, indicating RB can be considered a valuable tool for monitoring sleep metrics.

The current study is not without limitations, these include comparing AUTO vs MAN devices and associated algorithms in only an at-home environment and not in a laboratory condition, where differing results may have occurred between devices.[3] We also acknowledge that not comparing to PSG was a limitation, however, this comparison has already been made for both devices[3] [24] [25] and the main aim of the current study was to compare AUTO and MAN devices.

In conclusion, AUTO may provide an accurate and practical solution for use in healthy populations to give an indication of sleep/wake patterns. An automatically scored device such as the RB does not require any expertise, and when compared with the time required for manual scoring of actigraphy traces, the RB may provide a more time-efficient alternative, therefore allowing large groups of individuals to be monitored effectively with comparable accuracy.


#
#

Conflict of Interests

The authors declare that there is no conflict of interest associated with this work.

Acknowledgments

DE designed the study, collected the data, analyzed the data and wrote the final version of the manuscript. NG designed the study and contributed to the writing of the manuscript. CMB & MD designed the study, analyzed the data and contributed to the writing of the manuscript. JZ contributed to the data collection and analysis and writing of the manuscript.

  • References

  • 1 Chinoy ED, Cuellar JA, Huwa KE. et al. Performance of seven consumer sleep-tracking devices compared with polysomnography. Sleep 2021; 44 (05) 1-16
  • 2 Van de Water AT, Holmes A, Hurley DA. Objective measurements of sleep for non-laboratory settings as alternatives to polysomnography--a systematic review. J Sleep Res 2011; 20 (1 Pt 2, 1pt2) 183-200
  • 3 Dunican IC, Murray K, Slater JA. et al. Laboratory and home comparison of wrist-activity monitors and polysomnography in middle-aged adults. Sleep Biol Rhythms 2018; 16: 85-97 DOI: 10.1007/s41105-017-0130-x.
  • 4 O'Donnell S, Beaven CM, Driller MW. From pillow to podium: a review on understanding sleep for elite athletes. Nat Sci Sleep 2018; 10: 243-253
  • 5 Driller M, McQuillan J, O'Donnell S. Inter-device reliability of an automatic-scoring actigraph for measuring sleep in healthy adults. Sleep Sci 2016; 9 (03) 198-201
  • 6 Palotti J, Mall R, Aupetit M. et al. Benchmark on a large cohort for sleep-wake classification with machine learning techniques. NPJ Digit Med 2019; 2: 50 DOI: 10.1038/s41746-019-0126-9.
  • 7 Ancoli-Israel S, Cole R, Alessi C, Chambers M, Moorcroft W, Pollak CP. The role of actigraphy in the study of sleep and circadian rhythms. Sleep 2003; 26 (03) 342-392
  • 8 Noor Z, Smith A, Smith S, Nissen L. Feasibility and acceptability of wrist actigraph in assessing sleep quality and sleep quantity: A home-based pilot study in healthy volunteers. Health 2013; 5: 63-72 DOI: 10.4236/health.2013.58A2010.
  • 9 Wulff K, Dijk DJ, Middleton B, Foster RG, Joyce EM. Sleep and circadian rhythm disruption in schizophrenia. Br J Psychiatry 2012; 200 (04) 308-316
  • 10 Fullagar HH, Duffield R, Skorski S. et al. Sleep, travel, and recovery responses of national footballers during and after long-haul international air travel. Int J Sports Physiol Perform 2016; 11 (01) 86-95
  • 11 Arendt J. Shift work: coping with the biological clock. Occup Med (Lond) 2010; 60 (01) 10-20
  • 12 Tremaine R, Dorrian J, Paterson J. et al. Actigraph estimates of the sleep of Australian midwives: the impact of shift work. Biol Res Nurs 2013; 15 (02) 191-199
  • 13 McGillis Z, Dorman SC, Robertson A. et al. Sleep quantity and quality of Ontario wildland firefighters across a low-hazard fire season. J Occup Environ Med 2017; 59 (12) 1188-1196
  • 14 Adler AB, Gunia BC, Bliese PD, Kim PY, LoPresti ML. Using actigraphy feedback to improve sleep in soldiers: an exploratory trial. Sleep Health 2017; 3 (02) 126-131
  • 15 Sadeh A. The role and validity of actigraphy in sleep medicine: an update. Sleep Med Rev 2011; 15 (04) 259-267
  • 16 Chow CM, Wong SN, Shin M. et al. Defining the rest interval associated with the main sleep period in actigraph scoring. Nat Sci Sleep 2016; 8: 321-328
  • 17 Follesø HS, Austad SB, Olsen A, Saksvik-Lehouillier I. The development, inter-rater agreement and performance of a hierarchical procedure for setting the rest-interval in actigraphy data. Sleep Med 2021; 85: 221-229
  • 18 Dennis J, Dawson B, Heasman J, Rogalski B, Robey E. Sleep patterns and injury occurrence in elite Australian footballers. J Sci Med Sport 2016; 19 (02) 113-116
  • 19 Athavale Y, Krishnan S, Raissi A. et al. Actigraphic detection of periodic limb movements: development and validation of a potential device-independent algorithm. A proof of concept study. Sleep 2019; 42 (09) zsz117
  • 20 Arifin W. . A Web-based Sample Size Calculator for Reliability Studies. Educ Med J 2018
  • 21 Driller MW, O'Donnell S, Tavares F. What wrist should you wear your actigraphy device on? Analysis of dominant vs. non-dominant wrist actigraphy for measuring sleep in healthy adults. Sleep Sci 2017; 10 (03) 132-135
  • 22 Fowler P, Duffield R, Vaile J. Effects of simulated domestic and international air travel on sleep, performance, and recovery for team sports. Scand J Med Sci Sports 2015; 25 (03) 441-451
  • 23 Lee S, Bonnar D, Roane B, Gradisar M, Dunican IC, Lastella M, Maisey G, Suh S. Sleep characteristics and mood of professional esports athletes: A multi-national study. Int J Environ Res Public Health 2021; Jan 14; 18 (02) 664 DOI: 10.3390/ijerph18020664. . PMID: 33466788; PMCID: PMC7830734
  • 24 Russell C, Caldwell J, Arand D, Myers L, Wubbels P, Downs HJAL. Validation of the fatigue science readiband actigraph and associated sleep/wake classification algorithms. 2000. Archinoetics, LLC;
  • 25 Meltzer LJ, Walsh CM, Traylor J, Westin AM. Direct comparison of two new actigraphs and polysomnography in children and adolescents. Sleep 2012; 35 (01) 159-166
  • 26 Cole RJ, Kripke DF, Gruen W, Mullaney DJ, Gillin JC. Automatic sleep/wake identification from wrist activity. Sleep 1992; 15 (05) 461-469
  • 27 McHugh ML. Interrater reliability: the kappa statistic. Biochem Med (Zagreb) 2012; 22 (03) 276-282
  • 28 Chan YH. Biostatistics 104: correlational analysis. Singapore Med J 2003; 44 (12) 614-619
  • 29 Hopkins W. Spreadsheets for analysis of validity and reliability. Sports Science. 2017;21.
  • 30 Werner H, Molinari L, Guyer C, Jenni OG. Agreement rates between actigraphy, diary, and questionnaire for children's sleep patterns. Arch Pediatr Adolesc Med 2008; 162 (04) 350-358

Address for correspondence

David T. Edgar, PhD
Te Huataki Waiora, School of Health, University of Waikato
Hamilton, Waikato
New Zealand   

Publication History

Article published online:
06 July 2023

© 2023. Brazilian Sleep Association. This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes, or adapted, remixed, transformed or built upon. (https://creativecommons.org/licenses/by-nc-nd/4.0/)

Thieme Revinter Publicações Ltda.
Rua do Matoso 170, Rio de Janeiro, RJ, CEP 20270-135, Brazil

  • References

  • 1 Chinoy ED, Cuellar JA, Huwa KE. et al. Performance of seven consumer sleep-tracking devices compared with polysomnography. Sleep 2021; 44 (05) 1-16
  • 2 Van de Water AT, Holmes A, Hurley DA. Objective measurements of sleep for non-laboratory settings as alternatives to polysomnography--a systematic review. J Sleep Res 2011; 20 (1 Pt 2, 1pt2) 183-200
  • 3 Dunican IC, Murray K, Slater JA. et al. Laboratory and home comparison of wrist-activity monitors and polysomnography in middle-aged adults. Sleep Biol Rhythms 2018; 16: 85-97 DOI: 10.1007/s41105-017-0130-x.
  • 4 O'Donnell S, Beaven CM, Driller MW. From pillow to podium: a review on understanding sleep for elite athletes. Nat Sci Sleep 2018; 10: 243-253
  • 5 Driller M, McQuillan J, O'Donnell S. Inter-device reliability of an automatic-scoring actigraph for measuring sleep in healthy adults. Sleep Sci 2016; 9 (03) 198-201
  • 6 Palotti J, Mall R, Aupetit M. et al. Benchmark on a large cohort for sleep-wake classification with machine learning techniques. NPJ Digit Med 2019; 2: 50 DOI: 10.1038/s41746-019-0126-9.
  • 7 Ancoli-Israel S, Cole R, Alessi C, Chambers M, Moorcroft W, Pollak CP. The role of actigraphy in the study of sleep and circadian rhythms. Sleep 2003; 26 (03) 342-392
  • 8 Noor Z, Smith A, Smith S, Nissen L. Feasibility and acceptability of wrist actigraph in assessing sleep quality and sleep quantity: A home-based pilot study in healthy volunteers. Health 2013; 5: 63-72 DOI: 10.4236/health.2013.58A2010.
  • 9 Wulff K, Dijk DJ, Middleton B, Foster RG, Joyce EM. Sleep and circadian rhythm disruption in schizophrenia. Br J Psychiatry 2012; 200 (04) 308-316
  • 10 Fullagar HH, Duffield R, Skorski S. et al. Sleep, travel, and recovery responses of national footballers during and after long-haul international air travel. Int J Sports Physiol Perform 2016; 11 (01) 86-95
  • 11 Arendt J. Shift work: coping with the biological clock. Occup Med (Lond) 2010; 60 (01) 10-20
  • 12 Tremaine R, Dorrian J, Paterson J. et al. Actigraph estimates of the sleep of Australian midwives: the impact of shift work. Biol Res Nurs 2013; 15 (02) 191-199
  • 13 McGillis Z, Dorman SC, Robertson A. et al. Sleep quantity and quality of Ontario wildland firefighters across a low-hazard fire season. J Occup Environ Med 2017; 59 (12) 1188-1196
  • 14 Adler AB, Gunia BC, Bliese PD, Kim PY, LoPresti ML. Using actigraphy feedback to improve sleep in soldiers: an exploratory trial. Sleep Health 2017; 3 (02) 126-131
  • 15 Sadeh A. The role and validity of actigraphy in sleep medicine: an update. Sleep Med Rev 2011; 15 (04) 259-267
  • 16 Chow CM, Wong SN, Shin M. et al. Defining the rest interval associated with the main sleep period in actigraph scoring. Nat Sci Sleep 2016; 8: 321-328
  • 17 Follesø HS, Austad SB, Olsen A, Saksvik-Lehouillier I. The development, inter-rater agreement and performance of a hierarchical procedure for setting the rest-interval in actigraphy data. Sleep Med 2021; 85: 221-229
  • 18 Dennis J, Dawson B, Heasman J, Rogalski B, Robey E. Sleep patterns and injury occurrence in elite Australian footballers. J Sci Med Sport 2016; 19 (02) 113-116
  • 19 Athavale Y, Krishnan S, Raissi A. et al. Actigraphic detection of periodic limb movements: development and validation of a potential device-independent algorithm. A proof of concept study. Sleep 2019; 42 (09) zsz117
  • 20 Arifin W. . A Web-based Sample Size Calculator for Reliability Studies. Educ Med J 2018
  • 21 Driller MW, O'Donnell S, Tavares F. What wrist should you wear your actigraphy device on? Analysis of dominant vs. non-dominant wrist actigraphy for measuring sleep in healthy adults. Sleep Sci 2017; 10 (03) 132-135
  • 22 Fowler P, Duffield R, Vaile J. Effects of simulated domestic and international air travel on sleep, performance, and recovery for team sports. Scand J Med Sci Sports 2015; 25 (03) 441-451
  • 23 Lee S, Bonnar D, Roane B, Gradisar M, Dunican IC, Lastella M, Maisey G, Suh S. Sleep characteristics and mood of professional esports athletes: A multi-national study. Int J Environ Res Public Health 2021; Jan 14; 18 (02) 664 DOI: 10.3390/ijerph18020664. . PMID: 33466788; PMCID: PMC7830734
  • 24 Russell C, Caldwell J, Arand D, Myers L, Wubbels P, Downs HJAL. Validation of the fatigue science readiband actigraph and associated sleep/wake classification algorithms. 2000. Archinoetics, LLC;
  • 25 Meltzer LJ, Walsh CM, Traylor J, Westin AM. Direct comparison of two new actigraphs and polysomnography in children and adolescents. Sleep 2012; 35 (01) 159-166
  • 26 Cole RJ, Kripke DF, Gruen W, Mullaney DJ, Gillin JC. Automatic sleep/wake identification from wrist activity. Sleep 1992; 15 (05) 461-469
  • 27 McHugh ML. Interrater reliability: the kappa statistic. Biochem Med (Zagreb) 2012; 22 (03) 276-282
  • 28 Chan YH. Biostatistics 104: correlational analysis. Singapore Med J 2003; 44 (12) 614-619
  • 29 Hopkins W. Spreadsheets for analysis of validity and reliability. Sports Science. 2017;21.
  • 30 Werner H, Molinari L, Guyer C, Jenni OG. Agreement rates between actigraphy, diary, and questionnaire for children's sleep patterns. Arch Pediatr Adolesc Med 2008; 162 (04) 350-358

Zoom Image
Fig. 1 Level of agreement (Bland-Altman) plots showing ± 95% limits of agreement between automatically scored Readiband™ (AUTO) and manually-scored Micro Motionlogger® (MAN) for a) total sleep time (TST); b) sleep latency (SL); c) sleep efficiency (SE).