Open Access
CC BY 4.0 · Rev Bras Ortop (Sao Paulo) 2024; 59(01): e107-e118
DOI: 10.1055/s-0043-1775889
Artigo Original
Ortopedia Pediátrica

Comparison of Locked Plates and Blade Plates for Varus Osteotomy of the Proximal Femur by the Finite Element Method

Article in several languages: português | English
1   Grupo de Ortopedia e Traumatologia Pediátrica, Pontifícia Universidade Católica de Campinas, Campinas, SP, Brasil
,
2   Programa de Pós-Graduação em Ciência e Engenharia de Materiais, Universidade Federal de São Carlos, São Carlos, SP, Brasil
,
3   Grupo de Ortopedia e Traumatologia Pediátrica, Faculdade de Ciências Médicas, Santa Casa de São Paulo, São Paulo, SP, Brasil
,
3   Grupo de Ortopedia e Traumatologia Pediátrica, Faculdade de Ciências Médicas, Santa Casa de São Paulo, São Paulo, SP, Brasil
,
4   Ortopedia Pediátrica, Pontifícia Universidade Católica de Campinas, Campinas, SP, Brasil
,
4   Ortopedia Pediátrica, Pontifícia Universidade Católica de Campinas, Campinas, SP, Brasil
› Author Affiliations


Financial Support The present study received no specific financial support from public, commercial, or not-for-profit sources.
 

Abstract

Objective: The present study compared the difference in load and pressure distribution behavior of the blade plate and locked plate for varus osteotomy of the proximal femur per the finite element method.

Methods: Modeling was performed by scanning a medium-sized left femur with medial valgus deformity made of polyurethane.

Results: The stiffness of the locked plate is higher compared with that of the blade plate. However, this difference was not significant. In addition, the locked plate has proximal locking screws to ensure that the bending moments on the screws are smaller during loading.

Conclusion: In summary, both plates are well-established and effective. However, the study using the finite element method plays a fundamental role in understanding the load and pressure distribution of the implant. Moreover, it opens up new possibilities for further studies, including surgical proposals and customized implant materials.


Introduction

Varus osteotomies of the proximal femur are pediatric reconstructive surgeries widely performed in patients with neurological abnormalities, congenital hip diseases, sequelae, and acquired conditions.[1] [2] Fixation of these osteotomies may use several implants, including blade plates, dynamic compression plates (DCP), locked plates for the proximal femur, unilateral and circular external fixators, Kirschner wires, and screws.[1] [2] [3] The surgical complexity of these procedures resulted in advanced synthesis materials to facilitate surgery and improve outcomes.[1] [2] [4]

The most used materials in varus osteotomy of the proximal femur include locked and blade plates.[1] [2] [3] [4] Clinically, blade and locked plates have no statistical difference in the risk of failure (breakage),[3] which is the worst complication related to the choice of implant. Biomechanical tests with load application in experimental models show that locked plates with a support screw have higher axial resistance, lower resistance to torsion, and irreversible equivalent strain to deforming cycles compared with blade plates.[4]

The finite element method (FEM) is a mathematical tool used to solve problems in engineering because it explores the effects of load application on the bone and its biomechanical behavior. One of its main advantages is its potential use in solids with irregular geometry presenting heterogeneous material properties. The introduction of FEM in orthopedic biomechanics occurred in the 1970s; since then, the number of publications on load analysis in bones, arthroplasty, and osteosynthesis has been increasing.[5]

The present study aims to analyze, using FEM, the load and pressure distribution behavior when implanting blade or locked plates for varus osteotomy of the proximal femur.


Methods

Model Generation

Modeling was based on a computed tomography scanning of a left femur made from medium-sized polyurethane (Nacional Ossos, Brazil, reference number 2025 DMVL) and presenting a medial valgus deformity. The software used for plate scanning and modeling was SolidWorks (Dassault Systems SolidWorks Corp., Waltham, MA, USA) ([Fig. 1a]).

Zoom
Fig. 1 Valgus femur from Nacional Ossos for scanning. (A) Simplified femur and (B) varus osteotomy. Source: Data constructed by the authors using SolidWorks software.

For femur simplification, we sectioned the diaphysis and discarded the distal part since it would not be studied. A varus osteotomy was performed for 20° correction. This simplification provided a gain in calculation processing by FEM without result distortion.[6] The environment (femur) was the same for both studies, not distorting or favoring any data ([Fig. 1b]).

Fixation of the first femur employed a locked plate for the proximal femur with a 10-mm step, 100°, and three holes (Techimport, Rio Claro, SP, Brazil, reference number TI030.1003.100). Proximal fragment fixation used 3.5 × 50-mm diameter locked screws in two holes and a 3.5 × 40-mm locked screw in the third hole. Distal fragment fixation used 1 3.5 × 30-mm and 2 3.5 × 30-mm locked screws ([Fig. 2a]). Fixation of the second femur employed a blade plate for the proximal femur with a 10-mm step, 100°, 3 holes, 50-mm blade (Techimport, Rio Claro, SP, Brazil, reference number ref. TI030.1010.350). The proximal hole received 3.5 × 50 mm-diameter locked screws, while the distal non-locked plate holes received two 3.5 × 30-mm non-locked compression screws and a 3.5 × 30-mm locked screw for distal fixation ([Fig. 2b]).

Zoom
Fig. 2 Locked plate fixed to the femur with screws (A). Blade plate fixed to the femur with screws (B). Source: Data constructed by the authors using SolidWorks software.

Plate and screw assembly through visual positioning used the SolidWorks software (Dassault Systems SolidWorks Corp., Waltham, MA, USA). We created a 27-mm distance restriction between the proximal axis of the locked plate screw and the central axis of the blade plate and a 71-mm distance restriction between the medial face of the proximal femur and the inner face of the plates. These restrictions standardized the flexor moment generated by applying the axial force and the displacement stress assessment ([Fig. 2]).


Material Properties

Plates, screws, cortical, and cancellous bone models were homogeneous, linear, elastic, and isotropic, based on the properties described in the literature ([Table 1]). The yield stress was set at 795 MPa, the limit stress of the elastic region of the titanium alloy.[7] [8]

Table 1

Materials

Elastic modulus (E) [MPa]

Poisson ratio

Yield stress at 106 cycles [MPa]

Cortical bone

8,700

0.33

200

Cancellous bone

500

0.30

125

Titanium alloy

110,000

0.34

540


Simulation Parameters (Load, Mesh, and Contact Conditions)

[Fig. 2] shows the applied forces of 450, 500, 550, and 600 N considered a normal loading position, which assumes that the load vector has an angle of 8° of adduction with the hip longitudinal axis in the plane.[6] [9] [10] [11] [12] [13] [14] [15] [16] [17] In a clinical situation, both values are greater than those produced by touch support with crutches and should provide sufficient postoperative stability.[10] [11]

The force application frequency was 1 Hz, considering walking 1 step per second.[11] [12] Since this is a temporary fixation device, the plate must withstand at least 106 cycles, equivalent to ∼ 1 year, with this frequency.[13] We constrained the transverse face of the femoral shaft in all translational degrees of freedom ([Fig. 2]).

We merged the mesh models with three-dimensional quadratic tetrahedral elements in SolidWorks software of 1 mm for plates, 0.5 mm for screws, and 5 mm for bone.[14] Bonding contact occurred between bone tissue and implants, except for osteotomy interfaces ([Fig. 3]).

Zoom
Fig. 3 Mesh applied to three-dimensional locked plate models (A). Mesh applied to the three-dimensional blade plate models (B). Source: Data constructed by the authors using SolidWorks software.

Plate and screw contact surfaces had a friction coefficient of 0.34.[15] [17] The screw and plate contacted at the surface of the screw head and the surface of the countersunk in the plate holes, all with a restriction to not allow penetration between them. As for the contacts, locked screws in plate holes were deemed connected and fixed.[17]



Results

Von Mises stress and displacement for the blade plate were higher compared with the locked plate ([Tables 2] [3] [4] to [5]). [Fig. 4] demonstrates the displacement, in millimeters, of locked and blade plates after applying 450, 500, 550, and 600 N forces. [Fig. 5] shows the von Mises stress in MPa of locked and blade plates after applying 450, 500, 550, and 600 N forces. [Fig. 6] shows the von Mises stress in MPa on the bone fixed with locked or blade plates after applying 450, 500, 550, and 600 N forces.

Table 2

Variable

Group

Number of nodes (mesh)

Maximum value

Minimum value

Applied force [N]

Total displacement [mm]

Locked plate

473,114

5.6797

0

450

Blade plate

325,578

6.1151

0

Plate and screw stress [MPa]

Locked plate

473,114

492.1

73.58

Blade plate

325,578

510.12

93.222

Bone stress [MPa]

Locked plate

473,114

245.86

40.469

Blade plate

325,578

339.41

66.558

Table 3

Variable

Group

Number of nodes (mesh)

Maximum value

Minimum value

Applied force [N]

Total displacement [mm]

Locked plate

473,114

6.4123

0

500

Blade plate

325,578

6.9125

0

Plate and screw stress [MPa]

Locked plate

473,114

552.04

82.515

Blade plate

325,578

580.26

103.87

Bone stress [MPa]

Locked plate

473,114

270.61

45.442

Blade plate

325,578

365.99

73.925

Table 4

Variable

Group

Number of nodes (mesh)

Maximum value

Minimum value

Applied force [N]

Total displacement [mm]

Locked plate

473,114

7.1602

0

550

Blade plate

325,578

7.7405

0

Plate and screw stress [MPa]

Locked plate

473,114

628.5

91.55

Blade plate

325,578

647.06

114.58

Bone stress [MPa]

Locked plate

473,114

314.46

50.454

Blade plate

325,578

405.72

81.292

Table 5

Variable

Group

Number of nodes (mesh)

Maximum value

Minimum value

Applied force [N]

Total displacement [mm]

Locked plate

473,114

7.942

0

600

Blade plate

325,578

8.6027

0

Plate and screw stress [MPa]

Locked plate

473,114

657.72

100.69

Blade plate

325,578

716.19

125.35

Bone stress [MPa]

Locked plate

473,114

345.65

55.504

Blade plate

325,578

445.55

88.664

Zoom
Fig. 4 Locked and blade plate displacement versus applied forces. Source: Data constructed by the authors using SolidWorks software.
Zoom
Fig. 5 Locked and blade plate von Mises stress versus applied forces. Source: Data constructed by the authors using SolidWorks software.
Zoom
Fig. 6 Von Mises stress in the bone versus applied forces. Source: Data constructed by the authors using SolidWorks software.

The blade plate caused the highest displacement ([Table 2]). The highest displacement site for locked and blade plates was at the point of vertical force application at the femoral head ([Fig. 7]). The blade plate had the highest von Mises stress ([Table 2]). For the locked plate, the highest stress concentration was in the locking area between the proximal screw and the plate. For the blade plate, the region with the highest stress concentration was at the beginning of the advancement, at the osteotomy level ([Fig. 8]).

Zoom
Fig. 7 Locked plate total displacement, F = 450 N (A), blade plate total displacement, F = 450 N (B), locked plate total displacement, F = 500 N (C), blade plate total displacement, F = 500 N (D), locked plate total displacement, F = 550 N (E), blade plate total displacement, F = 550 N (F), locked plate total displacement, F = 600 N (G), blade plate total displacement, F = 600 N (H). Source: Data constructed by the authors using SolidWorks software.
Zoom
Zoom
Fig. 8 Von Mises stress for the locked plate, F = 450 N (A), von Mises stress for the blade plate, F = 450 N (B), von Mises stress for the locked plate, F = 500 N (C), von Mises stress for the blade plate, F = 500 N (D), von Mises stress for the locked plate, F = 550 N (E), von Mises stress for the blade plate, F = 550 N (F), von Mises stress for the locked plate, F = 600 N (G), von Mises stress for the blade plate, F = 600 N (H). Source: Data constructed by the authors using SolidWorks software.
Zoom

The bone fixed with the blade plate showed the highest von Mises stress ([Table 2]). The region with the highest bone stress concentration was the osteotomy, in the corner of the proximal part with the spongy bone surface in the distal part. The stress occurred at the same regions for all applied forces ([Fig. 9]).

Zoom
Fig. 9 Von Mises stress for locked plate, F = 450 N (A), von Mises stress for blade plate, F = 450 N (B). Source: Data constructed by the authors using SolidWorks software.

Fixation with the locked plate withstand more cycles compared with the blade plate for loads of 450, 500, 550, and 600 N in a 1 Hz frequency ([Table 6]).

Table 6

Applied force [N]

Group

Number of cycles

450

Locked plate

> 1,000,000

Blade plate

> 1,000,000

500

Locked plate

617,680

Blade plate

207,710

550

Locked plate

49,944

Blade plate

32,371

600

Locked plate

25,362

Blade plate

5,271

[Fig. 10] shows the number of cycles for locked and blade plates under 450, 500, 550, and 600 N loads. Furthermore, it demonstrates that after applying a load > 600 N, locked and blade plates tend to fail in the first cycle.

Zoom
Fig. 10 Number of cycles versus applied forces. Source: Data constructed by the authors using SolidWorks software.

Discussion

Conduction of clinical investigations on the forces required to stimulate bone healing is complex. The finite element method is widely used in the medical and orthopedic field because it provides a comprehensive view of vector dissolution in undermined structures and allows for accurate failure detection. Moreover, it may avoid unnecessary costs when failure is identifiable only after structural design or manufacturing. The finite element method also reduces the time from the first conceptual design to production, as the creation of a large number of experimental specimens becomes unnecessary. Finite element method analysis provides access to information difficult to obtain under laboratory conditions, such as predicted stress distribution and material strength, which are fundamental to assessing fatigue strength.[18]

In the present comparative biomechanical study, we investigated the flexural stiffness of the locked and blade plates for proximal femoral varus osteotomy. Our data suggest the locked plate presents higher stiffness compared with the blade plate. However, this difference was not significant ([Tables 2] [3] [4] to [5]). In addition, the locked plate has proximal locking screws to ensure that the bending moments acting on screws are lower during loading.

Locked plate failure occurs with significantly higher forces compared with the blade plate. The locked plate has significantly higher stiffness and load to failure values due to the nature of its design. At 19 mm, the locked plate is 8 mm wider in the proximal area, thus withstanding more stress than the blade plate. Both plates have approximately the same thickness of 3 mm. This results in a higher moment of inertia on the locked plate, which reflects in the proximal force results.

Consistent with our data, a comparative analysis by Radtke et al.[19] found mean values of 554 N for locked plates and 399 N for blade plates. These authors used synthetic bone and plates. Forward et al.[20] reported mean values of 620 N for locked plates and 450 N for blade plates in a study performed with cadavers.

Femoral stress distribution was consistent with a study by Sim et al.[21] reporting a higher stress concentration between the proximal and distal parts at the point of separation osteotomy.

Absolute displacement values were higher for the blade plate, which had two areas of load application with higher variation (femur head). However, there is a biomechanical advantage associated with the stress areas of the system. In locked plates, the stress area of the joint between the plate and the bone is at the proximal screw. On the other hand, in blade plates, this area is at the blade advance region. Thus, in blade plates, all the stress concentrates in the osteotomy region, which usually constitutes an obstacle to bone consolidation resulting from the Wolff law.[22] The success of biological bone healing depends on a favorable mechanical environment. In addition, the Wolff law and the Perren strain theory allow using several osteosynthesis systems to promote adequate stabilization and the differentiation of various cell types at the bone healing site.[22] [23]

The relative stability indicated for comminuted diaphyseal or extra-articular fractures allows for some controlled mobility at the fracture site and exuberant formation of bone callus, which characterizes an indirect or endochondral ossification. Direct or intramembranous ossification is recommended to avoid developing bulky bone calluses in joint fractures, following absolute fixation with greater rigidity.[24]

The locked plate system also presents a larger stress area at the osteotomy region but produced lower absolute values than the blade plate. Thus, we inferred that locked plates create a more favorable biomechanical situation for bone consolidation.

Our study has limitations regarding FEM, which considers structures as a gathering of small particles of finite quantity, the so-called finite elements, connected to a finite number of points, the nodes, or nodal points. These particles represent the approximate result of every discretized system.[25] The finite element method allows the evaluation of the approximate stress distribution in a structure, observing the element strain through visualization and image interpretation on a color chart.[5]

The present study demonstrated that the blade plate resisted fewer cycles when the loading forces were lower. However, locked and blade plates tend to fail under the highest applied force. Thus, despite being exposed to controlled load situations, the locked plate was more resistant to implant failure. In exacerbated load situations, locked and blade plates tend to fail.

Both osteosynthesis implants are consecrated and effective. Nevertheless, our study using FEM shows a fundamental role in understanding the biomechanical situation of the implant. It also opens up new possibilities for further studies, including surgical proposals and customized implant materials. Therefore, our study corroborates a hypothesis raised by common sense, that is, the superiority of the locked plate compared with blade plates for varus osteotomy of the proximal femur. Still, there is no unanimity in the literature, especially regarding clinical outcomes.

Our study may yield future models with laboratory biomechanical tests to prove the differences between locked and blade plate fixation systems.


Conclusion

Both osteosynthesis implants are well-established and effective in clinical practice. However, our study applying FEM demonstrated the biomechanical superiority of the locked plate compared with the blade plate for proximal femoral varus osteotomy in the proposed model.



Conflito de Interesses

Os autores declaram não haver conflito de interesses.

  • Referências

  • 1 Joeris A, Audigé L, Ziebarth K, Slongo T. The Locking Compression Paediatric Hip Plate: technical guide and critical analysis. Int Orthop 2012; 36 (11) 2299-2306
  • 2 Ruzbarsky JJ, Swarup I, Garner MR, Meyers KN, Edobor-Osula F, Widmann RF. et al. Biomechanical Comparison of Two Pediatric Blade Plate Designs in Proximal Femoral Osteotomies. HSS J 2020; 16 (01) 81-85
  • 3 Jain A, Thompson JM, Brooks JT, Ain MC, Sponseller PD. Implant-related Fractures in Children With Proximal Femoral Osteotomy: Blade Plate Versus Screw-side Plate Constructs. J Pediatr Orthop 2016; 36 (01) e1-e5
  • 4 Crist BD, Khalafi A, Hazelwood SJ, Lee MA. A biomechanical comparison of locked plate fixation with percutaneous insertion capability versus the angled blade plate in a subtrochanteric fracture gap model. J Orthop Trauma 2009; 23 (09) 622-627
  • 5 Soni JF, Santili C, Lancellotti CLP, Hecke MB, Almeida FR, Karam LZ. Análise comparativa em modelo computadorizado bidimensional com simulação do emprego de hastes flexíveis de aço e titânio na fratura do fêmur da criança utilizando o método dos elementos finitos. Rev Bras Ortop 2008; 43 (05) 183-192
  • 6 Nag P, Chanda S. Biomechanical design prognosis of two extramedullary fixation devices for subtrochanteric femur fracture: a finite element study. Med Biol Eng Comput 2021; 59 (02) 271-285
  • 7 Maurer P, Holweg S, Schubert J. Finite-element-analysis of different screw-diameters in the sagittal split osteotomy of the mandible. J Craniomaxillofac Surg 1999; 27 (06) 365-372
  • 8 Janeček M, Novy F, Harcuba P, Stráský J, Trško L, Mhaede M. et al. The Very High Cycle Fatigue Behaviour of Ti-6Al-4V Alloy. Acta Phys Pol A 2015; 128: 497-503
  • 9 Marco M, Giner E, Larraínzar R, Caeiro J, Miguélez H. Analysis of the variation of the mechanical behavior of the proximal extremity of the femur using the XFEM method (eXtended Finite Element Method). J Osteoporos Min Metabol 2016; 8: 61-69
  • 10 Perry J. Gait Analysis: Normal and Pathological Function. Thorofare, NJ: SLACK, Inc.; 1992: 11-12
  • 11 Wang J, Ma JX, Lu B, Bai HH, Wang Y, Ma XL. Comparative finite element analysis of three implants fixing stable and unstable subtrochanteric femoral fractures: Proximal Femoral Nail Antirotation (PFNA), Proximal Femoral Locking Plate (PFLP), and Reverse Less Invasive Stabilization System (LISS). Orthop Traumatol Surg Res 2020; 106 (01) 95-101
  • 12 Taddei F, Viceconti M, Manfrini M, Toni A. Mechanical strength of a femoral reconstruction in paediatric oncology: a finite element study. Proc Inst Mech Eng H 2003; 217 (02) 111-119
  • 13 Kainz H, Killen BA, Wesseling M, Perez-Boerema F, Pitto L, Aznar JMG. et al. A multi-scale modelling framework combining musculoskeletal rigid-body simulations with adaptive finite element analyses, to evaluate the impact of femoral geometry on hip joint contact forces and femoral bone growth. PLoS One 2020; 15 (07) e0235966
  • 14 Lewis GS, Mischler D, Wee H, Reid JS, Varga P. Finite Element Analysis of Fracture Fixation. Curr Osteoporos Rep 2021; 19 (04) 403-416
  • 15 Zhang QY, Zhou Y, Li XX, Wang L, Cui XH, Wang SQ. Accelerated Formation of Tribo-oxide Layer and Its Effect on Sliding Wear of a Titanium Alloy. Tribol Lett 2016; 63 (01)
  • 16 Adamus J, Lacki P, Motyka M. Kubiak KInvestigation of sheet-titanium drawability. In: Ti 2011 Proceedings of the 12th World Conference on Titanium, Bei- jing, China, june 19 to ] une 24, 2011: 337-341
  • 17 Eberle S, Gerber C, von Oldenburg G, Hungerer S, Augat P. Type of hip fracture determines load share in intramedullary osteosynthesis. Clin Orthop Relat Res 2009; 467 (08) 1972-1980
  • 18 Faria FF, Gruhl CEM, Ferro RR, Rached RN, Soni JF, Trevilatto P. Finite Element Analysis of a Controlled Dynamization Device for External Circular Fixation. Rev Bras Ortop 2021; 56 (01) 36-41
  • 19 Radtke K, Goede F, Schwarze M, Paes P, Ettinger M, Welke B. Fixation stability and stiffness of two implant systems for proximal femoral varization osteotomy. Appl Sci (Basel) 2020;•••;. Doi: 10.5867.10.3390/app10175867
  • 20 Forward DP, Doro CJ, OʼToole RV, Kim H, Floyd JCP, Sciadini MF. et al. A biomechanical comparison of a locking plate, a nail, and a 95° angled blade plate for fixation of subtrochanteric femoral fractures. J Orthop Trauma 2012; 26 (06) 334-340
  • 21 Sim E, Freimüller W, Reiter TJ. Finite element analysis of the stress distributions in the proximal end of the femur after stabilization of a pertrochanteric model fracture: a comparison of two implants. Injury 1995; 26 (07) 445-449
  • 22 Wolff J. Das Gesetz der Transformation der Knochen. Berlin: Hirschwald; 1892
  • 23 Perren S, Boitzy A. Cellular differentiation and bone biomechanics during the consolidation of a fracture. Clin Anat 1978; 1: 13-28
  • 24 Glatt V, Matthys R. Adjustable stiffness, external fixator for the rat femur osteotomy and segmental bone defect models. J Vis Exp 2014; 9 (92) e51558
  • 25 Zienkiewicz OC. The finite element method. 3rd expanded and rev. ed.. London; New York: McGraw-Hill; ; c 1977

Endereço para correspondência

Wilisson Ribeiro Filho
Departamento de Ortopedia e Traumatologia da Pontifícia Universidade Católica de Campinas (SP) – Grupo de Ortopedia e Traumatologia Pediátrica
Avenida John Boyd Dunlop, s/n°, Jardim Ipausssurama – 13060-904–Campinas
SP   

Publication History

Received: 28 November 2022

Accepted: 24 January 2023

Article published online:
21 March 2024

© 2024. The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution 4.0 International License, permitting copying and reproduction so long as the original work is given appropriate credit (https://creativecommons.org/licenses/by/4.0/)

Thieme Revinter Publicações Ltda.
Rua do Matoso 170, Rio de Janeiro, RJ, CEP 20270-135, Brazil

  • Referências

  • 1 Joeris A, Audigé L, Ziebarth K, Slongo T. The Locking Compression Paediatric Hip Plate: technical guide and critical analysis. Int Orthop 2012; 36 (11) 2299-2306
  • 2 Ruzbarsky JJ, Swarup I, Garner MR, Meyers KN, Edobor-Osula F, Widmann RF. et al. Biomechanical Comparison of Two Pediatric Blade Plate Designs in Proximal Femoral Osteotomies. HSS J 2020; 16 (01) 81-85
  • 3 Jain A, Thompson JM, Brooks JT, Ain MC, Sponseller PD. Implant-related Fractures in Children With Proximal Femoral Osteotomy: Blade Plate Versus Screw-side Plate Constructs. J Pediatr Orthop 2016; 36 (01) e1-e5
  • 4 Crist BD, Khalafi A, Hazelwood SJ, Lee MA. A biomechanical comparison of locked plate fixation with percutaneous insertion capability versus the angled blade plate in a subtrochanteric fracture gap model. J Orthop Trauma 2009; 23 (09) 622-627
  • 5 Soni JF, Santili C, Lancellotti CLP, Hecke MB, Almeida FR, Karam LZ. Análise comparativa em modelo computadorizado bidimensional com simulação do emprego de hastes flexíveis de aço e titânio na fratura do fêmur da criança utilizando o método dos elementos finitos. Rev Bras Ortop 2008; 43 (05) 183-192
  • 6 Nag P, Chanda S. Biomechanical design prognosis of two extramedullary fixation devices for subtrochanteric femur fracture: a finite element study. Med Biol Eng Comput 2021; 59 (02) 271-285
  • 7 Maurer P, Holweg S, Schubert J. Finite-element-analysis of different screw-diameters in the sagittal split osteotomy of the mandible. J Craniomaxillofac Surg 1999; 27 (06) 365-372
  • 8 Janeček M, Novy F, Harcuba P, Stráský J, Trško L, Mhaede M. et al. The Very High Cycle Fatigue Behaviour of Ti-6Al-4V Alloy. Acta Phys Pol A 2015; 128: 497-503
  • 9 Marco M, Giner E, Larraínzar R, Caeiro J, Miguélez H. Analysis of the variation of the mechanical behavior of the proximal extremity of the femur using the XFEM method (eXtended Finite Element Method). J Osteoporos Min Metabol 2016; 8: 61-69
  • 10 Perry J. Gait Analysis: Normal and Pathological Function. Thorofare, NJ: SLACK, Inc.; 1992: 11-12
  • 11 Wang J, Ma JX, Lu B, Bai HH, Wang Y, Ma XL. Comparative finite element analysis of three implants fixing stable and unstable subtrochanteric femoral fractures: Proximal Femoral Nail Antirotation (PFNA), Proximal Femoral Locking Plate (PFLP), and Reverse Less Invasive Stabilization System (LISS). Orthop Traumatol Surg Res 2020; 106 (01) 95-101
  • 12 Taddei F, Viceconti M, Manfrini M, Toni A. Mechanical strength of a femoral reconstruction in paediatric oncology: a finite element study. Proc Inst Mech Eng H 2003; 217 (02) 111-119
  • 13 Kainz H, Killen BA, Wesseling M, Perez-Boerema F, Pitto L, Aznar JMG. et al. A multi-scale modelling framework combining musculoskeletal rigid-body simulations with adaptive finite element analyses, to evaluate the impact of femoral geometry on hip joint contact forces and femoral bone growth. PLoS One 2020; 15 (07) e0235966
  • 14 Lewis GS, Mischler D, Wee H, Reid JS, Varga P. Finite Element Analysis of Fracture Fixation. Curr Osteoporos Rep 2021; 19 (04) 403-416
  • 15 Zhang QY, Zhou Y, Li XX, Wang L, Cui XH, Wang SQ. Accelerated Formation of Tribo-oxide Layer and Its Effect on Sliding Wear of a Titanium Alloy. Tribol Lett 2016; 63 (01)
  • 16 Adamus J, Lacki P, Motyka M. Kubiak KInvestigation of sheet-titanium drawability. In: Ti 2011 Proceedings of the 12th World Conference on Titanium, Bei- jing, China, june 19 to ] une 24, 2011: 337-341
  • 17 Eberle S, Gerber C, von Oldenburg G, Hungerer S, Augat P. Type of hip fracture determines load share in intramedullary osteosynthesis. Clin Orthop Relat Res 2009; 467 (08) 1972-1980
  • 18 Faria FF, Gruhl CEM, Ferro RR, Rached RN, Soni JF, Trevilatto P. Finite Element Analysis of a Controlled Dynamization Device for External Circular Fixation. Rev Bras Ortop 2021; 56 (01) 36-41
  • 19 Radtke K, Goede F, Schwarze M, Paes P, Ettinger M, Welke B. Fixation stability and stiffness of two implant systems for proximal femoral varization osteotomy. Appl Sci (Basel) 2020;•••;. Doi: 10.5867.10.3390/app10175867
  • 20 Forward DP, Doro CJ, OʼToole RV, Kim H, Floyd JCP, Sciadini MF. et al. A biomechanical comparison of a locking plate, a nail, and a 95° angled blade plate for fixation of subtrochanteric femoral fractures. J Orthop Trauma 2012; 26 (06) 334-340
  • 21 Sim E, Freimüller W, Reiter TJ. Finite element analysis of the stress distributions in the proximal end of the femur after stabilization of a pertrochanteric model fracture: a comparison of two implants. Injury 1995; 26 (07) 445-449
  • 22 Wolff J. Das Gesetz der Transformation der Knochen. Berlin: Hirschwald; 1892
  • 23 Perren S, Boitzy A. Cellular differentiation and bone biomechanics during the consolidation of a fracture. Clin Anat 1978; 1: 13-28
  • 24 Glatt V, Matthys R. Adjustable stiffness, external fixator for the rat femur osteotomy and segmental bone defect models. J Vis Exp 2014; 9 (92) e51558
  • 25 Zienkiewicz OC. The finite element method. 3rd expanded and rev. ed.. London; New York: McGraw-Hill; ; c 1977

Zoom
Fig. 1 Fêmur valgo para escaneamento Nacional Ossos. (A) fêmur simplificado e (B) osteotomia varizante. Fonte: Dados construídos pelos autores com o software SolidWorks.
Zoom
Fig. 2 Placa bloqueada fixada no fêmur com parafusos (A). Placa lâmina fixada no fêmur com parafusos (B). Fonte: Dados construídos pelos autores com o software SolidWorks.
Zoom
Fig. 3 Malha aplicada aos modelos tridimensionais da placa bloqueada (A). Malha aplicada aos modelos tridimensionais da placa lâmina (B). Fonte: Dados construídos pelos autores com o software SolidWorks.
Zoom
Fig. 1 Valgus femur from Nacional Ossos for scanning. (A) Simplified femur and (B) varus osteotomy. Source: Data constructed by the authors using SolidWorks software.
Zoom
Fig. 2 Locked plate fixed to the femur with screws (A). Blade plate fixed to the femur with screws (B). Source: Data constructed by the authors using SolidWorks software.
Zoom
Fig. 3 Mesh applied to three-dimensional locked plate models (A). Mesh applied to the three-dimensional blade plate models (B). Source: Data constructed by the authors using SolidWorks software.
Zoom
Fig. 4 Deslocamento da placa bloqueada e lâmina versus forças aplicadas. Fonte: Dados construídos pelos autores com o software SolidWorks.
Zoom
Fig. 5 Tensão de von-Mises da placa bloqueada e lâmina versus forças aplicadas. Fonte: Dados construídos pelos autores com o software SolidWorks.
Zoom
Fig. 6 Tensão de von-Mises no osso versus forças aplicadas. Fonte: Dados construídos pelos autores com o software SolidWorks.
Zoom
Fig. 7 Deslocamento total da placa bloqueada F = 450N (A), Deslocamento total da placa lâmina F = 450N (B), Deslocamento total da placa bloqueada F = 500N (C), Deslocamento total da placa lâmina F = 500N (D), Deslocamento total da placa bloqueada F = 550N (E), Deslocamento total da placa lâmina F = 550N (F), Deslocamento total da placa bloqueada F = 600N (G), Deslocamento total da placa lâmina F = 600N (H). Fonte: Dados construídos pelos autores com o software SolidWorks.
Zoom
Zoom
Fig. 8 Tensão de von-Mises da placa bloqueada F = 450N (A), Tensão de von-Mises da placa lâmina F = 450N (B), Tensão de von-Mises da placa bloqueada F = 500N (C), Tensão de von-Mises da placa lâmina F = 500N (D), Tensão de von-Mises da placa bloqueada F = 550N (E), Tensão de von-Mises da placa lâmina F = 550N (F), Tensão de von-Mises da placa bloqueada F = 600N (G), Tensão de von-Mises da placa lâmina F = 600N (H). Fonte: Dados construídos pelos autores com o software SolidWorks.
Zoom
Zoom
Fig. 9 Tensão de von-Mises no osso para placa bloqueada F = 450N (A), Tensão de von-Mises no osso para placa lâmina F = 450N (B). Fonte: Dados construídos pelos autores com o software SolidWorks.
Zoom
Fig. 10 Número de ciclos versus forças aplicadas. Fonte: Dados construídos pelos autores com o software SolidWorks.
Zoom
Fig. 4 Locked and blade plate displacement versus applied forces. Source: Data constructed by the authors using SolidWorks software.
Zoom
Fig. 5 Locked and blade plate von Mises stress versus applied forces. Source: Data constructed by the authors using SolidWorks software.
Zoom
Fig. 6 Von Mises stress in the bone versus applied forces. Source: Data constructed by the authors using SolidWorks software.
Zoom
Fig. 7 Locked plate total displacement, F = 450 N (A), blade plate total displacement, F = 450 N (B), locked plate total displacement, F = 500 N (C), blade plate total displacement, F = 500 N (D), locked plate total displacement, F = 550 N (E), blade plate total displacement, F = 550 N (F), locked plate total displacement, F = 600 N (G), blade plate total displacement, F = 600 N (H). Source: Data constructed by the authors using SolidWorks software.
Zoom
Zoom
Fig. 8 Von Mises stress for the locked plate, F = 450 N (A), von Mises stress for the blade plate, F = 450 N (B), von Mises stress for the locked plate, F = 500 N (C), von Mises stress for the blade plate, F = 500 N (D), von Mises stress for the locked plate, F = 550 N (E), von Mises stress for the blade plate, F = 550 N (F), von Mises stress for the locked plate, F = 600 N (G), von Mises stress for the blade plate, F = 600 N (H). Source: Data constructed by the authors using SolidWorks software.
Zoom
Zoom
Fig. 9 Von Mises stress for locked plate, F = 450 N (A), von Mises stress for blade plate, F = 450 N (B). Source: Data constructed by the authors using SolidWorks software.
Zoom
Fig. 10 Number of cycles versus applied forces. Source: Data constructed by the authors using SolidWorks software.