Ultraschall Med 1999; 20(4): 158-160
DOI: 10.1055/s-1999-8900
ORIGINALARBEIT
Georg Thieme Verlag Stuttgart ·New York

Dreidimensionaler Ultraschall und intraoperative Navigation: ein neuer Einsatz des Ultraschalltopometers bei Umstellungsosteotomie des proximalen Femurs

Three-Dimensional Ultrasound and Intraoperative Navigation: a new Application in Femoral Osteotomies T. Wallny,  J. Klose1 ,  G. Steffny1 ,  D. Schulze Bertelsbeck,  L. Perlick,  G. Schumpe1
  • 1Orthopädische Universitätsklinik Bonn, Bonn
  • 2Abteilung für Biomechanik und Biophysik in der Orthopädischen Universitätsklinik Bonn, Bonn
Further Information

Publication History

Publication Date:
31 December 1999 (online)

Zusammenfassung

Studienziel: Die Präzision von Umstellungsosteotomien kann mangels genauer intraoperativer Kontrollmechanismen nur abgeschätzt werden. Deshalb wurde mit dem neuentwickelten Einsatzbereich des Ultraschalltopometers eine on-line Möglichkeit zur Überprüfung der angestrebten Korrektur geschaffen. Methode: Mit je drei auf einem Dreieck befestigten Ultraschallsendern kann über Messung der Senderlaufzeiten zu Mikrophonen die Position der Dreiecke im Raum bestimmt werden. Bei Positionierung eines Referenzdreiecks distal der Osteotomie und auf dem Klingenmeißel kann der angestrebte Korrekturwinkel dreidimensional genau bestimmt werden. Es wurden Positionsbestimmungen und -änderungen im Laborversuch durchgeführt und an 5 Schweinefemura die Operation simuliert. Ergebnisse: Die Abweichung der vom Meßprogramm angegebenen Werte war kleiner als 0,5 °, bei Simulation der Operation war das System praktikabel. Fazit: Mit dem System lassen sich zuverlässig Raumpositionen bestimmen und die Genauigkeit von Umstellungsosteotomien unterstützen.

Aim: As precise operative control is difficult to achieve, the accuracy of femoral osteotomies can only be estimated. Therefore, a computer-assisted ultrasound navigation system was developed in order to apply an on-line control for femoral osteotomies. Method: Three ultrasound emitters were fixed on a triangle. The exact position of triangles could be determined by measuring the time the ultrasound beam takes to reach microphones positioned in a frame. With a reference triangle fixed distally to the osteotomy and a second triangle fixed on the surgical chisel the exact correction angle can be determined three-dimensionally. Results: A high degree of accuracy was found in both laboratory trials and in simulation trials using pig femurs. The deviation of measured values compared to a laser beam control was less than 0.5 degrees. Conclusion: The system was introduced into our operating theatre as an optimised control device that can provide excellent support to the surgical procedure.

Literatur

  • 1 Laine T, Schlenzka D, Mäkitalo K, Tallroth K, Nolte LP, Visarius H. Improved Accuracy of Pedicle Screw Insertion With Computer-Assisted Surgery.  Spine. 1997;  22 1254-1258
  • 2 Langlotz F, Stucki M, Bächler R. The first twelve cases of computer assisted periacetubular osteotomy.  Comp Aid Surg. 1997;  2 317-326
  • 3 Schumpe G. Ganguntersuchungen und funktionelle Wirbelsäulenvermessung mittels eines neu entwickelten Echtzeit-Stereo-Ultraschalltopometers. Stuttgart; Enke 1977
  • 4 Heisel J, Mittelmeier H, Schwarz B. Gelenkerhaltende Operationsverfahren bei der idiopathischen Hüftkopfnekrose.  Z Orthop. 1984;  122 705-715
  • 5 Mittelmeier H. Osteosynthesetechniken bei coxalen Femurosteotomien. Springorum HW, Katthagen D (Hrsg.) Aktuelle Schwerpunkte der Orthopädie. Stuttgart; Thieme 1992 Band 3: 90-99
  • 6 Nolte LP, Zamorano LJ, Jiang Z, et al. Image-guided insertion of transpedicular screws.  Spine. 1995;  20 497-500

Dr. T. Wallny

Orthopädische Universitätsklinik Bonn

Sigmund-Freud-Straße 25

D-53127 Bonn

    >