Initial clinical experience of using a newly designed preshaped microguidewire in acute endovascular thrombectomy
Mechanical thrombectomy has been widely used for the treatment of acute ischemic stroke. During this procedure, operators must navigate the microcatheter with a microguidewire (MGW) into vessels that cannot be visualized on fluoroscopy as rapidly as possible. In this study, we developed a modified pigtail-shaped MGW (MPMGW) for security and controllability. Moreover, the efficacy of the MPMGW for the treatment of acute ischemic stroke was assessed. The MPMGW was designed using 0.014 MGW. Because we created four MPMGWs during a clinical evaluation before the launch in the market, these wires were used in four consecutive patients with acute ischemic stroke in the single institution. The occluded arteries were the basilar artery (n = 1), middle cerebral arteries (M1 and M2, n = 2), and internal carotid artery (n = 1). All four procedures were conducted without any complications. The procedures included navigating the MGW and passing it through the clot. Complete recanalization was achieved in all cases. The average time between femoral artery puncture and recanalization was 15 min. The use of the preshaped MPMGW in acute thrombectomy was effective in terms of both security of procedure and reduction in recanalization time.
#
Key-words:
Acute ischemic stroke - endovascular treatment - microguidewire - modified pigtail-shaped - preshape - thrombectomyIntroduction
Rapid recanalization is the most effective treatment for acute ischemic stroke due to major cerebral artery occlusion.[[1]] Recently, endovascular technologies and techniques for acute ischemic stroke have continuously evolved. However, endovascular stent clot retrieval is still challenging to perform and has risks, particularly for beginners, because operators must navigate the microcatheter with a microguidewire (MGW) into far distal vessels that cannot be visualized on fluoroscopy as rapidly as possible. Some authors have reported the use of a modified pigtail-shaped MGW (MPMGW) for various endovascular treatments.[[2]],[[3]],[[4]],[[5]],[[6]] Although the MPMGW is effective for acute thrombectomy, the manual quick shaping of this wire requires some skills. The common measurement of the round tip is >3.0 mm. Therefore, we designed a preshaped MPMGW with a round tip measuring 2.5 mm. Herein, we report the initial clinical experience using preshaped MPMGW for the treatment of acute ischemic stroke.
#
Materials and Methods
Microguidewire
A preshaped MPMGW was designed using a 0.014 MGW (Eiger 0.014; Medical Innovation Co., Ltd., Tokyo, Japan) [[Figure 1]]. The tip was bent round, with a diameter of 2.5 mm, and a portion with a length of 2.0 mm was kept straight. At the 10.0 mm proximal side of the tip, the wire was bent at 135°. The round tip was designed to prevent vessel injury. The 2.0-mm straight portion facilitated the easy insertion of the wire into the microcatheter. The bent tip allowed operators to select branches with torque. The MGW was packed with an introducer sheath beforehand, as navigation into the microcatheter was quickly performed. Because we created four preshaped MPMGWs during a clinical evaluation before launching in the market, these wires were used in four consecutive patients with acute ischemic stroke in a single institution.


#
Endovascular procedure
The ASAP technique[[7]] is commonly used for the anterior circulation, and the simple stent retrieving is used for the posterior circulation. The guiding catheter, aspirator, microcatheter, and stent retriever were selected based on the operator's preference. The preshaped MPMGW was used in all procedures.
#
#
Results
A summary of the clinical data, devices used, and clinical outcomes are shown in [[Table 1]], [[Table 2]], [[Table 3]], respectively. The occluded arteries were the basilar artery (n = 1), middle cerebral arteries (M1 and M2, n = 2), and internal carotid artery (n = 1). The guiding catheters used were 6-Fr Envoy (90 cm; Codman, Raynham, MA, USA) for posterior circulation and 9-Fr Optimo (Tokai Medical Products, Aichi, Japan) for anterior circulation. Penumbra ACE 68 (Penumbra, Alameda, CA, USA) was used as an aspirator. Moreover, Rebar 18 and Marksman microcatheters (Medtronic, Minneapolis, MN, USA) were used. Revive (Codman) and Solitaire (Medtronic) stent devices were used as clot retrievers. The operative images of representative patients are shown in [[Figure 2]] and [[Figure 3]].










All four procedures have been conducted without any complications, and the procedures involved navigating the MPMGW and passing it through the clot. Complete recanalization was achieved in all cases. The average time between femoral artery puncture and recanalization was 15 min.
#
Discussion
Some authors have previously reported that the use of an MPMGW is an extremely safe and versatile method for neuroendovascular procedures.[[2]],[[3]],[[4]],[[5]],[[6]] The round tip of the wire prevents unintentional entry into the perforating branches or blebs of an aneurysm and reduces the risks of perforation if it enters them erroneously. Moreover, MPMGW is useful in acute thrombectomy. When a stent retriever is used, the MGW must pass through the occluded vessel, which is not observed during angiography. The MPMGW can be safely advanced across a lesion even if an unseen aneurysm is in the distal vessel. The wire also provides better vessel selectivity with torque. When a guide catheter cannot be advanced up to the common carotid artery, the MPMGW allows selection into the internal carotid artery and navigation into intracranial small arteries with the same tip shape. Moreover, this type of MGW facilitates better microcatheter trackability, because it acts as an anchor in the distal vessels compared to a conventional right-angled shape MGW.
Before we developed the preshaped MPMGW, we manually used Chikai 14 MGW (Asahi Intecc Co., Ltd., Aichi, Japan). Beginners often take a few minutes to shape the wire. Furthermore, making the small round tip is quite challenging for beginners and even experts. Therefore, the preshaped MPMGW was designed to facilitate a rapid procedure and recanalization.
In this case series, the MGW can be smoothly inserted into the microcatheter and navigated up to around the M2–M3 junction. The small round tip can always be navigated at the occluded area without any manipulations and negotiations. Moreover, it never migrated into the branching and perforating arteries.
In cases no. 1 and no. 4, the tip of the MGW appeared bent during penetration of the clots, which were found relative hard. As previously reported,[[3]] the configuration of the MGW provides information about the quality of the clot during penetration. The preshaped MPMGW was harder, and it can penetrate hard clots more smoothly than the Chikai 14 MGW without dislocation of clots distally. Since both patients presented with hard clots, the corkscrew penetrating method was used.[[4]]
The present study had some limitations. That is, a small sample was included, the same operator conducted all of the procedures, and the study was performed in a single institution. Moreover, the time between puncture and recanalization in our study was shorter than that of previous studies.[[8]] Thus, the recanalization rate and clinical outcomes in this study were good and comparable to those in large-scale studies, but we lacked sufficient statistical power to determine whether the MPMGW has an effect on outcomes. Despite these limitations, we believe that the results of this study can significantly contribute to the currently available clinical information about rapid recanalization in acute thrombectomy. However, large-scale, multicenter trials must be conducted to validate the results of the current study.
The use of the preshaped MPMGW in acute thrombectomy was effective in terms of both security of procedure and reduction in recanalization time.
#
#
Conflict of Interest
There are no conflicts of interest.
Acknowledgment
We would like to thank Seiichiro Watanabe, head of the Medical Innovation Co., Ltd., Tokyo, Japan, for the support and encouragement in carrying out this study.
Financial support and sponsorship
Nil.
-
References
- 1 Rha JH, Saver JL. The impact of recanalization on ischemic stroke outcome: A meta-analysis. Stroke 2007;38:967-73.
- 2 Sato M, Ohshima T, Ishikawa K, Goto S, Yamamoto T, Izumi T, et al. A novel technique of safe and versatile microguidewire shaping with neuroendovascular therapy: Modified pigtail method. J Neuroendovasc Ther 2017;11:266-71.
- 3 Ohshima T, Ishikawa K, Goto S, Yamamoto T. Relationship between clot quality and microguidewire configuration during endovascular thrombectomy for acute ischemic stroke. World Neurosurg 2017;107:657-62.
- 4 Ohshima T, Imai T, Sato M, Goto S, Yamamoto T, Nishizawa T, et al. A novel technique for higher success rates of recanalization with stent clot retriever: Corkscrew penetrating method. J Neuroendovasc Ther 2017;11:94-8.
- 5 Ohshima T, Shamim Ul Haq Siddiqi, Miyachi S, Matsuo N, Kawaguchi R, Kato Y, et al. Usefulness of modified pigtail-shaped microguidewire guidance for microcatheter navigation in difficult vasculatures during neuroendovascular interventions. Nagoya J Med Sci 2018;80:551-7.
- 6 Ato F, Ohshima T, Miyachi S, Matsuo N, Kawaguchi R, Takayasu M. Efficacy and safety of a modified pigtail-shaped microguidewire during endovascular thrombectomy. Asian J Neurosurg 2019;14:1040-3.
- 7 Goto S, Ohshima T, Ishikawa K, Yamamoto T, Shimato S, Nishizawa T, et al. A Stent-Retrieving into an Aspiration Catheter with Proximal Balloon (ASAP) technique: A technique of mechanical thrombectomy. World Neurosurg 2018;109:e468-75.
- 8 Goyal M, Menon BK, van Zwam WH, Dippel DW, Mitchell PJ, Demchuk AM, et al. Endovascular thrombectomy after large-vessel ischaemic stroke: A meta-analysis of individual patient data from five randomised trials. Lancet 2016;387:1723-31.
Address for correspondence
Publication History
Received: 17 December 2019
Accepted: 24 January 2020
Article published online:
16 August 2022
© 2020. Asian Congress of Neurological Surgeons. This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes, or adapted, remixed, transformed or built upon. (https://creativecommons.org/licenses/by-nc-nd/4.0/)
Thieme Medical and Scientific Publishers Pvt. Ltd.
A-12, 2nd Floor, Sector 2, Noida-201301 UP, India
-
References
- 1 Rha JH, Saver JL. The impact of recanalization on ischemic stroke outcome: A meta-analysis. Stroke 2007;38:967-73.
- 2 Sato M, Ohshima T, Ishikawa K, Goto S, Yamamoto T, Izumi T, et al. A novel technique of safe and versatile microguidewire shaping with neuroendovascular therapy: Modified pigtail method. J Neuroendovasc Ther 2017;11:266-71.
- 3 Ohshima T, Ishikawa K, Goto S, Yamamoto T. Relationship between clot quality and microguidewire configuration during endovascular thrombectomy for acute ischemic stroke. World Neurosurg 2017;107:657-62.
- 4 Ohshima T, Imai T, Sato M, Goto S, Yamamoto T, Nishizawa T, et al. A novel technique for higher success rates of recanalization with stent clot retriever: Corkscrew penetrating method. J Neuroendovasc Ther 2017;11:94-8.
- 5 Ohshima T, Shamim Ul Haq Siddiqi, Miyachi S, Matsuo N, Kawaguchi R, Kato Y, et al. Usefulness of modified pigtail-shaped microguidewire guidance for microcatheter navigation in difficult vasculatures during neuroendovascular interventions. Nagoya J Med Sci 2018;80:551-7.
- 6 Ato F, Ohshima T, Miyachi S, Matsuo N, Kawaguchi R, Takayasu M. Efficacy and safety of a modified pigtail-shaped microguidewire during endovascular thrombectomy. Asian J Neurosurg 2019;14:1040-3.
- 7 Goto S, Ohshima T, Ishikawa K, Yamamoto T, Shimato S, Nishizawa T, et al. A Stent-Retrieving into an Aspiration Catheter with Proximal Balloon (ASAP) technique: A technique of mechanical thrombectomy. World Neurosurg 2018;109:e468-75.
- 8 Goyal M, Menon BK, van Zwam WH, Dippel DW, Mitchell PJ, Demchuk AM, et al. Endovascular thrombectomy after large-vessel ischaemic stroke: A meta-analysis of individual patient data from five randomised trials. Lancet 2016;387:1723-31.