Key words:
Dental instruments - endodontics - Reciproc - root canal preparation - WaveOne
INTRODUCTION
The advent of rotary nickel–titanium (NiTi) instrumentation has brought significant
progress to root canal preparation in terms of preserving native root canal morphology
and preventing apical transportation.[1],[2] However, separation of instruments during continuous rotary motion is the most common
accident in clinical practice.[3]
NiTi instruments may exhibit premature fatigue failure,[4] thus reducing the life of the instrument or torsional stress failure, which is the
result of file binding to the canal walls.[5] Therefore, rotary instrumentation requires manual or mechanical cervical preflaring
to mitigate fracture risk.[6],[7]
In 2008, Yared[8] used a single rotating instrument (ProTaper) in reciprocating motion for canal preparation
and shaping and demonstrated that the technique provides greater speed, good cost‑benefit
ratio, and reduced fracture rates. A new approach to root canal preparation was introduced
in 2010 with the Reciproc (RC) system (VDW, Munich, Germany) and later by the WaveOne
(WO) system (Dentsply Maillefer, Ballaigues, Switzerland). Both use a single NiTi
M‑Wire alloy instrument driven by reciprocating motion, whereby the instrument cuts
the dentin wall of the root canal during the counterclockwise rotation, when taper
lock is imminent. The following clockwise motion then disengages the instrument from
dentin, reducing the incidence of separation.[1],[9]
According to manufacturer recommendations, a glidepath must be prepared using small
(#10 and #15) rotary or hand files, but cervical preflaring is unnecessary.[10] Within the context of widespread use of reciprocating systems, the present study
assessed the influence of multiple reuse (after resterilization) and cervical preflaring
on fracture strength, plastic deformation, and/or surface cracking in RC and WO reciprocating
instruments after root canal instrumentation in extracted human teeth. The null hypothesis
was that both instrument systems would be equally fracture resistant when driven by
reciprocating motion, with and without cervical preflaring.
MATERIALS AND METHODS
Ethical aspects and sample selection
The present study was approved by the local research ethics committee (CAAE: 13831213.4.0000.5052).
The sample comprised extracted human upper molars (length 18–22 mm) with fully formed
roots, constricted root canals, and root curvature ≤35°. For selection purposes, teeth
were identified and imaged by conventional radiography and cone beam computed tomography.
Only teeth in which the entire root canal system could be visualized were included
in the sample for experimental group allocation. All root canals of all selected teeth
were included in the study. Teeth with root curvature >35°, partially or completely
calcified canals, or excessively wide canals were excluded from the study. After application
of the selection criteria, 550 teeth were selected from a set of 1153 extracted human
molars. Of the 550 selected teeth, 485 were used in the study. [Figure 1] provides a schematic illustration of the study methods.
Figure 1: Study flowchart.
Preparation of specimens
In all selected teeth, endodontic access was established using high‑speed diamond
burs. Canal orifices were located with an endodontic explorer and special files under
×32 magnification (MC‑M1233 operating microscope, DFV Com. and Ind. Ltda, Valença,
RJ, Brazil), and teeth were then randomly allocated across the different study groups.
The presence of a fourth canal was identified in 41.2% of the teeth used in the sample
(n = 200). These teeth were distributed equally across the study groups.
Division of experimental groups
Forty 25‑mm reciprocating instruments were used in this study: 20 Reciproc® R25, tip
size 25, 0.08 variable taper (RC) (VDW GmbH, Munich, Germany) and 20 WaveOne® Primary,
tip size 25, 0.08 variable taper (WO) (Dentsply‑Maillefer, Ballaigues, Switzerland),
all powered by a Silver Reciproc Motor and Sirona ENDO 6:1 contra‑angle handpiece
(VDW GmbH, Munich, Germany). Instruments were driven according to the preset programs
defined by the motor manufacturer. RC files were driven by the RC ALL program and
WO files by the WO ALL program.
Four study groups (n = 10 each) were defined, according to root canal instrumentation procedure:
-
G1: RC;
-
G2: Cervical preflaring with Gates‑Glidden drills (#3 and #2) (Dentsply‑Maillefer,
Ballaigues, Switzerland) + RC;
-
G3: WO;
-
G4: Cervical preflaring with Gates‑Glidden drills (#3 and #2) + WO.
Root canal preparation
Instruments were used in alternating order. Procedures were performed by a single
operator, who followed the manufacturer‑issued instructions for each instrument. After
coronal access, teeth were irrigated with 5 mL sodium hypochlorite (NaOCl) 2.5% (Biodinâmica
Química e Farmacêutica Ltda, Ibiporã, PR, Brazil) and explored with hand files as
described elsewhere. The root canals were irrigated with 5 mL 2.5% NaOCl after each
instrumentation.
The root canal systems of all teeth were evacuated with a #10 K‑file, which was used
to identify patency. Working length determination was performed under operating microscope
magnification (MC‑M1233, DFV Com. & Ind. Ltda, Valença, RJ, Brazil), 1 mm short of
the apex. In all teeth, glidepaths were prepared with a #13/0.02 mm/mm PathFile (Dentsply‑Maillefer,
Ballaigues, Switzerland), which enabled standardized canal preparation.
In Groups G1 and G3, instrumentation was performed with a single file (RC and WO,
respectively) in reciprocating motion, advancing into each root third in 3‑mm increments
until the desired length was reached, as recommended by the manufacturers.
In Groups G2 and G4, instrumentation was preceded by cervical preflaring, performed
using #2 and #3 Gates‑Glidden drills (Dentsply‑Maillefer, Ballaigues, Switzerland),
in descending order, and driven by a Reciproc Silver motor programmed appropriately
for these instruments. After this step, RC and WO instruments were used as in Groups
G1 and G3. After each use, any dentin chip adhered to the instruments was removed
with moist gauze dressings, and the root canals were irrigated again with 5 mL of
NaOCl and recapitulated with a #10 K‑file to full length.
Once instrumentation of each tooth was complete, the instruments were decontaminated
and cleaned in an ultrasonic unit, examined under the operating microscope for potential
plastic deformation, and autoclaved for later reuse. This procedure was repeated for
all instruments. To enable reuse after sterilization, the identification ring of each
instrument was removed with a scalpel blade so as to allow reinsertion into the Sirona
contra‑angle handpiece. Each instrument was used repeatedly until fracture occurred,
at which time the instrument was identified, and the number of uses until separation
was recorded.
Scanning electron microscopy
The surfaces of all instruments were analyzed by scanning electron microscopy (SEM),
performed with a DSM‑940A system (Carl Zeiss, Jena, Germany), and checked for surface
changes potentially caused by irrigant exposure and for cracks or plastic deformation
caused by continued instrumentation. SEM images (magnification ×100, ×250, ×500, and
×1000) were obtained at three time points for each instrument: Before use [Figure 2]; after first use [Figure 3]; and after fracture [Figure 4].
Statistical analysis
Statistical procedures were performed with the Statistical Package for the Social
Sciences (SPSS 17.0 for Windows, SPSS Inc., Chicago, USA). Because values were normally
distributed across all groups, one‑way analysis of variance was used for comparative
purposes. The level of significance was set at 5%.
RESULTS
Overall study results are shown in [Table 1]. RC files exhibited significantly greater fracture strength than WO files (P < 0.001), under equal conditions. Comparison of groups in which the same instruments
were used revealed that cervical preflaring significantly increased the fracture strength
(P < 0.01) of both instrument systems. On comparison between different instrument systems
with and without preflaring, RC files exhibited the same fracture strength as WO files
did after preflaring, providing further evidence of the superiority of the RC system
over the WO system.
Table 1:
Mean and standard deviation number of teeth and root canals instrumented per instrument
per study group
Groups
|
Groups Teeth (mean±SD)
|
Total
|
Root canals (mean±SD)
|
Total
|
Different letters indicate results with significant statistically differences. SD:
Standard deviation, RC: Reciproc®, WO: WaveOne®, CP: Cervical preflaring
|
G1 (RC)
|
10.9±1.20A
|
109
|
37.1±3.9A
|
371
|
G2 (WO)
|
6.2±1.03B
|
62
|
21.2±3.4B
|
212
|
G3 (CP + RC)
|
19.7±1.34C
|
197
|
67.5±4.22C
|
675
|
G4 (CP + WO)
|
11.7±1.25A
|
117
|
39.9±3.67A
|
399
|
Figure 2: Scanning electron microscope images of Reciproc® (a) and WaveOne® (b) instruments before use.
Figure 3: Surface images of WaveOne® instruments after the first use. Note the presence of surface cracks on the WaveOne® instrument (arrows) (scanning
electron microscope images magnification: a ‑ ×250; b ‑ ×500; c ‑ ×1000).
On SEM, surface cracking was only seen in WO instruments [Figure 3]. It should be stressed that although these cracks were already apparent after the
first instrument use, the files remained fracture resistant and could be reused at
least five times before separation.
DISCUSSION
Major changes in the techniques and principles of endodontics have occurred in the
last 20 years.[2] This is demonstrated clearly by recent advances in instrumentation of root canal
systems, such as the development of safer techniques,[8] files made from more elastic and stable alloys,[6] and a lower rate of errors during endodontic instrumentation.[7]
The present study assessed the fracture strength and development of plastic deformations
and/or surface cracks on RC R25 and WO primary reciprocating instruments, with and
without cervical preflaring, by means of operating microscope and SEM images. The
null hypothesis of the study was rejected as the results demonstrated that cervical
preflaring had an impact on the fracture strength of the tested instruments.
The choice of instruments for this study was based on the fact that both were made
from the same material (NiTi M‑Wire alloy), are driven by reciprocating motion, and
share similar manufacturing processes.[11]
The method of reusing sterilized instruments on extracted human teeth simulates a
real‑world clinical setting, particularly in terms of the risk of instrument separation,
although not all anatomical features of the pulp cavity can be standardized.[5] However, the large sample size, random allocation of teeth across study groups,
and glidepath standardization with a #13 PathFile prevented this variable from having
any significant influence on results.
Figure 4: Reciproc® (used in 12 teeth ‑ GI) and WaveOne® (used in 6 teeth ‑ GII) instruments after separation (scanning electron microscope
images magnification: ×250, ×500, ×1000). Note cracks inside the metal mass of Reciproc®
(arrow).
Negotiating the apical foramen and determining apical patency with #15 or smaller
files are essential clinical steps and are firmly enshrined in current practice as
they prevent plugging and/or deviation of the root canal.[12] Hence, the inclusion of these steps in our methods brings our experiment even closer
to the reality of daily clinical practice as clearance of potential interferents from
the root canal system and determination of apical patency play a critical role in
ensuring safety and improving the prognosis of endodontic treatment.[13]
Likewise, the use of cervical preflaring – a clinical procedure widely employed in
endodontic practice – in Groups G2 and G4 of this study further demonstrates our concern
with making our experimental condition reproduce clinical aspects as closely as possible.
Preflaring ensures there is a sufficient space for penetration of the mechanical instrument,
prevents coronal binding, and reduces the risk of instrument separation.[7] In addition, it enhances the range of action of irrigating solution by enabling
advancement of the irrigation needle and penetration of the ultrasonic tip during
passive ultrasonic irrigation (PUI).[5]
Although the RC and WO instruments were conceived for single use only, these devices
have a high level of embedded technology, which led us to analyze their fracture strength
after repeated sterilization before each reuse. This analysis showed that both instruments
remained safe for reuse.
According to the current literature, the advent of reciprocating motion systems has
made root canal shaping and cleaning safer.[8] Several studies have shown that reciprocation provides increased cyclic fatigue
resistance for NiTi instruments.[14],[15] In addition, the past studies suggest the possibility of using only one instrument
to prepare the entire root canal with satisfactory results, particularly because a
single instrument has increased fracture strength as compared to full‑sequence rotary
instrumentation.[14],[16]
The influence of cross‑section, alloy type, manufacturing process, and type of driving
motion on the cyclic fatigue resistance of NiTi instruments has been the object of
several recent studies.[15],[17] Nevertheless, how and why the design of the instrument might influence its behavior
under cyclic fatigue stress remains unclear. Cheung and Darvell[18] showed, in various instruments, that fatigue resistance does not appear to be affected
by instrument design, suggesting that neither the cross‑section area of an instrument
nor its shape is the primary determinant of fatigue life.
However, other studies have suggested that difference in cross‑section appears to
be an important determinant of cyclic fatigue resistance across different files.[19] In a study conducted by Plotino et al.,[20] RC instruments exhibited significantly greater fatigue strength than WO instruments,
which suggests that this difference may be related to the differences in cross‑sectional
area and reciprocating motion between the two systems. Earlier studies also demonstrated
the superiority of RC instrumentation over the WO system in fatigue resistance testing[21] and have shown that the angles employed in instrument motion have a direct influence
on cyclic fatigue.[22] Kiefner et al.[14] believe another factor that enhances resistance in RC and WO instruments is the
type of alloy used in their manufacture. The authors showed that a synergistic effect
between the M‑Wire alloy and reciprocating motion provides a significant increase
in cyclic fatigue resistance. Recently, a new continuous rotary instrument was developed
by Coltene/Whaledent AG (Hyflex EDM). Manufactured using the electrical discharge
machining technique, this instrument has greater resistance to cyclic fatigue when
compared with RC and WO.
In our sample, the RC system was superior to the WO system regardless of cervical
preflaring, which suggests that clearance of the root canal system through glidepath
preparation before instrumentation is of the utmost clinical importance.[23] This step is recommended by the manufacturer of WO but deemed unnecessary when using
the RC system.[24] Perhaps, choosing not to prepare a glidepath could bias results significantly in
favor of the RC system as demonstrated elsewhere in the literature.[20]
In a study of discarded NiTi files, Alapati et al.[3] found cracks along the metal body of the instruments. In our sample, SEM analysis
of the instruments after the first use revealed cracks only on the WO files. We did
not observe any plastic deformation, which may be explained by the greater resistance
of the M‑Wire alloy as compared with other NiTi alloys.[17]
It bears stressing that the back‑and‑forth motion recommended by manufacturers and
followed during specimen preparation in this study is largely intended to reduce torsional
strain and, consequently, torsion fatigue, whereas its effect on flexural stresses
is probably less evident. To date, RC and WO have been marketed for single‑file root
canal instrumentation and are approved for single use only so as to prevent structural
weakening as a result of cyclic fatigue. However, “single use” means the same instrument
may still be used to instrument three or four root canals, each of which may have
complex, tortuous anatomy.[1] Therefore, single use reduces but does not altogether eliminate the risk of cumulative
fatigue damage and potential instrument separation. Hence, we conclude that cyclic
fatigue testing is as valuable for reciprocating instrumentation as for rotary instruments.
It should be noted that the high level of fatigue resistance of reciprocating instruments
may play an important role in improving root canal cleaning and shaping procedures
as shaping can be done with remarkable speed. Within this context, the principles
of endodontics are shifting away from simultaneous cleaning and shaping toward a paradigm,
whereby the root canal system is first shaped to enable effective cleaning by means
of enhanced irrigation procedures.[25] Evidence suggests that enhanced apical preparation provides a greater reduction
of bacterial load and dentin debris, primarily by enabling more effective irrigation,[4] which is made easier by cervical preparation.[26]
In the present study, cervical preflaring before instrumentation played a relevant
role in increasing the fatigue life of the tested instruments. Another relevant clinical
factor is localization, negotiation, and determination of apical patency, especially
as endodontic treatment of older adults – who often present with constricted pulp
chambers filled with dystrophic calcifications – has become more common. Therefore,
canal clearance and apical debridement are considered important procedures to reduce
instrument stress against the walls of the canal system and thus improve the safety
of endodontic treatment.[6] Nevertheless, in a recent study by De‑Deus et al.,[24] the RC system demonstrated satisfactory fracture strength even in the absence of
patent canals.
The present study sought to assess the fracture strength of RC and WO instruments
after repeated sterilization and reuse cycles. Our results suggest that both the RC
and WO systems can be safely reused several times, preceded by some preliminary manipulation
to prevent the instrument from binding and thus reduce the risk of separation. Based
on our findings, we suggest that files can be reused safely up to 3 times. This number
also facilitates control of the number of reuses of a single file. On the first use,
the instrument is taken directly from its original packaging and its identifier ring
is intact. After the first reprocessing, the identifier ring is enlarged and must
be removed before the instrument can fit into the handpiece; this serves as a milestone
for the second use of the instrument. Finally, at the time of the third use, the instrument
will already be missing its identifier ring, which demonstrates it should be discarded
after use.
Figure 5: Scanning electron microscope images (×100) of a Reciproc® R25 instrument showing corrosion and loss of metal mass.
Despite these conclusions, we believe that further research is needed to assess the
ability of these reciprocating systems to ensure adequate canal antisepsis, as the
duration of exposure to NaOCl irrigating solution tends to be reduced in single‑file
preparations, and to evaluate the influence of this clinical practice on treatment
success over time.
Notably, one of the instrument specimens produced dark discoloration of the irrigation
solution shortly after the start of its first use. The canal was irrigated with fresh
solution and dark discoloration occurred again. Upon removal of the instrument from
the root canal, it was found to have been markedly corroded by the 2.5% NaOCl solution,
with loss of material. On SEM examination, this instrument (lot #1108000846) exhibited
structural changes [Figure 5]. Energy dispersive X‑ray analysis did not show any difference in alloy properties
in relation to the other instruments used in the study [Figure 6]. It bears stressing that no instances of such damage, i.e., structural corrosion
of NiTi instruments by NaOCl solution, have ever been reported in the literature.[21] This event was probably due to a flaw in the manufacturing process as other instruments
from the same lot exhibited stable behavior.
Figure 6: Energy dispersive X‑ray analyses of the surface (a) and eroded area (b) of the instrument
damaged by 2.5% sodium hypochlorite solution.
CONCLUSION
Considering the proposed objective of this study, we conclude that both RC and WO
instruments exhibited good fracture strength after autoclave sterilization and reuse.
Cervical preflaring significantly increased the number of times RC and WO files could
be reused safely.
Financial support and sponsorship
Nil.