Introduction
In diagnostic imaging, the photoelectric effect is the most important mechanism by
which the x-ray photons interact with matter. The chances of this effect increase
as the energy of the x-ray beam approaches the Kedge of the particular atom.[[1]] Different atoms have different Kedge values and hence behave differently at different energies of the x-ray beam.
In dual-energy CT, two different x-ray spectra are used to acquire two image datasets
of the same anatomic region, allowing analysis of energy dependent changes in the
attenuation of different materials.[[2]] Every material shows a relatively specific change in attenuation between images
obtained with a high and a low-energy spectrum and this attenuation difference allows
better characterization of the tissues. Two different materials that show similar
attenuation on images acquired with one of the two energy spectra, may show substantial
differences in their attenuation on the images acquired with the other spectrum and
hence may be easily differentiated.[[3]]
There are multiple techniques available for dual energy CT imaging which can be broadly
classified as:
-
Dual source DECT
-
Single source DECT.
Single source DECT can be further divided based on the exact mechanism to generate
two different energy spectra as:
-
Fast kV switching
-
Dual layer detector
-
Slow kV switching
-
Dual spiral dual energy
-
Twin beam dual energy.
We use the Siemens Somatom Definition Edge CT scanner which is based upon a single
source twin beam dual energy technique [[Figure 1]].
Figure 1: In twin beam DECT, two different filters Tin (Sn) and Gold (Au) are used to split
the x-ray beam into two beams of different energies. The Gold (Au) filter, filters
the high energy photons and produces a low energy beam while Tin (Sn) filter, filters
the low energy photons and produces a high energy beam
In twin beam DECT, two different filters Tin [Sn] and Gold [Au] are used to split
the x-ray beam from a single source into two beams of different energies. The greatest
advantage of this technique over the many of the other techniques mentioned above
is eliminating temporal mis-registration.
Applications in Clinical Practice
Material differentiation
Bone removal in angiography
As DECT has the ability to differentiate different materials from one another, calcium
in the bones is readily separated from iodine in the contrast media. Hence the bones
in any angiography study can be easily removed. This is particularly useful in cerebral
[[Figure 2]] and lower limb angiography [[Figures 3] and [4]], where manual methods can be difficult and time consuming, especially when the
vessels lie close to bones.[[4],[5]] Similarly, calcified plaques can be removed for better visualisation of the lumen
[[Figure 5]].
Figure 2 (A-C): A routine cerebral CT angiography study (A) shows carotid and cerebral arteries overlapped
by bones making it difficult to visualise them. With DECT bone removal tool, all the
bones are effectively and selectively removed allowing clear visualisation of the
arteries with AVM (arrow) as shown in this MIP image (B). Similarly the arteries with
AVM (arrow) can also be evaluated in VRT image (C) with bone removal
Figure 3 (A-D): A routine lower limb CT angiography study (A) shows arteries overlapped by the long
bones. With DECT bone removal technique in MIP image (B), the right popliteal artery
occlusion (arrow) is clearly visualised with distal reformation of the tibial artery
seen via collateral circulation. Similarly left distal popliteal artery occlusion
(short arrow) just at the level of its division is also seen. Similarly VRT images
without (C) and with (D) bone removal also shows the pathologies (arrows) well
Figure 4 (A-C): A lower limb CT angiography study (A) shows arteries overlap by bones. DECT bone
removal technique VRT (B) and MIP (C) images shows the occlusion of the left superficial
femoral artery (arrow) well with extensive collateral circulation and distal reformation
of the popliteal and the tibial arteries (short arrow)
Figure 5 (A and B): Same case as shown in Figure 4, also shows diffuse atherosclerotic wall calcification
(arrows), compromising accurate lumen visualisation (A). With DECT calcium removal
technique applied on VRT images (B) the calcium in the wall of the arteries is effectively
removed, allowing the visualisation of the underlying lumen easily
Bone marrow edema
Bone marrow edema is best visualised on MRI imaging. However with DECT, it is now
possible to evaluate bone marrow edema on CT scan as well. As DECT can differentiate
calcium from other elements, it can also selectively remove calcium from the region
of interest. This is called virtual non-calcium (VNC) technique and it allows visualisation
of bone marrow edema.[[6]] Multiple studies have validated this technique for bone marrow edema detection.[[7],[8]]
It is especially useful in acute fractures where MRI may not be possible [[Figures 6] and [7]]. Kaup M et al.[[9]] demonstrated the utility of DECT for bone marrow edema evaluation in osteoporotic
vertebral compression fractures and found that its diagnostic performance approached
that of MRI.
Figure 6 (A-D): CT image in coronal reconstruction (A) and 3D reconstruction (B) shows fractures
(arrows) involving the neck and the greater tuberosity of the right humerus. With
DECT, the bone marrow edema (arrows) can be seen in the the neck and the greater tuberosity
of the humerus as seen in DECT coronal reconstruction (C) and DECT 3D reconstruction
(D)
Figure 7 (A-C): Another example of bone marrow edema seen with DECT, in a case of trauma to the right
knee, 1 week ago. CT scan in axial (A) and coronal (B) reconstructions shows a linear
undisplaced fracture (arrows) in the distal femur in the intercondylar region. DECT
images (C) also shows mild residual edema (arrows) at the fracture site
Material optimisation
Metal artefact reduction
Metal in prosthesis can cause significant artefacts and hamper evaluation of periprosthetic
soft tissue. These metal artefacts are caused by beam hardening artefacts.[[10]]
Radiographs are typically obtained post-operatively but they have low sensitivity
and specificity.[[11]] MRI is a great tool to assess bone marrow edema and soft tissue lesions but the
images are distorted by metal artefacts.[[12]]
With the conventional CT, multiple techniques are available to reduce the metal artefacts.[[13]] Increasing in tube current, increasing voltage and reducing slice thickness can
help reduce metal artefacts. Various modifications in reconstructions techniques like
changing the deconvolution kernel from bone to soft tissue or smooth kernel, using
interactive reconstructions and extending the Hounsfield unit scale from 4000 to 40000
HU can also help reduce metal artefacts.
Using DECT, the artefacts from metal can be significantly reduced to almost nil. As
the artefacts are more at lower energy x-ray beams, the monochromatic images at higher
keV settings show significantly less artefacts.
Additionally, iterative reconstruction protocols such as iMAR (interactive metal artefact
reduction) can be used that further reduce the metal artefacts. When both the techniques
are used together, the final result shows almost nil metal artefacts [[Figures 8] and [9]]. This allows the visualisation of periprosthetic soft tissue to identify complications
such as periprosthetic osteolyis, fracture, infection, aseptic loosening, focal particle
disease and tumor. Metal artefact reduction can also be used for evaluation of stents
and to detect endoleak after aneurysm repair.[[14]]
Figure 8: In a known case of left total hip replacement, the conventional CT images shows extensive
metal artefacts, compromising adequate visualisation of the bones and the soft tissue.
With DECT, multiple reconstructions at various keV can be obtained and with progressively
increasing keV, the metal artefacts keep reducing. Addition of iterative reconstruction
(iMAR), further reduces metal artefacts and when both techniques are combined, the
final image shows almost nil artefacts, allowing accurate evaluation of the bone and
the soft tissue
Figure 9: Another example of metal artefact reduction in a patient with bilateral total hip
replacement. Higher keV images shows less metal artefacts and addition of iterative
reconstruction (iMAR) further reduces the metal artefacts and when both techniques
are combined, the final image shows almost nil artefacts
Material separation
DECT has the ability to separate different materials based on their specific dual
energy ratio. Hence, uric acid crystal can be separated from calcium, both of which
otherwise appears hyperdense on conventional single energy CT.
Urinary stones
As different types of renal calculi are treated differently, knowledge of the composition
of stones may guide management decisions and predict the effectiveness of therapy.
For example, uric acid calculi can be managed medically with urine alkalinization
that facilitates dissolution, while non-uric acid calculi need other lines of treatment
such as lithotripsy or surgery.[[15]]
Using DECT, it is possible to differentiate uric acid from non-uric acid calculi.
Keeping the dual energy ratio at 1.05, the uric acid stones lie below the line/slope
and are coloured red, whereas the non-uric acid stones lie above the line/slope and
are differently colour-coded (blue) [[Figures 10] and [11]].
Figure 10 (A-D): The conventional CT images shows the calculi (arrows) in the kidney (A) and ureter
(B) as hyperdense structures and their composition cannot be determined. With DECT
uric acid calculus is shown in red (C) while non-uric acid calculus is shown as blue
(D), allowing their differentiation. The graph also helps shows the two calculi differently,
with the uric acid calculus lying below the line, and non-uric acid lying above
Figure 11 (A-C): Another example of urinary calculi (arrows), two calculi are seen in the kidney in
the CT scan image (A). DECT the calculus red (B) in colour and falling below the line
(C), thus confirming uric acid calculi
Gout
Gout is characterized by an inflammatory response to the deposition of monosodium
urate (MSU) crystals in the joints and soft tissue, which leads to acute or chronic
arthropathy and gouty tophi formation. Diagnosis can be made on clinical and biochemical
bases, and definite diagnosis requires microscopic demonstration of MSU crystals from
the aspiration of the joint fluid,[[16]] a method that is invasive and may have false negative results.
Keeping the dual energy ratio specific to uric acid crystals, the uric acid crystals
are coloured green, whereas calcium is coloured blue. The diagnosis of gout can be
easily made, non-invasively [[Figure 12]].
Figure 12 (A and B): A 35 years male presented with pain and swelling in right foot and wrist. Serum uric
acid-13 mg/dl. Radiographs of both feet, knees and wrists (A) shows juxta-articular
erosions (arrows) around right 1st metatarsophalangeal joint, right lateral femoral condyle, right distal ulna and 5th carpometacarpal joint with soft tissue. Diagnosis of gout was made. B. DECT (B) allows
visualisation of the urate crystals seen as green coloured areas (arrows), confirming
the diagnosis. Also, it shows involvement of the asymptomatic and radiographically
occult areas and displays total tophus volume
With DECT, it is also possible to quantify the overall tophus burden or volume of
urate deposition without any user variability. Thus, makes DECT an ideal tool for
evaluating any change in tophus burden and can be used for follow-up to document response
to treatment[[17]] [[Figure 13]].
Figure 13 (A and B): A 38 years female presented with bilateral knee pain with normal serum uric acid
levels. The DECT (A) showed small foci of urate crystal deposition (arrows) in both
the knees and the left foot, confirming the diagnosis of gout and the patient was
put on treatment with allopurinol. The follow up DECT scan (B) post 2 months of treatment
shows significant improvement and resolution of the urate crystals, suggestive of
response to treatment
Material decomposition
Iodine maps
Lung
Pulmonary Thromboembolism
The dual energy iodine maps allow us to evaluate lung perfusion and hence are useful
to demonstrate perfusion defects in cases of pulmonary thromboembolism. Additionally,
the dual energy iodine maps also improve visualisation of filling defects/thrombi
in a small segmental or subsegmental artery. Fink et al.[[18]] showed the sensitivity and specificity of dual-energy CT perfusion mapping for
the assessment of pulmonary embolism to be 100% on a per patient basis and 60-66.7%
and 99.5-99.8% on a per se gment basis, as compared to CT angiography.
In patients with pulmonary thromboembolism, the dual energy perfusion maps show peripheral
wedge-shaped areas of reduced perfusion representing perfusion defects [[Figure 14]]. They are also very useful for the follow up of such patients who are treated with
anticoagulants to demonstrate both a decrease in perfusion defects as well as the
thrombi [[Figures 15] and [16]].
Figure 14 (A and B): CT Pulmonary angiography (A) shows a saddle thrombus (arrows) involving both the
pulmonary arteries and also extending into the right upper and lower lobe branches.
The DECT perfusion maps (B) shows patchy perfusion defects (arrows) in the right upper
and lower lobes
Figure 15 (A-E): CT Pulmonary angiography axial (A) and coronal (B) images shows extensive bilateral
pulmonary thromboembolism (arrows). The HRCT axial (C) image shows a developing infarct
(arrow) in the posterior segment of the right upper lobe. The DECT images (D and E)
shows perfusion defects (arrows) in the lung parenchyma bilaterally with the infarct
in the right upper lobe
Figure 16 (A-E): The follow up CT Pulmonary angiography axial (A) and coronal (B) images, after 3
months of anticoagulation treatment shows complete resolution of the bilateral pulmonary
thromboembolism. The HRCT axial (C) image shows resolution of the infarct in the right
upper lobe. The DECT images (D and E) shows reversal of the perfusion defects in the
lung parenchyma bilaterally
One should keep in mind the areas where artefacts are commonly seen and should not
misinterpret them as areas of perfusion defects. These artefactual perfusion defects
are commonly seen in the upper lobes due to beam hardening artefacts, in the medial
segment of the right middle lobe and lingula due to cardiac pulsation and in the lung
bases due to diaphragmatic motion. The other common pitfall is apparent perfusion
defects in an area of lung parenchyma with systemic arterial supply. Also, parenchymal
diseases such as collapse, consolidation, masses and emphysema show perfusion defects.[[19]]
Lung nodule
With DECT a single contrast enhanced scan is sufficient to differentiate an enhancing
nodule from calcification with the use of virtual nonenhanced image reconstructions.[[20]] The enhancing and non-enhancing components of a heterogeneous lesion are better
appreciated on iodine maps, which also help target the appropriate area for biopsy,
to increase the yield [[Figure 17]].
Figure 17 (A-D): The plain HRCT (A) and contrast (B) axial images shows an ill-defined spiculated
enhancing mass (arrow) in the right upper lobe with mediastinal lymphadenopathy (short
arrow). The DECT axial images shows the more enhancing part with higher iodine uptake
in the medial part of the mass (C), which can be targeted for biopsy, while the lateral
part (D) shows relatively less enhancement
Degree of contrast enhancement is an important factor to differentiate a benign from
a malignant lesion.[[21]] Hence dual energy analysis with iodine maps is very useful in the evaluation of
solitary or multiple pulmonary nodules. The enhancement in a nodule is not always
easily appreciable on routine CT scan and DECT iodine maps help confirm or rule out
enhancement conclusively [[Figures 18] and [19]].
Figure 18 (A and B): The axial image (A) of the lungs shows a nodule (arrow) in the left lung with questionable
enhancement. The DECT iodine map (B) help rule out enhancing nodule by showing no
significant iodine uptake, hence suggestive of an old healed granuloma
Figure 19 (A-D): DECT iodine maps (A, B and D) shows no significant iodine uptake in most of the nodules.
Only one nodule (C) shows significant iodine uptake, worrisome for active nodule
Difference in the contrast wash out pattern of a metastatic nodule can be used to
evaluate various types of primary malignant tumors. For example, metastases from colorectal
carcinoma, malignant melanoma and thyroid carcinoma show a distinct wash out pattern
whereas those from lung cancer, salivary gland cancer and sarcoma show increased enhancement
in the delayed phase images.[[22]]
Liver lesions
A hypervascular lesion is more conspicuous on lower kV images on an arterial phase
of the study, as iodine is more attenuating at lower energies.[[23]] DECT may be used to identify fat, iron, calcium, or hemorrhage within the tumor.
Virtual non-enhanced images can be used to differentiate calcification in a lesion
from enhancement.
DECT iodine maps are especially useful in differentiating a simple cyst from a complex
cyst as they show the subtle enhancement in the lesion very well, which is often difficult
to appreciate on a single energy CT scan [[Figures 20] and [21]].
Figure 20 (A-D): A 53 years female, treated case of carcinoma breast, came for follow up scan. The
contrast enhanced abdomen image (D) shows a hypodense lesion (arrow) in the left lobe
of the liver. The DECT iodine maps (A-C) shows significant iodine uptake (arrow) suggestive
of an enhancing lesion, worrisome for metastatic disease, which was later proved with
biopsy
Figure 21 (A-D): The contrast enhanced abdomen image (D) shows a hypodense lesion (arrow) in the left
lobe of the liver. The DECT iodine maps (A-C) shows no significant iodine uptake (arrow)
suggestive of a simple cyst
Kidney lesions
DECT is useful in differentiating a hyperdense cyst from an enhancing solid tumor.
A hyperdense cyst containing proteinaceous material or blood products may mimic a
solid enhancing tumor on a contrast enhanced scan. In such situations, iodine maps
help as a hyperdense cyst will not show iodine uptake. Also, virtual non-enhanced
images will show hyperdense contents in the cyst. Similarly, enhancement in a hyperdense
lesion may be difficult to appreciate on routine CT images and DECT iodine maps show
the enhancement well even in hyperdense parts of the lesion [[Figure 22]].
Figure 22 (A-D): The contrast enhanced CT shows a bilobed complex cystic lesion arising exophytically
from the right kidney. The superior lobule shows thick walls with measurable enhancement
(arrow) (A) while the inferior lobule shows wall calcification (short arrow) (B).
The DECT iodine maps (C and D) shows definite iodine uptake in the thick walls of
the superior lobule, suggestive of bosniak class 3 cystic lesion
Adrenal lesions
The majority of the adrenal lesions seen on an abdomen CT scan are benign. On imaging
the diagnosis of a benign adrenal lesion can be confidently made if it shows a Hounsfield
value (HU) value of less than 10 on an unenhanced image (adenoma), shows macroscopic
fat (myelolipoma) or has attenuation of water (cyst). The appearance of an adenoma
on DECT is variable and may show decrease or increase in attenuation at lower kV images
than higher, depending on the amount of lipid within the lesion.[[24]] DECT iodine maps can help pick up enhancing focal lesions in an adrenal gland [[Figure 23]].
Figure 23 (A-E): The same case of carcinoma of lung (A and B) as shown in Figure 17, also shows an
enhancing nodule in the left adrenal gland (arrow) (C), worrisome for metastatic disease.
The DECT iodine maps confirms the iodine uptake in the left adrenal gland nodule (E),
and also shows another nodule with iodine uptake in the right adrenal gland (D) as
we
Bowel pathologies
DECT is helpful in the evaluation of inflammatory bowel pathologies and shows the
inflammation/enhancement of the wall better [[Figure 24]]. It is also useful in assessing the severity and evaluate extraintestinal manifestations.
Iodine maps demonstrate bowel wall enhancement/hyperemia in cases of inflammatory
bowel disease. Coexisting neoplastic lesion can also be better assessed with DECT
in such cases.
Figure 24 (A-D): 32 years male on treatment for inflammatory bowel disease presented with acute abdominal
pain and diarrhoea. The contrast enhanced CT axial (A and B) and coronal (C) images
shows multifocal areas of small bowel thickening (arrows) with mesenteric hypervascularity
(Comb’s sign) (short arrows), fat stranding and lymph nodes, suggestive of inflammatory
bowel disease. The DECT iodine maps (D) shows active areas with significant iodine
uptake (arrows) in the thickened bowel loops
Post ablation analysis
Radiofrequency ablation (RFA) is now frequently used to treat a focal lesion in the
liver, lung or kidney. It is important to detect residual tumor post RFA to decide
if re-ablation is necessary. Conventionally, contrast enhanced single energy CT has
been used to evaluate the success of RFA, and it may be difficult to pick up residual
tumor and differentiate it from post ablation changes. DECT allows better depiction
of the enhancing component in residual lesions using iodine overlay images, while
the necrotic post-ablation regions appear avascular and do not show iodine uptake
[[Figure 25]]. Monochromatic images also allow selection of the optimal keV that shows the best
contrast to noise ratio between the tissues, thus allowing better evaluation of the
interface between the ablation zone and surrounding tissue.[[25]]
Figure 25 (A-G): Axial HRCT (A) and contrast enhanced (B) images shows an enhancing speculated mass
in the left upper lobe (arrows), which on biopsy was proved to be adnocarcinoma. RFA
(C) was performed to treat the lesion. The preablation DECT iodine maps (D and E)
shows significant iodine uptake, which significantly reduced following RFA (F and
G) suggestive of successful ablation
Atomic number maps (Rho/Z)
The new dual energy CT application allows calculation of the atomic number in the
lesion, for separation of various materials. Mileto et al.[[26]] showed that non-enhancing renal cysts, including hyperattenuating cysts, can be
discriminated from enhancing masses on effective atomic number maps. They concluded
8.36 value was the optimal threshold with the enhancing masses showing higher values.
Our cases also showed similar results [[Figures 26] and [27]].
Figure 26 (A-D): The contrast enhanced CT (A and B) images shows 2 enhancing masses (arrows) in the
right kidney. The DECT atomic number maps shows Rho/Z values of 9.11 (C) and 10.5
(D) for the two masses, confirming solid enhancing masses
Figure 27 (A and B): The contrast enhanced CT (A) image shows a nonenhancing cystic lesion (arrow) in
the left kidney. The DECT atomic number maps shows Rho/Z value of 8.27 (B), confirming
non-enhancing cystic lesion