Safety of lumen-apposing stent with or without coaxial plastic stent for endoscopic ultrasound-guided drainage of pancreatic fluid collections: a retrospective study

Authors

  • Manuel Puga

    1   Department of Digestive Diseases, Complexo Hospitalario Universitario de Ourense, Ourense, Spain
    2   Endoscopy Unit, Department of Digestive Diseases, Hospital Universitari de Bellvitge, Bellvitge Biomedical Research Institute (IDIBELL), University of Barcelona, Barcelona, Spain
  • Claudia F. Consiglieri

    2   Endoscopy Unit, Department of Digestive Diseases, Hospital Universitari de Bellvitge, Bellvitge Biomedical Research Institute (IDIBELL), University of Barcelona, Barcelona, Spain
  • Juli Busquets

    3   Department of Surgery, Hospital Universitari de Bellvitge, Bellvitge Biomedical Research Institute (IDIBELL), University of Barcelona, Barcelona, Spain
  • Natàlia Pallarès

    4   Statistics Advisory Service, Bellvitge Biomedical Research Institute (IDIBELL), Barcelona, Spain
  • Lluis Secanella

    3   Department of Surgery, Hospital Universitari de Bellvitge, Bellvitge Biomedical Research Institute (IDIBELL), University of Barcelona, Barcelona, Spain
  • Nuria Peláez

    3   Department of Surgery, Hospital Universitari de Bellvitge, Bellvitge Biomedical Research Institute (IDIBELL), University of Barcelona, Barcelona, Spain
  • Joan Fabregat

    3   Department of Surgery, Hospital Universitari de Bellvitge, Bellvitge Biomedical Research Institute (IDIBELL), University of Barcelona, Barcelona, Spain
  • José Castellote

    5   Hepatology Unit, Department of Digestive Diseases, Hospital Universitari de Bellvitge, Bellvitge Biomedical Research Institute (IDIBELL), University of Barcelona, Barcelona, Spain
  • Joan B. Gornals

    2   Endoscopy Unit, Department of Digestive Diseases, Hospital Universitari de Bellvitge, Bellvitge Biomedical Research Institute (IDIBELL), University of Barcelona, Barcelona, Spain
Weitere Informationen

Publikationsverlauf

submitted 01. November 2017

accepted after revision 30. Januar 2018

Publikationsdatum:
28. März 2018 (online)

Preview

Abstract

Background The aim of this study was to evaluate whether the placement of a coaxial double-pigtail plastic stent (DPS) within a lumen-apposing metal stent (LAMS) may improve the safety of endoscopic ultrasound (EUS)-guided drainage of pancreatic fluid collections (PFCs).

Methods This was a retrospective cohort study including patients with PFCs and an indication for transmural drainage. Two strategies (LAMS alone or LAMS plus DPS) were used at the endoscopist’s discretion.

Results A total of 41 patients were treated (21 LAMS alone; 20 LAMS plus DPS). The characteristics of the PFCs, and the technical and clinical success rates did not differ between groups. The LAMS alone group had a significantly higher rate of adverse events than the LAMS plus DPS group (42.9 % vs. 10.0 %; P = 0.04). Bleeding was the most frequent adverse event observed.

Conclusions The addition of a coaxial DPS to LAMS was associated with a lower rate of adverse events in EUS-guided drainage of PFCs.