A novel tapered plastic stent with an ultrafine delivery system for one-step endoscopic ultrasound-guided hepaticogastrostomy

Authors

  • Saburo Matsubara

    Department of Gastroenterology and Hepatology, Saitama Medical Center, Saitama Medical University, Saitama, Japan
  • Keito Nakagawa

    Department of Gastroenterology and Hepatology, Saitama Medical Center, Saitama Medical University, Saitama, Japan
  • Kentaro Suda

    Department of Gastroenterology and Hepatology, Saitama Medical Center, Saitama Medical University, Saitama, Japan
  • Takeshi Otsuka

    Department of Gastroenterology and Hepatology, Saitama Medical Center, Saitama Medical University, Saitama, Japan
  • Masashi Oka

    Department of Gastroenterology and Hepatology, Saitama Medical Center, Saitama Medical University, Saitama, Japan
  • Sumiko Nagoshi

    Department of Gastroenterology and Hepatology, Saitama Medical Center, Saitama Medical University, Saitama, Japan
Preview

A 76-year-old woman was admitted to our hospital with obstructive jaundice due to pancreas head cancer ([Fig. 1]). Endoscopic transpapillary biliary drainage was attempted but was unsuccessful owing to duodenal invasion. We therefore tried one-step endoscopic ultrasound-guided hepaticogastrostomy (EUS-HGS) using a novel stent system (Crane stent; SB-Kawasumi, Kanagawa, Japan). The stent system comprises a straight-type, 7-Fr, plastic stent with a tapered tip of 4.7 Fr, an inner catheter with an extremely fine 2.5-Fr tip designed for use with a 0.025-inch guidewire, and a pusher connected to the stent allowing it to be pulled back ([Fig. 2]). A stent with a length of 15 cm between the flaps was selected.

Zoom
Fig. 1 Contrast-enhanced computed tomography showing: a dilated intrahepatic bile ducts; b pancreas head cancer (*) invading the second part of the duodenum (†).
Zoom
Fig. 2 A novel stent system (Crane stent; SB-Kawasumi, Kanagawa, Japan), which comprises a straight-type, 7-Fr, plastic stent with a 4.7-Fr tip, an inner catheter with a 2.5-Fr tip designed for use with a 0.025-inch guidewire, and a pusher connected to the stent. Inset: Image showing a 0.025-inch guidewire threaded through the delivery system. Source: SB-Kawasumi.

After puncturing B2 with a 19-gauge needle through the stomach, bile was aspirated as much as possible to reduce bile leakage. Following injection of a small amount of contrast, a 0.025-inch guidewire was inserted into the bile duct and the stent system was inserted over the guidewire without fistula dilation ([Video 1]). Finally, the stent was released in the stomach. The time from puncture to stent placement was 5 minutes. The post-procedure course was uneventful.

Video 1 One-step endoscopic ultrasound-guided hepaticogastrostomy with a novel, tapered, plastic stent with an ultrafine delivery system.

One-step EUS-HGS with a covered metal stent stowed in a thin delivery system has recently been reported to shorten procedure time and reduce bile leakage [1] [2]. Although the Crane stent is a plastic stent that cannot be stowed thinly, one-step stenting was possible by minimizing the gap between the guidewire and inner catheter, and between the inner catheter and stent ([Fig. 3]). The advantage of plastic stents is easy implantation [3], but bile leakage is more likely to occur than with covered metal stents [4]. However, as previously reported [5], bile leakage could be prevented even with plastic stents by aspirating bile sufficiently before stent insertion.

Zoom
Fig. 3 The novel stent system (Crane stent; SB-Kawasumi, Kanagawa, Japan) compared with other devices with a 0.025-inch guidewire threaded. a A conventional endoscopic retrograde cholangiopancreatography catheter (01 20 21 1; MTW Endoskopie, Düsseldorf, Germany). b A 7-Fr plastic stent (Flexima Plus; Boston Scientific Japan, Tokyo, Japan). c A 7-Fr plastic stent (Through & Pass TYPE IT; Gadelius Medical, Tokyo, Japan). d A 7-Fr plastic stent (SUZAKU; Kaneka Medix, Osaka, Japan). With the exception of the SUZAKU stent, which was also designed for a 0.025-inch guidewire, the gap between the tip of the devices and guidewire is larger than for the Crane stent. Additionally, the Crane stent is tapered more than the other stents and has the smallest gap with the inner catheter.

This novel plastic stent system has the potential to make EUS-HGS easier, faster, and safer. Further investigations in a large cohort are warranted.

Endoscopy_UCTN_Code_TTT_1AS_2AD

Endoscopy E-Videos is an open access online section, reporting on interesting cases and new techniques in gastroenterological endoscopy. All papers include a high quality video and all contributions are freely accessible online. Processing charges apply (currently EUR 375), discounts and wavers acc. to HINARI are available.

This section has its own submission website at https://mc.manuscriptcentral.com/e-videos



Publikationsverlauf

Artikel online veröffentlicht:
16. Januar 2023

© 2023. The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes, or adapted, remixed, transformed or built upon. (https://creativecommons.org/licenses/by-nc-nd/4.0/)

Georg Thieme Verlag KG
Rüdigerstraße 14, 70469 Stuttgart, Germany