Thermal changes in guidewires used during endoscopic retrograde cholangiopancreatography during electrical conductance: Ex-vivo safety experiment

Abstract
Risks of pancreaticobiliary tissue damage secondary to electrical conduction along cannulation wires in the vicinity of electrocautery have rarely been independently studied and remain mostly a theoretical issue. We aimed to evaluate heat generated by commercially available guidewires in the setting of currents used for sphincterotomy to assess their safety during endoscopic procedures. We tested heat generation from 11 common guidewires used in endoscopic retrograde cholangiopancreatography using an electrosurgical generator in combination with a thermocouple temperature probe in an ex vivo model. Thermal changes during electrical conductance were nominal except for minimally wrapped 0.018” wires where the maximum changes in temperature were 4.9°C with ENDO CUT Q settings and 6.8°C with ENDO CUT I settings. Also, the Glidewire 0.018” and the Visiglide 0.025” produced visible sparks along their distal ends with defects in insulation found later under microscopic evaluation. In our ex-vivo study, minimal heat was generated via electrical conductivity despite direct current, suggesting negligible risk of thermal injury during sphincterotomy.
Keywords
Pancreatobiliary (ERCP/PTCD) - ERC topics - Quality and logistical aspects - Performance and complicationsPublication History
Received: 20 August 2024
Accepted after revision: 31 January 2025
Accepted Manuscript online:
10 February 2025
Article published online:
14 March 2025
© 2025. The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution License, permitting unrestricted use, distribution, and reproduction so long as the original work is properly cited. (https://creativecommons.org/licenses/by/4.0/).
Georg Thieme Verlag KG
Oswald-Hesse-Straße 50, 70469 Stuttgart, Germany
Rahul Karna, Nicholas M McDonald, Jacob Weiner, Théodon I Netoff, Stuart K Amateau. Thermal changes in guidewires used during endoscopic retrograde cholangiopancreatography during electrical conductance: Ex-vivo safety experiment. Endosc Int Open 2025; 13: a25367997.
DOI: 10.1055/a-2536-7997
-
References
- 1
Cappell MS,
Friedel DM.
Stricter national standards are required for credentialing of endoscopic-retrograde-cholangiopancreatography
in the United States. World J Gastroenterol 2019; 25: 3468-3483
MissingFormLabel
- 2
Chandrasekhara V,
Khashab MA,
Muthusamy VR.
et al.
Adverse events associated with ERCP. Gastrointest Endosc 2017; 85: 32-47
MissingFormLabel
- 3
Moon JH,
Cha SW,
Cho YD.
et al.
Wire-guided endoscopic snare papillectomy for tumors of the major duodenal papilla.
Gastrointest Endosc 2005; 61: 461-466
MissingFormLabel
- 4
Han J,
Kim MH.
Endoscopic papillectomy for adenomas of the major duodenal papilla (with video). Gastrointest
Endosc 2006; 63: 292-301
MissingFormLabel
- 5
Young PE,
Darwin PE.
The conductivity of ERCP guide wires during direct contact with current source. Endoscopy
2008; 40: E36-E37
MissingFormLabel