Endoscopic salvage of a dislodged lumen-apposing metal stent in pancreatic pseudocyst drainage

Authors

  • Junfeng Zhang

    1   Institute of Hepatopancreatobiliary Surgery, Chongqing General Hospital, Chongqing University, Chongqing, China
    2   Chongqing Key Laboratory of Intelligent Medicine Engineering for Hepatopancreatobiliary Diseases, Chongqing, China
  • Fei Li

    3   Department of Gastroenterology, Yubei District Hospital of Traditional Chinese Medicine, Chongqing, China
  • Jinhua Deng

    1   Institute of Hepatopancreatobiliary Surgery, Chongqing General Hospital, Chongqing University, Chongqing, China
    2   Chongqing Key Laboratory of Intelligent Medicine Engineering for Hepatopancreatobiliary Diseases, Chongqing, China
  • Tao Zhang

    1   Institute of Hepatopancreatobiliary Surgery, Chongqing General Hospital, Chongqing University, Chongqing, China
    2   Chongqing Key Laboratory of Intelligent Medicine Engineering for Hepatopancreatobiliary Diseases, Chongqing, China

Gefördert durch: Chongqing Science and Health Joint Project 2025MSXM154
Gefördert durch: Natural Science Foundation of Chongqing, China CSTB2022NSCQ-MSX1339
Gefördert durch: Natural Science Foundation of Chongqing, China CSTB2024NSCQ-MSX0003
Preview

A 41-year-old man experienced abdominal pain after consuming greasy food and was diagnosed with a pancreatic pseudocyst following acute necrotizing pancreatitis ([Fig. 1] a), undergoing endoscopic ultrasound (EUS)-guided drainage with a lumen-apposing metal stent (LAMS, Boston Scientific). During the cystogastrostomy procedure, the LAMS inadvertently dislodged completely into the pseudocyst ([Fig. 1] b). Double pigtail stents and a nasocystic drainage tube were immediately placed intraoperatively to ensure continuous drainage of the cyst fluid.

Zoom
Fig. 1 The process of failed LAMS placement and the successful repositioning outcome. a Preoperative CT imaging of the pancreatic pseudocyst. b First intraoperative puncture showing no LAMS in the gastric cavity. c Second intraoperative ultrasound showing the LAMS dislodged into the cyst. d Postoperative CT after the second procedure showing successful repositioning of the LAMS and cyst shrinkage. Abbreviation: LAMS, lumen-apposing metal stent.

One week later, imaging demonstrated significant shrinkage of the pseudocyst ([Fig. 1] c), and the patient remained asymptomatic, with no signs of infection or bleeding. Given the favorable clinical outcome, we decided to attempt endoscopic repositioning of the dislodged LAMS ([Video 1]). First, the displaced stent was localized within the pseudocyst under EUS guidance. Subsequently, a needle knife was advanced over a guidewire through the original puncture tract, successfully entering the pseudocyst cavity. A 2 cm dilation balloon was then used to expand the tract between the gastric and cyst walls, preparing for the next step.

The process of endoscopically inserting a wire into the cyst, performing dilation, and repositioning the LAMS.Video 1

Using a therapeutic gastroscope (1.0 cm outer diameter) guided by the wire, the operator advanced into the pseudocyst, where extensive necrotic debris was encountered. After repeated careful exploration, the dislodged LAMS was identified. With the assistance of forceps, the position of the LAMS was adjusted. One end of the stent was gently pulled into the gastric lumen, while the other end was kept inside the pseudocyst. The stent was then repositioned to securely anchor between the gastric and cyst walls, successfully re-establishing an effective drainage pathway.

After repositioning, the gastroscope passed smoothly through the LAMS into the pseudocyst, and normal cyst drainage resumed (Fig. 1d). This case demonstrates that endoscopic repositioning of a dislodged LAMS is a feasible and safe strategy, offering an effective and minimally invasive alternative for the treatment of pancreatic pseudocysts.

Endoscopy_UCTN_Code_CPL_1AK_2AG

E-Videos is an open access online section of the journal Endoscopy, reporting on interesting cases and new techniques in gastroenterological endoscopy. All papers include a high-quality video and are published with a Creative Commons CC-BY license. Endoscopy E-Videos qualify for HINARI discounts and waivers and eligibility is automatically checked during the submission process. We grant 100% waivers to articles whose corresponding authors are based in Group A countries and 50% waivers to those who are based in Group B countries as classified by Research4Life (see: https://www.research4life.org/access/eligibility/).

This section has its own submission website at https://mc.manuscriptcentral.com/e-videos.



Publikationsverlauf

Artikel online veröffentlicht:
14. Mai 2025

© 2025. The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution License, permitting unrestricted use, distribution, and reproduction so long as the original work is properly cited. (https://creativecommons.org/licenses/by/4.0/).

Georg Thieme Verlag KG
Oswald-Hesse-Straße 50, 70469 Stuttgart, Germany