Abstract
Background Surveillance and monitoring of perforator flaps has been well established for many years as a reliable method to detect early changes in the blood flow of the flap. In this manner, early flap revisions could be performed in cases of probable thrombosis.
Methods We performed a pilot study that included the first 12 patients to use this device. A temperature sensor electrode was used for remote monitoring of perforator flaps using a mobile phone in real time via a general pocket radio service and short messages developed by the authors, allowing the surgeon to check the temperature from any location.
Results The device continually displays the temperature of the flaps in real time performing a curve with measurements every 5 minute using the web page. A significant difference (p < 0.05) is shown between flaps with sufficient flow and flaps with venous thrombosis.
Conclusion Real-time direct thermography via the surgeon's mobile phone is a reliable method for flap monitoring, facilitating monitorization during the time when the surgeon is away from the hospital.
-
References
- 1 Van Waes C, Seidenstuecker K, Mahajan A. , et al. The safety of the deep inferior epigastric artery perforator flap procedure: a prospective analysis in 1036 breast reconstruction patients. Plast Reconstr Surg 2011; 128 (4S): 57 . Doi: 10.1097/01.prs.0000406278.04351.a7
- 2 Furnas H, Rosen JM. Monitoring in microvascular surgery. Ann Plast Surg 1991; 26 (03) 265-272
- 3 Halvorson E. The limited utility and high cost of implantable dopplers for head and neck free flap monitoring. Plast Reconstr Surg 2012; 130 (5S): 13 . Doi: 10.1097/01.prs.0000421712.97633.9f
- 4 Setälä L, Koskenvuori H, Gudaviciene D, Berg L, Mustonen P. Cost analysis of 109 microsurgical reconstructions and flap monitoring with microdialysis. J Reconstr Microsurg 2009; 25 (09) 521-526
- 5 Neumeister MW. Free flap monitoring: keeping it simple and reliable. Plast Reconstr Surg 2006; 117 (03) 1073-1074
- 6 Goodson AM, Payne KF, Simmons J, Jain A. Free-flap telemonitoring to the surgeon's handheld device: proof of concept. Microsurgery 2014; 34 (05) 413-414
- 7 Keller A. Noninvasive tissue oximetry for flap monitoring: an initial study. J Reconstr Microsurg 2007; 23 (04) 189-197
- 8 Smit JM, Zeebregts CJ, Acosta R, Werker PM. Advancements in free flap monitoring in the last decade: a critical review. Plast Reconstr Surg 2010; 125 (01) 177-185
- 9 Abe Y, Hashimoto I, Goishi K, Kashiwagi K, Yamano M, Nakanishi H. Transcutaneous PCO2 measurement at low temperature for reliable and continuous free flap monitoring: experimental and clinical study. Plast Reconstr Surg Glob Open 2013; 1 (02) 1-8
- 10 Swan MC, Wilson HA, McGuiness CN. Free flap temperature homeostasis during microsurgery: benefits of the “microsurgical duvet”. J Reconstr Microsurg 2013; 29 (08) 561-564
- 11 Kruse AL, Luebbers HT, Grätz KW, Obwegeser JA. Free flap monitoring protocol. J Craniofac Surg 2010; 21 (04) 1262-1263
- 12 Hashimoto I, Nakanishi H, Takiwaki H, Takase MT, Yamano M, Sedo H. Flap monitoring by transcutaneous PO2 and PCO2: importance of transcutaneous PCO2 in determining follow-up treatment for compromised free flaps. J Reconstr Microsurg 2007; 23 (05) 269-274
- 13 Lohman RF, Langevin CJ, Bozkurt M, Kundu N, Djohan R. A prospective analysis of free flap monitoring techniques: physical examination, external Doppler, implantable Doppler, and tissue oximetry. J Reconstr Microsurg 2013; 29 (01) 51-56
- 14 Chen Y, Shen Z, Shao Z, Yu P, Wu J. Free flap monitoring using near-infrared spectroscopy: a systemic review. Ann Plast Surg 2016; 76 (05) 590-597
- 15 Sue GR, Kim R, Chen W, Narayan D. Real-time free flap monitoring using a standard video camera: a lower cost and simpler option?. Plast Reconstr Surg 2014; 133 (06) 905e-906e
- 16 Kempton SJ, Chen J, Poore S, Afifi A. Abstract 96: the venous anastomotic flow-coupler for free flap monitoring: a prospective analysis of 85 microsurgical breast reconstruction cases. Plast Reconstr Surg 2014; 133 (3, Suppl) 111 . Doi: 10.1097/01.prs.0000444920.45830.5d
- 17 Chen KT, Mardini S, Chuang DC. , et al. Timing of presentation of the first signs of vascular compromise dictates the salvage outcome of free flap transfers. Plast Reconstr Surg 2007; 120 (01) 187-195
- 18 Wei FC, Jain V, Celik N, Chen HC, Chuang DC, Lin CH. Have we found an ideal soft-tissue flap? An experience with 672 anterolateral thigh flaps. Plast Reconstr Surg 2002; 109 (07) 2219-2226 , discussion 2227–2230
- 19 Sakakibara S, Hashikawa K, Omori M, Terashi H, Tahara S. A simplest method of flap monitoring. J Reconstr Microsurg 2010; 26 (07) 433-434