J Am Acad Audiol 2022; 33(01): 029-035
DOI: 10.1055/s-0041-1735523
Research Article

Standardization of the TEOAE Contralateral Suppression Test in Terms of Stimulus Intensity and Contralateral Noise Duration in Individuals with Normal Hearing

Bahtiyar Celikgun
1   Department of Audiology, Institute of Health Sciences, Marmara University, Marmara Üsniversitesi Başıbüyük Sağlık Yerleşkesi, Sağlık Bilimleri Enstitüsü Müdürlüğü, Istanbul, Turkey
,
Ufuk Derinsu
2   Department of Odyoloji, Marmara Universitesi Cerrahi Tip Bilimleri, Istanbul, Turkey
› Institutsangaben
Preview

Abstract

Background A standard method and parameter study were performed for the contralateral suppression test.

Purpose Our study aimed to determine the contralateral transient-evoked otoacoustic emission (TEOAE) suppression test method and stimulus–noise parameters that have a standard procedure and will enable the efferent system to be easily evaluated in clinics.

Research Design: This study was conducted in two parts with two different groups of participants as a within-subjects design. In the first part, the signal-to-noise ratio at which maximum suppression obtained was investigated with 29 participants. In the second part, the optimal contralateral noise presentation method (in terms of noise-time or noise-sweep) was examined with 21 participants.

Study Sample In the first part 29 young adults aged between 18 and 32 (23.03 ± 2.84), 20 females and 9 males, and in the second part 21 young adults aged between 19 and 34 years (mean age: 23.71 ± 3.48 years), 16 females and 5 males, participated in the second part. All participants had normal hearing.

Data Collection and Analysis To obtain maximum OAE suppression, different parameters were tested with the Otodynamics ILO292-II OAE device at both parts of the study in a double-walled audiometric test booths. Multirepeated analysis of variance, pairwise comparison, Friedman test, and Wilcoxon signed-rank tests were used for statistical analysis.

Results In the first part, maximum suppression was achieved at 65 dB peSPL (decibel peak-equivalent sound pressure level) TEOAE stimulus and 65 dB SPL broadband noise. In the second part, maximum suppression was obtained in noise-time method with 30 seconds “duration.”

Conclusions To provide standardization in clinics, it can be recommended that the contralateral suppression of OAEs was measured at 65 dB peSPL TEOAE stimulus and 65 dB SPL broadband noise in the linear stimulation mode with Otodynamics ILO 292-II double-probe OAE device. To obtain maximum suppression, the noise-time method with 30 seconds duration can be used.

Disclaimer

Any mention of a product, service, or procedure in the Journal of the American Academy of Audiology does not constitute an endorsement of the product, service, or procedure by the American Academy of Audiology.




Publikationsverlauf

Eingereicht: 28. März 2021

Angenommen: 15. Juli 2021

Artikel online veröffentlicht:
05. Mai 2022

© 2022. American Academy of Audiology. This article is published by Thieme.

Thieme Medical Publishers, Inc.
333 Seventh Avenue, 18th Floor, New York, NY 10001, USA