Initial trimming followed by complete removal of an esophageal self-expandable metal stent for stent-related symptoms

Takeshi Tsujino
Division of Gastroenterology and Hepatology, H.H. Chao Comprehensive Digestive Disease Center, University of California, Orange, California, USA
,
John G. Lee
Division of Gastroenterology and Hepatology, H.H. Chao Comprehensive Digestive Disease Center, University of California, Orange, California, USA
,
Kenneth J. Chang
Division of Gastroenterology and Hepatology, H.H. Chao Comprehensive Digestive Disease Center, University of California, Orange, California, USA
› Author Affiliations
Further Information

Publication History

Publication Date:
23 March 2016 (online)

Placement of long, protruding self-expandable metal stents (SEMSs) into the gastrointestinal lumen may cause related symptoms. A few reports have described the usefulness of argon plasma coagulation (APC) for trimming or fenestrating a SEMS [1] [2] [3] [4]. We report a trimming technique for a covered SEMS in the esophagus using APC in a retrograde fashion, followed by its complete removal.

A 67-year-old woman presented with dysphagia. Esophagogastroduodenoscopy (EGD) showed a large ulcerated tumor in the esophagus with tumor excavation. A 12-cm partially covered SEMS was placed across the tumor. Subsequently the patient was able to resume eating solid food and underwent chemotherapy. However, 1 month after stent placement, she developed epigastric pain and dysphagia from impaction of the stent into the proximal stomach ([Fig. 1 a]). The distal portion of the stent was trimmed with APC using a generator at a setting of 80 W and a flow rate of 2 L/min ([Fig. 1 b]; [Video 1]). The procedure was performed with the scope in a retroflexed position to prevent esophageal mucosal injury. A length of the stent (approximately 4 cm) was completely severed in a circumferential manner and was successfully removed from the stomach ([Fig. 2]). After the procedure, the patient’s pain and dysphagia improved.

Zoom Image
Fig. 1 Endoscopic images showing: a the esophageal covered metal stent with its distal edge protruding into the stomach wall (retrograde view); b the esophageal covered metal stent being trimmed using argon plasma coagulation in retroflexed view.


Quality:
Endoscopic trimming of the esophageal covered metal stent using argon plasma coagulation: the esophageal covered metal stent is seen protruding into the stomach wall; the distal part of the stent is trimmed using argon plasma coagulation in retroflexed view; the transected stent is removed using a snare.

Zoom Image
Fig. 2 The transected portion of the covered metal stent.

After 3 months, however, she developed severe acid reflux and we decided to remove the remainder of the stent. Hyperplastic tissue at the uncovered proximal part of the stent was leveled using a stiff snare and APC to free up some of the mesh from the mucosa. The distal part of the stent was then grabbed with a rat-toothed forceps, and the endoscope was withdrawn in a steady rotational fashion, such that the mesh eventually inverted, was dislodged, and then was successfully removed en bloc ([Fig. 3]; [Video 2]). A subsequent esophagogram demonstrated improvement of the stricture without evidence of contrast extravasation ([Fig. 4]). All of the patient’s stent-related symptoms resolved after these interventions.

Zoom Image
Fig. 3 The remainder of the covered metal stent following its complete removal 3 months later.


Quality:
Endoscopic removal of the esophageal covered metal stent: the distal part of the stent is grabbed with a rat-toothed forceps, and the stent is removed completely using an inversion technique by rotating and withdrawing the endoscope.

Zoom Image
Fig. 4 Esophagogram showing a patent esophagus with no extravasation of contrast.

Endoscopy_UCTN_Code_TTT_1AO_2AZ

 
  • References

  • 1 Rerknimitr R, Naprasert P, Kongkam P et al. Trimming a metallic biliary stent using an argon plasma coagulator. Cardiovasc Intervent Radiol 2007; 30: 534-536
  • 2 Ishii K, Itoi T, Sofuni A et al. Endoscopic removal and trimming of distal self-expandable metallic biliary stents. World J Gastroenterol 2011; 17: 2652-2657
  • 3 Hamada T, Nakai Y, Isayama H et al. Trimming a covered metal stent during hepaticogastrostomy by using argon plasma coagulation. Gastrointest Endosc 2013; 78: 817
  • 4 Tieu AH, Saxena P, Singh VK et al. Fenestration of a covered metal stent during cystoduodenostomy using argon plasma coagulation. Endoscopy 2014; 46 (Suppl. 01) E512-E513