Subscribe to RSS
DOI: 10.1055/s-0045-1805275
Real-world evidence on the use of capnography monitoring during procedural sedation and analgesia for gastroenterology and bronchoscopy
Authors
Aims To measure and compare, in a real-world setting, the frequency of adverse events for patients undergoing procedural sedation and analgesia (PSA) during endoscopic procedures between standard of care monitoring (SOCM) and the additional use of capnography. Capnography measures exhaled carbon dioxide, providing real-time identification of airway obstruction and respiratory depression.
Methods A before-and-after, quality-improvement protocol was implemented at five centres in Leuven, Toronto, Ankara, Madrid, and Cambridge. Adverse events collected were escalation of care, mild (75–90%; up to 60 seconds (s)) and severe (< 75% or<90% for over 60 s) oxygen desaturation, prolonged apnoea (> 60 s), bradycardia, tachycardia, airway obstruction, cardiovascular shock or collapse, cardiac arrest, and death. Data were collected between 2017 and 2021 per procedure using SOCM or SOCM plus capnography (capnography). Patients could have more than one procedure performed. Data collection started without modification to each centre’s standard practice for PSA. SOCM during PSA included electrocardiogram, respiratory rate, blood pressure, and pulse oximetry. Capnography was then introduced, and staff were trained on its use. Following a 2–4-week wash-in period, data collection continued for the capnography cohort.
Results Data were collected for 6,801 procedures (n=3,431 for SOCM and n=3,370 for capnography). Average procedure times were 25 minutes without significant differences between both arms. Procedures were split into colonoscopy (n=2,773), gastroscopy (n=1,868), combined gastroscopy and colonoscopy (n=502), endoscopic ultrasound (n=676), bronchoscopy (n=449), endoscopic retrograde cholangiopancreatography (n=452), and others (n=81). When comparing SOCM versus capnography, the relative risk (RR) of at least one adverse event occurring per procedure was 0.66 (95% confidence interval: 0.56–0.78; 11.25% vs 7.19%). Escalation of care decreased with capnography RR 0.08 (0.01–0.65; 12 vs 1 events). Oxygen desaturation also decreased by RR 0.66 (0.54–0.80; 243 vs 158 events) for mild and RR 0.55 (0.32–0.96; 35 vs 19 events) for severe cases. Bradycardia and tachycardia were reduced using capnography by RR 0.42 (0.24–0.74; 41 vs 17 events) and RR 0.15 (0.07–0.33; 48 vs 7 events), respectively. Prolonged apnoea was increased by RR 4.33 (1.46–12.85; 4 vs 17 events) when using capnography vs SOCM, likely due to an easier detection of apnoea when using capnography. Differences in airway obstruction and cardiovascular events were not significant and no patients died.
Conclusions A significant decrease in adverse events occurring during PSA were observed when using capnography monitoring in addition to SOCM versus SOCM only in this initial analysis of a large, multi-centre, quality-improvement initiative in five countries. Additional analyses are required to limit the effect of confounders on results.
Publication History
Article published online:
27 March 2025
© 2025. European Society of Gastrointestinal Endoscopy. All rights reserved.
Georg Thieme Verlag KG
Oswald-Hesse-Straße 50, 70469 Stuttgart, Germany