Introduction
Capsule endoscopy is the first choice small-bowel examination for most gastroenterologists
worldwide in patients presenting with obscure (overt or occult) gastrointestinal bleeding
(OGIB), because of simple use, low risk, and high sensitivity compared to all other
small-bowel imaging modalities [1]
[2]. Capsule endoscopy is a very sensitive examination for the detection of small-bowel
lesions, as assessed by long-term follow-up of patients after negative capsule endoscopy:
only 10 % of patients present with ongoing bleeding related to a missed small-bowel
lesion [3]. The main limitation of capsule endoscopy is the substantial time needed for capsule
reading by highly qualified medical personnel. Most series [4]
[5]
[6], report a mean reading time of 50 to 60 minutes. In addition, the focus and attention
to detail that are necessary to efficiently read a film are difficult to maintain
during the whole reading [4]
[7]. For this reason, attempts have been made to transfer capsule reading to nurses
or other similarly qualified personnel, with relatively satisfying results [8]
[9]. Informatics approaches to solving the reading time dilemma have also been proposed,
with variable results. This approach needs clinical evaluation and validation [10]
[11].
Nowadays, very few informatics algorithms for accelerated reading have been validated,
although almost all capsule systems propose such informatics tools. This includes
the SBI (small-bowel blood indicator, Given Imaging/Medtronic), that proved to be
of poor sensitivity and specificity [12]
[13], and the Quick-view algorithm that was validated through a retrospective blinded
double-reading trial published in 2012 [14]. Since that study, very few evaluations of informatics reading algorithms have been
published, and no prospective evaluation [15]. Developing such complex algorithms requires specifically large image databases
that have been produced through international collaborations [16]. The Intromedic company recently proposed a new informatics algorithm named “express
view” to reduce the reading time of small-bowel capsule films by skipping repetitive
and non-significant small-bowel images on the film. To the best of our knowledge,
this algorithm has never been evaluated. We thus performed a prospective, multicenter
evaluation of the express view algorithm in unselected films from 10 French gastroenterology
centers, with the aim of evaluating the sensitivity and specificity of an express
view reading compared to the previous standard reading of the capsule film.
Materials and methods
Small-bowel capsule endoscopy and population
The study includes 83 small-bowel capsule films from 83 consecutive unselected patients
included prospectively in one of 10 experienced French centers belonging to the French
Society of Digestive Endoscopy (SFED). The indication for capsule endoscopy was obscure
(overt or occult) bleeding in all cases, with negative bidirectional endoscopy in
all cases. The capsule endoscopy (Mirocam, Intromedic, Seoul, Republic of Korea) procedure
was performed in each center following the usual local procedures including preparation
and prokinetics in some centers: as the main objective of the study was the reading
time, and as each capsule film was read in both standard mode and express view mode,
being its own control, differences in preparation quality or small-bowel transit time
could not modify the results. Characteristics of the Intromedic capsule system are:
a resolution of 320 × 320, a field of view of 170° and a frame rate of 3 images per
second. Each capsule film was read initially by one experienced local reader (standard
or initial reading). Information on patient characteristics, indication for capsule,
findings according to the SFED classification of relevance [17], and reading time, were recorded prospectively. All observed lesions were described
in the reports in terms of nature, location, and relevance. All 83 films were registered
on the same hard disk, and then dispatched to the different centers for express view
analysis.
Express view reading
For each center, one reader was in charge of capsule video analysis in express view
mode, and was blinded to any information about previous diagnosis. This reader was
thus in charge of the analysis of films from one other center. The readers were asked
to read only in express view mode and to select all pathological images, identify
their relevance, and define a final diagnosis corresponding to the most significant
image in the entire (including stomach and colon) capsule film. Readers were allowed
to choose their preferred image reading frame. The results of the express view reading
and of the initial reading were saved on a pro forma.
Interpretation of the results (see flow chart, [Fig.1])
A comparison was made for each case between the results of the initial reading and
that of the express view reading. In cases with concordant (positive or negative)
diagnosis between the standard and express view reading, no complementary reading
was performed. In case of a discordant final diagnosis and/or different lesions observed,
a secondary expert review was performed by three readers (JCS, PJ, GR), all three
having experience of more than 500 capsule readings, and a consensus decision was
made. The result of this consensus review was considered to be the gold standard result
and compared with the results of the standard reading and of the express view reading.
We performed a per patient analysis based on the main diagnosis obtained by capsule
endoscopy for each patient (i. e. small-bowel tumor, angiodysplasia, ulceration, etc.).
We performed a per lesion analysis by including all significant (P1 or P2) lesions
according to the SFED classification [17].
Fig. 1 Study flow chart showing the results of standard reading and express view reading,
then final results after expert consensus review.
Statistical analysis
Sensitivity values were presented as a percentage (95 % confidence interval). McNemar’s
test was used to compare the express view and standard reading sensitivity and specificity
results using the expert review as a gold standard. We used SPSS 12.0 version software
(SPSS Inc, Chicago, IL, United States) in a Windows XP (Microsoft, Seattle, WA, United
States) environment. A P value < 0.05 was considered to be statistically significant.
Results
Between February 2013 and February 2016, 83 patients with obscure/overt digestive
bleeding were included (43 women, mean age 62.5 years, range 21 – 85). The 83 CE films
were read in express view mode without technical issue, with a mean reading time of
18.3 minutes (range 3 – 41 minutes) versus 37.7 minutes (range 11 – 180 minutes, P < 0.0001, Student’s bilateral t test) for standard reading mode. Evaluation of landmarks in normal and express view
reading modes yielded comparable mean gastric and small-bowel transit times: 50.1
and 223.1 minutes (normal mode) and 49.6 and 210.4 minutes (express view mode). The
algorithm selection allowed a mean reduction in the number of images from a mean of
80 596 images (range 123 323 – 52 248) to a mean of 22 087 (5196 – 43 017) thus resulting
in a mean 70 % reduction in the number of images in one complete capsule film.
Per patient analysis
From the 83 patients prospectively included, 45 (54.2 %) had a main diagnosis after
consensus
review of the whole series ([Fig. 1]). The final diagnosis was
angiodysplasia in 29 cases, erosion-ulcerations in 10 cases, ongoing bleeding without
evidenced
diagnosis in four cases and tumor in two cases. Standard reading identified 42 (sensitivity
93.3 %) of these main diagnoses and express view reading identified 37 (82.2 %) of
these
diagnoses ([Table 1]). Standard reading missed one P2 gastric tumor,
one P2 cecal ongoing bleeding, and one P1 gastritis. Express view reading missed four
cases of
small-bowel angiodysplasia (including two cases where the image was skipped by the
express view
algorithm, [Fig. 2]), 1 jejunal ulcer ([Fig. 3] EV1), one P2 duodenal ulcer, and 2 gastric lesions (P1 gastritis and P2 ulcer).
Thus, six out of eight main diagnoses missed at express view reading were present
on the express view film but missed by the readers, and two were missed because of
the informatics selection.
Table 1
Sensitivity of initial reading and express view reading (95 % confidence interval)
in a per patient and per lesion analysis.
|
Normal reading
|
Express view reading
|
P value
|
Per patient sensitivity
|
93.3 % (0.88 – 0.98)
|
82.2 % (0.74 – 0.90)
|
0.388
|
Per lesion sensitivity
|
69.0 % (0.74 – 0.92)
|
77.2 % (0.69 – 0.88)
|
0.430
|
Fig. 2 Two P2 and P1 angiodysplasias skipped by the express view algorithm, thus absent from
the express view film.
Fig. 3 Different small-bowel diagnoses missed by the readers in express view (EV) or standard
reading (SR) mode but present on the corresponding films. EV1: P2 jejunal ulceration;
SR1: Aphthous P1 ulceration in the ileum; SR2: P2 angiodysplasia; EV2: P1 angiodysplasia.
Per lesion analysis
Seventy significant (P1 or P2) images were identified by the readers according to
the expert
review final evaluation. Standard reading (normal mode) identified 58 from these 70
images
(82.9 %), compared to 55 images identified in express view mode (78.6 %, P = 0.388). On
the express view film, 66/70 images were present (sensitivity of the algorithm 94.3 %).
Thus,
at initial reading, readers missed 12 images present on the entire capsule films (17.1 %)
and
express view readers missed 11 images present on the express view-selected capsule
film (16.6 %,
[Fig. 3]). The informatics algorithm skipped four significant images
from the initial film that were all small-bowel angiodysplasia, classified as highly
relevant in
two cases and of uncertain/intermediate relevance in two cases. [Fig. 2] shows two angiodysplasias skipped by the informatics algorithm.
Discussion
There is a high expectation of efficient algorithms allowing reading time reduction
for small-bowel capsule endoscopy films, as the mean reading time proved to be significant,
from 30 to 60 minutes [4]
[17]
[18]. Interestingly, numerous algorithms exist for different capsule endoscopy devices,
but very few prospective blinded evaluations have been performed at the present time.
One retrospective blinded multicentric evaluation validated the quick-view algorithm
of the Given Imaging (now Medtronic) small-bowel capsule endoscopy device [14]. That study showed a 93 % sensitivity of the informatics algorithm for significant
(P1 or P2) lesions in a setting of obscure bleeding. The present study shows that
the express view informatics algorithm developed for the Intromedic device and software
has a 94.2 % sensitivity for the detection of significant lesions. This sensitivity
for significant lesions should be the main objective of any prospective evaluation
of these algorithms, as the major consideration from the patient’s point of view is
that no important diagnostic should be missed (skipped) by the selection algorithm.
Thus, the express view algorithm, like the quick-view algorithm previously, seems
to be a very efficient and safe selection algorithm for clinical use, although confirmative
studies are needed.
The secondary end point of such reading studies is the sensitivity of the reading.
It is important, however, to differentiate this sensitivity from that of the informatics
algorithm, as this is “operator-dependent” and represents the efficacy of the reader
rather than the efficacy of the informatics algorithm or of the device itself. In
the present study, the sensitivity of reading was limited in express view mode, although
not significantly different from that of the standard reading. This represents real
life practice: we included readers from university hospitals, general hospitals, and
private practice. Although all centers had a large experience with capsule reading,
differences of reading interpretation is a characteristic of multicenter studies that
we observed in numerous previous studies [14]
[19]
[20]. On the other hand, image selection by an informatics algorithm may reduce the number
of consecutive images of an angiodysplasia, as an example, on the express view film
compared to the non-contrast film, which occurred in a few cases (not shown) in this
study. However, this did not explain the lower per patient sensitivity of express
mode reading in this study, as, in contrast, the per lesion sensitivity of express
view reading was slightly higher (but not significantly) compared to the per lesion
sensitivity of normal reading. In fact, this may actually represent the real life
variability of multicenter capsule reading and argues in favor of an evolution toward
automatic reading and selection by computer: this is certainly part of the future
of capsule endoscopy. In this setting, the main result of such a study is the theoretical
sensitivity of the informatics algorithm. The use of an expert review, considered
to be the final “gold standard”, may be criticized as it represents some bias of expert
judgment compared to the “real life” reading. However, we showed in previous studies
that the subjectivity of capsule reading is a constant feature of multicenter studies,
with a significant proportion of false positive/false negative images responsible
for confusing results when evaluating the sensitivity of reading, and especially when
evaluating an informatics algorithm for image selection [7]
[14]
[21].
Is the express view algorithm now ready for clinical use? The theoretical sensitivity
of more than 94 % is rather reassuring, although confirmation by an independent study
would be helpful. In this setting, one could use the express view mode to make a first
reading, that could be considered to be sufficient in those patients with significant
findings at the end of this first reading. In those patients with a negative express
view reading, a reading of the “complementary film” could be done, whenever available
on the Intromedic software. On the other hand, simple patient surveillance followed
by a complete reading in those with recurrent bleeding is an alternative. This approach
may yield precious time saving for gastroenterologists.