The first capsule endoscopy system available in the market was named M2A (M2A Capsule,
GivenImaging Ltd, Yokneam, Israel) [1]. This acronym reveals the philosophy behind this technology: the implementation
of a miniaturized system allowing to endoscopically evaluate the entire digestive
system, ideally from the mouth to the anus (which is what M2A stands for).
The technical barriers that had initially prevented the exploration of the entire
surface of the digestive system have been removed over time through the development
of batteries with longer life, the optimization of power consumption (e. g., through
the adaptive frame rate) and the increase in the number of cameras and in the angle
of view (to cover the entire mucosal surface of wider lumen organs). The study by
Vuik et al. [2] in the current issue of Endoscopy International Open, as well as several other publications,
[3] confirm that the endoscopic study of the entire surface of the digestive system
appears to be now technically possible.
However, does “technical feasible” also mean “clinically reliable”? Although the very
quick small bowel transit time reported in the present study (median small bowel transit
time: 47’) might negatively impact on small bowel capsule diagnostic yield, the key
role of capsule endoscopy in the evaluation of the small bowel mucosa has been confirmed
in several studies, and current guidelines endorse the use of this device as first-line
examination in many clinical scenarios [4]. The clinical reliability concerns, therefore, are related mostly to the evaluation
of the colonic, esophageal, and gastric mucosa, which would be the potential added
value of panenteroscopy.
In standard colonoscopy, there is a well-documented correlation between the degree
of cleanliness and diagnostic performance [5]. This is likely to have even greater relevance in the field of wireless endoscopy,
where the capsule is purely passive (e. g., it cannot wash or remove debris) and bowel
preparation also plays a key role in promoting device propulsion. In the paper from
Vuik et al. [2] the rate of patients with adequate colon cleansing (76.6 %) was below the ≥ 90 %
minimum standard recommended for optical colonoscopy [6], although involved subjects were healthy and highly compliant with the preparation
schedule (93.6 % of them took the entire bowel preparation). Although this may reflect
the specific preparation schedule adopted in the study, a recent systematic review
[3] showed that the rate of patients with adequate colon cleansing was highly variable,
but ≥ 90 % in only 54 % of studies, even when different schedules were used. Furthermore,
although no formal standard quality measure exists, the rate of patients with adequate
cleansing of the whole gastrointestinal tract (52.8 %) is far from being acceptable.
Finally, as far as the completion rate for colon evaluation is concerned, the 51 %
rate reported by Vuik et al. [2] is similar to those reported in other publications [7]
[8]
[9] and far below the recommended threshold of 90 % [6].
Although relevant gastric findings, missed by previous traditional gastroscopy, were
reported in patients undergoing small bowel capsule endoscopy [10], the parameter for assessing the efficacy of panenteroscopy in evaluation of gastric
mucosa (observation of more than 90 % of the mucosa) appears to be arbitrary, contrived,
and very difficult to measure. In the absence of a traditional examination as comparator,
no clinically relevant conclusions can be drawn about the diagnostic yield in the
stomach of such a non-steerable capsule. In addition, because the capsule was ingested
without a dedicated protocol for esophageal inspection [11], the Z line was identified in fewer than 50 % of the subjects included in the study,
underscoring that capsule endoscopy, when administered as a panenteroscopy tool, cannot
represent a viable alternative to regular gastroscopy.
Patient acceptability has always been emphasized as one of the major advantages of
capsule endoscopy over more invasive and less tolerated procedures, such as gastroscopy
and colonoscopy. However, Vuik et al. [2] demonstrated that moving from theory (pre-procedural questionnaire) to practice
(post-procedural questionnaire), patient expectations are somewhat disregarded. In
line with these results, in the TOPAZ study [12] more subjects who underwent both procedures under evaluation unexpectedly preferred
traditional colonoscopy over colon capsule endoscopy. Once again, the major determinant
behind this preference was the burden related to colon capsule endoscopy bowel preparation.
Taking into account all these issues, one might wonder if panenteric capsule endoscopy
evaluation is a journey worth undertaking and a goal worth pursuing. Taken individually,
each traditional endoscopic examination for evaluation of esophagus, stomach and colon
seems to be more reliable than panenteroscopy. However, the latter offers the unique
opportunity to combine evaluation of several segments of the digestive tract at the
same time, in a single examination. This becomes particularly relevant when the “timing
of the procedure” or “disease extent” are key factors. Indeed, there is growing evidence
supporting the use of a panenteroscopic approach in patients with digestive bleeding,
in whom an upper digestive tract lesion has been excluded [13]
[14]. Initial data suggest that in these patients, a timely panenteric evaluation has
relevant clinical (increased rate of diagnosis, crucial information for therapeutic
work-up) as well as organizational/economic consequences (reduction of the number
of unnecessary examinations and hospitalization time). Similarly, for staging and
surveillance in patients with Crohn's disease, especially in the pediatric setting,
panenteroscopy appears to be a reliable tool for assessing disease mucosal activity
and extent, compared to more invasive methods [15]
[16]
[17].
In light of all the pros and cons, with current technology and organizational/practical
issues (e. g. the preparation schedule), the window of opportunity for panenteroscopy
in everyday clinical practice currently is narrow. However, the most limiting factors
appear amenable to significant improvements, and the introduction of automated image
analysis systems (e. g. artificial intelligence-based software) [18], which could make the procedure even easier and more reliable, is expected soon.
These improvements are needed before implementing use of the capsule for whole-intestine
mucosal assessment in clinical practice.