Traction-assistance is very effective in facilitating endoscopic submucosal dissection
(ESD) of superficial colorectal lesions [1], notably in the case of difficult locations such as the ileocecal valve. We have
recently developed a multitraction strategy to increase the efficacy of traction for
lesions in such locations, using a rubber band to connect the different traction points
[2]. A new magnetic traction device (ProdiGI Traction Magnet, Medtronic) has been developed
that allows adjustment of the traction strategy during the procedure ([Fig. 1 a, b]; [Video 1]).
Fig. 1 a Schematic representation of a magnetic traction system used to facilitate endoscopic
submucosal dissection (ESD). b Appearance of the magnetic traction system after extraction from the patient's colon
at the end of the ESD procedure.
Video 1 Endoscopic submucosal dissection (ESD) of a laterally spreading tumor at the ileocecal
valve, using an adaptable magnetic traction system.
We report here the case of a 64-year-old patient with a 2-cm nongranular laterally
spreading tumor of the cecum, extending to the lower lip of the ileocecal valve. In
order to make the incision with satisfactory margins, the first step of the ESD procedure
consisted in exposing the inner part of the valve using a first clip as previously
described [3], by connecting the magnet attached to the first clip to the magnet of a second clip
placed on the opposite colonic wall ([Fig. 2]). Once the incision had been made, we had to expose the submucosa for the dissection
itself, with a third clip placed on the edge of the lesion, that was magnetically
connected to the clip on the opposite colonic wall. At this point, the first clip
used to pull the inner part of the valve to allow incision on the ileal side had become
troublesome because the traction angle was not optimal for submucosal exposure. Therefore,
we separated the magnet of the first clip from the two others using the cap of the
endoscope, in order to obtain optimal traction for the dissection itself ([Fig. 3]). This technique allowed a curative R0 resection of the lesion ([Fig. 4]).
Fig. 2 Using the magnetic traction system to obtain a satisfactory eversion of the ileocecal
valve before ESD of the laterally spreading tumor. The first magnetic clip has been
placed in the inner part of the valve and the second on the opposite colonic wall.
The attraction between the magnets is through the mucosa in this instance.
Fig. 3 Obtaining excellent submucosal exposure at the lateral edge of the lesion.
Fig. 4 Lesion measuring 25 × 35 mm, stretched on cork. The lesion was removed with satisfactory
margins. It corresponded to a high grade dysplasia adenoma.
To our knowledge, this is the first case in humans of the use of a fully internalized
magnetic traction system that can pass through the operating channel of the endoscope
[4]. This technique seems attractive in difficult locations where a change in traction
configuration is necessary during the procedure. Further studies are needed to confirm
its effectiveness.
Endoscopy_UCTN_Code_TTT_1AQ_2AD
Endoscopy E-Videos is an open access online section, reporting on interesting cases and new techniques
in gastroenterological endoscopy. All papers include a high quality video and all
contributions are freely accessible online. Processing charges apply (currently EUR
375), discounts and wavers acc. to HINARI are available.
This section has its own submission website at https://mc.manuscriptcentral.com/e-videos