Keywords
Endoscopy Upper GI Tract - Hygiene - Quality and logistical aspects - Performance
and complications
Correction: Performance of a single-use gastroscope for esophagogastroduodenoscopy:
Prospective evaluation
Koen van der Ploeg, Pieter J.F. de Jonge, Wim J. Lammers et al.
Endoscopy International Open 2024; 12: E428–E434. DOI: 10.1055/a-2271-2303
In the above-mentioned article Figure 1 was corrected. This was corrected in the online
version on 13.09.2024.
Introduction
Esophagogastroduodenoscopy (EGD) is an important tool in diagnosis and treatment of
upper gastrointestinal-disorders. The indications for EGD are diverse, comprising
both diagnostic and interventional purposes such as dysphagia, gastroesophageal reflux
disease, nasojejunal tube placement, dilatation of esophageal strictures, and treatment
of upper gastrointestinal bleeding [1]. EGD is a common procedure and is performed around 6.1 million times annually in
the United States [2]. Complications of EGD are rare, but include perforation, bleeding, aspiration, and
infection [3].
Endoscopy-associated infection (EAI) can be either endogenous, due to translocation
of the patient’s own microbial gut flora, or exogenous as a result of contaminated
equipment. The latter has received increased attention because of several outbreaks
reported in past decades [4]. Endoscopes require an extensive cleaning process in which reprocessing protocol
can be breached and endoscope damage or biofilm can prevent proper disinfection [5]
[6]. Endoscopes can only be sterilized with ethylene oxide (EtO), which has been shown
to end outbreaks but also carries carcinogenic and mutagenic risks [7]. Therefore, sterilization with EtO demands specific environmental and construction
prerequisites, making it time-consuming and limiting its applicability [8]. Also, in a randomized controlled trial, the addition of EtO sterilization to high-level
disinfection did not improve contamination rates [9]. Many of the published outbreaks involved contaminated duodenoscopes [4]. However, outbreaks due to contaminated gastroscopes have been described, resulting
in seven reported clinical infections [10]
[11]
[12]. The prevalence of contaminated patient-ready gastroscopes has been less frequently
studied compared with that in duodenoscopes. A recent meta-analysis reported a contamination
rate of 28.2%, based on six different studies [13].
The precise risk of EAIs in gastroscopes, however, is unknown. It is likely that EAIs
are underreported due to a lack of recognition of EAIs caused by susceptible microorganisms,
limited microbiological surveillance conducted after endoscopy procedures, and the
potential long duration between endoscopy and infection onset. Another explanation
for failure to report recognized EAIs could be the lack of a mandatory registration
system reduces the likelihood that an EAI will be reported once recognized.
In response to reported cases of EAI, the development of single-use endoscopes has
gained momentum. Utilization of single-use endoscopes eliminates risk of exogenous
EAIs. Their use also provides additional advantages, such as streamlining workflow
by eliminating the time required for endoscope reprocessing and enabling endoscopic
procedures outside regular working hours without the need for reprocessing personnel.
In addition, single-use endoscopes are associated with faster and enhanced product
refinement and development through successive iterations, along with the possibility
of tailoring devices for specific tasks. Furthermore, single-use endoscopes and their
processors are lighter and smaller, making them more portable. This streamlines endoscopic
procedures in settings beyond the Endoscopy Department, including the Intensive Care
Unit (ICU) and remote locations, eliminating the need for post-procedure endoscope
cleaning facilities. However, it is important to consider the potential benefits of
single-use endoscopes in the context of increased costs and their environmental impact
[14].
Multiple single-use duodenoscopes have already been introduced to the market, and
their performance seems to be comparable to that of reusable duodenoscopes [15]
[16]. Ambu is the first company that has produced a single-use gastroscope (Ambu aScope
Gastro). The aScope Gastro recently received European Certification (CE), but no studies
of its performance for a broad range of EGD indications have been published. The aim
of this study was to assess the performance of a single-use gastroscope in adult patients
undergoing EGD.
Patients and methods
Study design
This study employed a multicenter, prospective, observational, case series design
to assess performance of the CE-certified single-use gastroscope, the Ambu aScope
Gastro, in accordance with its intended use. The participating centers included the
Erasmus MC University Medical Center Rotterdam (Erasmus MC) and Oslo University Hospital
– Rikshospitalet (OUS).
Patient selection and informed consent
A total of 60 patients were included in this study, with 34 patients recruited from
the Erasmus MC and 26 patients from the OUS. At the Erasmus MC, all eligible patients
underwent EGD in the Endoscopy Department. At the OUS, only patients admitted to the
ICU were considered eligible. Eligible patients were individuals aged ≥ 18 years who
were scheduled for an EGD at the specified locations. Patients who were terminally
ill or those who had participated in other studies that could interfere with the outcomes
of this study did not qualify for participation. Patients scheduled for an EGD for
Barrett's esophagus surveillance were omitted per guideline requiring a high-definition
endoscope [17]. The study was approved by the Medical Research Ethics Committee of Erasmus MC (MEC-2022–0285).
All Dutch patients provided informed consent. In accordance with Norwegian law, the
study activities conducted at OUS were regarded as a quality assurance exercise and,
therefore, informed consent was not required.
Device description
The Ambu aScope Gastro is a sterile single-use gastroscope with a working length of
1030 mm and a 2.8-mm working channel that is compatible with commonly used endotherapy
instruments. The gastroscope is used in conjunction with the Ambu aBox 2 (aBox 2)
display and processing unit, which includes a touchscreen monitor. The aBox 2 features
a Full HD 12.8-inch color liquid-crystal display, with a total height of 27.8 cm width
of 33 cm, and weight of 8 kg.
Setting
At Erasmus MC, all EGDs were conducted in the same room in the Endoscopy Department.
Each procedure involved one endoscopist and two endoscopy nurses. The aBox 2 was linked
to a video distribution device, enabling transmission of the video signal to two extra
screens, typically employed during EGDs using reusable gastroscopes. Patients were
positioned on their left side and given the option of sedation (midazolam and fentanyl)
for the procedure. EGDs at the OUS were performed at in patient rooms in the ICU.
One endoscopist, one endoscopy nurse, and one ICU nurse were present. Only the screen
of the aBox 2 was used, which was positioned across the bed. Patients were mainly
treated in supine position with their heads tilted toward the left side; occasionally,
patients were placed on their left side. Most ICU patients were already deeply sedated;
in some cases, additional sedation was given for the procedure.
Procedure and evaluation
All EGDs were conducted in accordance with their respective indications, without any
additional study interventions except for using single-use gastroscopes. The procedures
were exclusively conducted by certified endoscopists, two at Erasmus MC and five at
OUS. Each endoscopist had performed more than 1000 EGDs during their career and possessed
expertise in advanced endoscopy techniques such as endoscopic retrograde cholangiopancreatography,
endoscopic ultrasound, or endoscopic mucosal resection. None of the endoscopists had
previous experience with the Ambu aScope Gastro. One endoscopist from Erasmus MC and
two from OUS had experience with single-use duodenoscopes (Ambu or Boston Scientific).
Furthermore, all endoscopists had prior experience with the SpyGlass (Boston Scientific),
a single-use cholangioscope. After each procedure, the performing endoscopist completed
a questionnaire to assess various aspects of single-use gastroscope performance compared
with prior experience with a reusable gastroscope. The endoscopist made the decision
if the aspect could be rated based on the EGD that had been performed. If the single-use
gastroscope could not fulfill the intended purpose of the EGD, the endoscopist switched
to a reusable endoscope to complete the procedure. The Erasmus MC used reusable gastroscopes
from either the Olympus 180 or the Olympus 190 series, while the OUS utilized gastroscopes
exclusively from the Olympus 190 series.
Study endpoints
The primary endpoint of the study was frequency with which an endoscopist successfully
achieved the intended diagnostic or therapeutic goals during an EGD. Diagnostic goals
included inspecting the upper digestive tract for esophageal dysphagia, ulcer follow-up,
or malignancy screening. Therapeutic goals encompassed varices treatment, nutritional
tube placement, and dilation of esophageal stenosis. Secondary endpoints encompassed
qualitative assessment of the single-use gastroscope in comparison with prior experiences
with reusable devices, utilizing a comprehensive 5-point Likert scale (ranging from
1 – “much worse” to 5 – “much better”). This assessment covered various dimensions
including image quality, handling aspects, and technical performance parameters such
as suction efficiency and passage of accessories through the working channel. Diagnostics
and therapeutics that were performed during the EGD were also rated. In addition,
procedure duration was monitored and instances of transitioning to reusable endoscopes
were recorded.
Statistical analyses
All statistical analyses were performed using R version 4.1.3 [18]. Categorical variables are presented as absolute and relative frequencies, whereas
continuous variables are reported as median with first quartile and third quartile
(Q1, Q3) or mean and standard deviation (SD). Due to the study design, differences
in populations between the study centers, and the ordinal nature of the data, no statistical
tests were conducted.
Results
Sixty patients were included in this study, of whom 34 were from the Endoscopy Department
at Erasmus MC and 26 from the ICU at OUS. Among them, 38 (63.3%) were male and the
median age was 61.5 years (range 53.5–73.3). Notably, patients recruited from OUS
were predominantly male (20 of 26; 76.9%) and younger (median age 57.0 years (range
42.3–68.5)) in comparison with those from Erasmus MC, where 18 of 34 (52.9%) were
male and the median age was 66.0 years (range 57.5–75.0) ([Table 1]). Furthermore, the patient cohort at Erasmus MC exhibited a higher prevalence of
relevant medical histories potentially influencing the EGD procedure. Specifically,
a greater proportion of these patients had a history of upper gastrointestinal surgery
(4 of 34; 11.8% vs. 1 of 26; 3.8%), esophageal stenosis (8 of 34; 23.5% vs. 0 of 26;
0%), and portal hypertension (4 of 34; 11.8% vs. 2 of 26; 7.7%). History of upper
gastrointestinal surgery included Roux-Y-gastrectomy (OUS, 1 of 26, 3.8%), esophagectomy with
gastric conduit reconstruction (Erasmus MC 3 of 34, 8.8%) and gastric bypass (Erasmus
MC, 1 of 34, 2.9%).
Table 1 Patient characteristics.
|
Total
|
Erasmus MC
|
Oslo University Hospital
|
Q, quartile.
|
Patients, n
|
60
|
34
|
26
|
Patient's age, years, median (Q1,Q3)
|
61.50 (53.50,73.25)
|
66.00 (57.5,75.0)
|
57.00 (42.3,68.5)
|
Male gender, n (%)
|
38 (63.3)
|
18 (52.9)
|
20 (76.9)
|
Medical history, n (%)
|
|
5 (8.3)
|
4 (11.8)
|
1 (3.8)
|
|
2 (3.3)
|
1 (2.9)
|
1 (3.8)
|
|
8 (13.3)
|
8 (23.5)
|
0 (0)
|
|
6 (10.0)
|
4 (11.8)
|
2 (7.7)
|
Sedation, n (%)
|
|
18 (30.0)
|
0 (0.0)
|
18 (69.2)
|
|
2 (3.3)
|
0 (0.0)
|
2 (7.7)
|
|
25 (41.7)
|
23 (67.6)
|
2 (7.7)
|
|
7 (11.7)
|
4 (11.8)
|
3 (11.5)
|
|
8 (13.3)
|
7 (20.6)
|
1 (3.8)
|
Gloucester Comfort Scale, n (%)
|
|
3 (5.0)
|
3 ( 8.8)
|
0 (0.0)
|
|
3 (5.0)
|
3 ( 8.8)
|
0 (0.0)
|
|
24 (40.0)
|
20 (58.8)
|
4 (15.4)
|
|
3 (5.0)
|
2 ( 5.9)
|
1 (3.8)
|
|
9 (15.0)
|
6 (17.6)
|
3 (11.5)
|
|
18 (30.0)
|
0 ( 0.0)
|
18 (69.2)
|
Performance
Seven endoscopists conducted the procedures; two at the Erasmus MC and five at the
OUS. In 58 of 60 cases (96.7%), the endoscopists achieved the intended diagnostic
or therapeutic goals, completing the EGD without the need for a crossover. In two
cases, the intended goals could not be achieved with a single-use gastroscope alone
due to its diameter. Consequently, a crossover to an ultra-slim endoscope was necessary
to reach or traverse the esophageal stenosis. Half of the EGD indications (30 cases)
were primarily therapeutic, with a higher proportion observed at OUS (18 of 26 cases,
69.2%) compared with the Erasmus MC (12 of 34 cases, 35.3%). EGD characteristics and
indications are listed in [Table 2]. After every EGD, the performance of the single-use gastroscope was rated. The results
of the qualitative assessment are illustrated in [Fig. 1].
Table 2 Esophagogastroduodenoscopy procedure characteristics and primary indication.
|
Total
|
Erasmus MC (n = 34)
|
Oslo University Hospital (n = 26)
|
EGD, esophagogastroduodenoscopy; LAMS, lumen-apposing metal stent; Q, quartile; SD,
standard deviation.
|
EGD location
|
|
Endoscopy Department
|
Intensive Care Unit
|
Number of endoscopists, n
|
7
|
2
|
5
|
Number of procedures per endoscopist (mean [SD])
|
8.6 [6.6]
|
17 [0]
|
5.2 [4.5]
|
Minutes per procedure, n (median [Q1,Q3])
|
10.00 [6.00,5.25]
|
7.00 [5.0,9.8]
|
15.50 [15.0,24.8]
|
Primary therapeutic indication EGD, n (%)
|
30 (50.0)
|
12 (35.3)
|
18 (69.2)
|
EGD indication, n (%)
|
|
10 (16.7)
|
0 (0.0)
|
10 (38.5)
|
|
11 (18.3)
|
5 (14.7)
|
6 (23.1)
|
|
9 (15.0)
|
9 (26.5)
|
0 (0.0)
|
|
8 (13.3)
|
5 (14.7)
|
3 (11.5)
|
|
6 (10.0)
|
5 (14.7)
|
1 (3.8)
|
|
3 (5.0)
|
3 (8.8)
|
0 (0.0)
|
|
2 (3.3)
|
2 (5.9)
|
0 (0.0)
|
|
2 (3.3)
|
1 (2.9)
|
1 (3.8)
|
|
2 (3.3)
|
2 (5.9)
|
0 (0.0)
|
|
7 (11.7)
|
2 (5.9)
|
5 (19.2)
|
Completed procedure with single-use gastroscope, n (%)
|
58 (96.7)
|
32 (94.1)
|
26 (100.0)
|
Failure to complete procedure intended EGD goal, n (%)
|
2 (3.3)
|
2 (5.9)
|
0 (0.0)
|
Fig. 1 Performance rating of single-use gastroscope according to a five-point Likert scale.
WC, working channel endoscope
In the Erasmus MC, 18 of 20 performance characteristics were rated as “comparable”
or “better” in at least 85% of EGDs. The weight of the single-use gastroscopes was
rated as better than the reusable scopes in 50% of cases. Duodenal intubation proved
more challenging in 35% of EGDs, whereas picture quality was deemed inferior in 47%
of cases. [Fig. 2] presents a comparative display of images captured using an Ambu aScope Gastro and
an Olympus 180 series gastroscope. At the OUS, 11 of 20 performance characteristics
were rated as “comparable” or “better” in 85% of EGDs. Notably, duodenal intubation
was rated as better than reusable gastroscopes in 33% of cases, and weight was scored
as better than reusable in all instances. Furthermore, fit to hand, maneuverability,
and distal tip deflection were considered as better than reusable scopes in over 50%
of cases. However, image size was rated as inferior in 100% of EGDs. Also, other characteristics
related to image quality, such as camera focus, lighting, color reproduction, and
overall picture and video quality, were rated as inferior compared with reusable scopes
in at least 35% of procedures. In the general remarks left by the endoscopists from
the Erasmus MC, common points of feedback were the absence of a freeze button (16
cases), lens cleaning problems (8 cases), and extensive loop formation in the stomach
(8 cases). The endoscopists from OUS reported that the screen size was too small (6
cases). Overall satisfaction was rated as “comparable” in 87% of the EGDs at Erasmus
MC and 96% at OUS. Importantly, there were no device failures or adverse events.
Fig. 2 Comparative images from a single patient, taken during two consecutive esophagogastroduodenoscopies
performed for the same indication: analysis of lymphadenopathies with progressive
clinical deterioration. The images on the left (a, c, e) were obtained with an Ambu aScope Gastro single-use gastroscope, whereas the images
on the right (b, d, f) were acquired using an Olympus 180 series gastroscope.
Discussion
In this series of 60 patients, upper gastrointestinal endoscopy with a single-use
gastroscope could be accomplished in a satisfactory percentage of almost 97% of cases.
There was no limitation in the execution of standard diagnostic and therapeutic procedures
while using a single-use endoscope except for two cases with a non-traversable esophageal
stricture. It is important to note that these two crossovers were necessitated by
the single-use gastroscope diameter, which matches that of reusable gastroscopes.
Therefore, these two cases do not indicate a performance failure. Overall, handling
of the single-use gastroscopes was rated at least comparable to reusable gastroscopes.
Performance characteristics related to video and picture quality, however, were rated
lower.
This study shows that single-use gastroscopes could be used for a myriad of standard
indications, offering a proper alternative to reusable gastroscopes. This is supported
by a recent publication, which demonstrated successful treatment of six cases of upper
gastrointestinal bleeding using the Ambu single-use gastroscope [19]. The use of single-use endoscopes for patient treatment is a topic of ongoing debate.
While they offer significant advantages, such as eliminating the need for reprocessing
and reducing gastroscope-associated infections, concerns have emerged regarding their
environmental impact. In an era in which climate change necessitates environmental
responsibility, it should also guide decisions in health care in general and, for
this subject in particular, the Endoscopy Department [14]. Furthermore, cost implications of implementing single-use gastroscopes require
further exploration. The current study solely focused on performance of single-use
gastroscopes and did not include environmental or cost aspects of implementing these
endoscopes in daily practice.
Single-use gastroscopes have certain areas for improvement. First, their image quality
was impacted by lens cleaning difficulties, and the absence of a freeze button impeded
documentation. Furthermore, lack of a narrow band imaging function impairs lesion
characterization and assessment of intestinal metaplasia [20]. However, the inherent benefit of single-use endoscopes lies in their adaptability,
allowing fast incorporation of design and functional improvements in newer versions.
This adaptability enables early availability of new functions in clinical practice
and for research purposes. Hospitals and clinics are not bound by the long-term investment
in reusable endoscopes. Instead, they can promptly adopt newer versions of single-use
gastroscopes when they become available.
There were several differences between the patient populations and procedure characteristics
at Erasmus MC and OUS. In OUS, patients were typically younger, predominantly male,
and more frequently under general anesthesia, contributing to their increased comfort
during the procedure. Regarding procedure characteristics, a larger number of endoscopists
performed the EGDs, procedure times were longer, and the primary indication was often
therapeutic. Erasmus MC also employed multiple screens during EGDs, as opposed to
just the aBox 2 screen. These distinctions can largely be attributed to the distinct
settings: ICU versus the standard Endoscopy Department. For instance, epidemiological
studies have indicated a higher ICU admission rate for men [21]. In addition, Erasmus MC required informed consent for participation, potentially
introducing inclusion bias. These numerous variations may have influenced evaluation
of single-use gastroscopes. Lower weight, for instance, potentially proved more advantageous
during longer procedures and use of multiple (larger) screens might explain the difference
in image quality rating between the study centers.
While the results of this case series are promising, this study has some limitations.
No control group was available and the performance rating was predominantly based
on subjective factors, which was underscored by the difference in ratings between
the study centers. Although we included patients from different departments and with
a broad range of indications, we cannot claim generalizability to the total population
of patients undergoing EGD. In addition, all EGDs were conducted by highly experienced
endoscopists. Consequently, our findings cannot be readily generalized to novice or
trainee endoscopists. Series from other groups are needed to confirm the effectiveness
and safety of single-use gastroscopes compared with reusable gastroscopes in clinical
practice.
Conclusions
This series shows that single-use gastroscopes can be used successfully for a broad
range of indications. Potential benefits are prevention of endoscope-associated infections,
absence of reprocessing time, and improved workflow when performing endoscopy in remote
locations. These benefits must be weighed against costs and environmental impact.
To make informed decisions regarding implementation of single-use gastroscopes in
endoscopy practice, a better understanding of their environmental impact in comparison
with reusable endoscopy is needed.