Continuous low water pressure dissection technique minimizing air bubbles during underwater endoscopic submucosal dissection
Underwater endoscopic submucosal dissection (UESD) in saline offers several advantages [1] [2]. However, air bubbles generated during UESD can impair the visual field, particularly when they accumulate in the hood. To address this issue, outlets can be introduced on the side of the hood to facilitate air bubble removal [3] [4]. Nevertheless, air bubbles accumulate in the hood in cases where the lesion is located opposite to gravity, the tissue obstructs the air bubble outlet, or large air bubbles form. This repeatedly necessitates the use of the waterjet function of the endoscope to flush out the air bubbles. Under saline, air bubbles are generated by the evaporation of saline caused by the energization of the electrosurgical knife, which creates a “vapor pocket” that enables the knife to perform dissection [5]. Continuous high water pressure can make dissection difficult by hindering the formation of a sufficient vapor pocket, whereas continuous low water pressure helps maintain dissection power while minimizing and removing air bubbles. Therefore, we propose the continuous low water pressure dissection technique (CDT), which helps maintain the visual field during UESD ([Fig. 1]).


[Video 1] demonstrates UESD using a hood (ST hood short type; DH-29CR; Fujifilm, Tokyo, Japan) with self-made air bubble outlets for a colorectal tumor. As half of the lesion was located against gravity, air bubbles accumulated in the hood. The flushing pump setting (OFP-2; Olympus, Tokyo, Japan) was adjusted from strong to weak after creation of the mucosal flap, and CDT was initiated. CDT removed air bubbles that entered the hood without the need to turn the waterjet on and off, thereby expediting the procedure. Without changing the output of the Swift Coagulation (Effect 3, 30 W) in the electrosurgical generator (VIO300D; Erbe, Tübingen Germany), dissection power was maintained ([Fig. 2]).
Underwater endoscopic submucosal dissection using a continuous low water pressure dissection technique and a hood with four self-made air bubble outlets for resection of a laterally spreading tumor in the cecum.Video 1

In conclusion, CDT helps maintain the visual field during UESD, enhancing the usefulness of a hood with air bubble outlets.
Endoscopy_UCTN_Code_TTT_1AQ_2AD_3AD
E-Videos is an open access online section of the journal Endoscopy, reporting on interesting cases and new techniques in gastroenterological endoscopy.
All papers include a high-quality video and are published with a Creative Commons
CC-BY license. Endoscopy E-Videos qualify for HINARI discounts and waivers and eligibility is automatically checked during the submission
process. We grant 100% waivers to articles whose corresponding authors are based in
Group A countries and 50% waivers to those who are based in Group B countries as classified
by Research4Life (see: https://www.research4life.org/access/eligibility/).
This section has its own submission website at https://mc.manuscriptcentral.com/e-videos.
#
Conflict of Interest
The authors declare that they have no conflict of interest.
-
References
- 1 Nagata M. Usefulness of underwater endoscopic submucosal dissection in saline solution with a monopolar knife for colorectal tumors (with videos). Gastrointest Endosc 2018; 87: 1345-1353
- 2 Nagata M. Underwater endoscopic submucosal dissection in saline solution using a bent-type knife for duodenal tumor. VideoGIE 2018; 3: 375-377
- 3 Nagata M. Tapered hood with wide holes in its sides for efficient air bubble removal during underwater endoscopic submucosal dissection. Dig Endosc 2022; 34: 654
- 4 Nagata M. Underwater endoscopic submucosal dissection using a tapered hood with air bubble outlets for a subcircumferential duodenal tumor. Dig Endosc 2024; 36: 225-227
- 5 Botto H, Lebret T, Barré P. et al. Electrovaporization of the prostate with the Gyrus device. J Endourol 2001; 15: 313-316
Correspondence
Publication History
Article published online:
07 August 2024
© 2024. The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution License, permitting unrestricted use, distribution, and reproduction so long as the original work is properly cited. (https://creativecommons.org/licenses/by/4.0/).
Georg Thieme Verlag KG
Rüdigerstraße 14, 70469 Stuttgart, Germany
-
References
- 1 Nagata M. Usefulness of underwater endoscopic submucosal dissection in saline solution with a monopolar knife for colorectal tumors (with videos). Gastrointest Endosc 2018; 87: 1345-1353
- 2 Nagata M. Underwater endoscopic submucosal dissection in saline solution using a bent-type knife for duodenal tumor. VideoGIE 2018; 3: 375-377
- 3 Nagata M. Tapered hood with wide holes in its sides for efficient air bubble removal during underwater endoscopic submucosal dissection. Dig Endosc 2022; 34: 654
- 4 Nagata M. Underwater endoscopic submucosal dissection using a tapered hood with air bubble outlets for a subcircumferential duodenal tumor. Dig Endosc 2024; 36: 225-227
- 5 Botto H, Lebret T, Barré P. et al. Electrovaporization of the prostate with the Gyrus device. J Endourol 2001; 15: 313-316