Interference Screw Fixation Using Bioabsorbable Screw as Void Filler
Publication History
Publication Date:
05 October 2011 (online)

ABSTRACT
We evaluated interference screw fixation in a plug-tunnel construct using bioabsorbable screws as void fillers with different percentages of the screw removed. Nine-millimeter tunnels in a closed-cell foam block were filled with a 10-mm bioabsorbable screw, and 10-mm revision tunnels were placed in parallel with tunnel overlap resulting in removal of 10%, 25%, or 50% of the screw diameter. Synthetic bone plugs were fashioned to fit 10-mm tunnels. In all groups, the plugs were secured in standard interference fixation with a 9-mm metal screw between the void-filling bioabsorbable screw and plug. Failure loads for the control group (no revision tunnel) averaged 926 ± 44 N, 10% (1024 ± 129 N) and 25% (932 ± 129 N) groups were not significantly different; failure load in the 50% diameter group (780 ± 72 N) was significantly lower than all other groups (p < 0.001). Using a bioabsorbable screw as void filler provided mean load to failure not different from that of standard reconstruction when 10 and 25% of the diameter of the void-filling screw was removed. Load to failure was significantly lower when 50% of the void-filling screw diameter group was removed. This may be applicable in anterior cruciate ligament reconstruction where a previous tunnel void has to be addressed.
KEYWORDS
Biomechanical - tunnel void - synthetic bone - anterior cruciate ligament reconstruction
REFERENCES
- 1 Brown Jr C H, Carson E W. Revision anterior cruciate ligament surgery. Clin Sports Med. 1999; 18 (1) 109-171
- 2 Johnson D L, Coen M J. Revision ACL surgery. Etiology, indications, techniques, and results. Am J Knee Surg. 1995; 8 (4) 155-167
- 3 Battaglia T C, Miller M D. Management of bony deficiency in revision anterior cruciate ligament reconstruction using allograft bone dowels: surgical technique. Arthroscopy. 2005; 21 (6) 767
- 4 Greis P E, Johnson D L, Fu F H. Revision anterior cruciate ligament surgery: causes of graft failure and technical considerations of revision surgery. Clin Sports Med. 1993; 12 (4) 839-852
- 5 O'Neill D B. Arthroscopically assisted reconstruction of the anterior cruciate ligament. A follow-up report. J Bone Joint Surg Am. 2001; 83-A (9) 1329-1332
- 6 Azar F M. Revision anterior cruciate ligament reconstruction. Instr Course Lect. 2002; 51 335-342
- 7 Wolf R S, Lemak L J. Revision anterior cruciate ligament reconstruction surgery. J South Orthop Assoc. 2002; 11 (1) 25-32
- 8 Noyes F R, Barber-Westin S D. Revision anterior cruciate ligament reconstruction: report of 11-year experience and results in 114 consecutive patients. Instr Course Lect. 2001; 50 451-461
- 9 Herbenick M, Gambardella R. Revision anterior cruciate ligament reconstruction using a unique bioabsorbable interference screw for malpositioned tunnels. Am J Orthop. 2008; 37 (8) 425-428
- 10 Allen C R, Giffin J R, Harner C D. Revision anterior cruciate ligament reconstruction. Orthop Clin North Am. 2003; 34 (1) 79-98
- 11 Getelman M H, Friedman M J. Revision anterior cruciate ligament reconstruction surgery. J Am Acad Orthop Surg. 1999; 7 (3) 189-198
- 12 Jaureguito J W, Paulos L E. Why grafts fail. Clin Orthop Relat Res. 1996; 325 (325) 25-41
- 13 Kaplan K S, Hayes W C, Keaveny T M et al.. Form and function of bone. In: Simon S R, ed. Orthopaedic Basic Science. Rosemont, IL: American Academy of Orthopaedic Surgeons; 1994: 127-184
- 14 Firoozbakhsh K K, Moneim M S, DeCoster T A. Pullout strength of power- and hand-driven staples in synthetic bone: effect of design parameters. J Orthop Trauma. 1992; 6 (1) 43-49
- 15 McNamara B P, Cristofolini L, Toni A, Taylor D. Evaluation of experimental and finite element models of synthetic and cadaveric femora for pre-clinical design-analysis. Clin Mater. 1994; 17 (3) 131-140
- 16 Lemos M J, Jackson D W, Lee T Q, Simon T M. Assessment of initial fixation of endoscopic interference femoral screws with divergent and parallel placement. Arthroscopy. 1995; 11 (1) 37-41
- 17 Kousa P, Järvinen T L, Kannus P, Järvinen M. Initial fixation strength of bioabsorbable and titanium interference screws in anterior cruciate ligament reconstruction. Biomechanical evaluation by single cycle and cyclic loading. Am J Sports Med. 2001; 29 (4) 420-425
- 18 Harvey A, Thomas N P, Amis A A. Fixation of the graft in reconstruction of the anterior cruciate ligament. J Bone Joint Surg Br. 2005; 87 (5) 593-603
- 19 Rowden N J, Sher D, Rogers G J, Schindhelm K. Anterior cruciate ligament graft fixation. Initial comparison of patellar tendon and semitendinosus autografts in young fresh cadavers. Am J Sports Med. 1997; 25 (4) 472-478
- 20 Kurosaka M, Yoshiya S, Andrish J T. A biomechanical comparison of different surgical techniques of graft fixation in anterior cruciate ligament reconstruction. Am J Sports Med. 1987; 15 (3) 225-229
- 21 Matthews L S, Lawrence S J, Yahiro M A, Sinclair M R. Fixation strengths of patellar tendon-bone grafts. Arthroscopy. 1993; 9 (1) 76-81
- 22 Seil R, Rupp S, Krauss P W, Benz A, Kohn D M. Comparison of initial fixation strength between biodegradable and metallic interference screws and a press-fit fixation technique in a porcine model. Am J Sports Med. 1998; 26 (6) 815-819
- 23 Woo S L, Gomez M A, Woo Y K, Akeson W H. Mechanical properties of tendons and ligaments. II. The relationships of immobilization and exercise on tissue remodeling. Biorheology. 1982; 19 (3) 397-408
- 24 Pavlik A, Hidas P, Czigány T, Berkes I. Biomechanical evaluation of press-fit femoral fixation technique in ACL reconstruction. Knee Surg Sports Traumatol Arthrosc. 2004; 12 (6) 528-533
Richard HintonM.D.
Department of Orthopaedics, Union Memorial Hospital, 3333 N. Calvert Street
Suite 400, Baltimore, MD 21218
Email: lyn.camire@medstar.net