Endoscopic removal of an impacted bile duct self-expanding metal stent (SEMS) using the SEMS-in-SEMS technique

N. González
1   Division of Gastroenterology and Hepatology, Basil Hirschowitz Endoscopic Center of Excellence, University of Alabama, Birmingham, Alabama, USA
2   Department of Gastroenterology, Hospital de Clínicas, Montevideo, Uruguay
,
J. Ramesh
1   Division of Gastroenterology and Hepatology, Basil Hirschowitz Endoscopic Center of Excellence, University of Alabama, Birmingham, Alabama, USA
,
C. M. Wilcox
1   Division of Gastroenterology and Hepatology, Basil Hirschowitz Endoscopic Center of Excellence, University of Alabama, Birmingham, Alabama, USA
,
K. Mönkemüller
1   Division of Gastroenterology and Hepatology, Basil Hirschowitz Endoscopic Center of Excellence, University of Alabama, Birmingham, Alabama, USA
› Author Affiliations
Further Information

Corresponding author

K. Mönkemüller, MD, PhD, FASGE
Division of Gastroenterology and Hepatology
Basil Hirschowitz Endoscopic Center of Excellence
Endoscopy Unit, JT 664
619 19th Street S
Birmingham, AL 35249
USA   
Fax: +1-205-934-1537   

Publication History

Publication Date:
05 September 2013 (online)

 

In contrast to covered self-expanding metal stents (c-SEMS), retrieval of uncovered SEMS (u-SEMS) from the bile duct may be quite difficult or impossible. We describe the endoscopic removal of an impacted u-SEMS using the SEMS-in-SEMS technique.

A 62-year-old man presented with acute pancreatitis, jaundice, and weight loss. Computed tomography (CT) revealed a “malignant-appearing” mass in the head of the pancreas causing bile duct dilation. The patient underwent placement of a u-SEMS to palliate the jaundice and as a bridge to surgery. Upon referral, endoscopic ultrasound was performed with fine-needle aspiration of the pancreatic head lesion. Cytology was negative for malignancy. In addition, follow-up CT 4 weeks later revealed resolution of the mass.

Over a period of 9 weeks, the clinical status of the patient remained stable, without weight loss or B symptoms. Removal of the SEMS was therefore planned. Despite using multiple utensils, including oval and hexagonal snares, rat tooth and alligator forceps, and changes in scope position and torquing maneuvers, it was impossible to remove the u-SEMS during three different ERCP sessions. During the third ERCP, a decision was made to place a c-SEMS inside the impacted u-SEMS ([Fig. 1]).

Zoom Image
Fig. 1 Radiograph showing the fully covered self-expanding metal stent within the uncovered metal stent.

The patient returned 10 days later for stent extraction. During ERCP, a snare was placed tightly around both stents, but removal was still impossible as the snare slipped off repeatedly. Therefore, a rat tooth forceps was used to grab the inner stent ([Video 1]). Constant traction was applied by pulling and torquing the scope for about 2 – 3 minutes. Both stents were retrieved completely ([Fig. 2]; [Video 1]). Cholangiogram did not reveal any bile duct damage. The patient was doing well 1 month later.

Zoom Image
Fig. 2 Both self-expanding metal stents (SEMS) could be retrieved completely. a The covered SEMS was inserted through the side mesh of the uncovered stent; b Close-up of both covered and uncovered SEMS.


Quality:
Rat tooth forceps was used to grab the uncovered self-expanding metal stent and then constant traction was applied by pulling and torquing both the rat tooth forceps and the scope for about 2 minutes. Under fluoroscopic view, careful attention was paid to observe stent dislodgement and its movement through the luminal upper gastrointestinal tract. Once retrieved through the mouth, the stents were placed on the table and examined for completeness of extraction.

The stent-in-stent technique has been reported to be useful for the removal of impacted SEMS in the esophagus [1]. Our report is of interest because it demonstrates that this salvage technique may be applicable to impacted u-SEMS in the bile duct. The mechanism may be similar to the one described for impacted esophageal stents [1]: the newly inserted c-SEMS exerts radial pressure inside the previous u-SEMS leading to sloughing or necrosis of the inner lining of the bile duct. Our report is also of potential clinical interest as there is a modern trend to use SEMS for benign pancreatobiliary disease. Knowledge of stent removal techniques is therefore mandatory. Reported techniques to remove SEMS include snare, rat tooth forceps, wire cutter, invagination and extraction in pieces [2] [3]. Our technique of intact snare retrieval using the SEMS-in-SEMS method may therefore be added to the armamentarium of the interventional pancreatobiliary endoscopist.

Endoscopy_UCTN_Code_TTT_1AR_2AZ


#

Competing interests: None

  • References

  • 1 Hirdes MM, Siersema PD, Houben MH et al. Stent-in-stent technique for removal of embedded esophageal self-expanding metal stents. Am J Gastroenterol 2011; 106: 286-293
  • 2 Kahaleh M, Tokar J, Le T et al. Removal of self-expandable metallic Wallstents. Gastrointest Endosc 2004; 60: 640-644
  • 3 Egan LJ, Baron TH. Endoscopic removal of an embedded biliary Wallstent by piecemeal extraction. Endoscopy 2000; 32: 492-494

Corresponding author

K. Mönkemüller, MD, PhD, FASGE
Division of Gastroenterology and Hepatology
Basil Hirschowitz Endoscopic Center of Excellence
Endoscopy Unit, JT 664
619 19th Street S
Birmingham, AL 35249
USA   
Fax: +1-205-934-1537   

  • References

  • 1 Hirdes MM, Siersema PD, Houben MH et al. Stent-in-stent technique for removal of embedded esophageal self-expanding metal stents. Am J Gastroenterol 2011; 106: 286-293
  • 2 Kahaleh M, Tokar J, Le T et al. Removal of self-expandable metallic Wallstents. Gastrointest Endosc 2004; 60: 640-644
  • 3 Egan LJ, Baron TH. Endoscopic removal of an embedded biliary Wallstent by piecemeal extraction. Endoscopy 2000; 32: 492-494

Zoom Image
Fig. 1 Radiograph showing the fully covered self-expanding metal stent within the uncovered metal stent.
Zoom Image
Fig. 2 Both self-expanding metal stents (SEMS) could be retrieved completely. a The covered SEMS was inserted through the side mesh of the uncovered stent; b Close-up of both covered and uncovered SEMS.