Introduction
We report how to prepare a cheap radiopaque filling material that sounds at first
like a curiosity. The imaging optimizations were developed especially to investigate
forensic tissue samples. Postmortem imaging of organs or tissue fragments has become
increasingly important for the evaluation of the cause of death and possible comorbidity
under circumstances that may be very specific ([1] Grabherr S et al. Am J Radiol 2007;
188: 832 – 838). Here, our contrast medium proved its functionality perfectly in postmortem
computed tomography of the vascular system.
Standard ex vivo CT imaging techniques of blood vessels use the Microfil® product line, offered by FlowTech, Carver, Massachusetts, USA. FlowTech products
enable the preparation of beautiful, long lasting casts with a huge benefit in research
and teaching. Radiopaque Microfil products enhance the contrast of the filled spaces
encapsulated by tissue of generally lower attenuation of X-rays, hence enabling CT
imaging. Thus, standard histological cutting methods are not performable when the
vascular system is filled with a plastic-like material. Therefore, we needed a radiopaque
filling material that enabled filling of a vessel system with minimal effect so that
a CT scan would image the original condition of the tissue of an organ. This implies
especially that the filling material has to be within a wide adjustable viscosity
range so that injections into the tissue samples are readily possible. The material
should be easy to handle, easy to prepare, and cheap, and contain no harmful materials.
Furthermore, it must be possible to perform histological examinations of the forensic
tissue directly after a CT scan. A radiopaque filling material containing curd, clinical
contrast agent, and water fulfilled all these desired requirements.
Materials, Methods, and Results
Materials, Methods, and Results
[Fig. 1] shows the experimental setup of the injection of the curd-contrast-agent mixture
under physiological pressure conditions on the scanner table. The CT investigation
was performed with a flat-panel-based eXplore Locus Ultra®, GE Healthcare, London, Ontario, Canada. 640 rows and 1024 columns were read out
from the detector in the used protocol. 1000 views were taken in one gantry rotation
of 16 s duration at 140 kVp and 10 mA. The CT data were reconstructed with a cone-beam
filtered back-projection algorithm into a 512 × 512 × 340 voxel matrix. The 12-bit
DICOM image data had isotropic voxel sizes of (0.295 mm)3. [Fig. 2] shows an image of the coronary vessels of a human heart under investigation in the
forensic department. [Fig. 3] shows a detailed view of the micro-vascularization. Visualization of the data was
performed with an Advantage Workstation®, Version 4.1, GE Medical Systems, Buc, France. Vessel filling and contrast enhancement
were achieved using a curd-contrast-agent mixture. [Fig. 4] shows a histological image of a sample of an organ that received vascular filling
with the curd-contrast-agent mixture and underwent a subsequent CT scan prior to histological
analysis. The histological examinations showed that no curd-contrast-agent could be
found in the extravasal space.
Fig. 1 Setup of the radiological investigation of a forensic sample. The hands, shown on
the left of the photograph, inject the curd-contrast-agent-water mixture. A syringe,
a pressure gauge (image foreground), and a catheter (image background), which goes
to the vessel of the tissue sample under investigation, are connected by a three-way
cock. After the injection of the contrast agent mixture, the three-way cock is closed
in the direction of the tissue sample.
Abb. 1 Aufbau zur radiologischen Untersuchung eines Weichteilpräparates aus der Rechtsmedizin.
Die Hände links im Bild applizieren das Kontrastmittelgemisch aus Quark, Röntgenkontrastmittel
und Wasser. Über einen Dreiwegehahn (im Bildzentrum) sind eine Spritze, ein Manometer
(Bildvordergrund) und ein Katheter (Bildhintergrund) zum Zugangsgefäß des zu untersuchenden
Präparates angeschlossen. Nach der Verabreichung des Kontrastmittelgemisches wird
der Dreiwegehahn zur Gewebeprobe hin verschlossen.
Fig. 2 Volume rendering image of the coronary arteries of a human heart. Image bottom border
length is 13.0 cm.
Abb. 2 Volumen-Rendering-Darstellung der Herzkranzgefäße eines menschlichen Herzens. Die
untere Bildkante ist 13,0 cm lang.
Fig. 3 Detail view of a heart’s microvascular system. Image bottom border length is 5.5 cm.
Abb. 3 Detailansicht der Mikrovaskularisation eines Herzens. Untere Bildkantenlänge ist
5,5 cm.
Fig. 4 Histological image of hematoxylin eosin staining of heart muscle tissue after the
application of the contrast media mixture. Within the tissue, an arteriole is visible
containing a contrast medium, b empty residual lumen, and c erythrocytes. The histological examination is not affected by the contrast media
injection.
Abb. 4 Histologisches Bild einer Hematoxylin-Eosin-Färbung von Herzmuskelgewebe nach der
Anwendung der Kontrastmittelmischung. Innerhalb des Gewebes ist eine Arteriole sichtbar,
die a Kontrastmittel, b leeres Residual-Lumen und c Erythrozyten enthält. Die histologische Untersuchung wird nicht durch die Kontrastmittelinjektion
beeinträchtigt.
To prepare the curd-contrast-agent mixture, we used low-fat curd (Ehrmann AG, Oberschönegg,
Germany), and Micropaque® CT suspension (Guerbet, Sulzbach, Germany) as the contrast agent. A typical mixture
contains e. g. 35 ml curd, 12 ml Micropaque CT suspension, and 6 ml water. The mixture
must be thoroughly shaken. Air bubbles have to be removed by vacuum, or by other appropriate
methods. An option to perform this is to put the contrast media mixture in a flask
that is connected via a three-way cook to a water-jet vacuum pump or another vacuum
source. Application of vacuum for a duration of 30 s is sufficient. When the vacuum
is shut off, the contrast media mixture must be slowly brought back to normal air
pressure by slowly opening the three-way cook. There may not be any air bubbles in
the mixture that is finally injected.
It is recommended to check the optimal amount of the different components of the recipe
in pre-evaluations with animal organs. High curd concentrations lead to mixtures with
high viscosities that can hardly be injected. Variations of the viscosity may be tried
out by changing the water concentration in a range from 10 to 20 ml – referring to
the amounts of the different components mentioned above. High concentrations of contrast
agent may lead to beam hardening artifacts in CT imaging. Low concentrations may produce
only insufficient contrast enhancement. Variations concerning the radiopaque imaging
properties of the contrast media mixture may be evaluated in a range from 10 to 20 ml
of the Micropaque CT suspension.
Discussion
In this study we introduced a contrast agent mixture and we showed CT imaging results
in [Fig. 2], [3] from the area of forensic medicine. A general introduction into the peculiarities
of postmortem imaging of human cadavers is given in the review article in ref. [1].
Those authors show e. g. that there was always an astonishing amount of imagination
concerning the materials being used to prepare vessel casts or to produce radiopaque
contrast media for X-ray-based vessel imaging. Currently used post mortem angiography
materials include diesel or paraffin oil with an addition of a lipophilic contrast
agent (Grabherr S et al. Am J Radiol 2006; 187:W515–W523). Grabherr and coworkers
developed and standardized a post mortem angiographic method where a modified heart
lung machine is used to establish an artificial paraffin circulation in a complete
human cadaver. A bolus injection of an oily contrast agent can then be used to conduct
an angiography investigation comparable to an angiography performed in living subjects
(Grabherr S et al. Int J Legal Med 2011; 125: 791 – 802). Such angiography methods
are possible in cadavers up to three days post mortem. Decomposition of the tissue
and the heterogeneous variation of the tissue stiffness after that make standardization
difficult or impossible. The controversy described in ref. [1] concerning the injection
methods, which range from manual injections to pressure controlled pumps, or gravity,
is therefore easy to understand. These issues must be seen in the background of the
basic challenge in forensic medicine not to alter an exam result by the investigation
method. Minimal invasiveness is therefore important and it is strongly recommended
to handle a subject or an organ as carefully as possible.
We conclude that there is no general optimal composition of the different components
of the contrast media mixture. Organs such as the heart, liver or kidneys can be rather
stiff, rigid, and hard, or they can be soft, mellow, or flexible in forensic investigations,
depending on the cause and circumstances of death, storage of specimen, and time of
postmortem imaging. Therefore, it is not easy to obtain reproducible vascular fillings
that give a similar visual result. Manometer control to gain a reproducible injection
pressure is only sometimes helpful, since organ conditions can vary substantially.
A certain degree of expertise is necessary to obtain acceptable results, which is
independent of the filling material used. We assume that this methodology can also
be applied successfully in experimental radiology, small animal imaging, and general
pre-clinical investigations. The standard method of vessel investigations from isolated
organs is in this area rather based on Microfil preparations. Typical applications
are the morphometrical analysis of the branching angles of porcine coronary arterial
tree from CT images by Wischgold and colleagues (Wischgoll T et al. Am J Physiol Heart
Circ Physiol 2009; 297: H1949 – H1955), the quantitative ex-vivo micro-computed tomographic
imaging of blood vessels and necrotic regions within tumors by Downey and coauthors
(Downey CM et al. PLoS ONE 2012; 7: e41 685), or the quantification of the vasa vasorum
density in MSCT coronary angiograms by Moritz and colleagues (Moritz R et al. J Comput
Assist Tomogr 2010; 34: 273 – 278). Such plastic cast producing methods have some
disadvantages compared to the suggested “curd-method”: plastic material may shrink
minimally during the hardening process; higher costs; and furthermore the difficulties
to perform histological examinations when a hard plastic material is embedded in the
tissue. A basic advantage of the use of plastic materials is the resulting cast, which
can be used to conduct anatomic measurements or visual presentations. Thus, current
developments in information technology make such casts often needless since anatomic
measurements can be performed on digital CT data. Similarly, the digital presentation
of CT data is currently standard and 3 D printing is also an option if casts are necessarily
used.