Introduction
            Endoscopic ultrasound-guided fine-needle aspiration (EUS-FNA) has become a mainstay
               diagnostic technique for the evaluation of lesions arising from the pancreas and upper
               gastrointestinal tract as well as adjacent structures, including lymph nodes and the
               liver. EUS-FNA is safe and has a diagnostic accuracy of 60 % to 90 % [1]
               [2]
               [3]. The diagnostic yield of EUS-FNA is influenced by several factors: the nature of
               the target lesion, experience of the endoscopist, presence of an on-site cytopathologist,
               and various technical aspects, such as needle size and number of passes. In order
               to limit the influencing parameters, a novel needle design was introduced in 2010 – 2011
               that features a hollowed-out reverse bevel to trap core samples (EchoTip ProCore;
               Cook Medical, Bloomington, Indiana, USA). A potential advantage of this type of needle
               is the acquisition of larger amounts of tissue with preserved architecture. Preliminary
               results on the performance of the 19-gauge (G), 22G, and 25G ProCore needles were
               promising, with reported high diagnostic accuracy rates of 89.6 % [4], 88.5 % [5], and 86 % [6], respectively. However, comparative studies of the 22G ProCore needle versus the
               22G standard cytology needle showed similar results for the cytology parameters, amount
               of blood contamination, and diagnostic adequacy and accuracy of the cell block material
               [7]
               [8]
               [9]. Fewer data are available on the diagnostic yield of the 25G ProCore needle. Preliminary
               data demonstrated an 83 % first-pass sensitivity for pancreatic malignancy, with a
               cumulative sensitivity of 96 % after three passes [6]. Histologic diagnosis was possible in 63 % after a single pass and in 80 % after
               multiple passes [6]. Taking into account the results of a recent meta-analysis [10] that shows a potential advantage of the 25G needle over the 22G needle for the diagnosis
               of pancreatic malignancy, the idea of a 25G needle that can obtain histologic specimens
               sounds attractive. Therefore, we designed a prospective study comparing the most commonly
               used 22G standard cytology needle versus the 25G EchoTip ProCore needle in the same
               group of lesions in terms of differences in accuracy, technical performance, and quality
               and quantity of the cytology and cell block specimen obtained.
         Materials and methods
            This pilot study included patients with pancreatic mass lesions or lymphadenopathy
               who were referred for EUS-FNA sampling at our hospital (Grand Hôpital de Charleroi,
               Charleroi, Belgium) between December 2012 and October 2013. Exclusion criteria included
               the following: cystic lesion, coagulation disorder (international normalized ratio
               > 1.5, platelet count < 50 000/mm3), pregnancy, age < 18 years, and refusal or inability to provide informed consent.
               Patients were monitored closely for possible complications after the procedure. The
               study was approved by the Grand Hôpital de Charleroi review board, and written informed
               consent was obtained from all patients for participation in the study. This trial
               was registered at ClinicalTrials.gov (B027201316271).
            A linear array echoendoscope (UCT160; Olympus Medical Systems, Tokyo, Japan) connected
               to a processor featuring color Doppler function (ProSound Alpha 7; Hitachi Medical
               Systems Europe [Aloka], Zug, Switzerland) was used for all procedures, which were
               carried out by four experienced endosonographers (A.S., P.W., H.H., and C.G.) and
               one gastroenterology fellow (G.M.) while the patients were under propofol sedation.
               The gastroenterology fellow was present during all procedures in order to ensure adherence
               to the study protocol. After the lesion had been carefully inspected and vessel interposition
               along the puncture route excluded by color Doppler, EUS-FNA was performed with both
               a 22G (EchoTip) needle and a 25G EchoTip ProCore needle ([Fig. 1]). The order in which the needles were used was according to computer-generated randomization.
               During each puncture, the needle went through the lesion, and 15 to 20 to-and-fro
               movements were made with minimal negative pressure simultaneously applied by pulling
               the needle stylet slowly and continuously (slow-pull technique) [6]. After the second puncture, if the acquired specimen was not adequate, EUS-FNA was
               repeated with one of the two needles until an adequate specimen had been obtained.
               No on-site cytopathologic evaluation was done. FNA material was recovered in a vial
               containing 15 mL of ThinPrep PreservCyt Solution (Hologic, Marlborough, Massachusetts,
               USA) by flushing with 5 mL of this solution and, if needed, by pushing the stylet
               through the needle. The specimens obtained were labeled according to the aspiration
               sequence and sent to a single experienced cytopathologist (B.W.), who was blinded
               to the types of needles used. All samples were processed in the same way. One monolayer
               was obtained with ThinPrep 2000 and stained with Papanicolaou stain (cytologic specimen);
               the leftover material was centrifuged and the clot retrieved in agar, fixed in formalin
               for at least 2 hours, and embedded in paraffin to obtain a cell block (histologic
               specimen). One 5-μm-thick slide stained with hematoxylin and eosin was available for
               every cell block. Immunocytochemistry was applied on the cell block if needed for
               the diagnosis. Only the results of the first and second punctures were compared.
 Fig. 1 The 25G EchoTip ProCore histology needle (left), which has a hollowed-out reverse
                  bevel, and the 22G EchoTip standard cytology needle (right). Photos supplied by Cook
                  Medical.
                  Fig. 1 The 25G EchoTip ProCore histology needle (left), which has a hollowed-out reverse
                  bevel, and the 22G EchoTip standard cytology needle (right). Photos supplied by Cook
                  Medical.
            
            The quality of the obtained cytologic and histologic specimens was reported by the
               cytopathologist according to the following grading system: insufficient material or
               contaminated material (score of 0), rare diagnostic cells (score of 1), diagnostic
               cells at every field with 10 times magnification (score of 2), or diagnostic cells
               at every field with 20 times magnification (score of 3). The primary outcome parameter
               was the percentage of cases in which the pathologist classified the quality of the
               sample as sufficient for histocytologic evaluation (score ≥ 1). This grading system,
               although not published as such, is very easy to apply to a cytologic slide, especially
               when the specimen has been obtained with a monolayer technique. It is also easily
               reproducible because one needs only to record the magnification at which diagnostic
               cells are seen filling the microscopic field.
            Secondary outcomes included comparisons of several performance parameters. The visibility
               of each needle was qualitatively scored as optimal or suboptimal and the ease of puncture
               as easy or difficult [6]
               [11]. The amount of blood contamination was scored qualitatively as no blood (score of
               1), small quantity of blood (score of 2), or large quantity of blood (score of 3)
               [12]. The quantity of the specimens was graded as small (score of 1) in the presence
               of fragments with no visible core, medium when one or two cores were visible (score
               of 2), or excellent when at least three cores were visible (score of 3).
            A final diagnosis of malignancy or benignancy was made according to one of the following
               reference methods: (1) definite benign or malignant histologic diagnosis based on
               surgical resection specimens from patients who had undergone surgery, (2) cytology
               or histology findings with definite proof of malignancy in patients with unresectable
               tumors according to imaging findings and compatible clinical follow-up, and (3) cytology
               or histology findings without proof of malignancy and a minimum clinical and radiologic
               follow-up of 7 months.
            For comparison of continuous data, a paired t test was performed if a normal distribution was shown, and the Wilcoxon rank sum
               test was carried out if normality could not be demonstrated. McNemar’s test was used
               for dichotomous categorical data. For all tests, a P value of less than 0.05 was regarded as statistically significant with SPSS 17.0
               for Windows (SPSS Inc., Chicago, Illinois, USA).
         Results
            A total of 28 EUS-FNA procedures targeting lesions of the pancreas (n = 19) and lymph
               nodes (n = 9) were performed in 27 patients (18 women, 9 men) with a median age of
               69 years (range, 38 – 88). The final diagnoses were pancreatic adenocarcinoma (n = 18),
               pancreatic neuroendocrine tumor (n = 1), malignant lymphadenopathy (n = 6), and benign
               lymphadenopathy (n = 3). No benign pancreatic lesions were encountered.
            In the subgroup of patients with pancreatic lesions, the median tumor size was 39 mm
               (range, 10 – 70). Of these pancreatic lesions, 10 were punctured through the duodenum
               and 9 were punctured through the stomach. 
            In the subgroup of patients with lymphadenopathy, the median lymph node size was 24 mm
               (range, 15 – 45). Of these lesions, 6 were punctured through the esophagus and 3 through
               the duodenum.
            Final diagnoses were made on the basis of surgery in 3 cases, positive FNA for malignancy
               with a compatible clinical course in 23 cases, and negative FNA for malignancy with
               at least 7 months of follow-up in 2 cases. No procedure-related complications were
               seen.
            In terms of EUS visualization, visualization was suboptimal in 16 % of punctures with
               the 25G needle versus 0 % of punctures with the 22G needle; however, this difference
               was not statistically significant (P = 0.125) ([Videos 1, 2]). No relevant differences were found regarding the ease of puncture (P = 0.688), amount of blood contamination (P = 0.705), macroscopic quantity of the material (P = 0.858), and quality of the cytologic (P = 0.438) and histologic (P = 0.220) specimens ([Table 1]). Subgroup characteristics regarding the adequacy of histocytologic material for
               the two needle types and different types of lesions are shown in [Table 2]. 
            
               
 Endoscopic ultrasound-guided fine-needle aspiration of a pancreatic mass with a 25G
               EchoTip ProCore needle.
            
               
 Endoscopic ultrasound-guided fine-needle aspiration of a pancreatic mass with a 22G
               EchoTip needle.
            
               
                  
                     Table 1
                     
                     Histocytologic and endoscopic assessment of the performance of the two needles.
                     
                  
                     
                     
                        
                        |  | 25G EchoTip ProCore, median (range) | 22G EchoTip, median (range) | 
                              P value | 
                     
                  
                     
                     
                        
                        | Cytologic score |  1.5 (0 – 3) | 2 (0 – 3) | 0.438[1]
                               | 
                     
                     
                        
                        | Histology score |  1 (0 – 2) | 1 (0 – 3) | 0.220[1]
                               | 
                     
                     
                        
                        | Total histocytologic score |  2.5 (0 – 5) | 3 (0 – 6) | 0.255[2]
                               | 
                     
                     
                        
                        | Macroscopic quantity |  2 (1 – 3) | 2 (1 – 3) | 0.858[1]
                               | 
                     
                     
                        
                        | Amount of blood contamination |  1 (1 – 2) | 1 (1 – 3) | 0.705[1]
                               | 
                     
                     
                        
                        | Difficulties in visualization | 16 % | 0 % | 0.125[3]
                               | 
                     
                     
                        
                        | Difficulties in use | 17.3 % | 8.6 % | 0.688[3]
                               | 
                     
               
               
               
               1 Wilcoxon rank sum test.
               
               
               2 Paired t test.
               
               
               3 McNemar’s test.
               
                
            
            
            
               
                  
                     Table 2
                     
                     Adequacy of material for cytologic and histologic assessment.
                     
                  
                     
                     
                        
                        |  | Adequate material for cytologic assessment | Adequate material for histologic assessment | 
                     
                  
                     
                     
                        
                        | Target lesion | 25G EchoTip ProCore | 22G EchoTip | 
                              P value[1]
                               | 25G EchoTip ProCore | 22G EchoTip | 
                              P value[1]
                               | 
                     
                     
                        
                        | Pancreatic mass | 16/19 | 15/19 | 1 | 15/19 | 17/19 | 0.68 | 
                     
                     
                        
                        | Lymph node |  8/9 |  7/9 | 1 |  7/9 |  7/9 | 1 | 
                     
                     
                        
                        | All lesions | 24/28 | 22/28 | 0.5 | 22/28 | 24/28 | 0.5 | 
                     
               
               
               
               1 McNemar’s test.
               
                
            
            
            In the subgroup of patients with pancreatic cancer, each needle missed two cases.
               The first case was a patient with adenocarcinoma of the head of the pancreas. Transduodenal
               puncture with the 25G needle showed rare benign cells (cytology score = 1, cell block
               score = 1), while puncture with the 22G needle was noncontributive (cytology score = 0,
               cell block score = 0). Because of the strong suspicion of malignancy, a follow-up
               EUS-FNA a few weeks later was performed with a standard 22G needle and confirmed the
               diagnosis of pancreatic adenocarcinoma. Each needle missed one other case of pancreatic
               adenocarcinoma of the head of the pancreas punctured through the duodenum because
               of insufficient histocytologic material (total histocytologic score = 0). Therefore,
               the single-pass sensitivity for pancreatic neoplasia for both needles was 89.5 % (95 %CI
               66.82 – 98.39). 
            In the subgroup of patients with lymphadenopathy, a total of two false-negative results
               for malignancy were obtained with both needles that concerned the same patients. The
               first patient had gallbladder cholangiocarcinoma and perihepatic lymph nodes suspicious
               for malignancy that were 2 cm in size. Transduodenal puncture with both needles resulted
               into noncontributive histocytologic material (total histocytologic score = 0). Follow-up
               imaging was compatible with metastatic lymph nodes. The second patient had lung cancer
               and mediastinal lymph nodes suspicious for malignancy (4 cm in size). EUS-FNA with
               both needles showed rare benign cells (cytology score = 1, histology score = 1). However,
               clinical and radiologic follow-up was compatible with metastatic lymph nodes. Therefore,
               the single-pass sensitivity, specificity, positive predictive value, negative predictive
               value, and accuracy for malignancy were equal for the needle types: 66 % (95 %CI 24.1 – 94),
               100 % (95 %CI 30.9 – 100), 44.4 % (95 %CI 39.5 – 100), 60 % (95 %CI 17 – 92.7), and
               84.8 % (95 %CI 67.3 – 94.2), respectively.
         Discussion
            Overall, EUS-FNA is highly effective for most pancreatic tumors and solid malignancies
               adjacent to the upper gastrointestinal tract, with reliable sensitivity, specificity,
               and overall diagnostic accuracy of 60 % to 90 % [1]
               [2]
               [3]. However, FNA cytology specimens may not be adequate in cases in which the diagnosis
               relies on tissue architecture, such as autoimmune pancreatitis, lymphomas, gastrointestinal
               stromal tumors, and well-differentiated adenocarcinomas [1]
               [13].
            In order to procure larger amounts of tissue with preserved architecture that would
               enable histologic analysis, a novel needle assembly with a reverse-bevel technology
               (EchoTip ProCore) was introduced in 2010 – 2011. According to the designer, the side
               notch (not Tru-Cut) should provide additional “cheese-grating” action, making it possible
               to obtain more specimens.
            Initially, a 19G version was introduced to the market. It had the ability to obtain
               full histology in 89.5 % of cases, with an overall diagnostic accuracy of 93 % [4]. However, because technical difficulties were encountered during transduodenal passes,
               the same needle was later introduced in a 22G platform. A PubMed search revealed only
               three studies of ProCore needles comparing the diagnostic yield of the most commonly
               used 22G FNA needle with that of the 22G ProCore needle ([Table 3]) [7]
               [8]
               [9]. No advantage of the ProCore design was reported in terms of cytology and histology.
               However, in the study by Witt et al. [8], fewer passes were needed to achieve diagnosis with the 22G ProCore needle (2.11
               with the 22G ProCore needle vs 2.94 with the 22G standard needle). Nevertheless, this
               result should be interpreted with caution because of the small number of different
               groups of lesions (n = 18). Interestingly, in the study by Strand et al. [9], technical failure was encountered in 5 of 32 patients because of significant resistance
               to advancement of the needle, with deflection of the echoendoscope (GF-UCT140, Olympus)
               while in the transduodenal position.
            
               
                  
                     Table 3
                     
                     Published comparative trials of the performance of the 22G EchoTip ProCore needle
                        versus the 22G standard EchoTip cytology needle.
                     
                  
                     
                     
                        
                        | Author | Type of study | Number of lesions | Target | Comparison of histology specimen | Overall diagnostic yield | Technical performance | Comment | 
                     
                  
                     
                     
                        
                        | Bang et al. 2012 [7]
                               | Randomized | 28 lesions per needle type | Pancreas | No significant difference  | Equivalent  | No significant difference | Needles of different manufacturers (Boston Scientific and Cook) | 
                     
                     
                        
                        | Witt et al. 2013 [8]
                               | Retrospective | 18 lesions per needle type | Pancreas, lymph nodes, other masses | No significant difference  | Equivalent; fewer passes needed with ProCore needle | No reported difficulties  | Retrospective study, small number per group | 
                     
                     
                        
                        | Strand et al. 2014 [9]
                               | Randomized | 32 lesions punctured by both needles | Pancreas | No significant difference  | Lower overall diagnostic yield for the ProCore needle  | Technical failure in 16 % cases with the ProCore needle | Only two passes permitted for the 22G ProCore group  | 
                     
               
             
            
            Most recently, the 25G ProCore needle has been introduced. Potential advantages of
               the smaller caliber are greater flexibility, less friction in the needle sheath, easier
               penetration of hard pancreatic tumors, and fewer bloody aspirates. This argument was
               suggested in a meta-analysis by Madhoun et al. [10] but not confirmed in a recent large randomized controlled trial [14]. Preliminary data on the performance of the 25G ProCore needle suggest an 83 % first-pass
               sensitivity for pancreatic malignancy, with a cumulative sensitivity of 96 % after
               three passes [6]. Histologic diagnosis was possible in 63 % after a single pass and in 80 % after
               multiple passes. In our study, both the 22G standard needle and the 25G ProCore needle
               had a slightly better single-pass performance. Cytologic and histologic diagnoses
               were possible in 84.2 % and 78.9 % of cases, respectively, with an 89.5 % overall
               sensitivity for pancreatic cancer.
To the best of our knowledge, this is the first randomized, single-group, prospective
               study to assess the diagnostic yield of the 25G ProCore versus the 22G standard cytology
               needle in pancreatic mass lesions and lymphadenopathy adjacent to the upper gastrointestinal
               tract. Our pilot study compared different needle characteristics and the adequacy
               of cytology and histology. Although the 25G needle was not  visualized optimally in
               16 % of punctures no effect on the quality of cytology or histology, amount of blood
               contamination, and technical difficulty was observed. We found no significant differences
               between the two needles in terms of diagnostic yield, with similar diagnostic rates
               and similar numbers of successful procedures.
            Our prospective, single-group, paired design has the intrinsic strength that performance
               comparisons were made on the same group of lesions, eliminating in this way the bias
               of differences in lesion type, size, and location. This is in contrast with most EUS-FNA
               studies, which compare different needles in different populations. The main limitation
               of our study is the small number of cases. A larger number of cases may be needed
               to detect subtle differences between the two needles. Furthermore, no benign pancreatic
               lesion was encountered during the study period, such as focal autoimmune pancreatitis,
               in which an EUS-FNA diagnosis is challenging.
            In conclusion, this pilot study demonstrates that the diagnostic yield of the new
               25G EchoTip ProCore needle is comparable with that of a 22G standard FNA assembly.
               Both needles performed equally in terms of the quality of cytologic and histologic
               specimens, blood contamination, and ease of puncture.
 Fig. 2 Macroscopic (top), cytologic (center), and histologic (bottom) images of endoscopic
                  ultrasound-guided fine-needle aspiration specimens of pancreatic adenocarcinoma punctured
                  with a 25G EchoTip ProCore (left) and a 22G EchoTip cytology needle (right).
                  Fig. 2 Macroscopic (top), cytologic (center), and histologic (bottom) images of endoscopic
                  ultrasound-guided fine-needle aspiration specimens of pancreatic adenocarcinoma punctured
                  with a 25G EchoTip ProCore (left) and a 22G EchoTip cytology needle (right).