Keywords hearing tests - mobile applications - automation
Introduction
Hearing impairment is one the most prevalent debilitating illnesses affecting the
worldwide populations.[1 ] The two primary populations impacted by hearing loss are those in occupational settings
and older adults (that is, those aged 65 years and older). It is estimated that approximately
one-third of Americans between the ages of 65 and 74 have some degree of hearing loss,
which increases to almost half of the population aged 75 years and older.[2 ] Additionally, the aging of the baby boomers is having a dramatic effect on the elderly
population in the U.S. Data indicate that from 2012–2050 the number of persons over
age 65 is expected to double.[3 ] With prevalence of hearing loss estimated at 40–80% of individuals between the ages
of 65–85 years,[4 ] early identification and proper management are often cited as the most important
factors in minimizing the effects of age-induced hearing loss in the patients' quality
of life.[5 ] These trends among the aging population will continue to place additional burden
on the population of audiologists, which is projected to have a shortage of qualified
professionals available in coming years.[6 ]
In addition to the strain that the aging population is placing on the community of
qualified audiologists, the rates of hearing loss among employees of governmental
agencies in the U.S. and Europe have led these groups to identify noise exposure as
a high-priority issue.[7 ]
[8 ] In the U.S., the National Institute for Occupational Safety and Health (NIOSH) has
designated occupational noise as a high priority research area in their National Occupational
Research Agenda (NORA).[8 ] Because occupationally-induced hearing loss is frequently targeted as a major concern
for employees,[9 ] the overall shortfall of qualified audiologists has created the demand for valid
hearing screening options for workers in general (regardless of the industry, be it
rural or urban, public or private) and for the aging populations.[10 ]
To address this need for alternatives of assessment methodologies that can accurately
screen the at-risk populations, the research team developed the hEAR mobile screening
application with the aim of producing quality automated screening results while increasing
the ease of access. This is in line with current research, in which mobile applications
and mobile software-based audiometers are becoming more commonplace.[11 ]
[12 ]
[13 ]
[14 ] Our previous work indicated the capacity of the hEAR mobile application to replicate
audiologist-collected screening data, but with a strong dependence upon headphone
reproduction capacity.[15 ] This study examined the use of commercially available headphones when combined with
a full-spectrum mobile application and their capability of reproducing audiologist-quality
hearing screenings.
Methods
Participants
Thirty participants who were university students, faculty, and staff were enrolled
in the study. Twelve of the 30 participants were female, and 18 (60%) were male. While
participants' ages ranged from 20–57 years, most were aged 25–32 years (n = 18). Participants had no previously diagnosed hearing loss and were required to
limit noise exposure 24 hours prior to all tests. All participant recruitment, consent,
data collection, and evaluation methodologies were reviewed and approved by the University
Institutional Review Board (IRB) for the Protection of Human Subjects.
Materials and Methods
A Samsung Galaxy Tab 3.0 (Samsung, Seoul, Korea), an Android device, was chosen to
test the hEAR mobile application because it is built upon the Android platform. The
hEAR application itself was designed based on the best practices for automated screenings
suggested by a variety of sources, such as the World Health Organization (WHO), who
recommend the Bekesy-style audiometry for self-administered hearing screenings.[16 ] As is best practice with these recommendations, the test tones initiate at inaudible
levels and subjects respond to the attenuator control once they hear the tone.
Four sets of headphones were chosen for this study:
Pioneer HDJ 2000 (professional DJ)
Bose Quiet Comfort 25 (noise cancelling) (Bose, Framingham, MA, USA)
Sony MDR 7506 (affordable option) (Sony, Minato, Tokyo, Japan)
Sennheiser HD280 Pro (professional audiologist-quality)
Working within the OSHA and WHO guidelines, the application automatically administered
a series of mini-trials based on the OSHA-designated frequencies (125, 250, 500, 1,000,
2,000, 4,000, and 8,000 Hz). Each frequency was administered a minimum of four times
bilaterally. Each of these mini-trials were administered randomly to the subject.
Each participant underwent at least 28 mini-trials; each individual frequency collection
period lasted from 27 to 33 seconds. The entirety of one full-spectrum collection
period was of ∼ 15 to 20 minutes.
With the potential for false positives/negatives in the user feedback, the application
has a series of algorithms to identify values that may be false positives/negatives
in the data collection. These series of algorithms are primarily based on the amount
of deviations from normal hearing responses (dB) of the general public at each test
frequency. In this regard, the application, upon identifying a potential false positive/negative
in the data collection, automatically and randomly re-administered the identified
frequency later in the test sequence.
The study herein described was the continuation of a previously conducted prospective
cohort pilot study.[15 ] The pilot data indicated the presence of confounders on basis of the hardware used,
which potentially impacted in the quality of the data collected. Thus, the purpose
of this study was to evaluate headphones with different frequency responses and sound
reproduction capacities for accuracy of app-based hearing screening data collection.
Each participant was assigned a participant ID, which was a 7-digit random number.
These were generated by a uniform distribution random number generator for data collection/analysis
purposes. Subject trials were randomized and counterbalanced so that half of the subjects
initiated data collection procedures in the laboratory (Group 1) and the other half
with the audiologist (Group 2). This was done to ensure unbiased estimates. The scheduling
of data collection was performed at an initial meeting with a member of the research
team in which subjects completed screening questionnaires. All communication between
the researchers and the local audiologist used this identification number to maintain
participant protection standards.
In order to meet the requirements of Appendix D of 29 CFR 1910.95, the testing procedures
were performed in the laboratories.[17 ] Ambient sound pressure level (SPL) on each testing cycle, for all 30 subjects, averaged
24 dBA.
Statistical Methods
The statistical analyses were performed using the software SAS Version 13.1 (SAS Institute,
Cary, NC, USA). Since the data was non-linear, generalized estimating equations (GEE)
were used, which are appropriate for longitudinal data, data with small sample size
and data with repeated measures.[18 ] The outcome variable of interest was the SPL in dB. The SPL measurements were differentiated
by both the hEAR mobile application and the pure-tone audiometry test, based on ear
side (that is., left and right ears). The preliminary analysis for the pilot data,
as well as the current data, showed that the ear side was not significant (t = 0.593,
df = 1,478, p = 0.5532).
The means per subject, per headphone, for the respective frequencies were calculated
for a preliminary comparison between the SPL measurements between the different chosen
headphones and those by the pure-tone audiometry test ([Fig. 1 ]). In [Figs. 2 ] and [3 ], we can observe that the counterbalanced group that the participants were assigned
to had only a marginal effect on the overall results.
Fig. 1 Boxplots of the headphones and the audiologist's test using summary statistics plotted
against the measured sound pressure levels (SPL) on the Y-axis.
Fig. 2 Sound Pressure Level (SPL) means per headphone for Group 1 calculated using headphone
means in SPL (dB) at each measured frequency on the X-axis. The plotted means for
headphones show similarities and differences with those measured by the audiologist's
test.
Fig. 3 Sound pressure level (SPL) means per headphone for Group 2 calculated using headphone
means in SPL (dB) at each measured frequency on the X-axis. The plotted means for
headphones show similarities and differences with those measured by the audiologist's
test. Counterbalanced Groups (Groups 1 and 2) had marginal effect (p = 0.08) on the results.
Each frequency was considered separately, and eight (8) agreement scores were generated
for each subject and headphone. Each agreement score was defined as the absolute difference
between the SPL response recorded by the hEAR application and the SPL response recorded
by the audiologist. The [Eq. 1 ] below was used for calculation of the agreement scores.
(1)
From these 8 scores, an average agreement score was generated for that ID at the particular
frequency. The smaller the agreement score, the better it is. Based on the agreement
scores, if Yi was a binary random variable, at a threshold tolerance value (Θ) of
8, defined by [Eq. (2) ]:
(2)
[Eq. 3 ] is the resulting logistic model.
(3)
The formula for the probability of success of a headphone is calculated based on the
[Eq. 1 ]/(1 + E^β0 ), whereas the probability of success of the subsequent headphones is calculated based
on the [Eq. 1 ]/(1 + E^β0+β1 ). Therefore, for headphones at a particular frequency, the higher the probability
of success, the better the headphones are when compared with the others.
Results
The results show that neither the group nor the order in which the two tests were
conducted (hEAR mobile application and audiologist's test) significantly impacted
the probability of success of the headphones (p = 0.0894) ([Table 1 ]). In general, Group 2 (audiologist test prior to hEAR mobile application testing)
had slightly higher probability values (Z = −1.70, p = 0.0894).
Table 1
Results of generalized estimating equation model analysis for the counterbalanced
headphones and audiologist's test of the test initiation
Analysis of GEE parameter estimates; empirical standard error estimates
Parameter
Estimate
SE
95% CI
Z
Pr > |Z|
Intercept
10.6054
1.199
8.254
12.957
8.84
< 0.0001
Pioneer Headphones (1)
0.0175
1.462
−2.849
−2.883
0.01
0.9905
Bose Headphones (2)
8.4433
1.434
5.634
11.253
5.89
< 0.0001[* ]
Sony Headphones (3)
6.0960
1.756
2.674
9.517
3.49
0.0005[* ]
Sennheiser Headphones (4)
−1.6569
1.403
−4.407
1.092
−1.18
0.2376
Control (Audiologist's test)
0
0
0
0
.
.
Frequency
0.0008
0.0001
0.0005
0.0011
5.53
< 0.0001
Group
−0.4885
0.2876
−1.0522
0.0752
−1.70[* ]
0.0894[* ]
Abbreviations: CI, confidence interval; GEE, generalized estimating equation; SE,
standard error.
GEE Fit Criteria: QIC = 1087.4051; QICu = 1054.0
* Indicates statistical significance
Analysis from the model, produced probability of successfully reproducing test results
similar to the audiologist control along with overall statistical significance (p -value). This was performed for every test frequency for each set of headphones. The
greater the probability, the more likely the headphones are to reproduce SPL responses
similar to the audiologist control when compared with the other sets of headphones.
[Table 2 ] indicates that, overall, the Pioneer HDJ 2000 performed the best, or it was statistically
similar to the audiologist-administered test, for all frequencies (all p > 0.05). Similarly, the Sennheiser HD280 Pro also performed well for all frequencies,
and it was not significantly different from the audiologist control (all p > 0.05). According to the frequency reproductions of the selected headphones, the
accuracy of reproduction decreases after 6,000 Hz, when used under daily, normal circumstances.
Additionally, in [Fig. 1 ] we can see that the boxplots for the Pioneer headphones and the audiologist's test
show noticeable similarity with strong overlap in the data. As results in [Table 2 ] indicate, the Bose Quiet Comfort 25 and the Sony MDR 7506 both had statistically
significant results overall, and across multiple frequencies when compared with the
audiologist control.
Table 2
Probability and statistical significance (p values) for test headphones
Frequency
Pioneer HDJ 2000
Bose Quiet Comfort 25
Sony MDR 7506
Sennheiser HD280 Pro
125
0.6792
(p = 0.0754)
0.2156
(p = 0.0072[* ])
0.2156
(p = 0.0013[* ])
0.4591
(p = 0.0598)
250
0.7079
(p = 0.0581)
0.6126
(p = 0.4396)
0.4328
(p = 0.0158[* ])
0.5788
(p = 0.1485)
500
0.6987
(p = 0.0913)
0.6035
(p = 0.4375)
0.6987
(p = 1.0000)
0.6676
(p = 0.6548)
1000
0.7372
(p = 0.0513)
0.6425
(p = 0.4043)
0.6745
(p = 0.5257)
0.7372
(p = 1.0000)
2000
0.6678
(p = 0.1549)
0.4140
(p = 0.1041)
0.1791
(p = 0.0006[* ])
0.5786
(p = 0.3150)
4000
0.2630
(p = 0.0716)
0.1150
(p = 0.0367[* ])
0.0387
(p = 0.0247[* ])
0.1526
(p = 0.1712)
8000
0.2500
(p = 0.646)
0.2500
(p = 1.0000)
0.2500
(p = 1.0000)
0.2500
(p = 1.0000)
Overall
0.6648
(p = 0.2897)
0.4122
(p = 0.0271[* ])
0.3607
(p = 0.0003[* ])
0.4954
(p = 0.0641)
* Indicates statistical significance.
Discussion
The results of the study depict interesting outcomes. Both sets of professional headphones
(that is, the Pioneer HDJ 2000 and the Sennheiser HD280 Pro) performed well in this
test, as both have broad frequency reproduction spectrums. However, the more day-to-day
use headphones (such as the Bose Quiet Comfort 25 and the Sony MDR 7506) did not perform
as well.
The probability of success indicates that the Pioneer HDJ 2000 are the best headphones
for use with the hEAR mobile application. The Sennheiser HD 280 Pro headphones were
designed specifically for audiometric testing in noise insulated or soundproof testing
rooms to reproduce frequencies in an accurate flat frequency response from 100 to
10,000 Hz in noise-insulted environments. This was not the case in our testing room,
where the ambient noise levels were at almost 25 dB, well within the requirement of
both OSHA and ANSI. It was also the researchers' experience that even though the Sennheiser
HD 280 Pro headphones are marketed as “over the ear,” their fit can best be described
as “on the ear” since several participants commented that the ear cups did not fully
cover their ears during the test. This lack of coverage could potentially have had
an effect on the overall results for the Sennheiser headphones.
Although noise cancelling, the Bose Quiet Comfort 25 headphones are specifically designed
for high bass music listening or daily use, and also for comfort rather than for audiometric
testing.[19 ] Therefore, even though these headphones were preferred by most participants due
to their comfort, they did not prove as effective for testing hearing. The same was
the case with the Sony MDR 7506.
All of the headphones used in this exercise had limitations in accurately reproducing
high frequency sounds (above 6,000 Hz). This property could be the reason for the
lower probability of success at the frequency of 8,000 Hz.
Overall, the Pioneer HDJ 2000 headphones were the best audiometry screening option
for use with the hEAR mobile application. Therefore, they may offer a portable option
for full-spectrum audiometric screening. The Sennheiser HD280 Pro headphones, while
potentially more capable than the Pioneer HDJ 2000, may be limited to very isolated
testing environments. The Bose Quiet Comfort 25 and the Sony MDR 7506 did not have
the capability of producing audiologist-quality data when paired with the hEAR application.
Conclusions
It is important to note that there are limitations to this study. The ambient noise
in the testing environment was not tested with an octave band analyzer, as is required
by most organizations, including OSHA.[17 ] However, it is not expected that the analysis with an octave band analyzer would
have produced results that would have altered the overall interpretation of the data,
as the SPL of the overall background noise was just 24 dBA. While there is no way
of accurately knowing if there was frequency-specific interference, the research team
does not expect that octave band analysis would have significantly changed the results
of the study. More evaluation is also needed for a broader test population. However,
even with these limitations, the Pioneer HDJ 2000 headphones paired with the self-administered
hEAR mobile application were able to reproduce overall and frequency-specific results
that were not significantly different than those of a certified audiologist in a controlled
testing environment. These results show a promising trajectory for mobile automated
hearing screening options.