    A 46-year-old woman presented with a large walled-off collection, likely related to a prior episode of pancreatitis. Percutaneous drainage had been attempted at an outside institution, but the collection recurred. She was transferred urgently to our hospital in multiorgan failure with sepsis and abdominal pain.

    Computed tomography (CT) scanning showed a large, mature left upper quadrant collection ([Fig. 1]). Endoscopic ultrasound (EUS)-guided transluminal drainage was performed. A transgastric location was accessed with a 19-gauge needle. A cystogastrostomy was created using a 15 × 10-mm cautery-assisted lumen-apposing metal stent (LAMS; Boston Scientific, Marlborough, Massachusetts, USA). The distal flange was deployed in the cavity; however, because of the mass effect from the stomach ([Fig. 2]), deployment of the proximal flange was difficult. It was clear both endoscopically and on EUS that the proximal flange had not deployed intraluminally ([Video 1]).

    Zoom Image
    Fig. 1 Contrast-enhanced computed tomography scan showing a large walled-off collection in the left upper quadrant that is compressing the stomach.
    Zoom Image
    Fig. 2 Endoscopic and fluoroscopic images during deployment of the lumen-apposing metal stent (LAMS) across the cystogastrostomy showing the limited intragastric space owing to the mass effect from the collection (red arrows) and the absence of the proximal portion of the LAMS endoscopically (red dashed circle) indicating its deployment in the cavity.

    Video 1 Endoscopic transluminal drainage complicated by maldeployment of a lumen-apposing metal stent into the cavity, which is managed by successful removal and salvage of the cystogastrostomy.


    Quality:

    Maintaining wire access to the collection allowed passage of a second LAMS to control the defect without creating a second cystogastrostomy. A single-channel therapeutic upper gastrointestinal endoscope was passed through the second LAMS into the collection. The maldeployed LAMS was grasped using a rat-tooth forceps under fluoroscopic guidance, suctioning the collection down to bring the stent closer and opening and closing the forceps like “Pac-man” ([Fig. 3]). To avoid dislodgement of the second LAMS, the maldeployed LAMS was pulled through the channel of the endoscope.

    Zoom Image
    Fig. 3 Fluoroscopic image showing a single-channel therapeutic upper gastrointestinal endoscope that has been advanced across the lumen-apposing metal stent at the cystogastrostomy and the maldeployed stent in the cavity being grasped with a rat-tooth forceps.

    In summary, we present a case of immediate retrieval of a maldeployed LAMS and salvage of the original cystogastrostomy. The use of multimodal imaging – white-light endoscopy, EUS, and fluoroscopy – during the deployment process is critical. If maldeployment occurs, immediate recognition will allow same-session rescue to be attempted. Maintaining wire access to the collection to facilitate deployment of a second LAMS and subsequent retrieval was crucial in this example. Finally, withdrawal of the LAMS through the channel of a therapeutic endoscope presents a safe removal option.

    Endoscopy_UCTN_Code_CPL_1AL_2AD

    Endoscopy E-Videos
    https://eref.thieme.de/e-videos

    Endoscopy E-Videos is a free access online section, reporting on interesting cases and new techniques in gastroenterological endoscopy. All papers include a high quality video and all contributions are freely accessible online.
    This section has its own submission website at https://mc.manuscriptcentral.com/e-videos


    #

    Competing interests

    Martin Freeman has acted as a consultant for Boston Scientific, Cook Medical, and Xlumena. Stuart Amateau has acted as a consultant for Merit Endoscopy, Boston Scientific, US Endoscopy, and Neurotronic, and has received research support from Cook Medical. Nabeel Azeem has no conflicts to report.


    Corresponding author

    Stuart K. Amateau
    University of Minnesota
    MMC 36
    420 Delaware St SE
    Minneapolis
    MN 55455
    USA   


    Zoom Image
    Fig. 1 Contrast-enhanced computed tomography scan showing a large walled-off collection in the left upper quadrant that is compressing the stomach.
    Zoom Image
    Fig. 2 Endoscopic and fluoroscopic images during deployment of the lumen-apposing metal stent (LAMS) across the cystogastrostomy showing the limited intragastric space owing to the mass effect from the collection (red arrows) and the absence of the proximal portion of the LAMS endoscopically (red dashed circle) indicating its deployment in the cavity.
    Zoom Image
    Fig. 3 Fluoroscopic image showing a single-channel therapeutic upper gastrointestinal endoscope that has been advanced across the lumen-apposing metal stent at the cystogastrostomy and the maldeployed stent in the cavity being grasped with a rat-tooth forceps.