Keywords bipolar - hemiarthroplasty - intraprosthetic dissociation - dislocation
Femoral neck fracture is the second most common hip fracture,[1 ] and its high prevalence in older adults is only increasing with increasing rates
of osteoporosis in the older population.[2 ] Fractures of the femoral neck can be treated with internal fixation if the fracture
is nondisplaced or the patient is very young, but hip arthroplasty remains the gold
standard treatment for most displaced fractures in older patients.[3 ] Among the arthroplasty options, hemiarthroplasty (HA) yields a higher reoperation
rate but lower postoperative dislocation rate compared with total hip arthroplasty
(THA).[4 ]
[5 ]
[6 ] As treatment trends and guidelines continue to change and tend to favor THA over
HA for more active patients, HA remains a viable and frequently used treatment option
especially for the most elderly patients with lower functional demands. However, the
choice between unipolar and bipolar HA implants remains controversial.[7 ]
While some studies have shown no differences in hip function, complications, and dislocation
risk[3 ]
[8 ]
[9 ] between unipolar and bipolar HA, bipolar HA is associated with better health-related
quality of life (HRQoL) and later onset of acetabular erosion compared with unipolar
HA.[10 ] However, the rare complication of intraprosthetic dissociation (IPD) continues to
occur in bipolar HA, frequently during dislocation or reduction of the dislocated
prosthetic hip. This complication has previously manifested in a variety of ways,
for example, as dislodgement of the femoral stem[11 ]
[12 ] or dissociation of the two articulating surfaces of the femoral ball head and polyethylene
(PE) lined cup.[13 ]
[14 ]
[15 ] IPD requires conversion from closed to open reduction to repair or replace the prosthesis
itself before reducing the hip dislocation, which may contribute to the increased
risk of open reduction in bipolar HA.[16 ]
A recent study has shown an IPD rate of 13% during manual reduction of dislocated
bipolar HA,[14 ] prompting the need for further study of this phenomenon. We herein describe a case
of bipolar prosthetic dissociation during manual reduction of a dislocated HA. We
also review the literature to discuss the consequences of this complication, including
risks with revision surgery and potential loosening of the prosthetic's mobile components.
Methods
We reviewed the PubMed database for the following keywords: “bipolar,” “hemiarthroplasty,”
“component,” “disassembly,” “dissociation,” and “dislocation.” Our inclusion criteria
included the publication formats of case reports, case series, observational studies,
and randomized trials. We also required publications to be available in English. Our
search yielded 18 case reports/series[11 ]
[12 ]
[13 ]
[15 ]
[17 ]
[18 ]
[19 ]
[20 ]
[21 ]
[22 ]
[23 ]
[24 ]
[25 ]
[26 ]
[27 ]
[28 ]
[29 ]
[30 ] collectively presenting 45 cases of IPD among bipolar HA patients, as well as 4
observational studies[14 ]
[31 ]
[32 ]
[33 ] presenting 27 more IPD cases. From each publication included in this literature
review, we categorized the described IPD cases both by the implant component involved
in the dissociation and by etiology ([Table 1 ]).
Table 1
Findings of the reviewed sources on cases of intraprosthetic dissociation
Author (publication year)
Study type
Patient sample (N )
Cases of IPD (N )
Cases per implant component involved (N )
Cases per etiology (N )
Inner head or outer head/cup
PE liner or locking ring
Femoral stem/ neck
Spontaneous
Trauma: Dislocation
Trauma: Other
Iatrogenic: Closed reduction
Unknown
Marcelino Gomes et al (2011)[11 ]
CS
1
1
1
1
Yun et al (2010)[12 ]
CS
1
1
1
1
Loubignac and Boissier (1997)[13 ]
CS
2
2
2
2
Georgiou et al (2006)[15 ]
CS
5
5
4
1
2
1
2
Barmada and Mess (1987)[17 ]
CS
2
3
3
3
Calton et al (1998)[18 ]
CS
5
5
5
5
Corteel and Putz (1996)[19 ]
CS
1
1
1
1
Guo et al (2008)[20 ]
CS
1
1
1
1
Sevinç (2021)[21 ]
CS
2
2
2
1
1
Kim (1986)[22 ]
CS
1
1
1
1
Lee et al (2008)[23 ]
CS
1
1
1
1
Tanaka et al (2002)[24 ]
CS
1
2
2
1
1
Shiga et al (2010)[25 ]
CS
1
1
1
1
Uruç et al (2017)[26 ]
CS
5
5
5
3
2
Moores et al (2013)[27 ]
CS
1
1
1
1
Star et al (1992)[28 ]
CS
3
3
1
1
1
3
Tabutin and Damotte (2004)[29 ]
CS
4
4
4
4
Herzenberg et al (1988)[30 ]
CS
3
6
6
5
1
Lee et al (2018)[14 ]
RO
55
7
7
7
Hasegawa et al (2004)[31 ]
RO
61
7
7
7
Bhuller (1982)[32 ]
RO
33
2
2
2
Figved et al (2006)[33 ]
RO
350
11
11
9
1
1
Total cases,
N (%)
72
44 (61.1)
26 (36.1)
2 (2.8)
35 (48.6)
13 (18.1)
4 (5.6)
19 (26.4)
1 (1.4)
Abbreviations: CS, case study or series; IPD, intraprosthetic dissociation; PE, polyethylene;
RO, retrospective observational study.
We encountered a case of IPD at our own institution in April 2021. A 76-year-old male
patient with a history of Parkinson's presented to our institution with a displaced
right femoral neck fracture after a fall. He was treated with cemented bipolar HA
using a posterolateral approach ([Figs. 1 ] and [2 ]) with proper restoration of leg length and offset. Two months postoperatively, this
patient flexed the hip upon standing from a low seat, and his right hip dislocated
posteriorly ([Fig. 3 ]). A closed reduction with propofol sedation in the emergency department setting
was attempted, which resulted in a reduction clunk and improvement in the leg's clinical
length and alignment, but further imaging showed IPD of the bipolar prosthesis ([Fig. 4 ]). The specific form of IPD in this patient's prosthesis was dissociation of the
outer femoral head from the inner head caused by closed reduction maneuvers and levering
of the outer shell against the pelvis during traction maneuvers. This required subsequent
open reduction and revision of the femoral head component. At the time of open reduction,
the femoral head was revised to a one-size-longer unipolar prosthesis ([Fig. 5 ]) and the patient has experienced no further dislocation events at 6-month follow-up.
The patient has provided their informed consent for their relevant medical information
to be published in this journal article.
Fig. 1 Immediately postoperative anteroposterior radiograph showing right bipolar hemiarthroplasty
implant with pelvic view.
Fig. 2 Radiographs postoperative 1.5 months showing right bipolar hemiarthroplasty implant
in correct position using anteroposterior views of the (A ) pelvis and (B ) lateral view of the right hip.
Fig. 3 Anteroposterior radiograph showing dislocation of the right bipolar hemiarthroplasty
implant.
Fig. 4 Anteroposterior radiograph showing intraprosthetic dissociation of the right bipolar
hemiarthroplasty implant following attempted closed reduction maneuvers. Dissociation
occurred between the outer and inner heads of the prosthesis.
Fig. 5 Immediately postoperative anteroposterior radiograph of the pelvis, showing right
femoral head revised with a unipolar hemiarthroplasty implant and the prior dislocation
reduced.
Results
IPD refers to the dissociation or misalignment of components within a bipolar HA prosthesis
that can arise from prosthetic dysfunction or damage, and is typically associated
with manual reduction after a dislocation event. From a review of the literature,
we have found that this is a broad and diverse group of complications. IPD can arise
spontaneously,[26 ] with trauma such as prosthesis dislocation,[11 ]
[13 ]
[15 ]
[23 ]
[24 ] and during reduction maneuvers for a dislocated prosthesis.[12 ] IPD has also been found to occur after erosion or fracture of the PE liner.[18 ]
[31 ] Dissociation can be further categorized according to the specific component that
has dissociated from its proper position, including the outer head, PE liner, locking
ring, and femoral stem. For instance, one research group has identified three different
subtypes of locking mechanism failure alone: detached ring, dissociation of inner
and outer heads, and the cooccurrence of both.[31 ]
We found that the majority (61.1%) of cases involved the inner head or outer head/cup
of the implant, while the PE liner or locking ring was involved in 36.1% of cases.
In contrast, the femoral stem or neck of the implant was rarely involved in IPD events.
The predominating etiology for IPD was spontaneous (48.6%), in which there was no
identifiable preceding trauma on the hip joint. Several IPD cases occurred due to
trauma, of which the majority was specifically hip dislocation (18.1%) when the outer
head/cup of the prosthesis was displaced from the acetabulum, while another quarter
of cases occurred iatrogenically as a result of closed reduction maneuvers (26.4%).
Discussion
Despite general consensus on choosing HA over THA in older patients with lower ambulatory
demand, the choice between unipolar and bipolar HA remains controversial. A 2005 international
survey of operative management of femoral neck fractures showed that surgeons had
varied preferences in their choice of arthroplasty for older patients with Garden
type IV fractures, with 32% preferring unipolar HA and 41% preferring bipolar.[7 ] Managing a patient using HA involves weighing a variety of implant options, including
cemented and uncemented femoral stems, fixed-neck and modular-neck, and unipolar and
bipolar femoral head designs.[34 ] With each category of implant options, several factors about the case such as the
patient's bone health, age, potential comorbidities, surgeon preferences and familiarity,
patient priorities when weighing quality of life against risk of revision, and implant
cost must all be reviewed in making a final decision on HA implant design. While surgeons
performing THA, as opposed to HA, may consider the risk of IPD in choosing a dual-mobility
implant due to the more robust body of work on this prosthetic's risk,[35 ] we reinforce that this phenomenon also has the potential to occur in HA when selecting
a bipolar implant and must be similarly considered as a potential risk to the patient
by treating physicians.
IPD of dual mobility implants in THA has been documented in the literature as early
as 2007,[36 ] nearly three decades after the first implant was introduced in Europe in 1979.[37 ] Some aspects of IPD in dual mobility THA implants are similar to those in bipolar
HA implants. Both implants have a femoral component that consists of a femoral stem
or neck with a small-diameter head that articulates within a PE outer head or liner,
and the outer head is then involved in a large-diameter articulation with the acetabulum.[35 ] With such similarities in the dual-articulation design, IPD of both implants frequently
manifests as a separation of the smaller femoral head from the outer head,[38 ] at times with migration of the outer head from the acetabular space as in our case.
However, IPD of bipolar implants has demonstrated its difference from dual mobility
IPD by also involving intraprosthetic interfaces other than the two articulating heads,
such as the femoral stem[25 ]
[28 ] and the locking ring within the PE liner.[31 ] IPD cases in dual mobility implants may occur due to mechanical loosening from wear
of the PE liner,[38 ] typically in late IPD, or from traumatic causes such as closed reduction attempts
in early IPD[35 ] which involve the “bottle-opener” mechanism.[13 ] While IPD in bipolar HA has occurred from these two causes ([Table 1 ]), it has also occurred spontaneously with no erosion of the PE liner.[26 ] The unique etiologies and components involved in IPD of bipolar implants highlight
the need for further study of IPD in HA as it has been widely studied in dual mobility
THA.
Close to half of reported IPD in bipolar HA has occurred due to hip dislocation or
during reduction maneuvers (44.5%, [Table 1 ]), which indicates that reducing the frequency of dislocation may be a critical step
toward decreasing the risk of IPD. There has been extensive study of hip HA surgical
approaches and their impact on postoperative dislocation rates. Surgeons commonly
use three surgical approaches in hip HA[39 ]: anterior approach (AA), in which access to the anterior joint capsule is obtained
using the plane between the sartorius, rectus femoris, and tensor fascia lata[40 ]
[41 ]
[42 ]; lateral approach (LA), in which the capsule is accessed by movement of the gluteus
medius insertion[43 ]
[44 ]; and posterior approach (PA), in which the gluteus maximus muscle is divided along
its fibers for posterior access to the joint.[45 ]
[46 ] The posterolateral approach (a type of PA) has been associated with a higher risk
of dislocation than the anterolateral approach (considered either AA or LA), even
when performing posterior repair to reattach the short external rotators and posterior
joint capsule.[47 ] Other studies and meta-analyses have had similar findings, with the PA found to
be associated with higher risk of dislocation,[48 ]
[49 ]
[50 ] higher risk of reoperation due to dislocation,[51 ] and higher risk of both dislocation specifically and reoperation in general[39 ] compared with other surgical approaches, particularly the anterolateral and direct
AAs. The PA may have advantages over other approaches such as decreased operative
time[49 ]; however, other proposed advantages such as decreased wound infections, less pain,
and better quality of life[52 ] have been less consistently reported.[39 ]
[50 ]
[53 ] With the PA introducing consistently higher rates of dislocation, we recommend surgeons
consider performing hip HA using the AA or LA to reduce postoperative dislocations,
and thus reduce the likelihood of IPD.
A significant proportion of IPD has occurred due to failure of the implant's locking
mechanism, consisting of a PE ring and, in some manufacturer's models, a metal locking
ring (36.1%, [Table 1 ]). Bipolar implant locking mechanisms may be divided into single and dual locking
categories, both within the classic design of a metallic bipolar shell/outer cup containing
a liner or insert of PE. In a single lock design, a PE ring is placed around the smaller-diameter
neck of the femoral head after it has been inserted into the outer shell ([Fig. 6A ]); this PE ring replaced the locking leaflets of older prosthetics such as the Bateman
UPF-I.[30 ] The PE ring then expands to fit into a matching groove near the peripheral rim of
the PE liner which serves to lock the femoral head within the shell during hip articulation.
In a dual locking mechanism, there is an additional metal locking ring anchored in
the outer surface of the PE liner that enhances its close fit within the metal shell
([Fig. 6B ]). Prosthetics such as the Bencox Bipolar Cup (Corentec, Cheonan, South Korea), Bencox
Bipolar Head (Aesculap, Tuttlingen, Germany), and Bencox Self-Centering Cup (DePuy,
Warsaw, IN) are equipped with a single locking mechanism, while others like the Multipolar
Cup and Modular Bipolar System (Zimmer, Warsaw, IN) and RINGLOC Cup (Biomet, Warsaw,
IN) have a dual locking mechanism.[14 ] In many cases of IPD, failure of the locking mechanism has largely been attributed
to macroscopic wear of the PE ring. The PE ring may degrade due to persistent impingement
of the femoral neck on the ring from wider oscillations of the hip,[31 ] for example, with deep flexion and extension, or following natural wear of a thinner
PE liner[18 ] that permits looser articulation of the femoral head within the outer cup construct.
As the PE ring degrades from the peripheral rim, there is further opportunity for
dislodgement of the metal locking ring in dual locking implants, with or without visible
deformity in the metal ring.[31 ] With PE wear comes the dispersal of debris within the acetabular space, which has
been observed alongside acetabular osteolysis[18 ] leading to implant failure.
Fig. 6 Manufacturer images of bipolar implant constructs, depicting (A ) single locking mechanism (Bencox Bipolar Cup, Corentec[54 ]) and (B ) dual locking mechanism with metal ring (Modular Bipolar System®, Zimmer[55 ]).
Based on these proposed etiologies for PE ring degradation, surgeons may mitigate
this process by cautioning their patients about wide hip oscillation movements and
selecting thicker PE liners as a part of their implant construct. Preliminary research
has also suggested that implants with a single locking mechanism are associated with
increased risk of IPD during closed reduction compared with those with a dual locking
mechanism (7 IPD cases out of 55 hips, 6 with single locking mechanism, p = 0.04).[14 ] While further studies are needed to control for potential confounding factors, evaluate
the biomechanics of locking mechanism failure, and compare failure rates between implant
manufacturers, the existing data suggest that surgeons may decrease the rate of locking
mechanism failure and thus decrease IPD risk by selecting dual locking implants.
Nearly, if not all, IPD cases require open surgery to correct the complication. This
potential intervention is risky for bipolar HA patients precisely because they are
typically preselected for being the oldest, least active adults with femoral neck
fractures, and thus are less tolerant of surgical procedures in general. The risk
of surgical intervention for IPD partially negates the potential benefit of slower
acetabular wear with bipolar HA prostheses (compared with unipolar), and surgeons
must weigh both factors in selecting the best prosthetic for a patient. In cases of
outer head dissociation, correction via intraoperative reassembly of the dissociated
component with the rest of the prosthesis may introduce an increased risk of future
IPD by using the same components for repair. For example, outer head dissociation
may indicate a degree of wear on the PE liner that is not clearly apparent on radiographic
imaging, but may predispose the same prosthetic components to dissociate again. Because
of this, the authors recommend replacing the entire femoral head construct when possible.
Conclusion
In conclusion, the risk of IPD must be considered when deciding whether a unipolar
or bipolar HA implant is the better treatment for a patient's femoral neck fracture.
The potential advantages of bipolar compared with unipolar HA, such as slower acetabular
erosion and better HRQoL, have been inconsistent across the literature. Other advantages
of bipolar HA are manufacturer-specific, in that some device companies offer more
customization options with bipolar versus unipolar in terms of head/neck length and
head circumference. Factors like the potentially higher cost and the complication
risk of IPD remain consistent disadvantages that are exclusive to bipolar HA. When
performing hip HA, surgeons may want to consider choosing unipolar prosthetics over
bipolar to reduce future operative risks to the patient related to IPD, and they should
be aware of this potential complication when choosing implants and especially when
performing closed reduction maneuvers on dislocated bipolar HA implants.