Subtrocar for Removing a Gallbladder from the Abdominal Cavity

S. Kakizoe 1 , Y. Kakizoe 1 , S. Kase 1 , M. Ishida 1 , H. Kakizoe 2 , T. Kakizoe 2 , K. Kakizoe 1
  • 1 Dept. of Surgery, Ilikai Medical INC. Kakizoe Hospital, Hirado, Nagasaki, Japan
  • 2 Dept. of Medicine, Ilikai Medical INC. Kakizoe Hospital, Hirado, Nagasaki, Japan
Further Information

S. Kakizoe, M.D., Ph.D. 

Dept. of Surgery
Kakizoe Hospital

Kagamigawa 278
Hirado
Nagasaki 859-5152
Japan

Fax: Fax:+ 81-950-225131

Email: E-mail:kakizoes@nagasaki.med.or.jp

Publication History

Publication Date:
31 December 2001 (online)

Table of Contents

Recent advances in laparoscopic cholecystectomy have enabled us to perform the procedure using three 3 mm or 5 mm trocars for operative instruments and a 10 mm trocar for the laparoscope [1] [2] . A small-diameter laparoscope is necessary to perform the procedure when removing a gallbladder from the abdominal cavity. Small-diameter laparoscopes, however, are expensive, and still give poor imaging [3]. We have developed a subtrocar which enabled us to extract the gallbladder without using a small-diameter laparoscope.

As shown in Figure [1], the tip of a subtrocar is designed to allow it to be inserted alongside a 10 mm reusable trocar without damaging the surrounding tissue. Thereafter, 5 mm grasping forceps are inserted through the subtrocar and used to grasp the gallbladder. The gallbladder is observed using the laparoscope (Figure [2]), and is pulled out from the abdominal cavity using the subtrocar and the 10 mm trocar used for the laparoscope.

Zoom Image

Figure 1The subtrocar is made of stainless steel and is 200 mm in length and 7 mm in diameter. An ordinary 5 mm forceps can be inserted. The tip of the subtrocar is designed as shown in the inset and is shaped around the 10 mm reusable trocar to reduce the size when it is inserted into the abdominal cavity. a The subtrocar is inserted alongside the 10-mm reusable trocar at the navel port site. b The navel site wound is bluntly dilated with the subtrocar and the subtrocar is positioned beside the 10 mm trocar. The air leak from the navel site wound is minimal

Zoom Image

Figure 2Extraction of the gallbladder from the abdominal cavity. A 5 mm forceps is inserted through the subtrocar (*) and used to grasp the gallbladder. The view can be obtained when only the subtrocar is used

We performed the procedure in 17 patients between 31 August 1998 and 9 August 2000. In one patient, the gallbladder slipped off from the grasping forceps during removal from the abdominal cavity, and the patient showed mild subcutaneous bleeding at the operative wound. Another two patients also had evidence of mild subcutaneous bleeding, which resolved without special treatment. In the other patients, the gallbladder was removed safely without any complications. We conclude that the subtrocar is a useful instrument for performing laparoscopic cholecystectomy with small-size trocars.

#

References

  • 1 Kimura T, Sakuramachi S, Yoshida M, et al.. Laparoscopic cholecystectomy using fine-caliber instruments.  Surg Endosc. 1998;  12 283-286
  • 2 Yuan R H, Lee W J, Yu S C.. Mini-laparoscopic cholecystectomy: a cosmetically better, almost scarless procedure.  J Laparoendosc. Adv Surg Tech. 1997;  7 205-211
  • 3 Berci G.. Laparoscopic cholecystectomy using fine-caliber instruments; smaller is not necessarily better.  Surg Endosc. 1998;  12 197

S. Kakizoe, M.D., Ph.D. 

Dept. of Surgery
Kakizoe Hospital

Kagamigawa 278
Hirado
Nagasaki 859-5152
Japan

Fax: Fax:+ 81-950-225131

Email: E-mail:kakizoes@nagasaki.med.or.jp

#

References

  • 1 Kimura T, Sakuramachi S, Yoshida M, et al.. Laparoscopic cholecystectomy using fine-caliber instruments.  Surg Endosc. 1998;  12 283-286
  • 2 Yuan R H, Lee W J, Yu S C.. Mini-laparoscopic cholecystectomy: a cosmetically better, almost scarless procedure.  J Laparoendosc. Adv Surg Tech. 1997;  7 205-211
  • 3 Berci G.. Laparoscopic cholecystectomy using fine-caliber instruments; smaller is not necessarily better.  Surg Endosc. 1998;  12 197

S. Kakizoe, M.D., Ph.D. 

Dept. of Surgery
Kakizoe Hospital

Kagamigawa 278
Hirado
Nagasaki 859-5152
Japan

Fax: Fax:+ 81-950-225131

Email: E-mail:kakizoes@nagasaki.med.or.jp

Zoom Image

Figure 1The subtrocar is made of stainless steel and is 200 mm in length and 7 mm in diameter. An ordinary 5 mm forceps can be inserted. The tip of the subtrocar is designed as shown in the inset and is shaped around the 10 mm reusable trocar to reduce the size when it is inserted into the abdominal cavity. a The subtrocar is inserted alongside the 10-mm reusable trocar at the navel port site. b The navel site wound is bluntly dilated with the subtrocar and the subtrocar is positioned beside the 10 mm trocar. The air leak from the navel site wound is minimal

Zoom Image

Figure 2Extraction of the gallbladder from the abdominal cavity. A 5 mm forceps is inserted through the subtrocar (*) and used to grasp the gallbladder. The view can be obtained when only the subtrocar is used