Digitale Mammographie mit hochauflösenden Speicherfolien (CR) versus digitaler Vollfeldmammographie (CCD) (DR): Mikrokalzifikationen und Herdbefunde - eine retrospektiv klinisch-histologische Analyse (n = 102)

Digital Mammography with High-resolution Storage Plates (CR) versus Full-field Digital Mammography (CCD) (DR) for Microcalcifications and Focal Lesions - A Retrospective Clinical Histologic Analysis (n = 102)R. Schulz-Wendtland1 , M. Lell1 , E. Wenkel1 , C. Böhner1 , M.-S Dassel1 , W. Bautz1
  • 1Radiologisches Institut (Direktor: Prof. Dr. med. W. Bautz), Universität Erlangen-Nürnberg
Further Information

Publication History

Publication Date:
09 November 2004 (online)

Preview

Zusammenfassung

Ziel: Bestimmung der diagnostischen Genauigkeit von digitaler Mammographie in hochauflösender Speicherfolientechnik (Hardcopy) und digitaler Vollfeldmammographie (CCD) bei Mikrokalzifikationen und Herdebefunden (n = 102) in einer retrospektiv klinisch-histologischen Studie. Material und Methode: Von 05 - 09/2003 untersuchten wir 102 Patientinnen mit digitaler Mammographie und Speicherfolientechnik. Nach Diagnosestellung und präoperativer Drahtmarkierung erfolgte eine digitale Vollfeldmammographie bei gleichen Aufnahmeparametern. 5 Untersucher stellten retrospektiv nach der Operation anhand der zufallsverteilten Hardcopies (jeweils mediolaterale Ebene) die Diagnosen (BI-RADS™ I-V), welche mit der endgültigen Histologie korreliert wurden. Ergebnisse: Mit der digitalen Mammographie in Speicherfolientechnik und der digitalen Vollfeldmammographie (CCD) wurden insgesamt (n = 102) diagnostische Genauigkeiten von 73 und 71 %, für Mikrokalzifikationen (n = 51) 73 bzw. 71 % sowie für Herdbefunde (n = 51) 72 und 70 % erzielt. Insgesamt zeigten die Ergebnisse keine Unterschiede. Schlussfolgerung: Auf der Basis unserer Untersuchung ist von einer Gleichwertigkeit von digitaler Mammographie in (hochauflösender) Speicherfolientechnik und digitaler Vollfeldmammographie (CCD) auszugehen.

Abstract

Purpose: To determine the diagnostic accuracy of microcalcifications and focal lesions in a retrospective clinical-histological study using high-resolution digital phosphor storage plates (hard copy) and full-field digital mammography (hard copy). Materials and Methods: From May 2003 to September 2003, 102 patients underwent digital storage plate mammography (CR), using a mammography unit (Mammomat 3000 N, Siemens) in combination with a high resolution (9 lp/mm) digital storage phosphor plate system (pixel size 50 μm) (Fuji/Siemens). After diagnosis and preoperative wire localization, full-field digital mammography (CCD) (DR) was performed with the same exposure parameters. The full-field digital mammography used a CCD-detector (SenoScan) (Fisher Imaging) with a resolution of 10 Ip/mm and a pixel size of 50 μm. Five investigators determined the diagnosis (BI-RADS™ I - V) retrospectively after the operation from randomly distributed mediolateral views (hard copy reading). These results were correlated with the final histology. Results: The diagnostic accuracy of digital storage plate mammography (CR) and full-field digital mammography (CCD) (DR) was 73 % and 71 % for all findings (n = 102), 73 % and 71 % for microcalcifications (n = 51), and 72 % and 70 % for focal lesions (n = 51). The overall results showed no difference. Conclusion: Our findings indicate the equivalence of high-resolution digital phosphor storage plate mammography (CR) and full-field digital mammography (CCD) (DR).