Keywords Adaptive statistic iterative reconstruction - automatic tube current modulation -
computed tomography optimization
Introduction
Computed tomography (CT) examinations are now common medical procedures, as the scans
facilitate rapid and more accurate diagnosis. Radiation-induced carcinogenesis is
a stochastic effect, whose probability increases with radiation dose. Several models
of radiation-induced cancer have been reported in the literature.[[1 ]] However, the dramatic rise in the use of CT has been responsible for the increased
medical exposure to ionizing radiation,[[2 ]] so there is a need to ensure that radiation protection methods for CT procedures
are optimized in order to keep dose levels to a minimum.
CT scans of the chest are one of the more common examinations used in the diagnosis
of chest disorders, such as infection, inflammation, and cancer, and are also being
considered in screening for lung cancer. A recommendation has been made that this
application of CT should be justified for dose levels with the volume-weighted CT
dose index (CTDIvol ) <3.0 mGy for standard size patients.[[3 ]],[[4 ]] Effective dose is the parameter related to the risk from exposure to ionizing radiation.
For CT, it can be calculated from multiplication of the dose length product (DLP)
by a DLP to effective dose conversion factor. DLP is based on the product of CTDIvol and irradiation length for the scan. There have been some studies that demonstrate
the possibility of reducing radiation exposure using low-dose CT protocols, to give
effective dose in the dose range of chest radiography.[[5 ]] These doses are now in the realm that could enable justification for using low-dose
CT instead chest radiography for some applications. A survey in Thailand has reported
that the average values of CTDIvol and effective dose for chest CT are about 10 mGy and 5 mSv, respectively.[[6 ]] For CT scans to be used for such purposes, a level of image quality must be maintained
to ensure that the scans provide the required diagnostic information.
To optimize radiation dose in CT, various technological strategies can be applied.
Modern CT scanners incorporate automatic tube current modulation (ATCM) systems to
reduce patient doses. These are designed to aid in maintaining optimized radiation
dose and consistent image quality throughout CT scans and for patients of varying
size.[[7 ]],[[8 ]],[[9 ]],[[10 ]] In principle, the tube current is adjusted automatically to that appropriate for
the X-ray attenuation of the patient cross-section being scanned. Current scanners
also incorporate iterative reconstruction (IR), a CT reconstruction algorithm that
is able to give a lower image noise level when compared to the filtered back projection
(FBP) reconstruction method.[[11 ]],[[12 ]],[[13 ]]
Use of a combination of ATCM and IR has the potential to further reduce the dose while
maintaining image quality. In the present study, a general electric (GE) system with
a Smart mA ATCM and adaptive statistic iterative reconstruction (ASIR) was used. The
goals of this study were to investigate the use of IR together with the ATCM to determine
how the two can best be applied together, and to establish a new protocol implementing
Smart mA and ASIR that can be used as a CT low-dose protocol providing an acceptable
level of image quality similar to that for the routine protocol used in the hospital.
Methods
Ethical approval (No. 0348/60) was obtained from the Institutional Ethics Committee.
Because the qualitative image quality assessment was performed by radiological technologists,
informed consent was obtained from human participants. Experiments were carried out
on an anthropomorphic phantom and a Catphan image quality phantom. An Alderson chest
phantom (RSD model RS330) was used for this study. It extends from the neck to below
the diaphragm, and is molded around a male skeleton, corresponding to a patient that
is 175 cm tall and 73.5 kg in weight. The materials are equivalent to natural bone
and soft tissues. Lungs are fixed in the inflated state and are molded to conform
to the pleural cavities of the phantom. The pulmonary arteries are injected with a
blood equivalent plastic.
Measurements were performed using a 128-slice CT scanner from GE Healthcare (Optima
660). In the present study, Smart mA which is x–y plane and z -axis modulation was used.[[14 ]] The standard protocol used by the hospital is a NI of 11.57 HU, 120 kVp, 0.8 s
rotation time, 1.375 pitch factor, 40 mm beam collimation, 5 mm slice thickness, and
standard filter. Image data set was reconstructed using 30% ASIR. The scanned region
covered the entire lung fields. The range of the tube current was between 10 and 350
mA. Scans were performed with the reference default NI of 11.57 and other NI settings
of 13.3, 15.04, 16.78, 18.52, 20.26, and 22 (NI was increased by 15 for each setting).
Each image data set was reconstructed using three different ASIR levels of 30, 50,
and 70%. Details of the standard chest protocol used by the hospital and the other
protocols developed are summarized in [[Table 1 ]].
Table 1
CT protocols and dose data
Protocol
Dose step
NI
ASIR level (%)
CTDIvol (mGy)
DLP (mGy cm)
Effective dose (mSv)
NI=Noise index, ASIR=Adaptive statistic iterative reconstruction, CTDIvol =Volume weighted computed tomography dose index, DLP=Dose length product
Standard
0
11.57
30
5.2
211
3.0
1
50
2
70
3
−3
13.3
30
3.9
160
2.2
4
50
5
70
6
−6
15.04
30
3.1
125
1.8
7
50
8
70
9
−9
16.78
30
2.5
100
1.4
10
50
11
70
12
−12
18.52
30
2.0
82
1.2
13
50
14
70
15
−15
20.26
30
1.7
69
1.0
16
50
17
70
18
−18
22
30
1.4
59
0.8
19
50
20
70
For every scan, CTDIvol in mGy and DLP in mGy cm were recorded. Values for effective dose were then calculated
as effective dose = DLP × EDLP , where EDLP is a DLP to effective dose conversion factor with a value for the chest of 0.015
mSv/mGy cm.[[15 ]],[[16 ]] The quantitative analysis of the images in terms of pixel value (signal) and standard
deviation of the pixel value (noise) were carried out using Image J.[[17 ]] The pixel value and standard deviation of the pixel value from images recorded
near the middle level of the scan for slice numbers 23-25, 28-33, and 36 were measured.
Ten regions of interest (ROIs) each 5 mm in diameter were placed within the lung and
soft tissue for each slice (three reconstructions for each acquisition protocol),
as shown in [[Figure 1 ]]. Brightness was adjusted for viewing by the radiological technologist, who was
asked to position the ROI for one image series, and the coordinates of individual
ROIs were recorded and the same positions used for all image series.
Figure 1: CT images of the chest phantom at the middle of the heart obtained from standard
protocol and ROI locations
The image quality was then evaluated by calculating the signal-to-noise ratios (SNRs)
in the lungs and soft tissue, and contrast-to-noise ratios (CNR) for the lung and
soft tissue as follows:
where σ corresponds to the standard deviation of pixel value in the ROIs.
The Statistical Package for the Social Sciences (SPSS) statistical package version
23 (SPSS Inc., Chicago, IL, USA) was used for statistical analysis. Analysis of variance
(ANOVA) was used to assess the differences in SNR and CNR between the group means,
at α = 0.05. If significant differences were found, post-hoc analysis would be used by dependent t -test to test the difference between protocols at α = 0.01 (Bonferroni adjustment
for five comparisons). Protocols having similar (not statistically different) or higher
values of SNR and CNR compared to the standard protocol were considered and were then
further evaluated by subjective image quality assessment.
For qualitative image quality analysis, images were evaluated by five observers each
with more than 7 years’ experience of interpreting CT images, and blinded to the CT
settings. Image quality was assessed using the following criteria: image noise, contrast,
sharpness and image artifact. Images for 5 mm slice thicknesses were assessed with
fixed windowing of lung images at a window width of 1500 HU and window level of −600
HU, and mediastinum images at a window width of 350 HU and window level of 50 HU.
The evaluation form was based on a five-level Likert scale (1–5); higher scores indicating
better image quality. Intra-class correlation (ICC) was calculated for evaluation
of the correlation between observers and the nonparametric Friedman test with significance
level at P < 0.01 was used to evaluate whether the subjective image quality for the newly developed
protocols differed from that for the standard protocol.
Image quality assessment was carried out using images acquired from a Catphan 700
phantom (The Phantom Laboratory, Inc., Salem, NY).[[18 ]] The scan parameters were 120 kV, field of view 240 mm, and slice thickness of 5
mm. Other scan settings were similar to those used for the anthropomorphic phantom.
As the Catphan is cylindrical in shape, it does not reflect mA modulation. Therefore,
the mA was selected to achieve values for the CTDIvol of 1.4, 1.7, 2.0, 2.5, 3.1, 3.9 and 5.2 mGy, according to the CTDIvol result from the chest phantom. Measurements of CT number from various materials such
as air, lung foam, polystyrene, 20 and 50% hydroxyapatite (CTP682), as well as low-contrast
detectability (LCD) (CTP 515), and spatial resolution at 50 and 10% levels of the
modulation transfer function (MTF) from 50 μm Teflon wire (CTP682) were performed.
The CNR was calculated using the contrast between the values of signal measured at
1% contrast of 15 mm in diameter at CTP515 and the signal measured at the background
next to the contrast object, and divided by the noise which was the mean of the standard
deviation of the three ROIs located on the background,[[19 ]] as shown in Equation (3).
Images were reconstructed using the standard kernel for FBP and 30, 50 and 70 ASIR.
Web-based software provided by Image owl ® was used for the analysis.[[20 ]]
Results
Dose and quantitative image quality assessment from anthropomorphic phantom
Changing the NI settings from 11.57 through to 22 resulted in reductions in DLP from
25 to 72% [[Table 1 ]], corresponding to the tube current reductions shown in [[Figure 2 ]]. Results for the SNR measured in the lung and soft tissue and CNR between the lung
and soft tissue are shown in [[Table 2 ]]. SNRsoft tissue, SNRlung , and CNR for the standard protocol were 3.3, 90.7, and 137.1, respectively. [[Figure 3 ]] and [[Table 2 ]] show the SNR and CNR values for each protocol for individual NI settings. The SNR
and CNR increased with the strength of ASIR, and decreased with higher NI settings
[[Figure 3 ]] and [[Table 2 ]]. Statistical analysis showed that values for the SNR and CNR for protocol numbers
1, 2, 4, 5, and 8 were similar to or higher than those of the standard protocol.
Table 2
Quantitative image analysis for different CT protocols
Protocol
NI
ASIR level (%)
SNR soft tissue
SNR lung
CNR
ASIR=Adaptive statistic iterative reconstruction, SNR=Signal-to-noise ratio, CNR=Contrast-to-noise
ratio
Standard
11.57
30
3.3
90.7
137.1
1
50
3.8
99.1
158.1
2
70
4.4
108.5
185.1
3
13.3
30
2.8
82.7
114.3
4
50
3.2
92.7
130.9
5
70
3.6
99.5
151.6
6
15.04
30
2.6
77.0
108.8
7
50
3.0
84.9
126.2
8
70
3.5
93.7
148.3
9
16.78
30
2.3
70.3
96.5
10
50
2.7
77.4
111.4
11
70
3.1
85.4
130.4
12
18.52
30
2.1
67.2
88.9
13
50
2.5
74.6
103.3
14
70
2.9
83.3
122.2
15
20.26
30
2.0
63.7
80.1
16
50
2.3
70.5
93.4
17
70
2.7
78.2
110.7
18
22
30
1.8
58.1
75.6
19
50
2.1
64.5
87.9
20
70
2.5
71.9
103.9
Figure 2: Tube current (mA) modulations for individual protocols
Figure 3 (A-C): SNR measured at the lungs (A) and soft tissue (B) and CNR (C) for individual protocols
Qualitative image assessment
Qualitative scoring of image noise, contrast, sharpness, and image artifact from the
five readers are shown in [[Table 3 ]], scores ranged between 3.0 and 4.6. A moderate degree of reliability was found
between scoring. The average measure of ICC was 0.52 with a 95% confidence interval
from 0.205 to 0.744 (P < 0.001). For the acceptability of the image noise level, all developed protocols
received similar or higher scores compared to the standard protocol, while those for
the contrast and sharpness received lower scores (from scores of 4–5 for each reader,
to scores of 2–4). However, the difference in the mean image quality score in each
category of image noise, contrast, sharpness, and image artifact was not statistically
significant.
Table 3
Overall scores of subjective image quality analysis
Protocol
Average qualitative image quality score (n =5)
Image noise
Sharpness
Contrast
Image artifact
P>0.01
Standard
3.20±0.84
4.20±0.84
4.60±0.55
3.60±0.55
1
3.40±0.55
3.80±0.45
4.00±0.00
3.20±0.45
2
3.60±0.55
3.00±0.71
3.20±0.84
3.20±0.45
4
3.80±0.84
3.60±0.55
3.60±0.55
4.40±0.55
5
3.20±0.45
3.20±0.84
3.80±0.45
4.00±0.71
8
3.40±0.55
3.80±0.45
3.80±0.45
3.40±0.55
Image quality assessment from Catphan
Values for CT number measured from air, lung foam, polystyrene, 20 and 50% hydroxyapatite
using the standard protocol were 947, −776, −33, 236, and 663 HU, respectively. CT
numbers varied slightly but were within ±2 HU for each material and reconstruction
method. CNR and noise analysis with the Catphan showed improvements when using ASIR
compared to the FBP [[Figure 4 ]]A and [[Figure 4 ]]B. CNR was 16–30, 30–57, and 47–89% higher, for higher dose settings, for 30, 50,
and 70% ASIR, respectively, and the noise was reduced by up to 5 HU when using 70%
ASIR.
Figure 4 (A and B): CNR (A) and noise (B) for different reconstruction methods and dose levels
Results for high-contrast spatial resolution are illustrated by the MTF curves for
the different reconstruction methods. At a CTDIvol of 1.4 mGy, values for the 50% level were 3.8, 3.9, 4.0, and 4.1 lp cm−1 , while the number of lp cm−1 increased to 6.7, 6.8, 6.9, and 7.0 at 10% MTF for FBP and 30, 50, and 70% ASIR,
respectively. The MTF improves slightly for higher dose levels and showed a steady
progression from FBP to higher strength of ASIR. [[Figure 5 ]] shows MTF for the recommended protocol of 3.1 mGy.
Figure 5: Variation of spatial resolution with different reconstruction methods for the recommendation
protocol of 3.1 mGy
Results for the LCD are shown in [[Figure 6 ]]. Lower dose settings with the same reconstruction methods and contrast level tended
to give lower abilities to identify objects. With the same dose and contrast level,
smaller detail diameters could be detected with ASIR compared to FBP and with higher
strengths of ASIR.
Figure 6: Variation of low-contrast detail detectability with dose levels and different reconstruction
methods, for contrast levels of 1, 0.5, and 0.3%
Discussion
As ATCM systems and IR have been widely used, the effectiveness of dose reduction
has been studied. Several investigators have published data relating to the efficiency
of ASIR for pathologic findings, such as reticulations, nodules, and bronchiectasis,
in reducing dose by 32–65%for patient examinations, without increasing the noise level
in the reconstructed images.[[12 ]],[[20 ]],[[21 ]],[[22 ]],[[23 ]] The efficiency of ASIR when used together with ATCM has been studied by Qi et al . who suggested that use of 50% ASIR with a NI setting of 15 allowed a dose reduction
of 57% for chest CT.[[24 ]]
Image quality improved when either higher dose or ASIR levels were set [[Table 2 ]] and [[Figure 3 ]]. Results obtained from the Catphan phantom confirmed those from the chest phantom.
The measured noise (standard deviation of the pixel value) from Catphan for the NI
setting of 11.57 (CTDIvol of 5.2 mGy) reconstructed with FBP was about 7 HU. This was slightly lower than the
targeted noise level. NI is defined as the standard deviation of pixel values in the
central region of an image of a uniform water phantom. As the location of measuring
noise was at the edge of the phantom, there was less attenuation of photons, therefore
lower noise compared to the center. The results indicated that a higher level ASIR
gave similar image quality compared to the lower level ASIR with a NI setting that
was one step lower. For the standard protocol, the NI setting of 11.57 with 30% ASIR
gave the measured noise of about 6 HU or 15% lower compared to the target value. From
[[Figure 4 ]]B, similar noise value to the NI setting of 11.57 with FBP would be obtained for
the NI setting of 15.04 with 30% ASIR, NI setting of 16.78 with 50% ASIR, and NI setting
of 18.52 with 70% ASIR. Results for CNR varied in a similar way. This indicated that,
with the same noise level, users can set higher NI values when ASIR was implemented.
The effect of ASIR on CT numbers for different materials was evaluated. All reconstruction
methods gave relatively constant values of CT numbers with variations of up to 2 HU
These results were similar to those reported by Dodge et al .[[19 ]]
Results of LCD showed improvement when using higher level of ASIR, especially for
lower NI settings or higher dose. At 0.5 and 1% contrast, similar values of LCD were
found from use of the higher NI setting of 15.04 with 70% ASIR compared to the lower
NI setting of 11.57 mGy with FBP [[Figure 6 ]]. For the 0.3% contrast object, there were no significant differences between ASIR
and FBP for the NI settings higher than 16.78. This may be because the higher noise
obscured the small differences between the target and background. Results imply that
different NI and ASIR settings should be implemented for different investigations.
Diffuse lung disease that is considered to be a low-contrast resolution task may not
be detected with NI settings higher than 16.78.
The ability of CT image to detect small lung nodules is crucial for lung cancer screening.
Sharpness in terms of MTF is capable of separating smaller objects from surroundings.
Results showed the values were relatively constant for various NI (or dose) settings
[[Figure 5 ]]. Comparisons of MTF between FBP and ASIR showed that higher level of ASIR improved
resolution slightly. Results also suggested that the use of ATCM together with a higher
level of ASIR has the potential for lung cancer screening because the results for
noise, CNR, LCD, and MTF improved. SNR and CNR measured from P8 were similar to or
better than those from the standard, P1, and P4 protocols but lower than those from
P2 and P5 protocols [[Table 2 ]] and [[Figure 3 ]]. However, image quality analysis based on noise and CNR measurements does not completely
encompass the differences in image texture which may affect patient diagnoses. For
the protocol P8, although the objective image quality was not significantly changed
compared to the standard one, the subjective image quality scores in terms of contrast
decreased from 4.6 to 3.8 [[Table 3 ]]. This may be because of a different appearance of the images from more statistical
reconstruction of the CT data sets, noise and texture are different from those in
the standard protocol, and this may give an unsatisfactory diagnostic result. Therefore,
care must be taken not to use too high degree of ASIR as this could result in an over-smoothing
artifact.[[25 ]],[[26 ]] In detail, all readers gave scores of 4–5 for the contrast obtained from the standard
protocol and 3–4 for that from the other protocols. However, there was one reader
who gave a score of 2 for contrast and sharpness of protocol P2. This mismatching
classification induced lower ICC among readers. In order to clarify whether the lower
scores obtained had implication for diagnosis, the readers have been asked and confirmed
that protocol P8 produced adequate image quality that is suitable for interpreting
diagnostic images.
The optimal combination of setting a higher NI together with a higher percent of IR
in order to achieve a lower dose while preserving acceptable image quality could be
a way of following the ALARA principle. The objective of this study was to develop
new protocols that have a similar level of image quality compared to the routine chest
CT protocol of the hospital (Smart mA technique with the noise index of 11.57 and
30% ASIR reconstruction). The result suggested that protocol P8 provides the best
CNR and SNR with the lowest effective dose. Doses obtained from the present study
were lower than those reported from a survey in Thailand.[[6 ]] This is because the majority of CT scanners do not use IR and ATCM. CTDIvol can be reduced from 5.2 to 3.1 mGy (and effective dose from 2.96 to 1.75 mSv) [[Table 1 ]], which was about a 40% reduction while preserving a similar level of image quality
with a NI of 15.04 and 70% ASIR. The image quality for both objective and subjective
assessments were not significantly different from the routine protocol used in the
hospital.
This study has several limitations. First, this was a phantom study and the results
cannot be automatically transferred to humans. A slice thickness of 5 mm may not be
adequate for major indications for chest CT such as interstitial lung disease, lung
nodule, and pulmonary embolus. It is possible that this increased slice thickness
enhanced measures such as SNR that could have otherwise produced more noise at thinner
slices and also influenced the ideal NI and ASIR combination.[[27 ]],[[28 ]],[[29 ]] Lung CT images can be reconstructed using different filter reconstructions based
on clinical indications and these will affect the image noise and MTF. Hhowever, the
effect of such reconstructions on the image quality using the different variables
has not been included in this study. Additional studies may be performed in order
to determine the most efficient use of ATCM and ASIR with filter reconstructions.
Second, CT dose optimization in this study was based on SNR and CNR only. Other advanced
image quality parameters that take noise texture into account, such as the noise power
spectrum, should be considered. Lastly, effective dose was calculated by using a generic
DLP to effective dose conversion factor established for a broad range of different
scanners and scan settings, and therefore can only be used to give a broad idea of
the effective dose.
In summary, this phantom study evaluated quantitative and qualitative image quality
of chest CT at different combinations of noise index and ASIR level. The new protocol
of NI 15.04 with 70% ASIR can reduced the dose by 41% compared to the standard protocol
of NI 11.57 with 30% ASIR without degradation of image quality. Because scan parameters
affect CT ATCM system operation, additional studies should be performed in order to
determine the most efficient use of ATCM and ASIR.
Acknowledgements
Financial support from Naresuan University. The authors wish to thank Udorn Medical
Supply, which provided necessary facilities for carrying out this project. We would
also like to thank Mr. Kanlapaphat Temphon for his help in operating the CT scanner.
Financial support and sponsorship
Naresuan University supports research grants. Udorn Medical Supply supports a CT scanner
used in this study.
Conflicts of interest
There are no conflicts of interest.