Detailerkennbarkeit bei Durchleuchtung und Photographie mit der Bildverstärkerröhre*

P. M. van Alphen, G. C. E. Burger, W. J. Oosterkamp, M. C. Teves, T. Tol
  • Philips Forschungslaboratorium, N. V. Philips' Gloeilampenfabrieken, Eindhoven-Niederlande
* Vortrag gehalten auf der 34. Tagung der Deutschen Röntgen-Gesellschaft in Wiesbaden vom 28. bis 30. April 1952.
Further Information

Publication History

Publication Date:
27 August 2009 (online)

Summary

A Bakelite Thorax phantom was used to test the performance of an X-ray image amplifier. The visual acuity during fluoroscopy in a room with fair illumination after 1 min. of adaptation and with only 1/5 of the dose rate is almost twice as good as during ordinary fluoroscopy.

Photographs of the amplified image can be made with 1/10 of the energy used for ordinary contactphotographs ans 1/50 of the energy used for ordinary camera photographs. These photographs show a resolution of detail which lies midway between contactphotography and ordinary fluoroscopy.

The results are in good agreement with calculated data based upon the photon-fluctuation theory of Morgan and Sturm.

Résumé

Les auteurs ont fait des investigations avec un fantôme de bakelite sur les qualités d'un tube amplificateur à Rayons X. L'acuité de vision dans une chambre illuminée modérément en utilisant 1/5 de l'intensité des Rayons X est à peu près le double en comparaison de la radioscopie normale.

On peut photographier l'image amplifiéc en utilisant seulement 1/50 de l'énergie nécessaire pour une Photographie à contact et 1/50 de l'énergie nécessaire pour une photographie indirecte ordinaire. La résolution de détail de ces photographies est au milieu de la Photographie à contact et de la radioscopie.

Les résultats sont en bon accord avec la théorie basée sur les fluctuations des photons de Morgan et Sturm.

    >