Nuklearmedizin 1990; 29(02): 24-27
DOI: 10.1055/s-0038-1629512
Originaler Artikel
Schattauer GmbH

Tomographic Scintigraphy Using a Pinhole Collimator and a Rotating Gamma Camera

L. E. Olsson
1   From the Department of Radiation Physics, University of Lund, Malmö University Hospital Malmö, Sweden
,
L. Ahlgren
1   From the Department of Radiation Physics, University of Lund, Malmö University Hospital Malmö, Sweden
› Author Affiliations
Further Information

Publication History

Received: 29 May 1989

Publication Date:
05 February 2018 (online)

Preview

When a pinhole collimator is to be used for tomographic studies the standard SPECT software has to be modified so as to avoid serious image distortion outside the axis of rotation. The MTF measured at the axis of rotation shows that the better resolution of the pinhole collimator as compared with that of a parallel-hole collimator improves the tomographic image. The low sensitivity of the pinhole collimator excludes the use of an aperture smaller than 6 mm for patient tomographic studies where the activity which may be administered is restricted.

Zusammenfassung

Wenn ein Einlochkollimator für tomographische Studien an Patienten verwendet wird, muß die Software abgeändert werden, um ernsthafte Bilddeformierungen außerhalb der Rotationsachse zu vermeiden. Die MTF, gemessen in der Rotationsachse, zeigt, daß das größere Auflösungsvermögen des Einlochkollimators im Vergleich zu dem eines Parallellochkollimators eine Verbesserung des tomographischen Bildes bewirkt. In Patientenstudien, bei denen die einzusetzende Aktivität beschränkt ist, schließt die geringe Sensitivität des Einlochkollimators eine Blendenöffnung aus, die kleiner als 6 mm ist.